K091364 · Oehm Und Rehbein GmbH · LLZ · Dec 11, 2009 · Radiology
Device Facts
Record ID
K091364
Device Name
DICOMPACS, MODEL 5.2, DICOMPACS DX-R, MODEL 1.4
Applicant
Oehm Und Rehbein GmbH
Product Code
LLZ · Radiology
Decision Date
Dec 11, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
dicomPACS® and dicomPACS®DX-R are software systems for the administration, archiving, processing, improvement and compression of medical image data for diagnosis. The images are either acquired from imaging modalities via DICOM or imported directly. All images are archived in a database as DICOM compliant files. The data is displayed on a computer monitor for diagnosis. dicomPACS® and dicomPACS® DX-R also provides services for administering the data. dicomPACS® DX-R can control X-ray generators. dicomPACS®DX-R is not approved for the acquisition of mammographic image data. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 MP resolution and meets other technical specifications reviewed and accepted by FDA. Functions to be carried out using dicomPACS® and dicomPACS® DX-R are, for example, but not limited to adjustment of window leveling, rotation, zoom, and measurements. dicomPACS® and dicomPACS®DX-R are meant to be used by qualified medical personnel only. All users must be qualified to create and diagnose radiological image data.
Device Story
Software systems for administration, archiving, processing, and compression of medical images. Inputs include DICOM images from modalities or direct imports. dicomPACS® DX-R includes additional functionality for X-ray generator control, CR/DR acquisition, and raw data processing. Used by qualified medical personnel in clinical settings to view and manipulate images (window leveling, rotation, zoom, measurements, MPR). Output displayed on computer monitors for diagnostic purposes. Supports clinical decision-making by providing tools for image enhancement and management.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-based PACS and radiological image processing system. Features include DICOM compliance, multi-planar reconstruction (MPR), generator control, and CR/DR acquisition. Operates on standard computer hardware. No specific materials or energy sources defined as it is a software-only device.
Indications for Use
Indicated for qualified medical personnel to administer, archive, process, and diagnose medical image data. Not approved for mammographic image acquisition. Mammographic images require 5 MP resolution monitors for interpretation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણ
dicomPACS® I dicomPACS® - DX-R Special 510(k)
November 10, 2009
510(k) Summary
# DEC 1 1 2009
# 1. Administrative Information
### Reason for Submission:
### 510(k) Notification for dicomPACS® 5.2 and dicomPACS DX-R® 1.6
#### Submitter:
:
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
.
| Address: | Oehm und Rehbein GmbH<br>Waldemarstr. 20 g/h<br>18057 Rostock<br>GERMANY |
|---------------------------------------|--------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 3006542593 |
| Submission contact person: | Mr. Wolfgang Moeller |
| Contact telephone: | +49 381 2036 1295 |
| Contact e-mail: | wolfgang.moeller@oehm-rehbein.de |
| Common device name: | PACS (Picture Archiving and Communications System) |
| Classification: | Class II |
| Regulation no.: | 892.2050 |
| Product code: | LLZ |
| US Agent: | Alan Schwartz |
| US Agent company: | mdi Consultants, Inc. |
| US Agent address: | 55 Northern Blvd. Suite 200, Great Neck, NY 11021 |
| US Agent telephone: | 516-482-9001 |
| Substantial equivalent device: | |
| Device Name(s): | dicomPACS® 5.2<br>dicomPACS® DX-R 1.6 |
| Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Common Product Name: | Radiological Image Processing System |
| Predicate device: | |
| Device Name(s): | dicomPACS® 5 |
| Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Common Product Name: | Radiological Image Processing System |
| 510(k) Number: | K070618 |
Executive Summary
{1}------------------------------------------------
## 2. Indications for Use
dicomPACS® DX-R can control X-ray generators. dicomPACS®DX-R is not approved for the acquisition of mammographic image data.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 MP resolution and meets other technical specifications reviewed and accepted by FDA.
Functions to be carried out using dicomPACS® and dicomPACS® DX-R are, for example, but not limited to, adjustment of window leveling, rotation, zoom, and measurements.
dicomPACS® and dicomPACS®DX-R are meant to be used by qualified medical personnel only. All users must be qualified to create and diagnose radiological image data.
## 3. Functional Comparison
The software comes in two varieties - as dicomPACS® 5.2. and as dicomPACS® DX-R 1.6. While dicomPACS® DX-R 1.6 provides some functionality like generator control, CR/DR acquisition and raw data processing, dicomPACS® does not have these functions. The following table gives you an overview which configuration provides which new functionality.
The functions as given in the previous Premarket Notification, K070618, remain valid. Please find the 510(k) for the previous Premarket Notification in Sec. 11-A of this submission.
There have been modifications since version 5 of the software. The following table provides a list of all modifications that have taken place between dicomPACS® in version 5 and dicomPACS® in version 5.2 and dicomPACS® DX-R in version 1.6.
| | Komponente: | dicomPACS | dicomPACS DX-R |
|-----|---------------------------------------------------|-----------|----------------|
| 1. | Viewer-dicomWeb | X | X |
| 2. | Viewer-Prosthesis Module | X | X |
| 3. | 100% magnifying glass | X | X |
| 4. | MPR (multi-planar reconstruction) | X | X |
| 5. | Hanging protocol section area | X | X |
| 6. | Generator Control<br>[Accession No.: 0910034-000] | - | X |
| 7. | CR/DR acquisition | - | X |
| 8. | Raw data processing | - | X |
| 9. | Image overlay configuration | - | X |
| 10. | X-ray assistant | - | X |
| 11. | Examination tree | - | X |
| 12. | Virtual keyboard | - | X |
Table 1: Modifications to different configurations
#### Executive Summary
{2}------------------------------------------------
## 4. Significant Change
As layed out in the "List of Modifications" in Sec. 10 of this submission some of the current changes are significant according to the FDA memorandum #K97-1 "Deciding When to Submit a 510(k) for a Change to an Existing Device". These significant changes are
No. 4 - Multiplanar Reconstruction (MPR),
No. 7 - CR/DR acquisition, and
No. 8 - Raw data processing.
## 5. Substantial Equivalence Conclusion
None of the modifications alter the Indications for Use in a significant way, nor the fundamental scientific technology, and do not introduce a fundamentally new scientific technology.
We therefore believe that the information presented in this Special 510(k) demonstrate that the product is safe for the patient, user, and bystander and does not raise any new questions regarding safety and effectiveness.
#### Executive Summary
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Oehm und Rehbein GmbH % Mr. Alan Schwartz Official Correspondent mdi Consultants. Inc. 55 Northern Blvd., Suite 200 GREAT NECK NY 11021
# DEC 1 1 2009
Re: K091364
Trade/Device Name: dicomPACS® 5.2 / dicomPACS® DX-R 1.6 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 18, 2009 Received: December 2, 2009
Dear Mr. Schwartz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical
{4}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): Device Name: __ dicomPACS® 5.2 / dicomPACS® DX-R 1.6
Indications for Use:
dicomPACS® and dicomPACS®DX-R are software systems for the administration, archiving, processing, improvement and compression of medical image data for diagnosis. The images are either acquired from imaging modalities via DICOM or imported directly. All images are archived in a database as DICOM compliant files. The data is displayed on a computer monitor for diagnosis. dicomPACS® and dicomPACS® DX-R also provides services for administering the data.
dicomPACS® DX-R can control X-ray generators. dicomPACS®DX-R is not approved for the acquisition of mammographic image data.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 MP resolution and meets other technical specifications reviewed and accepted by FDA.
Functions to be carried out using dicomPACS® and dicomPACS® DX-R are, for example, but not limited to adjustment of window leveling, rotation, zoom, and measurements.
dicomPACS® and dicomPACS®DX-R are meant to be used by qualified medical personnel only. All users must be qualified to create and diagnose radiological image data.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
lomi mzh
(Division Sign-Off)
-Division of Reproductive, Abdominal and Radiological Devices 510(k) Number K091364
Indications for Use Statement
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.