REUNION TSA SYSTEM-HUMERAL PRESS-FIT & CEMENTED STEMS MODEL 5569-YY-20XX (YY=P,C; XX=07-17,06L-12L) REUNION TSA SYSTEM N
Applicant
Howmedica Osteonics Corp.
Product Code
KWS · Orthopedic
Decision Date
May 5, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reunion® TSA System components are sterile, single-use devices intended for use in primary and revision total shoulder arthroplasty to alleviate pain and restore function. The subject humeral stems will be offered in both cementless applications. The self pressurizing glenoid components are intended for cemented use only.
Device Story
ReUnion® Total Shoulder Arthroplasty (TSA) System is a modular orthopedic implant for shoulder joint replacement. Components include humeral stems (titanium alloy), modular neck adapters (cobalt chrome), single radius heads (cobalt chrome), and self-pressurizing glenoids (X3® polyethylene). Humeral stems are available in cemented and cementless (plasma spray/hydroxyapatite coated) designs. System is used by orthopedic surgeons in clinical/hospital settings to replace damaged shoulder joints, alleviate pain, and restore function. Surgeons select appropriate component sizes and configurations based on patient anatomy and pathology. Implantation involves standard surgical arthroplasty techniques. Benefits include improved joint mobility and pain relief for patients with degenerative or traumatic shoulder conditions.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, specifically fatigue strength testing of humeral components and dynamic evaluation of glenoid loosening conducted in accordance with ASTM F2028-08.
Technological Characteristics
Materials: Titanium alloy (ASTM F136) for humeral stems; Cobalt chrome (ASTM F1537, ASTM F75) for neck adapters and heads; X3® polyethylene (ASTM F648) for glenoids. Design: Modular humeral stems (cemented/cementless), single radius heads (standard/eccentric), and pegged/keeled glenoids. Cementless stems feature circumferential plasma spray and hydroxyapatite (HA) coating. Sterilization: Sterile, single-use.
Indications for Use
Indicated for hemi or total shoulder replacement in patients with aseptic necrosis of the humeral head; painful, disabling joint disease (degenerative, rheumatoid, or post-traumatic arthritis); proximal humeral fractures/dislocation; clinical management problems where arthrodesis or alternative reconstruction is less likely to succeed; or revision of previous unsuccessful shoulder replacement/resurfacing. Glenoid components are for cemented use only.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Osteonics Solar Shoulder Humeral Bearing Head (K990598)
Depuy Global Advantage Shoulder, Global Advantage Humeral Stem, Global Advantage Eccentric Head (K992065)
Depuy Global Shoulder Crosslinked Glenoid (K052472)
DVO Extremity Solutions Total and Hemi Shoulder System (K060988)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a handwritten string of characters. The string appears to be "K103835". The characters are written in a cursive style, with some connections between the letters and numbers. The image is in black and white.
# 510(k) Summary (
| Sponser: | Howmedica Osteonics Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430 |
|----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Avital Merl-Margulies<br>Regulatory Affairs Specialist<br>Howmedica Osteonics Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-6365 Fax: (201) 831-3365 |
| Date Prepared: | May 3, 2011 |
| Proprietary Name: | ReUnion® Total Shoulder Arthroplasty (TSA) System |
| Common Name: | Artificial Shoulder Components |
| Classification Name | 21 CFR §888.3660: Shoulder Joint Metal/Polymer Semi<br>Constrained Cemented Prosthesis<br>21 CFR §888.3690: Shoulder Joint Humeral (hemi-<br>shoulder) metallic uncemented prosthesis |
| Proposed Regulatory Class: | Class II |
| Product Codes: | 87KWS: Prosthesis, shoulder, semi-constrained,<br>metal/polymer, uncemented<br>87HSD: Prosthesis, shoulder, hemi-humeral, metallic,<br>cemented |
| Legally marketed device to which substantial equivalence is claimed: | |
| | Osteonics Shoulder Humeral Components: K955731 |
| | Solar ReUnion Fracture Stem: K062113 |
| | Osteonics All Polyethylene Glenoid Shoulder Component: K950521 |
| | Osteonics All Polyethylene Glenoid Shoulder Component: K962082 |
| | Solar Shoulder Offset Humeral Head: K001419 |
| | Osteonics Solar Shoulder Humeral Bearing Head: K990598 |
| | Depuy Global Advantage Shoulder, Global Advantage Humeral Stem, Global Advantage<br>Eccentric Head: K992065 |
| | Depuy Global Shoulder Crosslinked Glenoid: K052472 |
| | DVO Extremity Solutions Total and Hemi Shoulder System: K060988 |
Device Description:
The subject Reunion® Total Shoulder Arthroplasty (TSA) System is intended for shoulder arthroplasty. The components of this system consist of humeral stems, a
Page 1 of 3
{2}------------------------------------------------
# K103835
modular neck adapter, single radius heads, and self pressurizing glenoids (SPG). The humeral stem components will be manufactured from titanium allov according to ASTM F136 and will be offered in both cemented and cementless designs. The cementless humeral stem design will feature a circumferential plasma spray and hydroxyapatite (HA) coating and be offered in 7mm-17mm distal diameters in 1mm increments. The cemented humeral stems will be offered in 6mm-15mm distal diameters in 1mm increments and will also include 4 longer stems with distal diameters 6mm-12mm in 2 mm increments. These humeral stems were designed to mate with the subject single radius heads or the modular neck adapter, manufactured from cobalt chrome according to ASTM F1537, for compatibility with other marketed humeral heads. The single radius heads will be manufactured from cobalt chrome according to ASTM F75 and will be offered in spherical diameter head sizes 40mm-56mm in 4mm increments for both standard and eccentric designs. The self pressuring glenoids (SPG) will be manufactured from X3% polyethylene according to ASTM F648 and will mate with the single radius heads. The SPGs will be offered in both pegged and keeled configurations ranging from 40mm-56mm spherical diameters in 4mm increments.
#### Intended Use:
The Reunion® TSA System components are sterile, single-use devices intended for use in primary and revision total shoulder arthroplasty to alleviate pain and restore function. The subject humeral stems will be offered in both cementless applications. The self pressurizing glenoid components are intended for cemented use only.
### Indications for Use:
#### For use as a Hemi or Total Shoulder Replacement:
- Aseptic necrosis of the humeral head .
- Painful, disabling joint disease of the shoulder resulting from: degenerative . arthritis, rheumatoid arthritis, or post-traumatic arthritis.
- Proximal humeral fractures and/or dislocation. .
- Clinical management problems where arthodesis or alternative reconstructive . techniques are less likely to achieve satisfactory results.
- Revision of previous unsuccessful total shoulder replacement, resurfacing or other . procedure.
The gleonid components are intended for cemented use only.
#### Summary of Technologies:
The technological characteristics (material, design, sizes, and operational principles) of the ReUnion® TSA System components are similar or identical to its predicate devices.
Non-Clinical Testing: Non-clinical laboratory testing was performed to determine substantial equivalence. The following mechanical tests were conducted according to ASTM F2028-08: Fatigue strength on humeral components and dynamic evaluation of glenoid loosening on the glenoids. The testing demonstrated that the ReUnion® TSA
Image /page/1/Picture/15 description: The image shows the text "Page 2 of 3" written in cursive. The text indicates that it is the second page of a three-page document. The handwriting is somewhat stylized, with distinct loops and curves forming the letters. The overall impression is that of a handwritten note or document page number.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a handwritten alphanumeric string. The string is "K1D3835". The characters are written in a cursive style, with some connections between the letters and numbers.
system components met performance requirements and are as safe and effective as their predicate devices.
# Clinical Testing:
None provided as a basis for substantial equivalence
## Conclusion
The ReUnion® TSA System is substantially equivalent to the predicate devices identified in this premarket notification.
Page 363
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Howmedica Osteonics Corp. % Ms. Avital Merl-Margulies Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430
MAY - 5 2011
Re: K103835
Trade/Device Name: ReUnion® Total Shoulder Arthroplasty (TSA) System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS, HSD Dated: April 29, 2011 Received: May 2, 2011
Dear Ms. Merl-Margulies:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
#### Page 2 - Ms. Avital Merl-Margulies
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Aig B. R h
Mark N. Melker Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known): K103835
Device Name: ReUnion® Total Shoulder Arthroplasty (TSA) System
# For use as a Henii of Total Shoulder Replacement:
- Aseptic necrosis of the humeral head o
- Painful, disabling joint disease;of the shoulder resulting from: degenerative arthritis, o rheumatoid arthritis, or post-traumatic arthritis.
- Proximal humeral fractures and/or dislocation. o
- Clinical management problems where arthrodesis or alternative reconstructive techniques o are less likely to achieve satisfactory results.
- Revision of previous unsuccessful total shoulder replacement, resurfacing or other o procedure.
The gleonid components are intended for cemented use only. The humeral stem components are intended for both cemented and cementless use.
| Prescription Use<br>(Part 21 CFR 801, Subpart D) | X |
|--------------------------------------------------|------------------------------------------------|
| AND/OR | Over-The-Counter Use<br>(21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Page 1 of 1 | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|---------------|-----------------------------------------------------------|
| | (Division Sign-Off) |
| | Division of Surgical, Orthopedic, and Restorative Devices |
| 510(k) Number | K103835 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.