The intended use of the modified humeral bearing heads is identical to that of the unmodified humeral bearing heads. As with the predicate humeral bearing heads, the modified humeral bearing heads are single use devices. They are intended for mating with the Osteonics® Solar Humeral Stems. The Osteonics® Solar Humeral Components are designed to be used in either cemented or cementless applications, as either a hemi-shoulder or a total shoulder component. For use as a total shoulder replacement, the modified and predicate humeral bearing heads may be used in conjunction with any legally marketed Osteonics glenoid component. The indications for the Osteonics® Solar Shoulder Components include: - Aseptic necrosis of the humeral head. - Painful. disabling joint disease of the shoulder resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis. - . Proximal humeral fracture and/or dislocation. - . Revision of previous unsuccessful total shoulder replacement, resurfacing or other procedure. - Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
Device Story
Osteonics® Solar Shoulder Humeral Bearing Head is a prosthetic component for shoulder arthroplasty; used as hemi-shoulder or total shoulder replacement; mates with Osteonics® Solar Humeral Stems; compatible with legally marketed Osteonics glenoid components. Device modification involves transition from wrought ASTM F-1537 cobalt chromium alloy to cast ASTM F-75 cobalt chromium alloy; includes minor dimensional changes (draft angles 0.5 to 1.5 degrees) to facilitate casting mold removal. Intended for single use; implanted by orthopedic surgeons in clinical/hospital settings. Provides structural replacement for humeral head to restore joint function and reduce pain in patients with degenerative or traumatic shoulder conditions.
Clinical Evidence
Bench testing only. Engineering analyses performed to demonstrate that mechanical properties of the cast cobalt chromium alloy are within acceptable limits for the device.
Technological Characteristics
Material: Cast ASTM F-75 cobalt chromium alloy (replacing wrought ASTM F-1537). Design: Humeral bearing head with 0.5 to 1.5-degree draft angles on internal walls for casting. Energy source: None (passive implant). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for patients with aseptic necrosis of the humeral head; painful, disabling shoulder joint disease (degenerative, rheumatoid, or post-traumatic arthritis); proximal humeral fracture/dislocation; revision of unsuccessful total shoulder replacement/resurfacing; or clinical management cases where arthrodesis or alternative reconstruction are less likely to achieve satisfactory results.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Osteonics® Solar Shoulder Humeral Bearing Head (K955731)
Submission Summary (Full Text)
{1}------------------------------------------------
3/15/99
K990598
# Special 510(k) - Device Modification Summary of Safety and Effectiveness for the Osteonics® Solar Shoulder Humeral Bearing Head
#### Submission Information
| Name and Address of the Sponsor<br>of the 510(k) Submission: | Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, NJ 07401-1677 |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Marybeth Naughton<br>Regulatory Affairs Team Member |
| Date of Summary Preparation: | February 23, 1999 |
| Device Identification | |
| Proprietary Name: | Osteonics® Solar Shoulder Humeral<br>Bearing Head |
| Common Name: | Shoulder Prosthesis |
| Classification Name and Reference: | Shoulder joint humeral (hemi-shoulder)<br>Metallic uncemented prosthesis<br>21 CFR §888.3690<br>Or<br>Shoulder joint metal/polymer semi-<br>constrained cemented prosthesis<br>21 CFR §888.3660 |
# Predicate Device Identification
The modified features of the Osteonics® Solar Shoulder Humeral Bearing Heads are substantially equivalent to features of the following Osteonics® predicate devices, which have been cleared for marketing via the 510(k) process:
- · Osteonics® Solar Shoulder Humeral Bearing Head (#K955731)
#### Device Description
The Osteonics® Solar Shoulder Humeral Bearing Heads are currently marketed devices that are being modified. The modification to the device involves a manufacturing process change in
{2}------------------------------------------------
manufacturing of the humeral bearing head from wrought ASTM (American Society for Testing and Materials) F-1537 cobalt chromium alloy to cast ASTM F-75 cobalt chromium alloy. Howmedica Osteonics intends to replace the current line of wrought cobalt chromium alloy humeral bearing heads with the subject cast cobalt chromium alloy humeral bearing heads. Minor dimensional changes will accompany the manufacturing process change. When changing from a machining process to a casting process. certain design features (dimensional) must be added to the cast part that is not necessary to the machined part when being manufactured. The internal geometry of the cast part requires draft angles to be placed on all walls of the device that are parallel to the parting line to facilitate the removal of the part from the casting mold. These angles range from approximately 0.5 degrees to 1.5 degrees per side.
## Intended Use:
The intended use of the modified humeral bearing heads is identical to that of the unmodified humeral bearing heads. As with the predicate humeral bearing heads, the modified humeral bearing heads are single use devices. They are intended for mating with the Osteonics® Solar Humeral Stems. The Osteonics® Solar Humeral Components are designed to be used in either cemented or cementless applications, as either a hemi-shoulder or a total shoulder component. For use as a total shoulder replacement, the modified and predicate humeral bearing heads may be used in conjunction with any legally marketed Osteonics glenoid component. The indications for the Osteonics® Solar Shoulder Components include:
- Aseptic necrosis of the humeral head.
- Painful. disabling joint disease of the shoulder resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis.
- . Proximal humeral fracture and/or dislocation.
- . Revision of previous unsuccessful total shoulder replacement, resurfacing or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
## Performance Data:
Engineering Analyses have been performed to demonstrate that mechanical properties of the cast cobalt chromium alloy to be well within the acceptable limits for the subject device.
## Statement of Technological Comparison:
- All features of the Osteonics® Solar Shoulder Humeral Bearing Head are unchanged with the exception of 1) the change in materials from wrought ASTM F-1537 cobalt chromium alloy to cast ASTM F-75 cobalt chromium alloy and 2)minor dimensional changes required for the casting procedure.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's logo, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 5 1999
Ms. Elizabeth A. Staub Vice President, Quality Assurance/Regulatory Compliance/Clinical Research Howmedica Osteonics Corporation 59 Route 17 Allendale, New Jersey 07401-1677
K990598 Re: Osteonics® Solar Shoulder Humeral Bearing Head Regulatory Class: III Product Codes: KWS and HSD Dated: February 23, 1999 February 24, 1999 Received:
Dear Ms. Staub:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate able beated in the enoices, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
#### Page 1 of 1
# 510(k) Number (if known): KY90598
Device Name: Osteonics® Solar Shoulder Humeral Bearing Head
Indications For Use:
The indications for the use of the Osteonics® Solar Shoulder Humeral Bearing Head , in keeping with those of other legally marketed Osteonics® shoulder components, are as follows:
#### Indications
- Aseptic necrosis of the humeral head. ●
- Painful, disabling joint disease of the shoulder resulting from: degenerative arthritis, ● rheumatoid arthritis or post-traumatic arthritis.
- Proximal humeral fracture and/or dislocation.
- Revision of previous unsuccessful total shoulder replacement, resurfacing or other ● procedure.
- Clinical management problems where arthrodesis or alternative reconstructive ● techniques are less likely to achieve satisfactory results.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109) OR
Over-The-Counter Use*_*_
(Optional Format 1-2-96)
Aisellifo
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K990898
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.