K103564 · Renal Solutions, Inc. · KDI · Feb 10, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K103564
Device Name
VENOFER PUMP
Applicant
Renal Solutions, Inc.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Feb 10, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Venofer® Pump Is an optional accessory for use on Fresenius 2008T Hemodialysis Machines and is intended to administer Venofer®, an iron sucrose, to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.
Device Story
Venofer Pump 1.1 is an optional module for Fresenius 2008T Hemodialysis Machines; designed to administer Venofer (iron sucrose) during dialysis. Device components include control panel, vial holder, fluid detector, and peristaltic pump. Operated by clinical staff in dialysis settings. Device automates iron sucrose delivery during hemodialysis treatment; output is controlled infusion of medication. Benefits include efficient, physician-prescribed iron supplementation for patients with chronic kidney disease-related anemia.
Clinical Evidence
No clinical data. Evidence consists of non-clinical verification testing, including performance testing, system-level hazard analysis, electromagnetic compatibility (IEC 60601-1-2: 2007), biocompatibility (AAMI/ANSI/ISO 10993-1:2009), and software verification/validation.
Technological Characteristics
Optional module for Fresenius 2008T Hemodialysis Machine; features peristaltic pump, fluid detector, vial holder, and control panel. Biocompatibility per AAMI/ANSI/ISO 10993-1:2009; EMC per IEC 60601-1-2:2007. Software-controlled operation.
Indications for Use
Indicated for patients with chronic kidney disease undergoing hemodialysis who require treatment for iron deficiency anemia via Venofer (iron sucrose) administration, as prescribed by a physician.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
# Response to FDA Questions K103564
Venofer Pump Version 1.1 Special 510(k)
FEB 1 0 2011
### Section 5: 510(k) Summary
# 510(k) Summary
# Submitter's Information
| Name | Renal Solutions, Inc. |
|-----------------------|------------------------------------------------------------------------------------------------|
| Address | 770 Commonwealth Drive<br>Warrendale, Pa 15086<br>Phone: (724) 720-2840<br>FAX: (724) 772-6925 |
| Official Contact | David J. Vanella<br>Senior Vice President, Quality Systems |
| Date Prepared | January 7, 2011 |
| Device Information | |
| Name | Venofer Pump, for use on the Fresenius<br>2008T Hemodialysis Machine |
| Common/Usual Name | Hemodialysis System |
| Product Code | KDI |
| Classification Name | Dialyzer, High Permeability With Or Without<br>Sealed Dialysate System |
| Regulation Number | 876.5860 |
| Proprietary Name | Venofer Pump |
| Unmodified Device | Venofer Pump (K093964) |
| Reason for Submission | Modification to existing device |
#### Substantial Equivalence
The Venofer Pump 1.1 is substantially equivalent to the unmodified device (Venofer Pump K093964) in terms its intended use, environment of use, operating principles, and technology.
Renal Solutions, Inc.
Venofer® Pump 1.1 Special 510(k)
{1}------------------------------------------------
# Venofer Pump Version 1.1 Special 510(k)
### Section 5: 510(k) Summary
# Intended Use
The intended use of the Venofer Pump 1.1 is identical to the unmodified device.
The Venofer® Pump Is an optional accessory for use on Fresenius 2008T Hemodialysis Machines and is intended to administer Venofer®, an iron sucrose, to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.
# Venofer Pump 510(k) Verification Testing Non-Clinical Summary
The verification (non-clinical) testing information consists of performance, safety, and software testing that was performed to verify the Venofer Pump meets its performance specifications and to demonstrate the device is substantially equivalent to the established predicate devices.
The following table summarizes the 510(k) verification testing activities performed. These include performance, safety and software testing, which demonstrates by technical examination that the Venofer Pump meets its performance specifications, the designated (FDA Consensus) standard requirements, and the software design input requirements.
| 510(k) Verification Testing | 510(K) Verification Testing Activities |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Venofer Pump Performance Testing<br>The performance (verification)<br>testing results support the<br>performance characteristics of the<br>Venofer Pump. | Functional testing that demonstrates that<br>the device performs as designed and<br>expected, includes the following:<br>The specific verification tests<br>conducted A description of all test protocol<br>including: objective of the tests test articles used in the tests test methods and procedures pre-defined acceptance or pass/fail<br>criteria. System-level hazard analysis that<br>confirms that the device does not<br>perform in an unexpected and/or<br>unsafe manner |
Renal Solutions, Inc.
Venofer® Pump 1.1 Special 510(k)
{2}------------------------------------------------
# Venofer Pump Version 1.1 Special 510(k)
# Section 5: 510(k) Summary
| 510(k) Verification Testing | 510(K) Verification Testing Activities |
|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Venofer Pump Safety Testing | |
| The verification testing results<br>support the safety characteristics of | Product safety testing that demonstrates that<br>the device performs per the FDA Consensus<br>Standards, as identified below: |
| the Venofer Pump.<br>Biocompatibility testing was<br>performed on all new materials that<br>are patient-fluid contacting | Electromagnetic compatibility (EMC)<br>testing (IEC 60601-1-2: 2007 - Medical<br>Electrical Equipment - Part 1-2: General<br>Requirements for Safety - Collateral<br>Standard: Electromagnetic Compatibility<br>- Requirements for Testing Biocompatibility Testing (AAMI / ANSI /<br>ISO 10993-1:2009 Biological evaluation<br>of medical devices Part 1: Evaluation &<br>testing ) |
| Venofer Pump Software Testing<br>The testing includes the required | Software testing that demonstrates the<br>device software meets the design input<br>requirements. The device was tested and<br>Includes the following |
| documentation as described in the<br>guidance titled Guidance for the<br>Content of Premarket Submissions<br>for Software | Structure chart of flow chart describing<br>software architecture Summary of software development<br>procedures, including changes made to<br>the software Software requirements specification with<br>traceability back to the hazard analysis Verification and validation test plans,<br>including pass/fail criteria and traceability<br>back to the requirements; System level test results Current software version number and<br>date of latest revision. |
Renal Solutions, Inc. Venofer® Pump 1.1 Special 510(k)
{3}------------------------------------------------
### Response to FDA Questions K103564
### Venofer Pump Version 1.1 Special 510(k)
#### Section 5: 510(k) Summary
The conclusions drawn from this testing demonstrates that the Venofer Pump is as safe, as effective, and performs at least as safely and effectively as the legally marketed devices identified as predicate devices to which it was compared.
#### General Safety and Effectiveness
The Venofer Pump 1.1 is an updated version of the Venofer Pump (K093964). The performance and technological characteristics of the modified device are equivalent to those of the unmodified device and raise no new types of safety or effectiveness questions.
#### Device Description
The Venofer Pump Module is an optional module for use on Fresenius 2008T Hemodialysis Machines. The module is designed to administer Venofer, an iron sucrose supplement, during dialysis treatments. The Venofer Pump Module is to be used in accordance with the approved Venofer Indications for Use and the physician's prescription.
The Venofer Pump Module fits into the module compartment of Fresenius 2008T Machines and consists of a control panel, vial holder, fluld detector, and a peristaltic pump.
Renal Solutions, Inc. Venofer® Pump 1.1 Special 510(k)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring. MD 20993-0002
Mr. David J. Vanella Senior Vice President, Quality Systems Renal Solutions®, Inc. 770 Commonwealth Drive, Suite 101 WARRENDALE PA 15086
FEB 10 201
Re: K103564
Trade/Device Name: Venofer Pump Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: January 7, 2011 Received: January 11, 2011
Dear Mr. Vanella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability … warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{5}------------------------------------------------
### Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm 17-10-11-11-15 the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolloffree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default:htm.
Sincerely yours,
Hubert Lemmer MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health.
#### Enclosure
{6}------------------------------------------------
KI03564
Venofer Pump Version 1.1 Special 510(k)
# Section 5: Indications for Use Statement
# Indications for Use Statement
510(k) Number (if known): N/A
Device Name: Venofer Pump
Indications for Use:
Fresenius 2008T is indicated for acute and chronic dialysis therapy.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
tive, Gastro-Renal, and SIU(K) NUME
> Renal Solutions, Inc. Venofer® Pump 1.1 Special 510(k)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.