AVEA VENTILATOR

K103211 · Care Fusion · CBK · May 12, 2011 · Anesthesiology

Device Facts

Record IDK103211
Device NameAVEA VENTILATOR
ApplicantCare Fusion
Product CodeCBK · Anesthesiology
Decision DateMay 12, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The AVEA is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It may be used on neonatal through adult patients. It should only be operated by properly trained clinical personnel, under the direction of a physician.

Device Story

AVEA is a servo-controlled, software-driven ventilator for neonatal to adult patients; used in institutional settings (hospitals) by trained clinical personnel. Device features a graphical user interface (UIM) with flat-panel color LCD, touch screen, and physical controls for parameter adjustment. Internal gas delivery system provides active inhalation/exhalation; internal batteries support inter-hospital transport and power backup. Device operates in conventional or non-invasive positive pressure ventilation (NPPV) modes. New software enhancements include 'Volume Guarantee' (automated regulation of inspiratory pressure to achieve clinician-specified target tidal volume in TCPL and pressure modes) and 'Nasal Intermittent Mandatory Ventilation' (time-triggered, time-cycled pressure control via nasal prongs). Clinicians monitor real-time charting and digital data on the UIM to adjust respiratory support, aiming to optimize patient ventilation and gas exchange.

Clinical Evidence

Bench testing only. Performance testing verified that the modified AVEA Ventilator meets all performance requirements and is substantially equivalent to legally marketed devices.

Technological Characteristics

Servo-controlled, software-driven ventilator. Features flat-panel color LCD UIM with touch screen, membrane keys, and dial. Internal gas delivery system with active inhalation/exhalation. Battery-powered for transport. Software-based control system. Connectivity via standard ventilator accessories.

Indications for Use

Indicated for continuous respiratory support in institutional settings for neonatal through adult patients. Contraindications not specified.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS | Manufacturing<br>Site: | CareFusion<br>22745 Savi Ranch Parkway<br>Yorba Linda, CA 92887<br>Tel. (714)922-7615 | |-------------------------------------------|---------------------------------------------------------------------------------------| | Contact: | Farokh Etemadieh (714) 922-7615 (phone); (714) 922-7615 (fax) | | Summary Date | May 12, 2011 | | Device Trade<br>Name: | AVEA Ventilator | | Device<br>Common/Classific<br>ation Name: | Classification name: 868.5895 Continuous Ventilator, 73 CBK | | Establishment<br>Registration<br>Number | 2050001 | | Device Class: | Class II | | Classification<br>Panel: | Anesthesiology | | Predicate Device: | The predicate devices are: | | | 1. K083693: 840 Ventilator (Covidien)<br>2. K974176: Baby log 8000 (Drager Inc.) | The AVEA is a servo-controlled, software-driven ventilator. It has a Device dynamic range of breathing gas delivery that provides for neonatal Description: through adult patients. Its graphical user interface module (UIM) has a flat panel color LCD with real time charting and digital monitoring capabilities, a touch screen for interaction, membrane kevs and a dial for changing settings and operating parameters. It also has an internal gas delivery system with servo controlled active inhalation and exhalation functions. Using internal batteries this provides inter-hospital transport as well as back up capability due to loss of AC power. The AVEA may be configured as a conventional ventilator or non-invasive positive pressure ventilator (NPPV). It has been designed to function using commonly available accessories. AVEA Volume Guarantee / Nasal intermittent positive pressure ventilation {1}------------------------------------------------ ### Intended Use: The AVEA is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It may be used on neonatal through adult patients. It should only be operated by properly trained clinical personnel, under the direction of a physician. ### Substantial Equivalence: The AVEA Ventilator is the same device as the AVEA Ventilator, which was cleared for market under 510(k) K013642, K022674, k062093, K073069, and K081837. #### Modifications to the AVEA Ventilator associated with this submittal: The purpose of this traditional 510K submission is to request authorization for enhancement to the AVEA Ventilator applicable only to changes that are necessary to software implement Volume Guarantee which is the automated requlation of inspiratory pressure to achieve a Clinician specified target tidal volume, and is applicable to TCPL and PRESSURE modes of Ventilation and Nasal Intermittent Mandatory Ventilation which is a time-triggered time-cycled mode of pressure control Ventilation provided via nasal prongs on cannula. All other AVEA requirements shall remain unchanged. The modified AVEA Ventilator have the following similarities to those which previously received 510(k) concurrence: - Have the same indicated use, ● - Use the same ventilation operating principle, . - Incorporate the same basic ventilator design with the exception of . modifications described in this submittal. - Incorporate the same basic electronic control system . - Are manufactured and packaged utilizing the same basic processes. . In summary, the AVEA Ventilator described in this submission is, in our opinion, substantially equivalent (Volume Guarantee and Nasal intermittent positive pressure ventilation) to the predicate devices. #### Summary of testing and validation: Performance testing verified that the AVEA Ventilator meets it's performance requirements and that this device is substantially equivalent to medical devices currently legally marketed in the United States. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Farokh Etemadieh Senior Regulatory Affairs Associate CareFusion 22745 Savi Ranch Parkway Yorba Linda, California 92887 MAY 1 2 2011 Re: K103211 Trade/Device Name: AVEA Ventilator Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: May 6, 2011 Received: May 11, 2011 Dear Mr. Etemadieh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2- Mr. Etemadieh Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Ph for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ ## Indications for Use 510(k) number (if known): K103211 Device Name: AVEA Ventilator Indications for Use: The AVEA is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It may be used on neonatal through adult patients. It should only be operated by properly trained clinical personnel, under the direction of a physician. Prescription Use ✔ AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED) L. Schulther (Division Sign-Off) Division of Anesthesiology, General Hospital nfection Control, Dental Devices 510(k) Number: k103211 Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...