PB 840 VENTILATOR SYSTEM WITH LEAK COMPENSATION OPTION, MODEL 840
Applicant
Covidien, Formerly Nellcor Puritan Bennett, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Apr 2, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The 840 Ventilator System with Leak Compensation Option is intended to provide continuous ventilation to patients requiring respiratory support. The device is intended for patients with an Ideal Body Weight (IBW) as low as 0.5 kg (with NeoMode option) to adult, and for use in a wide variety of clinical conditions. The 840 Ventilator System with Leak Compensation Options is intended for a wide range of patients ranging from neonate to adult (V- 5-2500 mL with NeoMode) or from infant to adult (V-25-2500 mL). The 840 Ventilator System with Leak Compensation is intended for use in hospital and hospital-type facilities. It may be used during hospital and hospital-type facility transport provided that electrical power and compressed gas are supplied.
Device Story
The 840 Ventilator System with Leak Compensation (LC) is a critical care ventilator for neonate-to-adult patients. The LC option adds software functionality to the existing 840 platform to detect and characterize gas flow leaks in the breathing circuit. It does not compensate for lost delivered volume but adjusts flow to optimize breath triggering/cycling and maintain set PEEP. The device displays leak-related parameters for clinician review. Used in hospitals and during transport by clinicians; the system is dual-microprocessor controlled. Clinicians use the output to manage ventilation settings for patients with circuit leaks, potentially improving patient-ventilator synchrony and therapy efficacy.
Clinical Evidence
Bench testing only. Functional testing confirmed the device meets performance specifications. Compliance with the July 1995 'Draft Reviewer Guidance for Ventilators' was verified. Software verification was conducted per FDA's 2005 guidance for software in medical devices.
Technological Characteristics
Dual-microprocessor controlled critical care ventilator. Software-based leak compensation algorithm. Operates on electrical power and compressed gas. Supports invasive and non-invasive ventilation. No hardware or firmware changes from the base 840 platform.
Indications for Use
Indicated for patients requiring continuous respiratory support, ranging from neonates (IBW 0.5 kg) to adults. Suitable for invasive or non-invasive ventilation in hospital and hospital-type facility settings, including transport.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
Puritan-Bennett Corp. 840 Ventilator System with NeoMode (K001646)
Puritan-Bennett Corp. 840 Ventilator System (K970460)
Puritan-Bennett Corp. 840 Ventilator System with Bilevel Ventilation (K984535)
Puritan-Bennett Corp. 840 Ventilator System with Volume Ventilation Plus (K021573)
Puritan-Bennett Corp. 840 Ventilator System with Respiratory Mechanics (K063650)
Puritan-Bennett Corp. 840 Ventilator System with Proportional Assist Ventilation (K053388)
Submission Summary (Full Text)
{0}------------------------------------------------
Kof 3693
2101 Faraday Avenue Carlsbad, CA 92008
Tele: 760-603-5300 Fax: 760-603-5355
# 510(k) Summary
| Submitted by: | Puritan-Bennett Corporation<br>2200 Faraday Avenue<br>Carlsbad, CA 92008 | APR - 2 2009 |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Company Contact: | Kim L. Bloom<br>Sr. Regulatory Affairs Specialist<br>Phone: (760) 603-5076<br>Fax: (760) 603-5350<br>kim.bloom@covidien.com | |
| Date Summary Prepared: | December 11, 2008 | |
| Product Name: | 840 Ventilator System with Leak Compensation | |
| Common Name: | Ventilator | |
| Classification: | Class II; Continuous Ventilator per 21 CFR §868.5895 | |
| Legally Marketed<br>(Unmodified) Device: | <ul><li>Puritan-Bennett Corp. 840 Ventilator System, K970460</li><li>Puritan-Bennett Corp. 840 Ventilator System with Bilevel Ventilation, K984535</li><li>Puritan-Bennett Corp. 840 Ventilator System with Neomode Option, K001646</li><li>Puritan-Bennett Corp. 840 Ventilator System with Volume Ventilation Plus, K021573</li><li>Puritan-Bennett Corp. 840 Ventilator System with Respiratory Mechanics, K063650</li><li>Puritan-Bennett Corp. 840 Ventilator System with Proportional Assist Ventilation, K053388</li></ul> | |
| Predicate Devices: | <ul><li>Puritan-Bennett Corp. 840 Ventilator System with NeoMode, K001646</li><li>Maquet Servo I, K073179</li><li>Hamilton Medical, G5 Ventilator, K070513</li></ul> | |
# Device Description:
The Leak Compensation Option enables the 840 Ventilator System to automatically detect and characterize gas flow that leaves the breathing circuit through air leak(s) and does not return to the 840's exhalation valve for inclusion in the displayed exhaled spirometry. LC does not makeup for lost delivered volume but does add the necessary additional flow into the breathing circuit to optimize breath triggering and cycling functions as well as maintain the set PEEP, and display clinically valuable leak-related parameters for consideration by the clinician. The Leak
Nellcor Puritan Bennett Incorporated Confidential 840 Ventilator System with Leak Compensation Option
Page 1 of 3
{1}------------------------------------------------
Compensation feature is implemented on the 840 Ventilator through additional functionality in software and by use of the existing User Interface panel. No hardware or firmware changes or additions were required. The 840 Ventilator is a dual-microprocessor controlled, critical care ventilator intended to provide continuous ventilation for neonate to adult (with NeoMode Option) or infant to adult (without NeoMode Option) patients who require either invasive ventilation or non-invasive ventilation (via face mask).
### Indications For Use:
The 840 Ventilator with Leak Compensation Option is intended to provide continuous ventilation to patients requiring respiratory support. The device is intended for patients with an Ideal Body Weight (IBW) as low as 0.5 kg (with NeoMode option) to adult, and for use in a wide variety of clinical conditions.
The 840 Ventilator with Leak Compensation Options is intended for a wide range of patients ranging from neonate to adult (V- 5-2500 mL with NeoMode) or from infant to adult (V-25-2500 mL).
The 840 Ventilator with Leak Compensation is intended for use in hospital and hospital type facilities. It may be used during hospital and hospital-type facility transport provided that electrical power and compressed gas are supplied.
### Determination of Substantial Equivalence:
The intended use of the 840 Ventilator Leak Compensation is the same as that for conventional, currently marketed, critical care ventilators with similar functions. The materials and design of this device are similar to those of the predicate devices. The technical characteristics of the device modification do not introduce new questions regarding safety or effectiveness of critical care ventilators. Furthermore, the labeling associated with the 840 Ventilator with Leak Compensation provides similar information as the predicate devices.
Information provided in this 510(k) submission provides comparative, predicate device information and describes development procedures that support the determination of substantial equivalence and assertion that the modified device is safe and effective for its intended use. Software design and development, (including verification testing, test and software quality procedures) were conducted using FDA's Guidance for the Content of Pre-market Submissions for Software contained in medical devices, dated May 11, 2005 as a guidance and per internal company requirements.
In summary, Puritan-Bennett Corporation has provided information that indicates the 840 Ventilator with Leak Compensation to be safe and effective. This device is considered to be substantially equivalent to currently marketed devices, incorporating similar functionality to those that have been previously cleared by FDA.
The term "Substantial Equivalence" as used in this 510(k) Premarket Notification submission is limited to the definition of Substantial Equivalence found in the Federal Food, Druq, and Cosmetic Act, 21 CFR § 807, Subpart E. A determination of substantial equivalency under this submission is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of, substantial equivalence in this submission shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
{2}------------------------------------------------
### Explanation of how intended use and fundamental scientific technology have not changed:
The 840 Ventilator with and without Leak Compensation is intended for use as continuous, critical care ventilator for patients requiring respiratory support. The modification, which adds Leak Compensation to the currently cleared 840 Ventilator, does not alter the intended use of the 840 Ventilator. The intended use and indication of the 840 Ventilator System with Leak Compensation Option, as described in its labeling, are the same as the intended use and indication for the 840 Ventilator System with NeoMode.
Based on the above comparative predicate information, the 840 Ventilator with Leak Compensation is substantially equivalent to the Puritan Bennett 840 Ventilator with NeoMode, the Maquet Servo I, and the Hamilton G5 Ventilator. The device modifications have not altered the fundamental scientific technology of the predicate device, the 840 Ventilator System with NeoMode. Furthermore, based on the descriptive information and risk analysis, the device modification does not raise significant or new questions of safety and effectiveness for the 840 Ventilator.
### Summary of Performance Testing:
- 1. Functional testing confirms that the 840 Ventilator with Leak Compensation Ventilator is capable of meeting its stated performance specifications. The device passed all tests.
- 2. Testing confirms that the 840 Ventilator with Leak Compensation Ventilator complies with the applicable portions of the July 1995 "Draft Reviewer Guidance for Ventilators" published by the Division of Cardiovascular, Respiratory, and Neurological Devices. The device passed all applicable tests.
- 3. All software is tested in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 2005. The devices passed all tests.
#### Conclusions:
We conclude that the 840 Ventilator with Leak Compensation meets the stated performance specifications and criteria referenced above and that the device and its accessories will operate safely in its intended environment and will be effective in fulfilling the intended use.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars above it, all enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Kim L. Bloom Senior Regulatory Affairs Specialist Covidien 2101 Faraday Avenue Carlsbad, California 92008
APR - 2 2009
Re: K083693
Trade/Device Name: PB 840 Ventilator System with Leak Compensation Option Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: March 17, 2009 Received: March 19, 2009
Dear Ms. Bloom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III ( (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Ms. Bloom
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Susan Runno
Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
#### 510(k) Number: K083693
Device Name: PB 840 Ventilator System with Leak Compensation Option
#### Indications for Use:
The 840 Ventilator System with Leak Compensation Option is intended to provide continuous ventilation to patients requiring respiratory support. The device is intended for patients with an Ideal Body Weight (IBW) as low as 0.5 kg (with NeoMode option) to adult, and for use in a wide variety of clinical conditions.
The 840 Ventilator System with Leak Compensation Options is intended for a wide range of patients ranging from neonate to adult (V- 5-2500 mL with NeoMode) or from infant to adult (V-25-2500 mL).
The 840 Ventilator System with Leak Compensation is intended for use in hospital and hospital-type facilities. It may be used during hospital and hospital-type facility transport provided that electrical power and compressed gas are supplied.
Prescription Use X (Part 21 CFR 801 Subpart D) Subpart C)
AND/OR
Over-The-Counter Use (21 CFR 801
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Susan Powers
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infectio
840 Ventilator system with Leak Compensation Option Covidien, formerly Nellcor Puritan Bennett Inc.
Concurrence of CDRH, Office of Device Evaluation (ODE)
CONFIDENTIAL
Page 772
510
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.