VERIGENE RESPIRATORY VIRUS NUCLEIC ACID TEST +(RVNAT+)
K103209 · Nanosphere, Inc. · OCC · Jan 10, 2011 · Microbiology
Device Facts
Record ID
K103209
Device Name
VERIGENE RESPIRATORY VIRUS NUCLEIC ACID TEST +(RVNAT+)
Applicant
Nanosphere, Inc.
Product Code
OCC · Microbiology
Decision Date
Jan 10, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3980
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K103209 · Jan 10, 2011
VERIGENE RESPIRATORY VIRUS NUCLEIC ACID TEST +(RVNAT+)
Nanosphere, Inc.
Residual nasopharyngeal swab specimens from symptomatic patients
Residual clinical specimens collected during routine respiratory seasons were used to establish the performance characteristics (sensitivity/specificity) of the Verigene RV+ test compared to culture/DFA and sequencing.
RV+ Methods Comparison Study; Prospective collection of residual clinical specimens; Follow-up/Duration: 2008-2009 and 2009-2010 respiratory seasons; Study Period: 2008-2010
Symptomatic patients presenting with respiratory tract infection; Sample Size: 1022; Number of Sites: 3
Culture-based methods confirmed with FDA-cleared DFA test; bidirectional sequencing for subtyping
Sensitivity and specificity for detection of Influenza A, Influenza B, and RSV (and subtypes)
Indications for Use
The Verigene® Respiratory Virus Plus Nucleic Acid Test (RV+) on the Verigene® System is a qualitative nucleic acid multiplex test intended to simultaneously detect and identify multiple respiratory virus nucleic acids in nasopharyngeal (NP) swab specimens from individuals with signs and symptoms of respiratory tract infection. The following virus types and subtypes are identified using the RV+: Influenza A, Influenza A subtype H1, Influenza A subtype H3, 2009 H1N1, Influenza B, Respiratory Syncytial Virus (RSV) subtype A, and RSV subtype B. The test is not intended to detect Influenza C virus. Detecting and identifying specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection, if used in conjunction with other clinical and laboratory findings. Negative results for Influenza A, Influenza B, or RSV do not preclude influenza virus or RSV infection and should not be used as the sole basis for diagnosis, treatment, or patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection. Performance characteristics for Influenza A Virus were established when Influenza A/H3, A/H1, and 2009 H1N1 were the predominant Influenza A viruses circulating. These characteristics may vary when other Influenza A viruses are emerging. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions used specifically for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Device Story
The Verigene System is a bench-top molecular diagnostics workstation comprising a Processor SP and a Reader. It processes nasopharyngeal swab specimens to detect respiratory viral nucleic acids. The workflow involves automated magnetic bead-based viral RNA extraction, multiplex RT-PCR amplification, and gold nanoparticle probe-based hybridization. The Processor SP performs fluid transfer and mixing via a robotic pipettor. The Reader manages test ordering, controls the Processor SP, performs automated imaging of the test cartridge, and generates results. The system is used in clinical laboratory settings by trained personnel. The output provides qualitative identification of specific viral targets, aiding clinicians in diagnosing respiratory infections. The device benefits patients by providing rapid, multiplexed viral identification to inform clinical management.
Clinical Evidence
Clinical methods comparison study of 1022 prospective nasopharyngeal specimens across three sites. RV+ results compared to culture/DFA and bidirectional sequencing. Influenza A sensitivity 98.7%, specificity 93.2%. Influenza B sensitivity 100%, specificity 99.7%. RSV sensitivity 97.2%, specificity 99.5%. Subtype-specific sensitivities were 100% for H3, H1, 2009 H1N1, RSV A, and RSV B. Reproducibility/precision study showed agreement ranging from 97.2% to 100% across sites.
Technological Characteristics
Bench-top molecular workstation (Processor SP and Reader). Uses silica-coated magnetic beads for extraction, multiplex RT-PCR for amplification, and gold nanoparticle probe-based hybridization for detection. Single-use disposable test cartridges with microfluidic channels. Connectivity: Reader controls up to 32 Processor units. Software: Embedded application on Micro-C/OS real-time operating system.
Indications for Use
Indicated for qualitative detection and identification of Influenza A (H1, H3, 2009 H1N1), Influenza B, and RSV (A, B) nucleic acids in nasopharyngeal swab specimens from symptomatic patients. Not for Influenza C detection. Results intended as an aid in diagnosis alongside clinical/laboratory findings; not for sole diagnostic or treatment decisions.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
Predicate Devices
Verigene® Respiratory Virus Nucleic Acid Test (RVNATSP) (K092566)
{1}------------------------------------------------
K103209
Image /page/0/Picture/1 description: The image contains the word "Nanosphere" in a bold, sans-serif font. To the left of the word is a circular logo that is partially filled with diagonal lines. The logo is black and white, and the text is black.
JAN 1 0 2011
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1900 and CFR 807.92.
"
| 510(k)<br>numbers: | K103209: Verigene® Respiratory Virus Plus Nucleic Acid Test on the Verigene® System (RV+) |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary<br>preparation<br>date: | December 28, 2010 |
| Submitted<br>by: | Nanosphere, Inc.<br>4088 Commercial Avenue<br>Northbrook, IL 60062<br>Phone: 847-400-9000 Fax: 847-400-9199 |
| Contact: | Gregory W. Shipp, M.D.<br>Chief Medical Officer<br>VP, Medical and Regulatory Affairs |
| Proprietary<br>names: | For instrument:<br>Verigene® System<br>For the assay:<br>Verigene® Respiratory Virus Plus Nucleic Acid Test on the Verigene® System (RV+) |
| Common<br>names: | For the instrument:<br>Bench-top molecular diagnostics workstation<br>For the assay:<br>Respiratory viral panel multiplex nucleic acid assay<br>Respiratory panel assay with sub-typing:<br>- Flu A (H3, H1, 2009H1N1)<br>- Flu B<br>- RSV (RSV A, RSV B) |
| Regulatory<br>information: | Regulation section:<br>866.3980 Respiratory viral panel multiplex nucleic acid assay<br>Class II<br>Panel:<br>Microbiology (83)<br>Product Code:<br>OCC (Respiratory viral panel multiplex nucleic acid assay) |
{2}------------------------------------------------
| Comparison<br>of RVNATSP<br>and RV+: | The instrumentation used with the RVNATSP (K092566) and the RV+ is identical. The Verigene® System<br>utilized for both assays allows sample preparation, target amplification, and hybridization test and analysis<br>using a single system. The RVNATSP and the RV+ reagents and substrate have been modified to allow<br>subtyping of Influenza A (H3, H1, 2009H1N1) and RSV (RSV A, RSV B). The safety and effectiveness of the<br>Verigene® Respiratory Virus Plus Nucleic Acid Test on the Verigene System (RV+) are demonstrated with the<br>analytical and methods comparison studies described below. | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The RV+ is designed to identify virus-specific nucleic acids for Influenza A virus, Influenza A virus subtype H1,<br>Influenza A virus subtype H3, Influenza A virus subtype 2009 H1N1, Influenza B virus, and respiratory syncytial<br>virus (RSV A. RSV B). The RVNATSP involves: | |
| | 1) Sample Preparation - magnetic bead-based viral RNA extraction from nasopharyngeal swab specimens<br>obtained from symptomatic patients; | |
| | 2) Target Amplification - multiplex RT-PCR-based amplification of the eluted viral RNA targets to generate<br>virus-specific amplicons; | |
| | 3) Verigene Hybridization Test and Analysis - detection and identification of virus-specific amplicons by using<br>gold nanoparticle probe-based technology. | |
| The entire RV+ test is performed on the Verigene® System, which is a bench-top molecular diagnostics<br>workstation that consists of two instruments, the Verigene Processor SP and the Verigene Reader. The<br>Verigene Processor SP performs the assay steps on each sample by using a robotic pipettor to transfer and<br>mix reagents within and between separate testing modules designed for nucleic acid extraction, target<br>amplification, and the Verigene Hybridization Test. The Verigene Hybridization Test module is the same as in<br>the original Verigene System with added modules for nucleic acid extraction and RT-PCR target amplification.<br>Key functions of the Verigene Processor SP include: | | |
| | 1) Reading of the barcode identification label on inserted Test Consumables to maintain positive identification<br>of patient samples throughout processing. | |
| | 2) Facilitation of nucleic acid extraction, multiplex RT-PCR target amplification, and the Verigene<br>Hybridization Test. | |
| Device<br>description: | 3) Real-time communication of test processing status to the Reader. | |
| | The Verigene Reader is the same instrument as in the FDA-cleared RVNATSP. It is a free-standing instrument<br>with a touch screen control panel and a wand-based barcode scanner. It utilizes a graphical user interface to<br>guide the user through the process of ordering tests and reporting results. There are no serviceable parts and<br>no user calibration is required. Interaction with the touch screen is minimized through barcode use. This<br>instrument also serves as the reader of the Test Cartridges using advanced optics. The key functions of the<br>Verigene Reader include: | |
| | 1) Entry and tracking of specimen identification numbers via manual keyboard input or via barcode-reader<br>wand. | |
| | 2) Test selection for each specimen. | |
| | 3) Automated transfer of specimen processing instructions on Test Cartridge-specific basis to linked<br>Processor SP unit(s). A single Reader unit can control up to 32 Processor units. | |
| | 4) Automated imaging and analysis of Test Cartridges. | |
| | 5) Results display. | |
| | 6) Results report generation. | |
| | RV+ consumables within each single-use disposable test kit include: (i) Tip Holder Assembly; (ii) Extraction<br>Tray; (iii) Amplification Tray; and (iv) RV+ Test Cartridge. The kit components are inserted into the<br>corresponding module of the Verigene Processor SP prior to each test, and the sample is added to the<br>Extraction Tray. Patient information is entered into the Reader to initiate the test procedure. | |
| | 1) Tip Holder Assembly - The robotic pipettor picks up pipettes from the Tip Holder Assembly. The pipettes<br>are used for mixing and transferring reagents within the test procedure. | |
| Intended<br>use: | 2) Extraction Tray – Nucleic acids are extracted from the sample by using magnetic bead-based methods<br>within the Extraction Tray. Each Tray contains reagents for a single extraction procedure. A robotic pipette<br>transfers reagents to designated wells within the Extraction Tray to affect the steps of lysis, capture of<br>nucleic acids onto the magnetic beads, washing, and eluting the isolated nucleic acids from the magnetic<br>beads. | |
| | 3) | Amplification Tray – The isolated nucleic acids are amplified by using multiplex RT-PCR within the<br>Amplification Tray. Each Tray contains reagents for a single multiplex RT-PCR procedure. A robotic<br>pipette transfers the reagents to a specific well within the Amplification Tray. A set thermal profile is then<br>initiated to perform all of the amplification related steps including UDG-based decontamination, reverse<br>transcription, and multiplex PCR in a single tube. Upon completion, an aliquot of the amplified sample is<br>mixed with hybridization buffer containing the virus specific mediator probes. The sample is then<br>transferred to the Test Cartridge. |
| | 4) | RV+ Test Cartridge for Verigene Hybridization Test – The virus-specific and subtype-specific amplicons are<br>detected and identified within a Test Cartridge by using specific nucleic acid probes in conjunction with gold<br>nanoparticle probe-based detection technology. Each Test Cartridge is a self-contained, laboratory<br>consumable that consists of two parts. The upper housing of each cartridge is called the "reagent pack"<br>and contains reservoirs filled with the detection reagents. When in place with the 'substrate holder', the<br>reagent pack creates an air-tight hybridization chamber surrounding the region of the substrate containing<br>a target-specific capture array. As each step of the test is completed, old reagents are moved out of the<br>hybridization chamber and new reagents are added from the reagent pack via microfluidic channels and<br>pumps. Once the test is complete, the Test Cartridge is removed from the Verigene Processor SP unit and<br>the reagent pack is snapped off and discarded. The remaining slide is now ready for imaging and analysis<br>in the Verigene Reader. |
| | 5) | End-point detection on the Verigene Reader: The test slide is inserted into the Verigene Reader wherein it<br>is illuminated along its side. The gold-silver aggregates at the test sites scatter the light, which is in turn<br>captured by a photosensor. The relative intensity arising from each arrayed test site is tabulated. Net<br>signals, defined as the absolute signal intensities with background signals subtracted, are compared with<br>thresholds determined by negative controls within the slide in order to arrive at a decision regarding the<br>presence or absence of target. These results are linked to the test and patient information entered at the<br>beginning of each test session to provide a comprehensive results file. |
| | The Verigene® Respiratory Virus Plus Nucleic Acid Test on the Verigene® System (RV+) is a qualitative nucleic acid<br>multiplex test intended to simultaneously detect and identify multiple respiratory virus nucleic acids in nasopharyngeal (NP)<br>swab specimens from individuals with signs and symptoms of respiratory tract infection. The following virus types and<br>subtypes are identified using the RV+: Influenza A, Influenza A subtype H1, Influenza A subtype H3, 2009 H1N1, Influenza<br>B, Respiratory Syncytial Virus (RSV) subtype A, and RSV subtype B. The test is not intended to detect Influenza C virus.<br>Detecting and identifying specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection<br>aids in the diagnosis of respiratory viral infection, if used in conjunction with other clinical and laboratory findings. | |
| | Negative results for Influenza A, Influenza B, or RSV do not preclude influenza virus or RSV infection and should not be<br>used as the sole basis for diagnosis, treatment, or patient management decisions. Conversely, positive results do not rule-<br>out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The<br>use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of<br>respiratory viral infection. | |
| | Performance characteristics for Influenza A Virus were established when Influenza A/H3, A/H1, and 2009 H1N1 were the<br>predominant Influenza A viruses circulating. These characteristics may vary when other Influenza A viruses are emerging. | |
| | If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria<br>recommended by public health authorities, specimens should be collected with appropriate infection control precautions<br>used specifically for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture<br>should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. | |
{3}------------------------------------------------
{4}------------------------------------------------
### Comparison between the Cleared and the New Systems
| Feature | Verigene RV+ | Verigene RVNAT SP | Prodesse ProFAST+ | Specimen | Nasopharyngeal swabs in sample matrix | Nasopharyngeal swabs in sample matrix | Nasopharyngeal swabs in sample matrix |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.