VERIGENE RESPIRATORY VIRUS NUCLEIC ACID TEST ON THE VERIGENE SP SYSTEM (RVNATSP)
K092566 · Nanosphere, Inc. · OCC · Oct 9, 2009 · Microbiology
Device Facts
Record ID
K092566
Device Name
VERIGENE RESPIRATORY VIRUS NUCLEIC ACID TEST ON THE VERIGENE SP SYSTEM (RVNATSP)
Applicant
Nanosphere, Inc.
Product Code
OCC · Microbiology
Decision Date
Oct 9, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3980
Device Class
Class 2
Indications for Use
The Verigene® Respiratory Virus Nucleic Acid Test on the Verigene SP System (RVNATsp) is a qualitative multiplex in vitro diagnostic test for the detection and identification of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids purified from nasopharyngeal swab specimens obtained from patients symptomatic for viral upper respiratory infection. The test is intended to be used on the Verigene® SP System as an aid in the differential diagnosis of Influenza A, Influenza B, and RSV infections. The test is not intended to detect Influenza C virus. Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative test results be confirmed by culture. Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Device Story
Verigene SP System provides sample-to-result testing for Influenza A, Influenza B, and RSV. Input: nasopharyngeal swab specimens. Process: automated on-board sample extraction, multiplex RT-PCR amplification, and hybridization of amplicons to gold-labeled probes on a microarray-based chip. Detection: signal enhancement with elemental silver via Verigene Processor. Output: qualitative detection/identification of viral targets. Used in clinical settings; operated by laboratory personnel. System automates fluid transfer and mixing previously performed manually. Results aid clinicians in differential diagnosis of respiratory infections; negative results require confirmation by culture.
Clinical Evidence
Bench testing only. Analytical sensitivity (LoD) compared to predicate using serial dilutions of Influenza A, B, and RSV strains; results were identical. Precision/reproducibility studies conducted at three sites over multiple days showed high agreement (98.7-100%) with expected results. Method comparison study (n=558 tests) demonstrated 97.9% PPA and 100% NPA compared to the predicate system. Discordant results were limited to low-positive Influenza A samples, which resolved upon repeat testing.
Technological Characteristics
System includes Verigene Reader and Processor SP. Uses magnetic bead-based isolation, multiplex RT-PCR, and gold nanoparticle probe hybridization on a microarray chip. Automated fluid transfer via internal pipettor. Connectivity: up to 32 SP units per Reader. Software version 2.0 (Processor) and 1.5.6 (Reader).
Indications for Use
Indicated for symptomatic patients with suspected viral upper respiratory infection to detect and identify Influenza A, Influenza B, and RSV nucleic acids in nasopharyngeal swab specimens.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
Predicate Devices
Verigene® Respiratory Virus Nucleic Acid Test on the Verigene System (K083088)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number: K092566
B. Purpose for Submission: New System comprising of a reader unit and a Processor (SP) containing a hybridization module with modules for sample extraction and target amplification added.
C. Measurand: Influenza virus and Respiratory Syncytial Virus (RSV) nucleic acids target sequences. Influenza A, Influenza B, and RSV are detected.
D. Type of Test: Multiplex nucleic acid assay for qualitative detection and identification of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) target sequences in nasopharyngeal swab specimens including nucleic acid isolation, multiplex RT-PCR amplification, capture of the amplicons hybridized to gold-labeled probes on a microarray-based chip, and detection of the signal enhanced with elemental silver using the Verigene® System
E. Applicant: Nanosphere, Inc.
F. Proprietary and Established Names:
For instrument:
Verigene® SP System
For the assay:
Verigene® Respiratory Virus Nucleic Acid Test on the Verigene® SP System (RVNAT SP)
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| OCC and NSU | Class II | 21CFR 866.3980 Respiratory viral panel multiplex nucleic acid assay; 21 CFR 862.2570 Instrumentation for clinical multiplex test systems | Microbiology (83) |
H. Intended Use: Same as K083088 described below:
The Verigene® Respiratory Virus Nucleic Acid Test is a qualitative multiplex in vitro diagnostic test for the detection and identification of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids purified from nasopharyngeal swab specimens obtained from patients symptomatic for viral upper respiratory infection. The test is intended to be used on the Verigene® System as an aid in the differential diagnosis of Influenza A, Influenza B, and RSV infections. The test is not intended to detect influenza C virus.
Negative results do not preclude influenza virus or RSV infection and should not be used
{2}
as the sole basis for treatment or other management decisions. It is recommended that negative test results be confirmed by culture.
Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.
If infections with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
2. Indication(s) for use: Same as Intended Use
3) Special conditions for use statement(s): For Prescription use only
4) Special instrument requirements: Verigene® System consisting of the Verigene Processor (software version 2.0) and the Verigene Reader (software version 1.5.6).
I. Device Description:
The Verigene SP consists of two core instruments, the Verigene Reader and the Verigene SP that offers 'sample-to-result' testing capability. The Verigene SP System utilizes single-use disposable test units. Up to 32 SP units may be connected to a single Verigene Reader. Each SP unit can be independently accessed through the Reader to run different tests. The Verigene Reader in the Verigene SP is the same as in the cleared system (K083088).
The SP unit is a reconfiguration of the Processor within the cleared Verigene System to allow on-board Sample Extraction and Target Amplification – steps that were previously performed externally. Added in the SP's package are an Extraction Module to extract nucleic acids from clinical samples and an Amplification Module to amplify the extracted nucleic acids. The amplified target is then detected in the Verigene Test as in the cleared Verigene System. All fluid transfers and mixing operations in the SP are performed by an automated pipettor.
Assay technology is same as described below in K083088: Magnetic bead-based isolation technology for sample preparation, multiplex RT-PCR for target amplification, gold nanoparticle probe technology for the Verigene test. The VRNAT also includes control materials as described in K083088.
The Verigene® Respiratory Virus Nucleic Acid Test (VRNAT) is a set of reagents and supplies designed to be utilized in RT-PCR and hybridization assay using the Verigene® instrument SP system for detection and identification of human influenza A, influenza B, and RSV viral RNA. The VRNAT primers target matrix (M) gene of influenza A, non
2
{3}
structural (NS) and M genes of influenza B, and L gene and F gene of RSV. The amplicons are hybridized to gold nanoparticle probes through a mediator oligonucleotide and target-specific capture oligonucleotides on a microarray-based chip in a disposable test cartridge. The signal enhanced with elemental silver at the test site is detected using the Verigene Processor.
J. Substantial Equivalence Information:
a) Predicate device name (s): Verigene® Respiratory Virus Nucleic Acid Test on the Verigene System (VRNAT)
b) Predicate Numbers (s): K083088
Comparison with predicate:
| Verigene® (Cleared System) K083088 | Verigene® SP (New System) K092566 |
| --- | --- |
| DEVICE | |
| Processor | Processor SP |
| 1) Sample Extraction is performed externally on the NucliSENS EasyMAG (bioMerieux) | 1) Sample Extraction is performed within the added Extraction Module using Roche reagents. NOTE: Roche licenses extraction technology from bioMerieux |
| 2) Target Amplification by RT-PCR is performed on an external thermocycler | 2) Target Amplification by RT-PCR is performed within the added Amplification Module using same reagents and procedure |
| 3) Verigene Test (hybridization) is performed in the hybridization module housed in the Processor of the Verigene® System | 3) No change in reagent or procedure |
| 4) Pipetting between the three steps is performed by the user. | 4) Pipetting module added to automate fluid transfer steps |
| Reader – provides the user interface, controls the Processor, performs image analysis, and provides results. | Reader – No change |
| Verigene® System is built under QSR | Verigene® SP is built under QSR |
| SOFTWARE | |
| | Software – no change in the architecture. Software programming to control the Extraction and Amplification Modules has been added. |
| ASSAY PARAMETERS | |
| Sample Extraction – parameters as on the NucliSENS EasyMAG (bioMerieux) | Sample Extraction – parameters are similar to the NucliSENS EasyMAG (bioMerieux) |
| Target Amplification – parameters for multiplex RT-PCR were developed to work on thermocyclers. | RT-PCR – no change |
{4}
4
| Verigene Test - for the detection of amplified DNA targets on the Verigene System | No change |
| --- | --- |
## K. Standard/Guidance Document Referenced (if applicable):
1. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005. http://www.fda.gov/cdrh/ode/guidance/337.pdf
2. Draft Guidance for Industry and FDA Staff: Assay Migration Studies for In Vitro Diagnostic Devices - http://www.fda.gov/OHRMS/DOCKETS/98fr/FDA-2008-D-0642-gdl.pdf
3. Draft Guidance for Industry and FDA Staff: Establishing the Performance Characteristics of In Vitro Diagnostic Devices for the Detection or Detection and Differentiation of Influenza Viruses - http://www.fda.gov/cdrh/oivd/guidance/1638.pdf
L. Test Principle: Same as K083088
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
### Limit of Detection (Analytical Sensitivity)
The analytical sensitivity of the RVNAT$_{SP}$ was compared to the cleared VRNAT (K083088) by determining the Limit of Detection (LoD) of Influenza A, Influenza B, RSV A, and RSV B viruses. Strains with established titers were used for each virus. Each virus stock was serially diluted into a sample matrix (Universal Transport Media, Copan), and each concentration was tested in quadruplicate using the RVNAT$_{SP}$. The LoD was confirmed by performing an additional 20 replicates for each strain in order to demonstrate that the virus was detected $\geq 95\%$ of the time. The LoD for each virus was identical to the LoD observed with the same strains on the cleared VRNAT (K083088)
Limit of Detection
| Limits of Detection | Concentration |
| --- | --- |
| Influenza A (A/Wisconsin/67/05) | 2 TCID_{50}/mL |
| Influenza B (B/Florida/04/2006) | 50 TCID_{50}/mL |
| RSV A (Strain Long) | 10 TCID_{50}/mL |
| RSV B (B-1 Wild Type (B WV/14617/85)) | 2 TCID_{50}/mL |
{5}
Carry-over contamination studies: High positive samples of Influenza A, Influenza B, RSV A, and RSV B were alternated with high negative samples for all four viruses. Based on the collective data, there was no evidence of cross-contamination from any of the test steps including sample preparation, multiplex RT-PCR step, and the Verigene Hybridization Test.
A. Precision/Reproducibility Studies Comparison between the RVNAT $_{SP}$ and the cleared VRNAT (K083088)
The Precision/Reproducibility Studies were performed at each of three sites. At Site 1, the reproducibility study was part of a larger precision study.
Precision/Reproducibility Studies for the RVNAT $_{SP}$ were conducted exactly as for the previously cleared VRNAT (K083088) to allow equivalency comparisons. As before, eight unique samples were created by diluting known concentrations of viral particles with Viral Transport Medium. Since the Analytical Sensitivity of the RVNAT $_{SP}$ was identical to the cleared VRNAT, the same strains and levels were used in the studies. Each strain was represented at 3 distinct concentrations: high negative, low positive, and moderate positive.
Sample panel for the Precision/Reproducibility Studies.
| Unique Samples | Viral Strains and Levels |
| --- | --- |
| 1 | Influenza A - High Negative; Influenza B - High Negative |
| 2 | RSV A - High Negative; RSV B - High Negative |
| 3 | Influenza A - Low Positive |
| 4 | Influenza B - Low Positive |
| 5 | RSV A - Low Positive |
| 6 | RSV B - Low Positive |
| 7 | Influenza A - Moderate Positive; RSV A - Moderate Positive |
| 8 | Influenza B - Moderate Positive; RSV B - Moderate Positive |
The Precision Study (Site 1) tested the sample set over 12 non-consecutive days. On each test day, two operators performed the RVNAT $_{SP}$ in duplicate for each sample (i.e., 4 sample sets per day total). In the reproducibility study performed by sites 2 and 3, the sample set was tested in triplicate daily by 2 operators on each of five non-consecutive days.
Performance characteristics of the RVNAT $_{SP}$ in the Precision/Reproducibility Studies were equivalent to those for the cleared VRNAT as shown in the following table.
{6}
Comparison of Precision/Reproducibility data from the new RVNAT $_{SP}$ to the cleared VRNAT
| | Agreement of 'Observed Results' to 'Expected Results' | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | New RVNATSP | | | | | | Cleared VRNAT | | | | | |
| Panel Member | | Site 1NSPH | Site 2MPLN | Site 3MCW | All Sites | % Agreement | 95% Score CI | Site 1 | Site 2 | Site 3 | All Sites | % Agreement | 95% Score CI |
| Influenza A | High Negative | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% | 46/48 | 14/15 | 15/15 | 75/78 | 96% | 89.3 - 98.7% |
| | Low Positive | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% | 48/48 | 14/15 | 15/15 | 77/78 | 98.7% | 93.1 - 99.8% |
| | Moderate Positive | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% | 48/48 | 15/15 | 15/15 | 78/78 | 100.0% | 95.3 - 100% |
| Influenza B | High Negative | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% |
| | Low Positive | 47/48 | 15/15 | 15/15 | 77/78 | 98.7% | 93.1 - 99.8% | 47/48 | 15/15 | 15/15 | 77/78 | 98.7% | 93.1 - 99.8% |
| | Moderate Positive | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% |
| RSV A | High Negative | 48/48 | 15/15 | 14/15 | 77/78 | 99% | 93.1 - 99.8% | 48/48 | 13/15 | 15/15 | 76/78 | 97% | 91.1 - 99.3% |
| | Low Positive | 47/48 | 15/15 | 15/15 | 77/78 | 98.7% | 93.1 - 99.8% | 47/48 | 15/15 | 15/15 | 77/78 | 98.7% | 93.1 - 99.8% |
| | Moderate Positive | 48/48 | 15/15 | 15/15 | 78/78 | 100.0% | 95.3 - 100% | 48/48 | 15/15 | 15/15 | 78/78 | 100.0% | 95.3 - 100% |
| RSV B | High Negative | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% | 48/48 | 14/15 | 14/15 | 74/78 | 95% | 87.5 - 98.0% |
| | Low Positive | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% |
| | Moderate Positive | 48/48 | 15/15 | 15/15 | 78/78 | 100% | 95.3 - 100% | 48/48 | 15/15 | 15/15 | 78/78 | 100.0% | 95.3 - 100% |
# B. Method Comparison Studies between RVNAT $_{SP}$ and the cleared VRNAT (K083088)
A sample set representing Influenza A, Influenza B, and RSV was prepared by diluting culture positive nasopharyngeal swab samples with negative samples. Dilutions were aimed to yield viral load levels close to the low positive levels for each virus type. A total of 62 unique samples were diluted, aliquoted, and frozen. The sample set was tested at the internal site (Site 1) using the cleared VRNAT, and at all three sites using the RVNAT $_{SP}$ . This sample testing yielded a total of $62 \times 4 = 248$ unique tests.
Each sample set yielded a total of 186 decisions from 62 unique samples as each test provides a decision of 'Detected' or 'Not Detected' for each of the 3 viruses, Influenza A, Influenza B, and RSV. Of the 62 samples, 3 samples had dual infections where 2 viruses were present.
{7}
Study Sample Set
| Sample Set | Positives | Negatives | Totals |
| --- | --- | --- | --- |
| Influenza A | 15 | 47 | 62 |
| Influenza B | 16 | 46 | 62 |
| RSV A/B | 34 | 28 | 62 |
| | | | |
| Total | 65* | 121 | 186 |
*3 samples had 2 viruses raising the total number of positives from 62 to 65
Decisions on the RVNAT $_{SP}$ for each site were compared to the cleared VRNAT. The data is shown in the tables below:
$\mathrm{RVNAT}_{SP}$ Method Comparison Data collected at Site 1
| All Viruses | VRNAT(Old System) | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | |
| RVNATSP(New System) | Positive | 65 | 0 | 65 | PPA 100.0% (95%CI=94.4% - 100.0%) |
| | Negative | 0 | 121 | 121 | NPA 100.0% (95%CI=96.9% - 100.0%) |
| | Total | 65 | 121 | 186 | |
$\mathrm{RVNAT}_{SP}$ Method Comparison Data collected at Site 2
| All Viruses | VRNAT (Old System) | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | |
| RVNATSP(New System) | Positive | 64 | 0 | 64 | PPA 98.5 % (95%CI=91.8% - 99.7%) |
| | Negative | 1a | 121 | 122 | NPA 100.0% (95%CI=96.9% - 100.0%) |
| | Total | 65 | 121 | 186 | |
$^{\mathrm{a}}$ Low positive discordant for Influenza A. Repeat tests were positive and gave the expected result.
$\mathrm{RVNAT}_{SP}$ Method Comparison Data collected at Site 3
| All Viruses | VRNAT (Old System) | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | |
| RVNATSP(New System) | Positive | 62 | 0 | 62 | PPA 95.4 % (95%CI=87.3% - 98.4%) |
| | Negative | 3a | 121 | 124 | NPA 100.0% (95%CI=96.9% - 100.0%) |
| | Total | 65 | 121 | 186 | |
aAll discordant samples were Influenza A low positive. Repeat tests were positive and gave the expected result.
{8}
RVNAT $_{SP}$ Method Comparison Data - Combined data from all 3 sites
| All Viruses/All Sites | VRNAT (Old System) | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | |
| RVNATSP(New System) | Positive | 191 | 0 | 191 | PPA 97.9% (95%CI=94.8% - 99.0%) |
| | Negative | 4a | 363 | 367 | NPA 100.0% (95%CI=99.0% - 100.0%) |
| | Total | 195 | 363 | 558 | |
${}^{a}$ All discordant samples were Influenza A low positive. Repeat tests were positive and gave the expected result.
N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.