The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemiarthroplasty is determined by the surgeon to be the preferred method of treatment. The cemented primary humeral stem, long/revision stem, fracture stems and all Equinoxe glenoids are intended for cemented fixation. The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the discretion the surgeon. The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon. Humeral Heads are intended for use in cemented and press-fit applications.
Device Story
Equinoxe Shoulder System comprises humeral stems, replicator plates, humeral heads, and glenospheres for hemi- and total-shoulder arthroplasty. Replicator plates connect to humeral stems via spherical bowl/torque-defining screw and to humeral heads via morse taper. System used in OR by orthopaedic surgeons to restore joint function in patients with degenerative disease or fractures. Replicator plates provide offset and angular adjustability for component positioning. Humeral heads and glenospheres provide articulating surfaces. Device benefits patients by restoring mobility and reducing pain associated with glenohumeral joint pathology. System supports cemented or press-fit fixation depending on component and surgeon preference.
Clinical Evidence
Bench testing only. Fatigue testing conducted to verify safety and effectiveness of modified replicator plates and material changes for humeral heads and glenospheres. No clinical data provided.
Technological Characteristics
Replicator plates: Wrought titanium alloy (Ti-6Al-4V E.L.I, ASTM F136). Humeral heads/Glenospheres: Cast Cobalt-Chrome alloy (Co-28Cr-6Mo, ASTM F75). Semi-constrained design. Fixation: Cemented or press-fit. Sterilization: Standard industry processes.
Indications for Use
Indicated for skeletally mature individuals with glenohumeral joint degenerative diseases or fractures (rheumatoid arthritis, osteoarthritis, osteonecrosis, post-traumatic degeneration, congenital abnormalities, primary/secondary necrosis, humeral head fractures with tuberosity displacement, 3/4-part proximal humerus fractures, mid-humerus spiral fractures, failed previous reconstructions, or previous fusion). Reverse system indicated for degenerative disease with grossly deficient, irreparable rotator cuff or failed joint replacement with superior humeral head migration.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2016
Exactech, Incorporated Mr. Graham L. Cuthbert Regulatory Affairs Specialist II 2320 N.W. 66th Court Gainesville, Florida 32653
Re: K102951
N 102731
Trade/Device Name: Equinoxe Fixed-Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWT, KWS Dated: October 1, 2010 Received: October 5, 2010
Dear Mr. Cuthbert:
This letter corrects our substantially equivalent letter of November 4, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device wo nave rollowed your wo determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the enactment date of the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{2}------------------------------------------------
# Page 2 - Mr. Graham L. Cuthbert
(21 CFR Part 807); labeling (21 CFR-Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Exactech® Equinoxe Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - Indications for Use
# 510(k) Number (if known): K102951
Device Name: Equinoxe Fixed-Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres.
### INDICATIONS FOR USE:
Sign-Oft)
Sign-Oft)
Prative Devices
The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemiarthroplasty is determined by the surgeon to be the preferred method of treatment.
- The cemented primary humeral stem, long/revision stem, fracture stems and all 0 Equinoxe glenoids are intended for cemented fixation.
- The press-fit humeral stems are intended for press-fit applications but may be o used with bone cement at the discretion the surgeon.
- The reverse humeral components are intended to be used in cemented applications 0 or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon.
- 0 Humeral Heads are intended for use in cemented and press-fit applications.
Clinical indications for the PRIMARY (P), LONG/REVISION (L/R) and FRACTURE (F) humeral components are as follows:
| | | | Indications |
|-------------------------------------------------------------------------------------------------|---|---|---------------------------------------------------------------------------------------------------------------------------|
| <span style="writing-mode: vertical-rl; text-orientation: upright;">Stock Number K102951</span> | √ | √ | rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic<br>degenerative problems |
| | √ | √ | congenital abnormalities in the skeletally mature |
| | √ | √ | primary and secondary necrosis of the humeral head. |
| | √ | | humeral head fracture with displacement of the tuberosities |
| | √ | √ | pathologies where arthrodesis or resectional arthroplasty of the<br>humeral head are not acceptable |
| | √ | √ | revisions of humeral prostheses when other treatments or devices have<br>failed (where adequate fixation can be achieved) |
| | - | | displaced three-part and four-part upper humeral fractures |
| | √ | | spiral and other fractures of the mid-humerus (in combination with<br>glenohumeral degenerative diseases) |
| | √ | | revision of failed previous reconstructions when distal anchorage is<br>required |
| | √ | √ | to restore mobility from previous procedures (e.g. previous fusion) |
The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed
{4}------------------------------------------------
## Exactech® Equinoxe® Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - Indications for Use
glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemi-arthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sonitu for mxn
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
and/or
510(k) Number K102951
{5}------------------------------------------------
K102951
1/5
### Exactech® Equinoxe® Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - Substantial Equivalence
| I. Company: | Exactech, Inc.<br>2320 N.W. 66 <sup>th</sup> Court<br>Gainesville, FL 32653 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------|
| | Phone: (352) 377-1140<br>Fax: (352) 378-2617 |
| Contact Person: | Graham L. Cuthbert, Regulatory Affairs Specialist II |
| Date: | October 1 <sup>st</sup> , 2010 |
| II. Proprietary Name: | Exactech Equinoxe Fixed Angle Replicator Plate<br>Exactech Equinoxe Humeral Heads<br>Exactech Equinoxe Glenosphere |
| Common Name: | Total Shoulder Prosthesis<br>Reverse Total Shoulder Prosthesis |
NOV - 4 2010### Classification Name:
- Shoulder joint metal/polymer non-constrained cemented prosthesis (21 CFR . 888.3650, Class II, Product Code KWT)
- Prosthesis, Shoulder, Semi-constrained, metal/polymer cemented (21 CFR . 888.3660.
Class II, Product Code KWS)
Legally Marketed Devices to Which Substantial Equivalence Is Claimed: III.
| 510(k) Number | Trade of Proprietary Model Name | Manufacturer |
|---------------|----------------------------------------|---------------|
| K042021 | Equinoxe Total Shoulder System | Exactech, Inc |
| K063569 | Equinoxe Reverse Total Shoulder System | Exactech, Inc |
#### IV. Device Description:
The Equinoxe Shoulder System is comprised of primary, fracture and reverse product lines for use in hemi- and total-shoulder joint replacement procedures. The Equinoxe Primary and Fracture system utilize both cemented and press-fit, semi-constrained glenohumeral prostheses for use in hemi-shoulder and totalshoulder joint replacement procedures, and a cemented semi-constrained glenohumeral fracture prosthesis for use in fractures of the proximal humerus.
The Equinoxe Reverse Shoulder System includes a reverse semi-constrained prosthesis for use in total-shoulder joint replacement procedures in cases with an irreparable or nonfunctional rotator cuff. The Equinoxe Reverse Shoulder System is designed to function with the Equinoxe primary press-fit, primary cemented, and long/revision humeral stems.
{6}------------------------------------------------
K102951 2/5
### Exactech® Equinoxe® Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - 510(k) Summary of Safety and Effectiveness
The proposed fixed angle replicator plates are manufactured from wrought titanium alloy (Ti-6Al-4V E.L.I) per ASTM F136 and connect to the humeral stem via a spherical bowl and are locked in place using the previously cleared primary torque-defining screw. The replicator plate attaches to the humeral head using a morse taper. The mating features on the proposed replicator plates (specifically the morse taper geometry and the spherical bowl geometry) are identical to the predicate part, only the offset and angular adjustability have been altered.
The proposed humeral heads are manufactured from Cast Cobalt-Chrome Alloy (Co-28Cr-6Mo) per ASTM F75 and provide an articulating surface for hemi and total joint replacement. They feature female taper geometry for mating with proposed and predicate replicator plate designs. No design changes have been made to this device. It is a material change only.
The proposed glenospheres are manufactured from Cast Cobalt-Chrome Alloy (Co-28Cr-6Mo) per ASTM F75and connect to the plate via an apical hole screw and slip fit with oval shaped plate. No design changes have been made to this device. It is a material change only.
#### Intended Use of the Device V.
The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemiarthroplasty is determined by the surgeon to be the preferred method of treatment.
- The cemented primary humeral stem, long/revision stem, fracture stems . and all Equinoxe glenoids are intended for cemented fixation.
- The press-fit humeral stems are intended for press-fit applications but may . be used with bone cement at the discretion the surgeon.
- The reverse humeral components are intended to be used in cemented ● applications or in revision cases when the humeral component is wellfixed/stable, as deemed by the orthopaedic surgeon.
- Humeral Heads are intended for use in cemented and press-fit . applications.
Clinical indications for the PRIMARY (P), LONG/REVISION (L/R) and FRACTURE (F) humeral components are as follows:
| P | L/R | F | Indications |
|---|-----|---|---------------------------------------------------------------------------------------------------------------------------|
| √ | √ | | rheumatoid arthritis, osteoarthritis, osteonecrosis or post-<br>traumatic degenerative problems |
| √ | √ | | congenital abnormalities in the skeletally mature |
| √ | √ | | primary and secondary necrosis of the humeral head. |
| √ | | √ | humeral head fracture with displacement of the tuberosities |
| √ | √ | | pathologies where arthrodesis or resectional arthro plasty of the<br>humeral head are not acceptable |
| √ | √ | | revisions of humeral prostheses when other treatments or devices have<br>failed (where adequate fixation can be achieved) |
| √ | | √ | displaced three-part and four-part upper humeral fractures |
{7}------------------------------------------------
K102951 3/5
# Exactech® Equinoxe® Total Shoulder - Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - 510(k) Summary of Safety and Effectiveness
| P | L/R | F | Indications |
|---|-----|---|-----------------------------------------------------------------------------------------------------------|
| | √ | | spiral and other fractures of the mid-humerus (in combination with<br>glenohumeral degenerative diseases) |
| | √ | | revision of failed previous reconstructions when distal anchorage is<br>required |
| √ | √ | | to restore mobility from previous procedures (e.g. previous fusion) |
The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemiarthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.
#### Rationale for Substantial Equivalence VI.
Equinoxe Fixed Angle Replicator Plate:
- Intended Use / Indications for Use. The Equinoxe Fixed Angle Replicator ● Plates and predicate devices are intended for use in total shoulder joint replacement and have identical indications for use.
- The Equinoxe Fixed Angle Replicator Plates and predicate � Materials. devices are composed of equivalent biocompatible materials conforming to recognized industry standards for permanent implants.
- Design Features. The Equinoxe Fixed Angle Replicator Plates and predicate ● devices have the same design features.
- The Equinoxe Fixed Angle Replicator Plates and predicate ● Dimensions. devices have identical geometry for mating with the previously cleared humeral stems and humeral heads. The only difference is a new offset option and a reduction in the through hole diameter to prevent angulation of the plate. This has no effect on the mechanical strength of the design.
- Packaging and Sterilization. The Equinoxe Fixed Angle Replicator Plates . and predicate devices are packaged and sterilized using the same materials and processes.
- Device Shelf Life. The Equinoxe Fixed Angle Replicator Plates and predicate . devices have the same shelf life.
{8}------------------------------------------------
K102951 4/5
Exactech® Equinoxe® Total Shoulder – Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) - 510(k) Summary of Safety and Effectiveness
Cast Humeral Heads:
- Intended Use / Indications for Use. The Equinoxe Cast Humeral Heads and . predicate devices are intended for use in total shoulder joint replacement and have identical indications for use.
- Materials. The Equinoxe Cast Humeral Heads and predicate devices are . composed of Co-28Cr-6Mo Alloys which are biocompatible materials conforming to recognized industry standards for permanent implants.
- Design Features. The Equinoxe Cast Humeral Heads and predicate devices ● have the same design features.
- Dimensions. The Equinoxe Cast Humeral Heads and predicate devices have . identical geometry.
- Packaging and Sterilization. The Equinoxe Cast Humeral Heads and ● predicate devices are packaged and sterilized using the same materials and processes.
- Device Shelf Life. The Equinoxe Cast Humeral Heads and predicate devices ● have the same shelf life.
Cast Glenospheres:
- Intended Use / Indications for Use. The Equinoxe Cast Glenospheres and . predicate devices are intended for use in reverse total shoulder joint replacement and have identical indications for use.
- The Equinoxe Cast Glenospheres and predicate devices are Materials. . composed of Co-28Cr-6Mo Alloys which are biocompatible materials conforming to recognized industry standards for permanent implants.
- Design Features. The Equinoxe Cast Glenospheres and predicate devices . have the same design features.
- Dimensions. The Equinoxe Cast Glenospheres and predicate devices have . identical geometry.
- Packaging and Sterilization. The Equinoxe Cast Glenospheres and predicate ● devices are packaged and sterilized using the same materials and processes.
- Device Shelf Life. The Equinoxe Cast Glenospheres and predicate devices ● have the same shelf life.
VII. Summary of Non-Clinical Performance Data
> Fatigue testing was conducted to verify the safety and effectiveness of the proposed Exactech Equinoxe Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres. These tests were done to mitigate the risks that were affected by the proposed changes. Risks that were not affected by the new design were not re-verified.
### Substantial Equivalence Conclusion
Results from these tests and analyses provided within this 510(k) demonstrate that the Exactech Equinoxe Fixed Angle Replicator Plates, Cast Humeral Heads, and
{9}------------------------------------------------
K102951 5/5
Exactech® Equinoxe® Total Shoulder – Fixed Angle Replicator Plates, Cast Humeral Heads, and Cast Glenospheres Special 510(k) -- 510(k) Summary of Safety and Effectiveness
·
Cast Glenospheres are substantially equivalent to the identified predicate devices and will be safe for clinical use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.