K102650 · Accuray Incorporated · IYE · Nov 17, 2010 · Radiology
Device Facts
Record ID
K102650
Device Name
CYBERKNIFE ROBOTIC RADIOSURGERY SYSTEM
Applicant
Accuray Incorporated
Product Code
IYE · Radiology
Decision Date
Nov 17, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CyberKnife and CyberKnife VSI Systems are intended to provide treatment planning and image-quided stereotactic radiosurgery and precision radiotherapy for tumors, lesions and conditions anywhere in the body when radiation treatment is indicated.
Device Story
CyberKnife and CyberKnife VSI are computer-controlled robotic radiosurgery systems for stereotactic radiosurgery and precision radiotherapy. System inputs include diagnostic X-ray images of the treatment area. The system utilizes a 6 MV linear accelerator mounted on a robotic manipulator to deliver radiation. Dynamic positioning and pointing are achieved via target locating subsystems, including skull tracking, fiducial tracking, Xsight Spine Tracking, Xsight Lung Tracking, and Synchrony Respiratory Tracking. Used in clinical settings by physicians and trained staff to treat tumors and lesions throughout the body. The system provides image-guided precision, allowing for real-time tracking of target motion during treatment. Output is high-energy radiation delivered to the target, enabling minimally invasive treatment of various oncological and neurological conditions. Benefits include precise dose delivery to targets while sparing surrounding healthy tissue.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing and performance verification of the tracking software. Tracking performance was validated using archived patient lung treatment images, demonstrating equivalence to the predicate system.
Technological Characteristics
Robotic radiosurgery system featuring a 6 MV linear accelerator, robotic manipulator, and X-ray-based target locating subsystem. Supports multiple tracking modalities: skull, fiducial, Xsight Spine, Xsight Lung, and Synchrony Respiratory. Dose rate up to 1000 MU/minute. System is computer-controlled for treatment planning and delivery.
Indications for Use
Indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy for tumors, lesions, and conditions anywhere in the body. Applicable to patients with astrocytoma, glioma, skull base tumors, metastases (brain/bony), nasopharyngeal carcinoma, meningioma, acoustic neuroma, schwannoma, pituitary adenoma, hemangioblastoma, craniopharyngioma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, and tumors of the neck, spine, pancreas, liver, lungs, ovary, prostate, and bladder. Patient candidacy determined by physician team.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.
### Name. Address, Phone and Fax number of the Applicant
Accuray Incorporated 1310 Chesapeake Terrace Sunnyvale, California 94089 Ph: (408) 716-4600 Fax: (408) 789-4260
#### Contact Person
Anne Schlagenhaft
Date Prepared September 30, 2010
#### Device Name
Trade Names: CyberKnife VSI™ Robotic Radiosurgery System CyberKnife® Robotic Radiosurgery System Common Name: Radiosurgery/radiotherapy treatment planning and delivery system Classification Name: Medical charged particle radiotherapy device
## Device Description
The CyberKnife Robotic Radiosurgery System and CyberKnife VSI Robotic Radiosurgery System are computer controlled medical systems for planning and performing minimally invasive stereotactic radiosurgery and precision radiotherapy using a treatment radiation generator, linear accelerator, manipulator (robot), and a target locating subsystem to accurately deliver radiation to the treatment target. The target locating system provides X-rays of the treatment area taken by a diagnostic X-ray system that lets the user know the position of the target. The CyberKnife and CyberKnife VSI Systems use skull tracking, fiducial tracking, Xsight® Spine Tracking, Xsight® Lung Tracking, and Synchrony® Respiratory Tracking for dynamic positioning and pointing of the linear accelerator. The system uses a 6 MV linear accelerator to provide a dose rate up to 1000 MU/minute.
#### Intended Use
The CyberKnife and CyberKnife VSI Systems are intended to provide treatment planning and image-quided stereotactic radiosurgery and precision radiotherapy for tumors, lesions and conditions anywhere in the body when radiation treatment is indicated.
NOV 1 7 2010
{1}------------------------------------------------
The CyberKnife Systems may be used to treat astrocytoma, glioma, skull base tumors, metastases (brain and bony), nasopharyngeal carcinoma, meningioma, acoustic neuroma, schwannoma, pituitary adenoma, hemangioblastoma, craniopharngioma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, and tumors of the neck, spine, pancreas, liver, lungs, ovary, prostate, and bladder. Patients should be examined by a team of physicians to determine if they are candidates for CyberKnife treatment.
## Substantial Equivalence
The CyberKnife and CyberKnife VSI Systems are substantially equivalent to the predicate CyberKnife System. The source energy, beam properties, design, materials and other physical properties are the same or equivalent to the predicate CyberKnife Testing included in the premarket notification demonstrated that the System. performance characteristics of the device are equivalent to the predicate CyberKnife System.
Testing was done to verify that the new tracking software, performs as designed, as well as regression testing to verify integrity of existing features. Archived images from CyberKnife patient lung treatments were used to test tracking performance which proved to be equivalent to the predicate CyberKnife tracking systems.
The intended use, principles of operation, technological characteristics and labeling are the same or equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircling the eagle. The text is arranged in a circular fashion around the eagle, with "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the left side and "USA" on the right.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Anne Schlagenhaft Senior Regulatory Affairs Associate Accuray Incorporated 1301 Chesapeake Terrace SUNNYVALE CA 94089
NOV 1 7 2010
Re: K102650
Trade/Device Name: CyberKnife® Robotic Radiosurgery System and CyberKnife VSI™ Robotic Radiosurgery System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 13, 2010 Received: September 23, 2010
Dear Ms. Schlagenhaft:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Signature
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K102650 Device Name: CyberKnife® Robotic Radiosurgery System and CyberKnife VSI™ Robotic Radiosurgery System Indications For Use:
The CyberKnife® Robotic Radiosurgery System and CyberKnife VSI™ Robotic Radiosurgery System are indicated for treatment planning and image guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
*(Division Sign-Off)*
Division of Radiological Devices
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K 5102650
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.