The HB1C assay on the Dimension® clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long-term glucose control in individuals with diabetes mellitus. The HB1C CAL is an in vitro diagnostic product for the calibration of the Hemoglobin A1c (HB1C) method on the Dimension® clinical chemistry system.
Device Story
Dimension® HB1C kit is an in vitro diagnostic assay for use on Dimension® clinical chemistry systems. It measures HbA1c and total hemoglobin in human EDTA-anticoagulated whole blood. The device uses a turbidometric inhibition immunoassay (TINIA) for HbA1c and a modified alkaline hematin reaction for total hemoglobin. The system samples whole blood directly and performs on-board dilution. The kit includes five-level lyophilized calibrators. Results are reported in %HbA1c and mmol/mol. The device is intended for use by laboratory professionals in clinical settings to assist in monitoring long-term glucose control in diabetic patients.
Clinical Evidence
Method comparison study conducted using 126 human EDTA-preserved whole blood samples. Comparison between Dimension® HB1C kit and BIO-RAD Variant® II HPLC. Deming regression analysis for %HbA1c yielded a slope of 0.90 (95% CI: 0.85-0.94) and intercept of 0.44 (95% CI: 0.12-0.75). For mmol/mol, slope was 0.90 (95% CI: 0.85-0.95) and intercept was 2.17 (95% CI: -0.35-4.70).
Technological Characteristics
Uses turbidometric inhibition immunoassay (TINIA) for HbA1c and modified alkaline hematin reaction for total hemoglobin. Reagents are liquid, ready-to-use. Calibrators are five-level lyophilized human whole blood hemolysate. Designed for use on Dimension® clinical chemistry system platforms. Traceable to NGSP and IFCC reference methods.
Indications for Use
Indicated for quantitative determination of HbA1c in human EDTA-anticoagulated whole blood to monitor long-term glucose control in patients with diabetes mellitus.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
{0}------------------------------------------------
Siemens Healthcare Diagnostics Inc. Dimension® HB1C method and calibrator 510(k) Notification
## MAY 13 2011
### 510(k) Summary for
### Dimension® HB1C Kit
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: Ki02510
#### 1. Manufacturer's Name, Address, Telephone, and Contact Person, Date of Preparation:
| Manufacturer: | Siemens Healthcare Diagnostics Inc<br>Newark, Delaware 19714-6101 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Information: | Siemens Healthcare Diagnostics Inc.<br>500 GBC Drive<br>P.O. Box 6101<br>Newark, Delaware 19714-6101<br>Attn: A. Kathleen Ennis<br>Tel: 302-631-9352<br>Fax: 302-631-6299 |
| Preparation date: | August 24, 2010 |
| Device Name: | Dimension® HB1C Kit containing HB1C Flex® reagent cartridges and HB1C Calibrator |
Classification: Class II
2
Product Code: LCP/KRZ
Panel: Hematology
#### 3. ldentification of the Legally Marketed Device:
BIO-RAD Variant® II Hemoglobin A16 - K070452
#### 4. Device Descriptions:
The Dimension® HB1C kit contains Flex® reagent cartridges and calibrator. Each cartridge contains reagents used to measure total hemoglobin and hemoglobin A1 . The reagents are liquid and ready to use on the instrument. The calibrator in the kit is a five level lyophilized product. Each level is hydrated with 2.0ml, of reagent grade water. The lot matched reagents and calibrator product are for use on all models of the Dimension® clinical chemistry system.
#### ર્ . Device Intended Uses:
### CONFIDENTIAL
{1}------------------------------------------------
The HB1C assay on the Dimension® clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long-term glucose control in individuals with diabetes mellitus.
#### 6. Summary of the devices technological characteristics
The Dimension ® HB1C kit has the same characteristics as the BIO-RAD Variant® II Hemoglobin Are. A comparison of features is provided.
| Feature | Predicate Device: | New Device: |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BIO-RAD Variant® II HbA₁c | Dimension® HB1C kit |
| | K070452 | |
| Similarities | | |
| Intended Use | Both kits are for <i>in vitro</i> diagnostic use for the quantitative determination of hemoglobin A1c in human whole blood. | Both kits are for <i>in vitro</i> diagnostic use for the quantitative determination of hemoglobin A1c in human whole blood. |
| Sample Type | Both devices are for use with human anticoagulated whole blood treated with EDTA. | Both devices are for use with human anticoagulated whole blood treated with EDTA. |
| Calibrator | Both devices are calibrated with whole blood hemolysate provided in lyophilized form. | Both devices are calibrated with whole blood hemolysate provided in lyophilized form. |
| Certification | Both methods certified by the NGSP as traceable to the Diabetes Control and Complications Trial (DCCT). | Both methods certified by the NGSP as traceable to the Diabetes Control and Complications Trial (DCCT). |
| Traceability | Both methods are traceable to the IFCC reference method. | Both methods are traceable to the IFCC reference method. |
| Sample Preparation | Both methods sample directly and dilute on board. | Both methods sample directly and dilute on board. |
| Packaging | Both kits contain the reagents and calibrator. | Both kits contain the reagents and calibrator. |
| Differences | | |
| Instrument | BIO-RAD Variant® II HPLC | All models of the Dimension® clinical chemistry system |
| Reporting Units | Reports in %HbA1c. | Reports in both %HbA1c and mmol/mol. |
| Analytical Measuring<br>Range | 3.1 – 18.5 % HbA1c | 3.6 – 16.0 % [17 – 151 mmol/mol] |
| Calibrator levels | Provides two calibrator levels plus a diluent for preparation of additional levels. | Includes five calibrator levels. |
| Technology | Uses ion-exchange high-performance liquid chromatography (HPLC). | Uses turbidometric inhibition immunoassay (TINIA) for the HbA1c measurement and both devices use a modification of the alkaline hematin reaction for the total hemoglobin portion of |
#### 7. Method Comparison
A split sample method comparison was conducted using the new device, Dimension ® HB1C kit vs. the predicate, BIO-RAD Variant® II Hemoglobin Ay. One hundred and twenty- six (126) human whole blood samples preserved with EDTA ranging from 4.4 to 16.18 homo and the month whole blood analyzed using Deming regression analysis. The analysis is as follows:
| | | %HbA1c | | mmol/mol | | | |
|-----------|--|-------------|--------|----------|-------------|--------|------|
| | | Coefficient | 95% CI | | Coefficient | 95% CI | |
| Intercept | | 0.44 | 0.12 | 0.75 | 2.17 | -0.35 | 4.70 |
| Slope | | 0.90 | 0.85 | 0.94 | 0.90 | 0.85 | 0.95 |
| n | | 1241 | | | 1232 | | |
{2}------------------------------------------------
Siemens Healthcare Diagnostics Inc.
Dimension® HB1C method and calibrator 510(k) Notification
¹ Two samples, out of range by one or both of the methods, were not included in the analysis
² Three samples, out of range by one or both of the methods, were not included in the analysis
#### 8. Conclusion
Based on a review of the devices technological features and the method comparison study, the new Dimension® HB1C kit is substantially equivalent to the legally marketed device, BIO-RAD Varian® II Hemoglobin A1c.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around the perimeter. In the center is a stylized symbol resembling a caduceus or a bird-like figure, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Siemens Healthcare Diagnostics Inc.
c/o Ms. Anna Marie Ennis
500 GBC Drive
PO Box 6101
Newark, DE 19714
MAY 1 3 2011
Re: k102510
Trade/Device Name: Dimension® HB1C kit (Model DF 105A), Dimension® HB1C Calibrator Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated Hemoglobin Assay Regulatory Class: Class II Product Code: LCP & JJX Dated: April 5, 2011 Received: April 7, 2011
Dear Ms. Ennis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed
{4}------------------------------------------------
### Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (01) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli fire (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
C.C.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
: 172 :
Enclosure
{5}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K 102510
Dimension(r)HB1C Kit
Device Name:
### Indications for Use:
The HB1C assay on the Dimension® clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin Alc (HbAlc) in human anticoagulated whole blood. Measurements of hemoglobin Alc are effective in monitoring long term glucose control in individuals with diabetes mellitus.
XX Prescription Use Over-The-Counter Use Prescription Ose _____________________________________________________________________________________________________________________________________________________________
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol C Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102510
Page 1 of
{6}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K102510
Dimension(r)HB1C Calibrator
Device Name:
### Indications for Use:
The HB1C CAL is an in vitro diagnostic product for the calibration of the Hemoglobin A1c (HB1C) method on the Dimension® clinical chemistry system.
XX Prescription Use Over-The-Counter Use Prescription Ose
(Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Cawf C. Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102510
Page 2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.