DIMENSION VISA HBA1C, MODEL K3105A

K102045 · Siemens Heatlhcare Diagnostics, Inc. · LCP · Feb 14, 2011 · Hematology

Device Facts

Record IDK102045
Device NameDIMENSION VISA HBA1C, MODEL K3105A
ApplicantSiemens Heatlhcare Diagnostics, Inc.
Product CodeLCP · Hematology
Decision DateFeb 14, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7470
Device ClassClass 2

Intended Use

The HbAlc assay used on the Dimension Vista® system is an in vitro diagnostic assay for the quantitative determination of per cent Hemoglobin A1c in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long term glucose control in individuals with diabetes mellitus. HbA 1c calibrator is intended for use in the calibration of the Hemoglobin A 1c (HbA 1c) method on the Dimension Vista® system.

Device Story

The Dimension Vista® HbA1c kit is an in vitro diagnostic assay for use on Dimension Vista® systems. It utilizes EDTA-treated human whole blood samples. The device employs a turbidometric inhibition immunoassay (TINIA) to measure HbA1c and a modified alkaline hematin reaction for total hemoglobin. The system processes 1 uL of sample to provide quantitative results in both %HbA1c and mmol/mol units. It is operated by laboratory personnel in a clinical setting. The output assists healthcare providers in assessing long-term glycemic control in diabetic patients, facilitating clinical decision-making regarding diabetes management.

Clinical Evidence

Bench testing and method comparison studies were performed. A split-sample comparison (n=124) against the predicate device using Passing-Bablok regression showed a slope of 0.93 (95% CI: 0.89-0.97). A second split-sample comparison (n=91) against a Tosoh Ion Exchange HPLC analyzer demonstrated a mean bias of -0.04 %HbA1c (SD=0.26 %HbA1c), with results falling within the Tosoh limits of ± 0.75 %HbA1c.

Technological Characteristics

The kit includes Flex® reagent cartridges and a five-level lyophilized calibrator. It utilizes turbidometric inhibition immunoassay (TINIA) and modified alkaline hematin reaction. The system is designed for use on all Dimension Vista® models. It provides results in %HbA1c and mmol/mol, with traceability to NGSP and IFCC standards.

Indications for Use

Indicated for the quantitative determination of percent Hemoglobin A1c in human anticoagulated whole blood to monitor long-term glucose control in individuals with diabetes mellitus.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 1 4 2011 ## 510(k) Summary for Dimension Vista® HbA1c Kit This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807,92. | The assigned 510(k) number is: | K102045 | |--------------------------------|---------| |--------------------------------|---------| - 1. Manufacturer's Name, Address, Telephone, and Contact Person, Date of Preparation: | Manufacturer: | Siemens Healthcare Diagnostics Inc | |---------------|------------------------------------| | | Newark, Delaware 19714-6101 | Contact Information: Siemens Healthcare Diagnostics Inc. 500 GBC Drive P.O. Box 6101 Newark, Delaware 19714-6101 Attn: A. Kathleen Ennis Tel: 302-631-9352 Fax: 302-631-6299 Preparation date: November 9, 2010 #### 2. Device Name: Dimension Vista® HbA1c Kit Classification: Class II Product Code: LCP Panel: Hematology #### 3. Identification of the Legally Marketed Device: Dimension Vista® HA1C Kit - K062128 #### 4. Device Descriptions: The Dimension Vista® HbA1c kit contains Flex® reagent cartridges and calibrator. Each cartridge contains reagents used to measure total hemoglobin and hemoglobin A1c. the reagents are liquid and ready to use on the instrument. The calibrator in the kit is a five level lyophilized product. Each level is hydrated with 2.0mL of reagent grade water. The - lot matched reagents and calibrator product are for use on the Dimension Vista Systems. #### ട. Device Intended Uses: The HbA1c assay on the Dimension Vista System is an in vitro diagnostic assay for the quantitative determination of hemoclobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long term glucose control in individuals with diabetes mellitus. #### Summary of the devices technological characteristics રં. The Dimension Vista® HbA1c kit has the same technological characteristics as the Dimension Vista® HA1C kit. A comparison of features is provided. {1}------------------------------------------------ | Feature | Predicate Device:<br>Dimension Vista® HA1C<br>kit (K062128) | New Device:<br>Dimension Vista® HbA1c kit | |-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Similarities | | | | Intended Use | Both kits are for <i>in vitro</i> diagnostic use for the quantitative<br>measurement of hemoglobin A1c in human anticoagulated whole<br>blood on the Dimension Vista® systems. | Both kits are for <i>in vitro</i> diagnostic use for the quantitative<br>measurement of hemoglobin A1c in human anticoagulated whole<br>blood on the Dimension Vista® systems. | | Sample Type | Both devices are for use with human anticoagulated whole blood<br>treated with EDTA. | Both devices are for use with human anticoagulated whole blood<br>treated with EDTA. | | Technology | Both devices use turbidometric inhibition immunoassay (TINIA) for<br>the HbA1c measurement and both devices use a modification of the<br>alkaline hematin reaction for the total hemoglobin portion of the<br>assay. | Both devices use turbidometric inhibition immunoassay (TINIA) for<br>the HbA1c measurement and both devices use a modification of the<br>alkaline hematin reaction for the total hemoglobin portion of the<br>assay. | | Sample size | Both assays use 1 uL from the sample cup. | Both assays use 1 uL from the sample cup. | | Reagents | The methods use similar reagents. | The methods use similar reagents. | | Calibrator | Both calibrator products are lyophilized human whole blood<br>hemolysates. | Both calibrator products are lyophilized human whole blood<br>hemolysates. | | Instrument | Both methods are used with all models of the Dimension Vista®<br>Systems. | Both methods are used with all models of the Dimension Vista®<br>Systems. | | Differences | | | | Reporting Units | Reports in %HbA1c. | Reports in both %HbA1c and<br>mmol/mol. | | Analytical<br>Measuring Range | 1.0 - 30.0 g/dL Hb<br>0.2 - 2.9 g/dL HbA1c | 3.5 - 16.0 % [ 15 - 151 mmol/mol]<br>5.0 - 25.0 g/dL [3.1 - 15.5 mmol/L] Hb<br>0.3 - 2.6 g/dL [0.2 - 1.6 mmol/L] HbA1c | | Calibrator levels | Uses four calibrator levels<br>plus saline. | Uses five calibrator levels. | | Calibrator<br>Traceability | Traceable to NGSP | Traceable to NGSP and IFCC | #### 7. Method Comparison A split sample method comparison was conducted using the new device, Dimension Vista ® HbA1c kit vs. the predicate, Dimension Vista®HA1C kit. one hundred and twenty- four (124) fresh, EDTA-treated whole blood samples with HbA1c values ranging 3.8 – 16.0 %HbA1c by the Dimension Vista®HA1C kit . The data was analyzed using Passing-Bablok regression analysis. The analysis is as follows: | | %HbA1c | | mmol/mol HbA1c | | |-----------|-------------|--------------|----------------|--------------| | | Coefficient | 95% CI | Coefficient | 95% CI | | Intercept | 0.71 | 0.43 to 0.96 | 6.13 | 3.95 to 7.83 | | Slope | 0.93 | 0.89 to 0.97 | 0.93 | 0.89 to 0.97 | | N = 124 | | | | | ## CONFIDENTIAL {2}------------------------------------------------ A split sample method comparison was also performed with the new device, Dimension Vista ® HbA1c kit and the Tosoh Ion Exchange HPLC analyzer using ninety- one (91) human whole blood samples preserved with EDTA. The HbA1c values from 4.6 to 9.6% HbA1c were used in a Bland-Altman bias analysis. The analysis is as follows: > Mean Bias = - 0.04 %HbA1c SD = 0.26 %HbA1c Upper 95% Cl =0.47 %HbA1c Lower 95% CI = -0.56%HbA1c Tosoh Limits = ± 0.75 %HbA1c Range of Values 4.6 - 9.6 %HbA1c n = 91 #### 8. Conclusion Based on a review of the devices technological features and the method comparison study, the new Dimension Vista® HbA1c kit is substantially equivalent to the legally marketed device, Dimension Vista® HA1C kit. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three curved lines resembling a bird or a wing-like shape. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 Siemens Healthcare Diagnostics Inc. c/o Ms. Anna-Marie K. Ennis Regulatory Affairs Manager 500 GBC Drive, M/S 514 Newark, DE 19714-6101 Re: k102045 Trade/Device Name: Dimension VISTA HbAlc Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated Hemoglobin Assay. Regulatory Class: Class II Product Code: LCP, JIT Dated: January 14, 2011 Received: January 18, 2011 IED 1 & Chi ### Dear Ms. Ennis We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 - lf you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollyfre mo (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, C.C. Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indication for: Use 510(k) Number (if known): k102045 Device Name: Dimension Vista HbAlc Indication For Use: The HbAlc assay used on the Dimension Vista® system is an in vitro diagnostic assay for the quantitative determination of per cent Hemoglobin A1c in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long term glucose control in individuals with diabetes mellitus. HbA 1c calibrator is intended for use in the calibration of the Hemoglobin A 1c (HbA 1c) method on the Dimension Vista® system. Prescription Use x (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD) ( and Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 1.0(k) k102045
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...