A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Sterile, powder-free, ambidextrous nitrile examination gloves; available in 9.5-inch and 12-inch lengths; packaged as singles or pairs. Designed for use by healthcare professionals to provide a barrier against contamination. Chemotherapy-labeled versions tested for permeation resistance against a specific list of chemotherapy drugs per ASTM D6978-05. Gloves function as a physical barrier; no active components or software. Used in clinical settings to protect both patient and examiner during examinations and handling of hazardous drugs.
Clinical Evidence
No clinical testing was required to determine substantial equivalence. Evidence is based on bench testing, including physical properties, dimensions, freedom from pinholes, powder content, and biocompatibility (ISO 10993-5/10). Chemotherapy-labeled variants include bench-based permeation testing per ASTM D6978-05.
Technological Characteristics
Material: Nitrile. Form factor: 9.5-inch or 12-inch ambidextrous gloves. Standards: ASTM D6319-00a (physical/dimensional), ASTM D5151-06 (pinholes), ASTM D6124-06 (powder content), ASTM D6978-05/F739-07 (chemotherapy permeation), ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization). Sterilization: Sterile. Connectivity: None. Algorithm: None.
Indications for Use
Indicated for use as a disposable patient examination glove worn by clinicians to prevent cross-contamination between patient and examiner. Chemotherapy-labeled variants are indicated for use during handling of specific chemotherapy drugs as tested per ASTM D6978-05.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
K102032
## Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Gloves - 12" Pairs
### Section 5 - 510(k) Summary
| Preparation Date: | July 06, 2010 |
|----------------------------|-----------------------------------------------------------------------------------------------------------|
| Applicant: | Kimberly Clark Corporation<br>1400 Holcomb Bridge Road<br>Roswell, GA 30097 |
| Contact Person: | Lester F. Padilla<br>Tel. No.: 678-352-6766 |
| Trade/Proprietary Name(s): | Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free<br>Exam Gloves - 12" Pairs |
| Common Name(s): | Powder-Free Nitrile Patient Examination Gloves (Sterile) |
| Classification Name: | Patient Examination Glove (21 CFR Part 880.6250),<br>Polymer Patient Examination Glove (Product Code LZA) |
## Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:
3 Kimberly-Clark Corporation
- Safeskin Sterile PURPLE NITRILE* Examination Gloves (Powder-Free) K992062 】.
- Safeskin PURPLE* Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic 2. Drugs Labeling Claim - K992162
- 3. Kimberly-Clark Sterling Nitrile & Nitrile Xtra Powder-Free Exam Gloves with Chemotherapy Drug Use Claim - K081089
- Kimberly-Clark Sterling Nitrile Powder-Free Exam Glove K081027 4.
#### Device Description(s):
Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Gloves are 12-inch long, sterile, purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D 6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. The gloves are packaged as two gloves per sterile pouch.
{1}------------------------------------------------
## Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Gloves - 12" Pairs
#### Intended Use(s):
A powder-free pațient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
#### Summary of Technologies:
The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.
#### Non-Clinical Testing:
| Characteristics | Applicable<br>Recognized Standards | FDA- | Performance Results |
|------------------------------------------------|------------------------------------|------|-------------------------|
| Dimensions | ASTM D 6319-00a | | Meets ASTM Requirements |
| Physical Properties | ASTM D 6319-00a | | Meets ASTM Requirements |
| Freedom from pinholes | ASTM D 6319-00a | | Meets ASTM Requirements |
| | ASTM D 5151-06 | | |
| Powder Free (Powder Content) | ASTM D 6319-00a | | Meets ASTM Requirements |
| | ASTM D 6124-06 | | |
| ISO Skin Irritation Study and<br>Sensitization | ISO 10993, Part 10 | | Meets ASTM Requirements |
| ISO Cytotoxicity Study | ISO 10993, Part 5 | | |
#### Clinical Testing:
No Clinical testing was required to determine substantial equivalence of these devices.
#### Conclusion:
The results of the non-dinical testing demonstrate that the gloves meet the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.
{2}------------------------------------------------
## Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves - 9.5" Pairs
| Preparation Date: | July 06, 2010 | |
|----------------------------|-----------------------------------------------------------------------------------------------------------|--|
| Applicant: | Kimberly Clark Corporation<br>1400 Holcomb Bridge Road<br>Roswell, GA 30097 | |
| Contact Person: | Lester F. Padilla<br>Tel. No.: 678-352-6766 | |
| Trade/Proprietary Name(s): | Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam<br>Gloves - 9.5" Pairs | |
| Common Name(s): | Powder-Free Nitrile Patient Examination Glove (Sterile) | |
| Classification Name: | Patient Examination Glove (21 CFR Part 880.6250),<br>Polymer Patient Examination Glove (Product Code LZA) | |
## Section 5 - 510(k) Summary
### Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:
- Safeskin Sterile PURPLE NITRILE* Examination Gloves (Powder-Free) K992062 1.
- Safeskin PURPLE* Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic 2. Drugs Labeling Claim - K992162
- 3. Kimberly-Clark Sterling Nitrile & Nitrile Xtra Powder-Free Exam Gloves with Chemotherapy Drug Use Claim - K081089
- 4. Kimberly-Clark Sterling Nitrile Powder-Free Exam Glove - K081027
#### Device Description(s):
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves are 9.5-inch long, sterile, purplecolored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D 6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. The gloves are packaged as two gloves per sterile pouch.
{3}------------------------------------------------
## Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves - 9.5" Pairs
#### Intended Use(s):
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
#### Summary of Technologies:
The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.
#### Non-Clinical Testing:
| Characteristics | Applicable<br>Recognized Standards | Performance Results |
|---------------------------------------------|------------------------------------|-------------------------|
| Dimensions | ASTM D 6319-00a | Meets ASTM Requirements |
| Physical Properties | ASTM D 6319-00a | Meets ASTM Requirements |
| Freedom from pinholes | ASTM D 6319-00a<br>ASTM D 5151-06 | Meets ASTM Requirements |
| Powder Free (Powder Content) | ASTM D 6319-00a<br>ASTM D 6124-06 | Meets ASTM Requirements |
| ISO Skin Irritation Study and Sensitization | ISO 10993, Part 10 | Meets ASTM Requirements |
| ISO Cytotoxicity Study | ISO 10993, Part 5 | Meets ASTM Requirements |
#### Clinical Testing:
No Clinical testing was required to determine substantial equivalence of these devices.
#### Conclusion:
The results of the non-clinical testing demonstrate that the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.
{4}------------------------------------------------
#### Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves - 9.5 Singles
| Preparation Date: | July 06, 2010 |
|----------------------------|-----------------------------------------------------------------------------------------------------------|
| Applicant: | Kimberly Clark Corporation<br>1400 Holcomb Bridge Road<br>Roswell, GA 30097 |
| Contact Person: | Lester F. Padilla<br>Tel. No.: 678-352-6766 |
| Trade/Proprietary Name(s): | Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam<br>Gloves - 9.5" Singles |
| Common Name(s): | Powder-Free Nitrile Patient Examination Glove (Sterile) |
| Classification Name: | Patient Examination Glove (21 CFR Part 880.6250),<br>Polymer Patient Examination Glove (Product Code LZA) |
### Section 5 - 510(k) Summary
#### Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:
- Safeskin Sterile PURPLE NITRILE* Examination Gloves (Powder-Free) K992062 1.
- Safeskin PURPLE* Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic 2. Drugs Labeling Claim - K992162
- Kimberly-Clark Sterling Nitrile & Nitrile Xtra Powder-Free Exam Gloves with Chemotherapy Drug 3. Use Claim - K081089
- 4. Kimberly-Clark Sterile Sterling Nitrile Powder-Free Exam Glove - K081027
#### Device Description(s):
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves are 9.5-inch long, sterile, purplecolored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D 6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. The gloves are packaged as one glove per sterile bouch.
{5}------------------------------------------------
## Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves - 9.5 Singles
#### Intended Use(s):
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
#### Summary of Technologies:
The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.
#### Non-Clinical Testing:
| Characteristics | Applicable Recognized Standards | FDA- | Performance Results |
|------------------------------------------------|-----------------------------------|------|-------------------------|
| Dimensions | ASTM D 6319-00a | | Meets ASTM Requirements |
| Physical Properties | ASTM D 6319-00a | | Meets ASTM Requirements |
| Freedom from pinholes | ASTM D 6319-00a<br>ASTM D 5151-06 | | Meets ASTM Requirements |
| Powder Free (Powder Content) | ASTM D 6319-00a<br>ASTM D 6124-06 | | Meets ASTM Requirements |
| ISO Skin Irritation Study and<br>Sensitization | ISO 10993, Part 10 | | Meets ASTM Requirements |
| ISO Cytotoxicity Study | ISO 10993, Part 5 | | |
#### Clinical Testing:
No Clinical testing was required to determine substantial equivalence of these devices.
#### Conclusion:
The results of the non-clinical testing demonstrate that the gloves meet the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.
{6}------------------------------------------------
#### K102032 Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) - 12" Pairs
| Preparation Date: | August 18, 2010 |
|----------------------------|--------------------------------------------------------------------------------------------------------------|
| Applicant: | Kimberly Clark Corporation<br>1400 Holcomb Bridge Road<br>Roswell, GA 30097 |
| Contact Person: | Lester F. Padilla<br>Tel. No .: 678-352-6766 |
| Trade/Proprietary Name(s): | Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free<br>Exam Gloves (Chemotherapy Glove) - 12" Pairs |
| Common Name(s): | Powder-Free Nitrile Patient Examination Chemotherapy Use Glove<br>(Sterile) |
| Classification Name: | Patient Examination Glove (21 CFR Part 880.6250),<br>Patient Examination Glove, Specialty (Product Code LZC) |
#### Section 5 - 510(k) Summary
#### Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:
- 1. Safeskin Sterile PURPLE NITRILE* Examination Gloves (Powder-Free) - K992062
- Safeskin PURPLE* Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic 2. Drugs Labeling Claim - K992162
- Kimberly-Clark Sterling Nitrile Xtra Powder-Free Exam Gloves with Chemotherapy Drug 3. Use Claim - K081089
- Kimberly-Clark Sterling Nitrile Powder-Free Exam Glove K081027 4.
#### Device Description(s):
Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) are 12-inch long, sterile, purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D 6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. In addition these gloves were tested for use with the Intended Use(s) section below, per ASTM D6978-05 "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs." The gloves are packaged as two gloves per sterile pouch.
{7}------------------------------------------------
#### K102032 Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) – 12" Pairs
#### Intended Use(s):
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:
The following drugs had NO breakthrough detected up to 240 minutes:
| Bleomycin sulfate (15.0mg/ml) | Idarubicin (1.0mg/ml) |
|----------------------------------------|--------------------------------|
| Busulfan (6.0mg/ml) | Ifosfamide (50mg/ml) |
| Carboplatin (10.0mg/ml) | Irinotecan (20.0mg/ml) |
| Cisplatin (1 mg/ml) | Mechlorethamine HCl (1.0mg/ml) |
| Cyclophosphamide (20 mg/ml) | Melphalan (5.0mg/ml) |
| Cytarabine HCl (100.0mg/ml) | Methotrexate (25.0mg/ml) |
| Dacarbazine (10 mg/ml) | Mitomycin (0.5mg/ml) |
| Daunorubicin Hcl (5.0mg/ml) | Mitoxantrone (2 mg/ml) |
| Docetaxel (10.0mg/ml) | Paclitaxel (Taxol) (6 mg/ml) |
| Doxorubicin HCI (Adriamycin) (2 mg/ml) | Paraplatin (10.0mg/ml) |
| Ellence (Epirubicin) (2.0mg/ml) | Rituximab (10.0mg/ml) |
| Etoposide (20 mg/ml) | ThioTEPA (10.0mg/ml) |
| Fludarabine (25.0mg/ml) | Trisenox (0.1mg/ml) |
| Fluorouracil (adrucil) (50 mg/ml) | Vincristine Sulfate (1 mg/ml) |
| Gemcitabine (38.0mg/ml) | |
Please note that the following drug has extremely low permeation times of less than 60 minutes: Carmustine (3.3 mg/ml) 48.6 minutes
#### Summary of Technologies:
The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.
{8}------------------------------------------------
#### K102032 Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) - 12" Pairs
#### Non-Clinical Testing:
| Characteristics | Applicable<br>FDA-<br>Recognized Standards | Performance Results |
|------------------------------------------------|--------------------------------------------|-----------------------------------------------------|
| Dimensions | ASTM D 6319-00a | Meets ASTM Requirements |
| Physical Properties | ASTM D 6319-00a | Meets ASTM Requirements |
| Freedom from pinholes | ASTM D 6319-00a<br>ASTM D 5151-06 | Meets ASTM Requirements |
| Powder Free (Powder Content) | ASTM D 6319-00a<br>ASTM D 6124-06 | Meets ASTM Requirements |
| ISO Skin Irritation Study and<br>Sensitization | ISO 10993, Part 10 | Meets ASTM Requirements |
| ISO Cytotoxicity Study | ISO 10993, Part 5 | Meets ASTM Requirements |
| Resistance to Permeation | ASTM D 6978-05 and/or<br>ASTM F 739-07 | Meets ASTM Requirements<br>See Intended Use Section |
#### Clinical Testing:
No Clinical testing was required to determine substantial equivalence of these devices.
#### Conclusion:
The results of the non-clinical testing demonstrate that the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Kimberly-Clark Corporation. The logo consists of a circular symbol with a stylized cross inside, followed by the company name in a bold, sans-serif font. The text is black, and the overall design is simple and corporate.
### K102032 Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) - 9.5" Pairs
| Preparation Date: | August 18, 2010 |
|----------------------------|--------------------------------------------------------------------------------------------------------------|
| Applicant: | Kimberly Clark Corporation<br>1400 Holcomb Bridge Road<br>Roswell, GA 30097 |
| Contact Person: | Lester F. Padilla<br>Tel. No .: 678-352-6766 |
| Trade/Proprietary Name(s): | Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam<br>Gloves (Chemotherapy Glove) - 9.5" Pairs |
| Common Name(s): | Powder-Free Nitrile Patient Examination Chemotherapy Use Glove<br>(Sterile) |
| Classification Name: | Patient Examination Glove (21 CFR Part 880.6250),<br>Patient Examination Glove, Specialty (Product Code LZC) |
## Section 5 - 510(k) Summary
#### Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:
- Safeskin Sterile PURPLE NITRILE* Examination Gloves (Powder-Free) K992062 1.
- Safeskin PURPLE* Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic 2. Drugs Labeling Claim - K992162
- 3. Kimberly-Clark Sterling Nitrile & Nitrile Xtra Powder-Free Exam Gloves with Chemotherapy Drug Use Claim - K081089
- Kimberly-Clark Sterling Nitrile Powder-Free Exam Glove K081027 4.
#### Device Description(s):
. ..
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) are 9.5-inch long, sterile, purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D 6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. In addition these gloves were tested for use with the drugs listed in the Intended Use(s) section below, per ASTM D6978-05 "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs." The gloves are packaged as two gloves per sterile pouch.
{10}------------------------------------------------
#### K102032 Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) - 9.5" Pairs
#### Intended Use(s):
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:
| Bleomycin sulfate (15.0mg/ml) | Idarubicin (1.0mg/ml) |
|----------------------------------------|--------------------------------|
| Busulfan (6.0mg/ml) | Ifosfamide (50mg/ml) |
| Carboplatin (10.0mg/ml) | Irinotecan (20.0mg/ml) |
| Cisplatin (1 mg/ml) | Mechlorethamine HCI (1.0mg/ml) |
| Cyclophosphamide (20 mg/ml) | Melphalan (5.0mg/ml) |
| Cytarabine HCl (100.0mg/ml) | Methotrexate (25.0mg/ml) |
| Dacarbazine (10 mg/ml) | Mitomycin (0.5mg/ml) |
| Daunorubicin Hcl (5.0mg/ml) | Mitoxantrone (2 mg/ml) |
| Docetaxel (10.0mg/ml) | Paclitaxel (Taxol) (6 mg/ml) |
| Doxorubicin HCl (Adriamycin) (2 mg/ml) | Paraplatin (10.0mg/ml) |
| Ellence (Epirubicin) (2.0mg/ml) | Rituximab (10.0mg/ml) |
| Etoposide (20 mg/ml) | ThioTEPA (10.0mg/ml) |
| Fludarabine (25.0mg/ml) | Trisenox (0.1mg/ml) |
| Fluorouracil (adrucil) (50 mg/ml) | Vincristine Sulfate (1 mg/ml) |
| Gemcitabine (38.0mg/ml) | |
The following drugs had NO breakthrough detected up to 240 minutes:
Please note that the following drug has extremely low permeation times of less than 60 minutes: Carmustine (3.3 mg/ml) 48.6 minutes
#### Summary of Technologies:
The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.
{11}------------------------------------------------
#### K102032 Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) - 9.5" Pairs
#### Non-Clinical Testing:
| Characteristics | Applicable<br>FDA-<br>Recognized Standards | Performance Results |
|--------------------------------------------------------------------------|--------------------------------------------|-----------------------------------------------------|
| Dimensions | ASTM D 6319-00a | Meets ASTM Requirements |
| Physical Properties | ASTM D 6319-00a | Meets ASTM Requirements |
| Freedom from pinholes | ASTM D 6319-00a<br>ASTM D 5151-06 | Meets ASTM Requirements |
| Powder Free (Powder Content) | ASTM D 6319-00a<br>ASTM D 6124-06 | Meets ASTM Requirements |
| ISO Skin Irritation Study and<br>Sensitization<br>ISO Cytotoxicity Study | ISO 10993, Part 10<br>ISO 10993, Part 5 | Meets ASTM Requirements |
| Resistance to Permeation | ASTM D 6978-05 and/or<br>ASTM F 739-07 | Meets ASTM Requirements<br>See Intended Use Section |
#### Clinical Testing:
No Clinical testing was required to determine substantial equivalence of these devices.
#### Conclusion:
The results of the non-clinical testing demonstrate that the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.
{12}------------------------------------------------
#### K102032 Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) - 9.5" Singles
### Preparation Date: August 18, 2010 Kimberly Clark Corporation Applicant: 1400 Holcomb Bridge Road Roswell, GA 30097 Lester F. Padilla Contact Person: Tel. No.: 678-352-6766 Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Trade/Proprietary Name(s): Gloves (Chemotherapy Glove) - 9.5" Single Powder-Free Nitrile Patient Examination Chemotherapy Use Glove Common Name(s): (Sterile) Patient Examination Glove (21 CFR Part 880.6250), Classification Name: Patient Examination Glove, Specialty (Product Code LZC)
### Section 5 - 510(k) Summary
#### Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:
- Safeskin Sterile PURPLE NITRILE* Examination Gloves (Powder-Free) K992062 l .
- 2. Safeskin PURPLE* Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs Labeling Claim - K992162
- Kimberly-Clark Sterling Nitrile Xtra Powder-Free Exam Gloves with Chemotherapy Drug 3. Use Claim - K081089
- Kimberly-Clark Sterling Nitrile Powder-Free Exam Glove K081027 4.
#### Device Description(s):
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) are 9.5-inch long, sterile, purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D 6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. In addition these gloves were tested for use with the Intended Use(s) section below, per ASTM D6978-05 "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs." The gloves are packaged as one glove per sterile pouch.
{13}------------------------------------------------
#### K102032 Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) - 9.5" Singles
#### Intended Use(s):
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:
The following drugs had NO breakthrough detected up to 240 minutes:
| Bleomycin sulfate (15.0mg/ml) | Idarubicin (1.0mg/ml) |
|----------------------------------------|--------------------------------|
| Busulfan (6.0mg/ml) | Ifosfamide (50mg/ml) |
| Carboplatin (10.0mg/ml) | Irinotecan (20.0mg/ml) |
| Cisplatin (1 mg/ml) | Mechlorethamine HCl (1.0mg/ml) |
| Cyclophosphamide (20 mg/ml) | Melphalan (5.0mg/ml) |
| Cytarabine HCl (100.0mg/ml) | Methotrexate (25.0mg/ml) |
| Dacarbazine (10 mg/ml) | Mitomycin (0.5mg/ml) |
| Daunorubicin Hcl (5.0mg/ml) | Mitoxantrone (2 mg/ml) |
| Docetaxel (10.0mg/ml) | Paclitaxel (Taxol) (6 mg/ml) |
| Doxorubicin HCI (Adriamycin) (2 mg/ml) | Paraplatin (10.0mg/ml) |
| Ellence (Epirubicin) (2.0mg/ml) | Rituximab (10.0mg/ml) |
| Etoposide (20 mg/ml) | ThioTEPA (10.0mg/ml) |
| Fludarabine (25.0mg/ml) | Trisenox (0.1mg/ml) |
| Fluorouracil (adrucil) (50 mg/ml) | Vincristine Sulfate (1 mg/ml) |
| Gemcitabine (38.0mg/ml) | |
Please note that the following drug has extremely low permeation times of less than 60 minutes: Carmustine (3.3 mg/ml) 48.6 minutes
#### Summary of Technologies:
The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo of Kimberly-Clark Corporation. The logo consists of a circular symbol with a stylized cross inside, followed by the company name in a bold, sans-serif font. The text is all in black, and the logo appears to be a vector graphic.
#### K102032 Traditional 510(k) Notification (Bundled): Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Gloves (Chemotherapy Glove) - 9.5" Singles
#### Non-Clinical Testing:
| Characteristics | Applicable<br>FDA- Recognized Standards | Performance Results |
|------------------------------------------------|-----------------------------------------|-----------------------------------------------------|
| Dimensions | ASTM D 6319-00a | Meets ASTM Requirements |
| Physical Properties | ASTM D 6319-00a | Meets ASTM Requirements |
| Freedom from pinholes | ASTM D 6319-00a<br>ASTM D 5151-06 | Meets ASTM Requirements |
| Powder Free (Powder Content) | ASTM D 6319-00a<br>ASTM D 6124-06 | Meets ASTM Requirements |
| ISO Skin Irritation Study and<br>Sensitization | ISO 10993, Part 10 | Meets ASTM Requirements |
| ISO Cytotoxicity Study | ISO 10993, Part 5 | Meets ASTM Requirements |
| Resistance to Permeation | ASTM D 6978-05 and/or<br>ASTM F 739-07 | Meets ASTM Requirements<br>See Intended Use Section |
#### Clinical Testing:
No Clinical testing was required to determine substantial equivalence of these devices.
#### Conclusion:
The results of the non-clinical testing demonstrate that the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of three human figures, represented by flowing lines, symbolizing health and well-being.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Kimberly-Clark Corporation C/O Mr. Ned Divine Responsible Third Party Official Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
SEP 0 8 2010
Re: K102032
Trade/Device Name: Kimberly-Clark Purple NITRILE-XTRA "Sterile Powder-Free
Examination Glove with Tested For Use with Chemotherapy Drug Labeling Claim (12"Pairs)
Labeling Claim (12"Pairs).
Kimberly-Clark Purple NITRILE Sterile Powder-Free Examination Glove with Tested For Use with Chemotherapy Labeling Claim (9.5"Pairs)
Kimberly-Clark Purple NITRILE Sterile Powder Free Examination Glove with Tested For Use with Chemotherapy Drug Labeling Claim (9.5"Single)
Kimberly-Clark Purple NITRILE-XTRA Sterile Powder-Free Examination Glove (12"Pairs)
Kimberly-Clark Purple NITRILE Sterile Powder-Free Examination Glove (9.5"Pairs)
Kimberly-Clark Purple NITRILE Sterile Powder-Free Examination Glove (9.5"Single)
Regulation Number: 21 CFR 880.6250
Regulation Name: Patient Examination Glove
Regulatory Class: I
Product Code: LZA, LZC .
Dated: August 18, 2010
Received: August 20, 2010
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act.
{16}------------------------------------------------
#### Page 2- Mr. Devine
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting; (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
{17}------------------------------------------------
Page 3- Mr. Devine
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
(-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-{-
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{18}------------------------------------------------
## Indications for Use
510(k) Number (if known): K102032
Device Name(s):
Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Glove (Chemotherapy Glove) - 12" Pairs
#### Indications for Use:
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:
The following drugs had NO breakthrough detected up to 240 minutes:
| Bleomycin sulfate (15.0mg/ml) | Idarubicin (1.0mg/ml) |
|----------------------------------------|--------------------------------|
| Busulfan (6.0mg/ml) | Ifosfamide (50mg/ml) |
| Carboplatin (10.0mg/ml) | Irinotecan (20.0mg/ml) |
| Cisplatin (1 mg/ml) | Mechlorethamine HCl (1.0mg/ml) |
| Cyclophosphamide (20 mg/ml) | Melphalan (5.0mg/ml) |
| Cytarabine HCl (100.0mg/ml) | Methotrexate (25.0mg/ml) |
| Dacarbazine (10 mg/ml) | Mitomycin (0.5mg/ml) |
| Daunorubicin Hcl (5.0mg/ml) | Mitoxantrone (2 mg/ml) |
| Docetaxel (10.0mg/ml) | Paclitaxel (Taxol) (6 mg/ml) |
| Doxorubicin HCl (Adriamycin) (2 mg/ml) | Paraplatin (10.0mg/ml) |
| Ellence (Epirubicin) (2.0mg/ml) | Rituximab (10.0mg/ml) |
| Etoposide (20 mg/ml) | ThioTEPA (10.0mg/ml) |
| Fludarabine (25.0mg/ml) | Trisenox (0.1mg/ml) |
| Fluorouracil (adrucil) (50 mg/ml) | Vincristine Sulfate (1 mg/ml) |
| Gemcitabine (38.0mg/ml) | |
Please note that the following drug has extremely low permeation times of less than 60 minutes: Carmustine (3.3 mg/ml) 48.6 minutes
Page 1 of 2
Elizabeth St. Claire-Will
(Division Sian-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
. . . .
{19}------------------------------------------------
## Indications for Use (cont'd)
510(k) Number (if known): K102032
Device Name(s):
Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Glove (Chemotherapy Glove) – 12" Sterile Pairs
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _2_ of _2_
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K | O2032
{20}------------------------------------------------
### Indications for Use
510(k) Number (if known): K102032
Device Name(s):
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Glove (Chemotherapy Glove) - 9.5" Pairs
#### Indications for Use:
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:
The following drugs had NO breakthrough detected up to 240 minutes:
| Bleomycin sulfate (15.0mg/ml) | Idarubicin (1.0mg/ml) |
|----------------------------------------|--------------------------------|
| Busulfan (6.0mg/ml) | Ifosfamide (50mg/ml) |
| Carboplatin (10.0mg/ml) | Irinotecan (20.0mg/ml) |
| Cisplatin (1 mg/ml) | Mechlorethamine HCl (1.0mg/ml) |
| Cyclophosphamide (20 mg/ml) | Melphalan (5.0mg/ml) |
| Cytarabine HCl (100.0mg/ml) | Methotrexate (25.0mg/ml) |
| Dacarbazine (10 mg/ml) | Mitomycin (0.5mg/ml) |
| Daunorubicin HCl (5.0mg/ml) | Mitoxantrone (2 mg/ml) |
| Docetaxel (10.0mg/ml) | Paclitaxel (Taxol) (6 mg/ml) |
| Doxorubicin HCl (Adriamycin) (2 mg/ml) | Paraplatin (10.0mg/ml) |
| Ellence (Epirubicin) (2.0mg/ml) | Rituximab (10.0mg/ml) |
| Etoposide (20 mg/ml) | ThioTEPA (10.0mg/ml) |
| Fludarabine (25.0mg/ml) | Trisenox (0.1mg/ml) |
| Fluorouracil (adrucil) (50 mg/ml) | Vincristine Sulfate (1 mg/ml) |
| Gemcitabine (38.0mg/ml) | |
Please note that the following drug has extremely low permeation times of less than 60 minutes: Carmustine (3.3 mg/ml) 48.6 minutes
Page 1 of 2
Elita I Clanto-Wille
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K1102032
{21}------------------------------------------------
## Indications for Use (cont'd)
510(k) Number (if known): K102032
Device Name(s):
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Glove (Chemotherapy Glove) - 9.5" Sterile Pairs
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _2_ of _2_
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _K 102032
{22}------------------------------------------------
### Indications for Use
510(k) Number (if known): K102032
Device Name(s):
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Glove (Chemotherapy Glove) -- 9.5" Singles
#### Indications for Use:
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:
The following drugs had NO breakthrough detected up to 240 minutes:
| Bleomycin sulfate (15.0mg/ml) | Idarubicin (1.0mg/ml) |
|----------------------------------------|--------------------------------|
| Busulfan (6.0mg/ml) | Ifosfamide (50mg/ml) |
| Carboplatin (10.0mg/ml) | Irinotecan (20.0mg/ml) |
| Cisplatin (1 mg/ml) | Mechlorethamine HCl (1.0mg/ml) |
| Cyclophosphamide (20 mg/ml) | Melphalan (5.0mg/ml) |
| Cytarabine HCl (100.0mg/ml) | Methotrexate (25.0mg/ml) |
| Dacarbazine (10 mg/ml) | Mitomycin (0.5mg/ml) |
| Daunorubicin Hcl (5.0mg/ml) | Mitoxantrone (2 mg/ml) |
| Docetaxel (10.0mg/ml) | Paclitaxel (Taxol) (6 mg/ml) |
| Doxorubicin HCl (Adriamycin) (2 mg/ml) | Paraplatin (10.0mg/ml) |
| Ellence (Epirubicin) (2.0mg/ml) | Rituximab (10.0mg/ml) |
| Etoposide (20 mg/ml) | ThioTEPA (10.0mg/ml) |
| Fludarabine (25.0mg/ml) | Trisenox (0.1mg/ml) |
| Fluorouracil (adrucil) (50 mg/ml) | Vincristine Sulfate (1 mg/ml) |
| Gemcitabine (38.0mg/ml) | |
Please note that the following drug has extremely low permeation times of less than 60 minutes: Carmustine (3.3 mg/ml) 48.6 minutes
Page 1 of 2
Elyatt P. Laurent-Well
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Nur
{23}------------------------------------------------
Image /page/23/Picture/0 description: The image shows the logo for Kimberly-Clark Corporation. The logo consists of a circular symbol with a stylized cross inside, followed by the company name in bold, sans-serif font. The text is black, and the overall design is simple and corporate.
## Indications for Use (cont'd)
510(k) Number (if known): K102032
Device Name(s):
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Glove (Chemotherapy Glove) - 9.5" Sterile Single
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 2_
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _ K/62032
{24}------------------------------------------------
## Indications for Use
510(k) Number (if known): K102032
Device Name(s):
Kimberly-Clark PURPLE NITRILE-XTRA* Sterile Powder-Free Exam Glove (12" Pairs)
Indications for Use:
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Elijabeth P. Claveria-Wilber
(Division Sign-Off) (Division Sign-on)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Reviens Division of Antool, Dental Devices
510(k) Number: K102032
{25}------------------------------------------------
## ද්යා Kimberly-Clark Corporation
## Indications for Use
510(k) Number (if known): K | 02032
## Device Name(s):
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Glove (9.5" Pairs)
## Indications for Use:
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 (Division Sign-Off)
(Division Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K70 2032
{26}------------------------------------------------
Image /page/26/Picture/0 description: The image shows the logo for Kimberly-Clark Corporation. The logo consists of a stylized symbol to the left of the company name. The text "Kimberly-Clark Corporation" is written in a bold, sans-serif font, with "Kimberly-Clark" being larger than "Corporation". The overall design is simple and professional.
## Indications for Use
510(k) Number (if known): K | 02032
### Device Name(s):
Kimberly-Clark PURPLE NITRILE* Sterile Powder-Free Exam Glove (9.5" Singles)
### Indications for Use:
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Edith P. Clauss-Willis
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K102032
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.