WAVESENSE DIABETES MANAGER MODEL VERSION 1.3.4

K101597 · Agamatrix · NBW · Oct 18, 2010 · Clinical Chemistry

Device Facts

Record IDK101597
Device NameWAVESENSE DIABETES MANAGER MODEL VERSION 1.3.4
ApplicantAgamatrix
Product CodeNBW · Clinical Chemistry
Decision DateOct 18, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The AgaMatrix WaveSense Diabetes Manager (WDM) application (app) is intended for use in the home and professional settings to aid individuals with diabetes and their healthcare professionals: in the review, analysis and evaluation of blood glucose test results to support an effective diabetes management program. It is an optional data management software accessory for use with the WaveSenseenabled blood glucose meter (BGM) with a mini-USB port. The WaveSense Diabetes Manager allows users to download Blood glucose reading automatically from the meter to an iPhone Operating System platform.

Device Story

WaveSense Diabetes Manager (WDM) is a digital logbook/diabetes management application for iPhone OS. Inputs: blood glucose readings downloaded automatically from WaveSense-enabled BGM via mini-USB cable; manual entry of insulin and carbohydrate intake. Operation: software processes and stores data to provide a digital logbook for review and analysis. Used in home/professional settings by patients and healthcare providers. Output: visual data logs/analysis of glucose, insulin, and carbs. Clinical impact: supports effective diabetes management programs by facilitating data review and evaluation.

Clinical Evidence

Clinical evaluation conducted in-house and in a clinical setting with persons with diabetes. Studies focused on usability and demonstrated the ease of operating the application as intended.

Technological Characteristics

Software application for iPhone OS platform. Connectivity via mini-USB cable to WaveSense-enabled BGM. Functions as a digital logbook for blood glucose, insulin, and carbohydrate data.

Indications for Use

Indicated for individuals with diabetes and their healthcare professionals in home and professional settings to review, analyze, and evaluate blood glucose readings to support diabetes management.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 10 Manor Parkway Salem, NH 03079 Office: +1 (603) 328-6000 Fax: +1 (603) 893-4191 ## 510(k) SAFETY AND EFFECTIVENESS SUMMARY OCT 1 8 2010 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. - 1. Submitter's Name, Address, Telephone Number, Contact Person, and date the summary was prepared. Submitter's Name: Address: AgaMatrix, Inc. 10 Manor Parkway Salem, NH 03079 (603) 328-6000 Contact Person: Connie Hertel Director Quality & Regulatory Affairs chertel @agamatrix.com Date the summary prepared: August 31, 2010 2. Device Name Trade/Proprietary Name: Common/Usual Name: Classification Name: Class AgaMatrix WaveSense Diabetes Manager WaveSense Diabetes Manager None unclassified Trade/Proprietary Name: Common/Usual Name: Classification Name: Class II AgaMatrix WaveSense BGM meters AgaMatrix WaveSense BGM meters NBW, JQP - Predicate Device 3. Zero-Click Data Management System 510(k) number: k062434 4. Device Description The WaveSense Diabetes Manager (WDM) application (app) is a digital logbook and diabetes management tool for the iPhone operating system platform. The application can be used alone or with the WaveSense Direct Connect Cable and a WaveSense-enabled Blood Glucose Meter (BGM) with a mini-USB port. Image /page/0/Picture/20 description: The image shows the word "AgaMatrix" in a simple, sans-serif font. A curved line underlines the word, starting below the "A" and ending near the "t". To the right of the word, there is a series of four squares, decreasing in size from right to left, forming a curved line above the word. K101597 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "AgaMatrix" in a simple, sans-serif font. A curved line underlines the text, starting below the "A" and ending near the "M". To the right of the word, there are four small, solid squares and circles arranged in a curved line, decreasing in size from right to left. - 5. Intended Use: The AgaMatrix WaveSense Diabetes Manager (WDM) application (app) is intended for use in the home and professional settings to aid individuals with diabetes and their healthcare professionals: in the review, analysis and evaluation of blood glucose test results to support an effective diabetes management program. It is an optional data management software accessory for use with the WaveSenseenabled blood glucose meter (BGM) with a mini-USB port. The WaveSense Diabetes Manager allows users to download Blood glucose reading automatically from the meter to an iPhone Operating System platform. - 6. Assessment of Performance: An evaluation of the WaveSense Diabetes Manager was studied in house and in a Clinical setting by persons with diabetes. The studies demonstrated the ease of operating the WaveSense Diabetes Manager application as intended. | | Zero-Click Data Management Software | WaveSense Diabetes<br>Management Application | |--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------| | Intended Use | The AgaMatrix WaveSense Diabetes<br>Manager (WDM) application (app) is<br>intended for use in the home and<br>professional settings to aid individuals<br>with diabetes and their healthcare<br>professionals: in the review, analysis and<br>evaluation of blood glucose test results to<br>support an effective diabetes<br>management program | same | | Accessory to | WaveSense Blood Glucose Monitoring<br>Meters | same | | Log book for | Blood glucose readings | Blood glucose readings<br>and Insulin and<br>Carbohydrates intake | | Use on | PC | iPhone Operating System<br>platform | - 7. Comparison to Predicate device ## 8. Conclusions The results of clinical evaluations of the WaveSense Diabetes Manager application demonstrate that the application is equivalent in performance to the predicate device and suitable for its intended use. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle faces right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Agamatrix c/o Connie Hertel 10 Maor Parkway, Salem, NH 03079 Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002 OCT 18 2010 Re: k101597 Trade/Device Name: WaveSense Diabetes Manager-application Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system. Regulatory Class: II Product Code: NBW, JQP Dated: September 22, 2010 Received: September 24, 2010 Dear: Ms. Hertel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. CA Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for AgaMatrix. The logo consists of the word "AgaMatrix" in a simple, sans-serif font. There is a curved line underneath the word, and a series of squares to the right of the word, which appear to be getting larger as they move to the right. 10 Manor Parkway Salem, NH 03079 Office: +1 (603) 328-6000 Fax: +1 (603) 893-4191 ## Indications for Use 510(k) Number (if known): KID|597 Device Name: WaveSense Diabetes Manager application OCT 1 8 2010 Indications for Use: The WaveSense Diabetes Manager (WDM) application (app) is intended for use in the home and professional settings to aid individuals with diabetes and their healthcare professionals; in the review, analysis and evaluation of blood glucose readings to support an effective diabetes management program. The WaveSense Diabetes Manager application is a digital logbook and diabetes management tool designed to operate using the iPhone Operating System platform. The application can be used alone or with the WaveSense Direct Connect Cable and a WaveSense-enabled blood glucose meter (BGM) with a mini-USB port. Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use X . (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Ciis Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 1101597 Page 1 of 1
Innolitics
510(k) Summary
Decision Summary
Classification Order