The SB100II Fingertip Pulse Oximeter is a non-invasive device intended for the spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate of adult and pediatric patients in home and hospital environments (including clinical use in internist/surgery, anesthesia, intensive care etc.). The SB100II is not intended for continuous monitoring.
Device Story
SB100II is a portable, battery-powered fingertip pulse oximeter. It uses an integral sensor with red (660nm) and infrared (905/880nm) LEDs and a photoelectric cell to measure light absorption through the finger. The device calculates functional SpO2 and pulse rate based on the differential absorption of light by oxyhemoglobin. It features a single On/Off control and displays digital SpO2 and pulse rate values, along with a vertical bar representing pulse amplitude. Used by clinicians or patients for spot-checking in homes and hospitals; not for continuous monitoring or transport. It provides no alarms. The device benefits patients by providing non-invasive, immediate physiological status checks.
Clinical Evidence
Performance testing included bench testing and clinical testing per ISO 9919:2005. The device met all requirements for electrical safety (IEC 60601-1) and electromagnetic compatibility (IEC 60601-1-2). Clinical testing confirmed accuracy within ±3% for SpO2 (70-99%) and ±3 bpm for pulse rate, meeting FDA guidance for non-motion pulse oximeters.
Technological Characteristics
Transmittance pulse oximeter; integral spring-loaded finger sensor; red LED (660nm) and IR LED (905/880nm); max optical output 4mW; powered by 2 AAA batteries; Type BF electrical safety; IPX1 ingress protection; standalone device; no connectivity; reusable.
Indications for Use
Indicated for spot-checking SpO2 and pulse rate in adult and pediatric patients in home and hospital settings. Not for continuous monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
| | Page 1 of 6<br>02-Jun-2010 | |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Official Contact: | CY Wang - RD VP<br>Atlantean<br>1F, No. 789, Bo-ai Street<br>Chubay City, TAIWAN<br>Tel - 886-3-555-8340<br>Fax - 886-2-2736-3250 | K10156<br>SEP 20 2010 |
| Proprietary or Trade Name: | SB100II Finger Pulse Oximeter | |
| Common/Usual Name: | Pulse oximeter | |
| Classification Name: | Oximeter<br>DQA - 870.2700 | |
| Predicate Devices: | Contec Medical Systems, K082641, CMS-50DL<br>Smith's Medical, K013171, BCI3420 | |
510(k) Summary
### Device Description:
The SB10011 pulse oximeter is an instrument for photoelectrically determining the oxygenation of blood in a translucent part of the body, in this case the finger.
This SB1001 contains the electronics to interface to the integral sensor. This sensor consists of a light emitting diode and a photoelectric cell placed on either side of the finger. Red and infrared light are transmitted through oxyhemoglobin and are sensed in the photoelectric cell. The red and infrared light is absorbed in different amounts depending on the oxygenation of the blood.
The SB10011 has the ability to determine both the percent of saturated hemoglobin and the pulse rate. It performs these functions on adult and pediatric patient populations. It is designed for spot checking both the SpO2 and Pulse rate information. To perform this, the SB100Il display digital values for both the SpO2 and Pulse.
Pulse amplitude is displayed graphically by means of a vertical bar, which elevates in synch with the pulse cycle. The SB1001 is powered by two "AAA" batteries. The wavelength of red LED is 660nm and Infrared LED is 905/880nm with maximum optical output power of 4mW.
The device incorporates a battery state indicator and provides a visual indication of low battery.
## Summary Description:
- The SB10011 is not life-supporting or life-sustaining, not for implant. .
- The SB100II is not sterile .
- ◆ The SB100H is reusable.
- The SB10011 is for prescription use. .
- The SB100II does not contain drug or biological products ●
- The SB100H is a class II device, product code DQA regulation 21 CFR 870.2700. .
{1}------------------------------------------------
# 510(k) Summarv Page 2 of 6 02-Jun-2010
- The SB100II is a standalone device it is not part of a multi-parameter module and does . not connect to any other devices
- The SB100H is for use in spot-checking it is not intended for continuous monitoring. .
- The SB10011 is not intended for out of hospital transport .
- . The SB100II can be used in homes.
- The SB10011 is a transmittance device that calculates functional SpO2 .
- . The SB100II does not provide audio or visual alarms
- The SB100I1 has only one control the On / Off key. When the device is on it can . display the following:
- ୍ତ SpO2 values
- Pulse Rate o
- An indication of pulse amplitude. O
- . There are no accessories for use with the SB1001. The sensor is integral to the device and is spring loaded so that it does not need to be adhered to the skin by tape or other measures.
- The SB100H is a standalone device that does not interface to any other devices so there . are no patient cables, extender cables, sensors or external power supplies.
- . The SB100II is not provided sterile is not intended to be sterilized.
- . The SB100II is not a reprocessed device.
- . The SB10011 does not have multiple modes. It is on or off, when it is on, it measures SpO2 and pulse rate as described above.
- . There are no claims relative to motion tolerance
- . There are no claims relative to low perfusion
### Indications for Use:
The SB10011 Fingertip Pulse Oximeter is a non-invasive device intended for the spot-checking of oxygen saturation of arterial hemoglobin (SpO-) and the pulse rate of adult and pediatric patients in home and hospital environments (including clinical use in internist/surgery, anesthesia, intensive care etc.). The SB10011 is not intended for continuous monitoring.
### Patient Population:
Adult and pediatric patient populations
Environment of Use: Home and hospital
Contraindications: None
{2}------------------------------------------------
Premarket Notification 510(k)
Section 5 -- 510(k) Summary
# 510(k) Summary
.
| Attribute | Smiths Medical<br>BC13420<br>510(k) K013171 | Contee Medical Systems<br>CMS50 DL<br>510(k) K082641 | Atlantean SB10011<br>(This submission) | Discussion of<br>Differences |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The BCI 3420 Digit Pulse Oximeter is<br>a handheld, finger mounted pulse<br>oximeter that combines the monitor<br>and sensor into one assembly that<br>measures SpO2, pulse rate and pulse<br>strength. It may be used as a spot<br>check device in the home, hospital or<br>clinical environments, including<br>patient ground transport in clinical and<br>EMS (Emergency Medical Services)<br>settings. The BCI 3420 Digit Pulse<br>Oximeter will provide reliable<br>measurements on patient ranging from<br>pediatric to adults. | The Fingertip Pulse Oximeter is<br>a non-invasive device intended<br>for the spot-check of oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and the pulse rate of<br>adult and pediatric patients in<br>home and hospital environments<br>(including clinical use in<br>internist/surgery, anesthesia,<br>intensive care etc.). This device<br>is not intended for continuous<br>monitoring. | The SB10011 Fingertip Pulse Oximeter is<br>a non-invasive device intended for the<br>spot-checking of oxygen saturation of<br>arterial hemoglobin (SpO2) and the pulse<br>rate of adult and pediatric patients in<br>home and hospital environments<br>(including clinical use in internist /<br>surgery, anesthesia, intensive care etc.).<br>The SB10011 is not intended for<br>continuous monitoring. | Same as K082641 |
| Environments of use | home, hospital or clinical<br>environments, including patient<br>ground transport in clinical and EMS<br>(Emergency Medical Services)<br>settings | home and hospital environments<br>(including clinical use in<br>internist/surgery, anesthesia,<br>intensive care etc. | home and hospital environments<br>(including clinical use in<br>internist/surgery, anesthesia, intensive<br>care etc. | Same as K082641 |
| Intended application<br>site | Finger | Finger | Finger | Same |
| Low perfusion claims | No | No | No | Same |
| Motion Claims | No | No | No | Same |
| Prescriptive | Yes | Yes | Yes | Same as K082641 |
| Patient population | Pediatric / adult | Pediatric / adult | Pediatric / adult | Same |
| Reusable | Yes | Yes | Yes | Same |
| Display Type | LED | LED | LED | Same |
| Attribute | Smiths Medical<br>BC13420<br>510(k) K013171 | Contec Medical Systems<br>CMS50 DL<br>510(k) K082641 | Atlantean SB10011<br>(This submission) | Discussion of<br>Differences |
| Display Parameter | SpO2, Pulse Rate, Pulse Strength | SpO2, Pulse Rate, Pulse Strength | SpO2, Pulse Rate, Pulse Strength | Same |
| Controls | On | On | On / off | Same |
| Alarms | Unknown | Unknown | None | No alarms on SB10011 |
| SpO2<br>Range | 0-99% | 35-99% | 35-99% | Same as K082641 |
| Resolution | 1% | 1% | 1% | Same |
| Accuracy | +/-2% 70-99% | +/-2% 70-99% | +/-3% 70-99% | Similar, actual Arms 2.5<br>Meets FDA guidance for<br>non-motion ≤ 3.0 %<br>Meets ISO 9919 for non-<br>motion <4.0 % |
| Type | <70% unspecified | <70% unspecified | 35-69% unspecified | Same as K082641 |
| Pulse<br>Range | 30-254 bpm | 30-250 bpm | 30-250 bpm | Same |
| Accuracy | ±2% or 2 bpm whichever is greater | ±2% or 2 bpm whichever is<br>greater | ±3 bpm | Similar |
| Resolution | 1 bpm | 1 bpm | 1 bpm | Same |
| Power | 2 "AAA" batteries | 2 "AAA" batteries | 2 "AAA" batteries | Same |
| Battery Life | 16 hours continuous<br>1400 spot checks | 30 hours | 16 hours continuous | Same time duration as<br>K013171 |
| Attribute | Smiths Medical<br>BC13420<br>510(k) K013171 | Contec Medical Systems<br>CMS50 DL<br>510(k) K082641 | Atlantean SB10011<br>(This submission) | Discussion of Difference |
| Environmental | | | | |
| Dimensions | 57.2 x 43.3 x 38.1 mm | unknown | 63.5x34x35mm | Similar to K013171 |
| Weight | 85g | Unknown | 37g | Lighter than K013171 |
| Operating<br>Temperature | 10°C-55°C | Unknown | 5°C to 40°C | Similar to K013171 |
| Storage Temperature | -40°C-75°C | Unknown | -20°C to 70°C | Similar toKk013171 |
| Operating Humidity | 15%-95% | Unknown | 15%-95% | Similar to K013171 |
| Storage Humidity | 10-95% | unknown | 15%-95% | Similar to K013171 |
| Standards | | | | |
| IEC 60601-1 | Yes (undated) | Yes (undated) | Yes IEC 60601-1: 1988 +A1+A2 | SB10011 meets current revision<br>of the recognized consensus<br>standard |
| IEC 60601-1-2 | Yes EN 60601-1-2: 1993 | Yes (undated) | IEC 60601-1-2: 2001 +AI | SB100II meets current<br>recognized consensus standard |
| ISO 9919 | Unknown | ISO 9919: 2005 | ISO 9919: 2005 | Same as k082641 (current<br>revision of the recognized<br>consensus standard) |
| General | | | | |
| Protection against<br>ingress of water | Unknown | Unknown | IPX1 and ISO 9919 Clause 44 | Meets FDA guidance |
| Protection Class<br>(electrical safety) | Unknown | Unknown | Type BF | Protection class is suitable for |
Page 5.20
{3}------------------------------------------------
Premarket Notification 510(k) Summary
# 510(k) Summary
Page 5.21
.
{4}------------------------------------------------
Premarket Notification 510(k)
Section 5 = 510(k) Summary
# 510(k) Summary
Page 5.22
{5}------------------------------------------------
# 510(k) Summary Page 6 of 6 . 02-Jun-2010
#### Performance Testing:
The following performance tests were done and the SB100II meet all requirements.
- . ISO 9919 including clinical testing
- 1EC 60601-1-2 .
- . IEC 60601-1
# Substantial Equivalence:
The SB100II Pulse Oximeter is viewed as substantially equivalent to the predicate devices because it performs the same basic functionality by similar technical means.
#### Indications -
The SB10011 Fingertip Pulse Oximeter is a non-invasive device intended for the spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate of adult and pediatric patients in home and hospital environments (including clinical use in internist/surgery, anesthesia, intensive care etc.). The SB10011 is not intended for continuous monitoring.
# Technology -
The SB100H pulse oximeter is an instrument for photoelectrically determining the oxygenation of blood in a translucent part of the body, in this case the finger.
This SB100II contains the electronics to interface to the integral sensor. This sensor consists of a light emitting diode and a photoelectric cell placed on cither side of the finger. Red and infrared light are transmitted through oxyhemoglobin and are sensed in the photoelectric cell. The red and infrared light is absorbed in different amounts depending on the oxygenation of the blood.
# Materials - - --
The materials in contact with the patient as common to medical devices.
## Environment of Use -
Home and hospital environments (including clinical use in internist/surgerv, anesthesia, intensive care etc.).
Patient Population -Adult and pediatric
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol on the right side. To the left of the bird symbol is a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Atlantean C/O Mr. Paul E. Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
SEP 2 0 2010
Re: K101568
Trade/Device Name: SB100II Finger Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: September 15, 2010 Received: September 16, 2010
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2- Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Jame O. Eastland
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Indications for Use Statement
K101568
SEP 2 0 2010
Page 1 of 1
510(k) Number:
(To be assigned)
Device Name:
SB100II Finger Pulse Oximeter
Indications for Use:
The SB100II Fingertip Pulse Oximeter is a non-invasive device intended for the spotchecking of oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate of adult and pediatric patients in home and hospital environments (including clinical use in internist/surgery, anesthesia, intensive care etc.). The SB100II is not intended for continuous monitoring.
Prescription Use XX (Part 21 CFR 801 Subpart D)
or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schuta
Division Sign-Off) Division of Anesthesiology. General Hospital Infection Control, Dental Devices
510(k) Number: K101568
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.