K013171 · Bci, Inc. · DQA · Jan 3, 2002 · Cardiovascular
Device Facts
Record ID
K013171
Device Name
3420 DIGIT PULSE OXIMETER, MODEL 3420
Applicant
Bci, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Jan 3, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BCI® 3420 Digit™ Pulse Oximeter is a handheld, finger-mounted pulse oximeter that combines the monitor and sensor into one assembly that measures SpO2, pulse rate, and pulse strength. It may be used as a spot check device in the home, hospital or clinical environments, including patient ground transport in clinical and EMS (Emergency Medical Services) settings. The BCI® 3420 Digit™ Pulse Oximeter will provide reliable measurements on patients ranging from pediatric to adults.
Device Story
Handheld, finger-mounted pulse oximeter; integrates monitor and sensor into single assembly. Powered by two AAA batteries; features LED display. Used for spot-check measurements of SpO2, pulse rate, and pulse strength. Operated by clinicians or patients in home, hospital, clinical, or EMS transport environments. Provides real-time physiological data to assist healthcare providers in assessing patient oxygenation and pulse status.
Clinical Evidence
Clinical testing of SpO2 parameter and software validation performed. Results demonstrated device performance within specifications. No specific sensitivity/specificity metrics provided.
Technological Characteristics
Handheld, finger-mounted assembly; LED display; battery-powered (2x AAA). Complies with EN 60601-1 (1990) and EN 60601-1-2 (1993).
Indications for Use
Indicated for pediatric to adult patients requiring spot-check monitoring of SpO2, pulse rate, and pulse strength in home, hospital, clinical, or EMS transport settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
KO13/7!
Image /page/0/Picture/1 description: The image shows the logo for BCI, which is a company that specializes in battery charging and testing equipment. The logo consists of a circle with vertical lines on the left, followed by the letters BCI in bold, sans-serif font. The circle is meant to represent a battery, and the letters BCI stand for Battery Charging and Testing International.
JAN 0 3 2002
BCI, Inc. N7 W22025 Johnson Road Waukesha, WI 53186-1856 USA Tel: 262-542-3100 Fax: 262-542-3325 w.smiths-bci.com
# Summary of Safety and Effectiveness
| Submitter: | BCI, Inc. |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | N7 W22025 Johnson Road<br>Waukesha, WI 53186 |
| Telephone: | (262) 542-3100 |
| Contact: | VP Regulatory Affairs |
| Prepared: | September 21, 2001 |
| Proprietary Name: | BCI® 3420 Digit™ Pulse Oximeter |
| Common/Classification Name: | Pulse Oximeter |
| Predicate Devices: | BCI 3301® Hand-Held Pulse Oximeter (K943404)<br>BCI® 3401 Handheld Pulse Oximeter (K980714)<br>Nonin® Onyx , Model 9500 (K942248, K001085) |
### New Device Description:
The BCI® 3420 Digit™ Pulse Oximeter consists of an pulse oximeter monitor and sensor contained on a single assembly. The device includes an LED display and power on key, and is powered by two "AAA" batteries.
#### Intended Use:
The BCI® 3420 Digit™ Pulse Oximeter is a handheld, finger-mounted pulse oximeter that combines the monitor and sensor into one assembly that measures SpO2, pulse rate, and pulse strength. It may be used as a spot check device in the home, hospital or clinical environments, including patient ground transport in clinical and EMS (Emergency Medical Services) settings. The BCI® 3420 Digit™ Pulse Oximeter will provide reliable measurements on patients ranging from pediatric to adults.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for BCI, which is a company that specializes in business continuity. The logo is black and white and features a circle made of vertical lines on the left, followed by the letters BCI in bold font. There is a registered trademark symbol to the right of the I.
#### Performance Data:
The design of this device utilizes currently available technology found in legally marketed devices. The major difference between the new device and the predicate BCI 3301 Handheld Pulse Oximeter is the combination of the pulse oximeter monitor and sensor, similar to the Nonin® Onyx®. Testing was done to ensure that the BCI® 3420 Digit™ Pulse Oximeter would perform safely and accurately within the environment(s) for which it is to be marketed.
Safety and environmental testing was conducted in accordance with the Reviewer's Guidance for Respiratory Devices, 1993, EN 60601-1: 1990, and EN 60601-1-2: 1993. Electromagnetic compatability (EMC), electrical, mechanical durability, safety (operator and patient), and temperature/humidity testing have been completed. The results demonstrate that the BC1® 3420 Digit™ pulse oximeter is in compliance with the guidelines and standards referenced in the reviewer's guides and that it performs within its specifications and functional requirements.
Testing of device performance included clinical testing of the SpO2 parameter and overall software validation. The results demonstrated that the BCI® 3420 Digit™ pulse oximeter performed within its specifications.
The testing described above indicate that there is no functional difference between the operation of the BCI® 3420 Digit™ pulse oximeter and the predicate devices. Based on these results, it is our determination that the device is safe, effective, and performs as well as the legally marketed predicate device(s).
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Respectfully,
Donald Alexander
Donald Alexander VP Regulatory Affairs
BCI and Digit are a BCI trademarks. The symbol ® indicates it is registered in the U.S. Patent and Trademark Office and certain other countries. Nonin and Onyx are Nonin trademarks.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### JAN 0 3 2002
Donald J. Alexander Vice President of Regulatory Affairs BCI, Inc. N7 W22025 Johnson Road Waukesha, WI 53186-1856
Re: K013171
3420 Digit™ Pulse Oximeter Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: Class II (two) Product Code: DQA Dated: November 20, 2001 Received: November 21, 2001
Dear Mr. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Donald J. Alexander
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Dea G. Tell
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications For Use
510(k) Number (if Known): `K013171
Device Name: BCI® 3420 Digit™ Pulse Oximeter
Indications For Use:
Intended Use:
The BCI® 3420 Digit™ Pulse Oximeter is a handheld, finger-mounted pulse oximeter that combines the monitor and sensor into one assembly that measures SpO2, pulse rate, and pulse strength. It may be used as a spot check device in the home, hospital or clinical environments, including patient ground transport in clinical and EMS (Emergency Medical Services) settings. The BCI 3420 Digit™ Pulse Oximeter will provide reliable measurements on patients ranging from pediatric to adults.
( PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED ) ============================================================================================================================================================================
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The Counter Use
Division of Cardiovascular & Respiratory Devices
510(k) Number K013171
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.