LC Bead Microspheres & Bead Block™ Compressible Microspheres is intended for embolization of hypervascular tumors and arteriovenous malformations.
Device Story
LC Bead and Bead Block are preformed, soft, deformable microspheres derived from polyvinyl alcohol (PVA) macromer. Used for arterial embolization of hypervascular tumors and arteriovenous malformations (AVMs). Microspheres are ~90% water, compressible by 20-30% diameter, and supplied in sizes 70-1200µm. At point-of-care, clinician mixes beads with nonionic contrast agent (30-50% by weight) to form suspension. Delivered via microcatheters (1.8-5Fr). Bead Block is blue-dyed for visualization and packaged in polycarbonate syringes; LC Bead is available in natural or blue and packaged in glass vials. Device occludes target vessels to block blood flow. Benefits include durable embolization effect and ease of delivery. Used by physicians in clinical settings.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993), size distribution, compressibility, catheter deliverability, and stability. Pre-clinical swine renal artery embolization model compared LC Bead (n=36) to Embosphere (n=36); results showed LC Bead had more durable embolization effect, fewer catheter clogs (1 vs 6), and similar tissue reaction.
Indicated for embolization of hypervascular tumors and arteriovenous malformations in patients requiring arterial occlusion.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}------------------------------------------------
# 510K Summary (K094018)
# Submitter:
Biocompatibles UK Ltd. Chapman House Weydon Lane, Farnham, Surrey, GU9 8QL United Kingdom
APR 1 6 2010
・・
+44 1252732732
# Contact:
Dr. Alistair Taylor, Director of Quality and Regulatory Affairs
### Common name, Trade name(s) & Classification 1
Trade name(s): LC Bead Microspheres & BeadBlock Microspheres Common name(s) & Codes:
> Vascular Embolization Device, embolization, arterial (Code: KRD) Neurovascular Embolization Device, artificial embolization (Code: HCG)
# 510(k) Numbers and Product Codes of equivalent devices. ଧ
Biocompatibles UK Ltd GelSpheres Microspheres Bead Block Compressible Microspheres 510K Number: K033761 Product Code: HCG/KRD CFR Section: 882.5950
Biocompatibles UK Ltd. GelSpheres Microspheres Bead Block Compressible Microspheres 510K Number(s): K042231/K083091 Product Code: HCG/KRD CFR Section: 870.3300/882.5950
{1}------------------------------------------------
## Indications for Use and Intended Population e
# "LC Bead Microspheres & Bead Block Compressible Microspheres is intended for embolization of hypervascular tumors and arteriovenous malformations."
## Device Description 4
LC Bead/Bead Block are preformed, soft, deformable microspheres that occlude arteries for the purpose of blocking the blood flow to a target tissue, such as a hypervascular tumor or arteriovenous malformations (AVM's). LC Bead/Bead Block consists of a macromer derived from polyvinyl alcohol (PVA). The fully polymerized microsphere is approximately 90% water and is compressible to approximately 20-30% by diameter. Bead Block is dyed blue (LC Bead are available as blue and in natural color) to aid in the visualization of the microspheres in the delivery syringe. The microspheres can be delivered through typical microcatheters in the 1.8-5Fr range.
LC Bead is supplied sterile and packaged in sealed glass vials. Bead Block is supplied sterile and packaged in polycarbonate syringes. The product configurations are described in the table. LC Bead/Bead Block are supplied in several unit sizes covering the range from 100-1200um diameter. At the time of use, LC Bead/Bead Block is mixed with a nonionic contrast agent, e.g. Omnipaque™, to make a 30-50% by weight solution.
| Product | Volume of beads (mL) | Volume PBS (mL) | Total volume (mL) |
|--------------------------------------------|----------------------|-----------------|-------------------|
| LC Bead<br>Microspheres | 1 | 7 | 8 |
| | 2 | 6 | 8 |
| Bead Block<br>Compressible<br>Microspheres | 1 | 5 | 6 |
| | 2 | 4 | 6 |
LC Bead/Bead Block product configurations.
## Similarities and Differences to Predicates 5
The intended use of LC Bead/Bead Block and the predicate device are the same and unchanged. Biocompatibles UK Ltd intend to market LC Bead with an additional SKU in the size range of 70-150μm. Only minor process modifications were made to allow for the
{2}------------------------------------------------
production of this size range. Other than the additional size range, there are no differences when comparing L.C Bead/Bead Block to the predicate device.
### 6 Physical Properties and Characteristics
LC Bead & Bead Block are preformed, soft, deformable microspheres which consist of a macromer derived from polyvinyl alcohol (PVA). The fully polymerized microsphere is approximately 90% water and is compressible to approximately 20-30% by diameter. Compressed beads will recover to their original size (e.g. when compressed passing though a catheter, the beads will return to their original size after exiting the catheter). This Pre-Market notification adds the size range of 70-150μm for the blue dyed version of LC Bead. Both products are supplied in a variety of size ranges as follows:
| Product Code | Size Range (µm) | Quantity Bead Block (mL) | Quantity PBS (mL) |
|--------------|-----------------|--------------------------|-------------------|
| EB1S103 | 100-300 | 1 | 5 |
| EB1S305 | 300-500 | 1 | 5 |
| EB1S507 | 500-700 | 1 | 5 |
| EB1S709 | 700-900 | 1 | 5 |
| EB1S912 | 900-1200 | 1 | 5 |
| EB2S103 | 100-300 | 2 | 4 |
| EB2S305 | 300-500 | 2 | 4 |
| EB2S507 | 500-700 | 2 | 4 |
| EB2S709 | 700-900 | 2 | 4 |
| EB2S912 | 900-1200 | 2 | 4 |
Bead Block available size ranges
| Product Code | Size Range (μm) | Quantity LC Bead (mL) | Quantity PBS (mL) |
|--------------|-----------------|-----------------------|-------------------|
| UB1V103 | 100-300 | 1 | 7 |
| UB1V305 | 300-500 | 1 | 7 |
| UB1V507 | 500-700 | 1 | 7 |
| UB1V709 | 700-900 | 1 | 7 |
| UB1V912 | 900-1200 | 1 | 7 |
| UB2V103 | 100-300 | 2 | 6 |
| UB2V305 | 300-500 | 2 | 6 |
| UB2V507 | 500-700 | 2 | 6 |
| UB2V709 | 700-900 | 2 | 6 |
| UB2V912 | 900-1200 | 2 | 6 |
LC Bead (undyed) available size ranges
{3}------------------------------------------------
| Product Code | Size Range (μm) | Quantity LC Bead (mL) | Quantity PBS (mL) |
|--------------|-----------------|-----------------------|-------------------|
| VE110GS | 70-150 | 1 | 7 |
| VE210GS | 100-300 | 1 | 7 |
| VE410GS | 300-500 | 1 | 7 |
| VE610GS | 500-700 | 1 | 7 |
| VE810GS | 700-900 | 1 | 7 |
| VE1010GS | 900-1200 | 1 | 7 |
| VE120GS | 70-150 | 2 | 6 |
| VE220GS | 100-300 | 2 | 6 |
| VE420GS | 300-500 | 2 | 6 |
| VE620GS | 500-700 | 2 | 6 |
| VE820GS | 700-900 | 2 | 6 |
| VE1020GS | 900-1200 | 2 | 6 |
LC Bead (dyed) available size ranges
#### 6.1 Differences between LC Bead and Bead Block
Bead Block is dyed blue using an FDA approved dye (used in contact lenses) to aid in the visualization of the microspheres in the delivery syringe (LC Bead are available as blue and in natural color). Bead Block is provided in a polycarbonate sterile syringe, LC Bead is provided in a sterile glass vial. The primary difference between LC Bead and Bead Block products, aside from the packaging relates to the degree of functionalisation of the macromer and the ratios of initiators used in the reaction which results in differences in the degree of crosslinking of the polymer in the microspheres.
This pre-market notification relates only to the addition of a size fraction for LC Bead in the range of 70-150µm which is a subgroup of the currently marketed LC Bead 100-300µm product and the 70-150µm size specification falls within that of the cleared 100-300 µm LC Bead size range. Please refer to Section 8: In-Vitro testing for further product characterization information. There is no change to the product supplied under the Bead Block trade name.
### Summary of Non-clinical data 7
LC Bead and Bead Block have been tested in pre-clinical models for biocompatibility and safety in accordance with the FDA Guidance for Industry and staff; Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.
#### Tests of Biocompatibility 7.1
Tests for biocompatibility were conducted in accordance with ISO 10993 parts 1, 3, 4, 6, 10 and 11 (listed in section 9), the products conform to the relevant requirements of these standards.
{4}------------------------------------------------
| Biocompatibility Test | Pass/Fail |
|-------------------------------------------------------------------------------------|-----------|
| Genotoxicity: In Vitro Chromosomal Aberration Study in Mammalian<br>Cells | Pass |
| Mouse Bone Marrow Micronucleus Study | Pass |
| In Vitro Hemolysis Study (Modified ASTM-Direct Contact Method) | Pass |
| ISO Muscle Implantation Study in the Rabbit | Pass |
| Cytotoxicity Study using the ISO Elution Method | Pass |
| ISO Sensitization Study in the Guinea Pig | Pass |
| ISO Acute Intracutaneous Reactivity Study in the Rabbit | Pass |
| Chronic Toxicity Study in the Rat following Subcutaneous Implantation<br>(13 weeks) | Pass |
| Subchronic Intravenous Toxicity Study in the Rat (14 day, saline<br>extract) | Pass |
| Genotoxicity: Bacterial Reverse Mutation Study | Pass |
| ISO Acute Systemic Toxicity Study in the Mouse (liquid/chemical) | Pass |
| ISO Surgical Muscle Implantation in the Rabbit (26 weeks) | Pass |
#### Pre-clinical testing in a large animal model 7.2
# Summary of the Evaluation of LC Bead (formerly Gelspheres) Embolic Agent in a Swine Embolization Model
The purpose of this study was to evaluate, characterize and compare the performance of LC Bead Embolic Agent (n=36) and Embosphere® microspheres (n=36) in a swine bilateral partial renal artery embolization model in order to assess the ability of these agents to occlude the vessel.
The primary outcomes for this study were assessment of:
- (1) recanalization of the vessels, and,
- (2) local and systemic foreign body tissue reactions.
The secondary outcomes were assessment of:
- (1) ease of delivery of the embolic agent,
- (2) the occurrence of blood vessel rupture
- (3) non-target embolization/device migration.
LC Bead Embolic Agent and Embospheres microspheres performed in a substantially equivalent manner at 2, 7 and 28 days for all parameters except recanalization, where LC Bead appears to have an advantage of having a more durable embolization effect. The tissue reaction for both LC Bead and Embospheres was very mild and was essentially the same. Both embolic agents delivered easily, but Embospheres had six cases of catheter clogging out of 36 cases. There was
{5}------------------------------------------------
only one case of catheter clogging with LC Bead. There were no incidents of blood vessel rupture during the embolization procedures. There was one case of unexplained non-target embolization with Embospheres and none with LC Bead. Alternatively, there was one potential case of device migration with LC Bead and none with Embospheres.
#### Summary of In-Vitro testing 8
Both LC Bead and Bead Block have been extensively tested and subject to product and process validation and verification testing. A summary of key characteristics for which test data has been provided in this 510K, are described in this section.
#### 8.1 Size distribution
Data was provided in this pre market notification regarding the verification and validation of the new size range of LC Bead. The table and illustration below provide the results of these tests and demonstrate that all product met specification with respect to bead size.
| Product | Sizing Specification | Fibres Specification |
|------------------------------|----------------------|----------------------|
| Current LC Bead<br>100-300μm | Pass | Pass |
| | LC Bead 70-150μm | Pass |
{6}------------------------------------------------
Image /page/6/Figure/0 description: The figure is a graph titled "70-150µm Bead Size Distribution". The x-axis is labeled "Bead size (µm)" and ranges from 0 to 350. The y-axis is labeled "% Frequency" and ranges from 0 to 30. There are three lines plotted on the graph, labeled "Lot A", "Lot B", and "Lot C", which all show a similar distribution.
### Compressibility 8.2
LC Bead has equivalent compressibility to other marketed embolic agents.
### Catheter Delivery 8.3
Catheter delivery characteristics have been tested in accordance with a written protocol to assure performance with typical microcatheters. The table below provides a summary of the test results for the current marketed LC Bead product and the 70-150μm size fraction.
{7}------------------------------------------------
| Catheter ID | | Microcatheter | LC Bead/ size ranges (µm) | | | | |
|-------------|------|------------------------------|---------------------------|---------|---------|---------|---------|
| (inches) | (µm) | Name | 70-150 | 100-300 | 300-500 | 500-700 | 700-900 |
| 0.024 | 610 | 5Fr. Angio<br>Dynamics | ✓ | ✓ | ✓ | ✓ | ✓ |
| 0.024 | 610 | FasTracker®<br>325 | ✓ | ✓ | ✓ | ✓ | ✓ |
| 0.021 | 540 | FasTracker®<br>18 | ✓ | ✓ | ✓ | ✓ | ✓ |
| 0.021 | 540 | Cook 3.0 Fr | ✓ | ✓ | ✓ | ✓ | |
| 0.016 | 420 | Prowler® 14 | ✓ | ✓ | ✓ | | |
| 0.022 | 570 | 2.4Fr<br>Progreat™<br>Terumo | ✓ | ✓ | ✓ | | |
| 0.018 | 457 | Spinnaker<br>Elite1.8 | ✓ | ✓ | ✓ | | |
Catheter can be used for the effective delivery of the LC Bead product.
#### 8.4 Other tests
<
Additional bead characterization data has been provided in this pre-market notification with respect to other attributes of the device. A summary of this additional test data is provided below.
{8}------------------------------------------------
| Test | Pass/Fail |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Residual starting materials<br>The residual starting materials present in the final packaged device. | Pass |
| Residual solvents<br>The residual solvent levels present in the final packaged device. | Pass |
| Product visual inspection for presence of fibres<br>The visual assessment of a sample of the final product to determine the level of fibres present. | Pass |
| Product catheter deliverability<br>Bead Aggregation/Clogging:<br>The incidence of any unintended bead aggregation in the syringe resulting in catheter blockage is assessed during catheter delivery testing.<br>Ease of Delivery:<br>The ease of delivery is assessed as part of catheter delivery testing and must be considered "not difficult" in order to pass this test.<br>Shape after embolic after injection:<br>The shape of the embolic agent is evaluated after catheter delivery using optical microscopy.<br>Bead Deliverability:<br>The ability to deliver the whole vial of beads mixed with contrast agent through a catheter as described in the Instructions for Use.<br>Levels of broken or bead fragments after catheter delivery:<br>The presence of broken is evaluated after catheter delivery using optical microscopy. | Pass |
| Time to Suspension Studies<br>The time taken for the beads to form a stable homogeneous suspension when mixed with the recommended ratio of contrast agent and saline/water | Pass |
| Bead aspiration from vial<br>The ease of removing the beads from the primary packaging using standard syringes and needles as described in the Instructions for Use. | Pass |
| Bead sizing<br>The size of the beads after packaging and sterilisation. | Pass |
| pH Testing<br>The pH of the final packing solution after sterilisation. | Pass |
.
. . . . . . .
.
ﻪ
{9}------------------------------------------------
#### Performance Standards 9
LC Bead/Bead Block Compressible Microspheres meet the following Performance Standards:
- Guidance For Industry; 2004: FDA Guidance for Neurological Embolization Products. .
- ISO/EN 10993-1; 1997 Biological Evaluation of Medical Devices, Part I: Evaluation and . Testing.
- ISO/EN 10993-3; 1993 Biological Evaluation of Medical Devices, Part 3: Tests for . genotoxicity, carcinogenicity and reproductive toxicity.
- ISO/EN 10993-4; 1993 Biological Evaluation of Medical Devices, Part 4: Selection of . tests for interaction with blood.
- ISO/EN 10993-6; 1995 Biological Evaluation of Medical Devices, Part 6: Test for local . effects after implantation.
- ISO/EN 10993-10; 1995 Biological Evaluation of Medical Devices, Part 10: Tests for . Irritation and Sensitization.
- ISO/EN 10993-11; 1993 Biological Evaluation of Medical Devices, Part 11: Tests for . Systemic Toxicity.
- ISO/EN 11607: 1997 Packaging for terminally sterilized products. .
- AAMI 17665-1: 2006 Sterilization of Health Care Products Requirements for validation . and routine control - Industrial moist heat sterilization 2nd edition.
- ANSI/AAMI/ISO 14937; 2009 Sterilization of Health Care Products Characterization of . a Sterilizing Agent and the Development, Validation and Routine Control of a Sterilization Process for Medical Devices.
- ISO 14971; 2007 Medical Devices Application of Risk Management .
#### Conclusion 9.1
There are more similarities than differences between the predicate device and the LC Bead/Bead Block products. This Premarket Notification explains the minor revisions made to the manufacturing process to enable production of the additional smaller diameter SKU which is a subset of the currently cleared 100-300 LC Bead product. The primary packaging, indications for use, specifications and chemistry are unchanged from K033761/K042231/K083091. The predicate device and L.C Bead/Bead Block products have the same intended use, warnings and contraindications. The predicate device and LC Bead/Bead Block products are identical other than the added size range, in design, and unchanged from the predicate device. When used in
{10}------------------------------------------------
accordance with the instructions for use, by qualified personnel, the LC Bead/Bead Block products are safe and effective, as indicated, for the intended use.
.
.
:
.
:
:
:
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of the text, which is arranged in a circular pattern around the left side of the seal. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
# APR 1 8 2010
Biocompatibles UK Ltd. c/o Mr. John Greenbaum President Generic Devices Consulting, Inc. 20310 SW 48th Street Ft. Lauderdale, FL 33332
Re: K094018
Trade/Device Name: LC Bead/Bead Block" Compressible Microspheres Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG Dated: March 12, 2010 Received: March 17, 2010
Dear Mr. Greenbaum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{12}------------------------------------------------
# Page 2 - Mr. John Greenbaum
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Zuh Thm for
Malvina B. Eydelman, M Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
510(k) Number(if known): K094018
Device Name:
# LC Bead Microspheres Bead Block™ Compressible Microspheres
Indications For Use:
"LC Bead Microspheres & Bead Block™ Compressible Microspheres is intended for embolization of hypervascular tumors and arteriovenous malformations."
OR Prescription Use_X (Per 21 CRF 801.109)
Over-The-Counter Use_
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 1-2-96)
Quynh Hoang
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K0014018
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.