SMPLE PLATE AND SCREWS

K093983 · Biomet Trauma (Aka Ebi, Lp) · HRS · Mar 19, 2010 · Orthopedic

Device Facts

Record IDK093983
Device NameSMPLE PLATE AND SCREWS
ApplicantBiomet Trauma (Aka Ebi, Lp)
Product CodeHRS · Orthopedic
Decision DateMar 19, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The SMPle™ Pediatric Submuscular Plating System is intended for adult or pediatric patients as indicated for long bone fracture fixation of bones that have been surgically prepared (osteotomy) for correction of deformity, malunions, non-union or arthrodesis. Indications for use include internal fixation of the tibia, fibula, femur, humerus, ulna, and radius. In pediatric femoral applications, the SMPle® Pediatric Submuscular Plating System is intended for use in the treatment of femoral fractures, malunions or non-unions, femoral osteotomies required for the correction of deformity, and arthrodesis. Indications for buttressing multi-fragmentary distal femoral fractures include: supracondylar, intra-articular and extra-articular condylar, periprosthetic fractures and fractures in normal or osteopenic bone, non-unions and mal-unions, and osteotomies of the femur.

Device Story

SMPle Pediatric Submuscular Plating System consists of metallic bone plates and screws for internal fixation of long bones. Device used by orthopedic surgeons in clinical settings via standard open or submuscular surgical techniques. System provides mechanical stabilization of fractures, osteotomies, or non-unions to facilitate bone healing. Plates may be used with other treatment modalities for deformity correction. Benefit to patient includes structural support for bone alignment and healing.

Clinical Evidence

No clinical data provided; substantial equivalence based on engineering analysis and comparison to predicate devices.

Technological Characteristics

Metallic bone fixation system comprising plates and screws. Designed for standard open or submuscular surgical placement. Technological characteristics are consistent with existing legally marketed metallic plating systems.

Indications for Use

Indicated for adult or pediatric patients requiring long bone fracture fixation (tibia, fibula, femur, humerus, ulna, radius) for deformity correction, malunions, non-unions, or arthrodesis. Pediatric femoral applications include fracture treatment, osteotomies, and buttressing of multi-fragmentary distal femoral fractures (supracondylar, intra-articular, extra-articular condylar, periprosthetic) in normal or osteopenic bone.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### Biomet Trauma Traditional 510(k) Premarket Notification h093983 (pg 1/2) Image /page/0/Figure/22 description: The image shows the logo for Biomet Trauma. The logo is in black and white and features the word "BIOMET" in a bold, sans-serif font. Below the word "BIOMET" is the word "TRAUMA" in a smaller, sans-serif font. # MAR 1 9 2010 ## 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.93 Submitter: Biomet Trauma 100 Interpace Parkway Parsippany, NJ 07054 2242816 Establishment Registration Number: Contact: Margaret F. Crowe Regulatory Affairs Project Manager Biomet Trauma 100 Interpace Parkway Parsippany, NJ 070654 Tel .: 973-299-9300, ext. 2260 Fax: 973-257-0232 E-mail: margaret.crowe@biomet.com Date Prepared: December 22, 2009 SMPle" Pediatric Submuscular Plating System Trade/Proprietary Name: Common/Usual Name: Plates/Screws Single/multiple metallic bone fixation appliances and Classification Name: accessories (21 CFR 888.3030) Device Panel/Product Code: Orthopedics HRS/HWC Device Description: #### Indications for Use: The SMPle" Pediatric Submuscular Plating System is intended for adult or pediatric patients as indicated for long bone fracture fixation of bones that have been surgically prepared (osteotomy) for correction of deformity, malunions, non-union or arthrodesis. Indications for use include internal fixation of the tibia, fibula, femur, humerus, ulna, and radius. {1}------------------------------------------------ ### Biomet Trauma Traditional 510(k) Premarket Notification In pediatric femoral applications, the SMPle™ Pediatric Submuscular Plating System is intended for use in the treatment of femoral fractures, malunions or non-unions, femoral osteotomies required for the correction of deformity, and arthrodesis. Indications for buttressing multi-fragmentary distal femoral fractures include: supracondylar, intra-articular and extra-articular condylar, periprosthetic fractures and fractures in normal or osteopenic bone, non-unions and mal-unions, and osteotomies of the femur. #### Summary of Technologies: The technological characteristics of the SMPle™ Pediatric Submuscular Plating System are the same, or similar to, other legally marketed predicate devices. This plating system may be placed via a standard open surgical technique or via a submuscular technique. These plates may be used in conjunction with other devices and treatment modalities for deformity correction at the discretion of the treating physician. ### Substantial Equivalence: The SMPle™ Pediatric Submuscular Plating System is substantially equivalent to its predicate devices with respect to intended use and indications, technological characteristics, and principles of operation and does not present any new issues of safety or effectiveness. Examples of predicates include: Synthes 3.5 and 4.5 Locking Compression Plates for expanded indications (K082807); Orthopediatrics PediLoc™ Plating Systems (K083286) and the Biomet IQL Plating System (K020221), and the Biomet OptiLock® Periarticular Plating System (K061098). Based upon the engineering analysis and the other information presented in this premarket notification, the SMPle" Pediatric Submuscular Plating System is equivalent for its intended use to other legally marketed plate and screw systems. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle's head and neck, depicted with three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 Biomet Trauma (aka EBI, LP) % Ms. Margaret Crowe Regulatory Affairs Project Manager 100 Interpace Parkway Parsippany, NJ 070654 MAR 1 9 2010 Re: K093983 Trade/Device Name: SMPle™ Pediatric Submuscular Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS. HWC Dated: December 23, 2009 Received: December 24, 2009 Dear Ms. Crowe: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours for Mark N. Malle Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Biomet Trauma Traditional 510(k) Premarket Notification ## Indications for Use 510(k) Number (if known): _in093983 (pa 1/1 ) Device Name: SMPle™ Pediatric Submuscular Plating System The SMPle™ Pediatric Submuscular Plating System is intended for adult or pediatric patients as indicated for long bone fracture fixation of bones that have been surgically prepared (osteotomy) for correction of deformity, malunions, non-union or arthrodesis. Indications for use include internal fixation of the tibia, fibula, femur, humerus, ulna, and radius. In pediatric femoral applications, the SMPle® Pediatric Submuscular Plating System is intended for use in the treatment of femoral fractures, malunions or non-unions, femoral osteotomies required for the correction of deformity, and arthrodesis. Indications for buttressing multi-fragmentary distal femoral fractures include: supracondylar, intra-articular and extra-articular condylar, periprosthetic fractures and fractures in normal or osteopenic bone, non-unions and mal-unions, and osteotomies of the femur. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) . ### Concurrence of CDRH, Office of Device Evaluation (ODE) Temperature Sonta mxm (Division Sign-Off) Division of Surgica!, Orthopedic, and Restorative Devices 510(k) Number K693983 Page 4-2
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...