SMART-READ EZTEST-STEAM SCBI AND TEST PACK

K093794 · Sgm Biotech, Inc. · FRC · May 14, 2010 · General Hospital

Device Facts

Record IDK093794
Device NameSMART-READ EZTEST-STEAM SCBI AND TEST PACK
ApplicantSgm Biotech, Inc.
Product CodeFRC · General Hospital
Decision DateMay 14, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2800
Device ClassClass 2

Intended Use

Smart-Read EZTest – Steam self-contained biological indicators (SCBI) are for monitoring the efficacy of pre-vacuum saturated steam sterilization processes. Performance characteristics are established in accordance with USP 31 for the 121°C steam process. Additional pre-vacuum saturated steam sterilization temperatures are also included in the Certificate of Analysis. Smart-Read EZTest – steam SCBIs are also appropriate for use in processes of 132°C, 134°C and 135°C. The Smart-Read EZTest - steam SCBI is placed into a sterilizer load in a horizontal position in a location that is judged to be the most difficult to sterilize. A sterilization cycle appropriate for the particular type of load is run. Following the sterilization cycle, the operator removes the SCBI from the load and waits approximately 10 minutes until the device has cooled so it can be comfortably handled. The operator then activates the self-contained ampoule by placing it into the activation well in the Smart-Well incubator and gently pulling forward. This action mechanically flexes the side wall of the plastic vial to break the glass ampoule containing the culture medium. The activated Smart-Read EZTest – steam SCBI is then placed in an incubator at 60 ± 2°C for 10 hours which provides conditions conducive to growth of any bacterial spores which might have survived the sterilization cycle. Microbial growth of the indicator organisms will produce acid in the microbiological culture medium containing bromcresol purple, thus lowering the pH of the solution and invoking a yellow color change of the media components. If sterilization conditions are met the spores have been killed and the media will remain clear and purple at the conclusion of the 10 hour incubation time.

Device Story

Self-contained biological indicator (SCBI) for steam sterilization monitoring; contains Geobacillus stearothermophilus spores on paper carrier and glass ampoule of bromcresol purple culture medium within polypropylene vial. Device placed in sterilizer load; after cycle, operator cools device, then activates by flexing vial to break internal ampoule. Activated unit incubated at 60°C for 10 hours. If spores survive, microbial growth produces acid, turning medium yellow; if sterilized, medium remains purple. Provides visual confirmation of sterilization efficacy to healthcare personnel.

Clinical Evidence

Bench testing only. Reduced incubation time (10 hours) validated by exposing 100 biological indicators to steam processes reducing spore population to ~1 spore/unit. Performance met USP 31 and ANSI/AAMI/ISO 11138-1 requirements, with 97% of results available at 10 hours compared to 7-day incubation.

Technological Characteristics

Self-contained biological indicator; polypropylene vial (1.75" x 0.34"); Geobacillus stearothermophilus spores (ATCC 7953) on Whatman filter paper; Type I borosilicate glass ampoule with soybean casein digest medium and bromocresol purple; autoclavable filter paper vent. Complies with USP 31 and ANSI/AAMI/ISO 11138-1.

Indications for Use

Indicated for monitoring the efficacy of pre-vacuum saturated steam sterilization processes (121°C, 132°C, 134°C, and 135°C) in healthcare settings.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K093794 p185 . . # Tab 11 510(k) Summary #### Section 21 CFR §807.92: MAY 1 4 2010 510(k) Notification Smart-Read 11 EZTest® - Steam biological indicator monitor for steam sterilizers Submitted by: SGM Biotech, Inc. 10 Evergreen Drive Bozeman, MT 59715 Contact: John R. Gillis, Ph.D. President Tel: 406-585-9535 Fax: 406-585-9219 Email: jgillis@sgmbiotech.com This summary was prepared on 26 Apr 2010 [as required by Section 807.92(a)(1)] DEVICE: Classification Name: Indicator, Biological Sterilization Process Trade name: Smart-Read™ EZTest® - Steam biological indicator Common name: Self-contained biological indicator for steam [as required by Section 807.92(a)(2)] CLASSIFICATION: Class II (General Hospital/General Controls) SUBSTANTIALLY EQUIVALENT OT THE FOLLOWING LEGALLY MARKETED DEVICES: EZTest® - steam - SGM Biotech, Inc. (K930682) EZTest® - steam - SGM Biotech, Inc. (K963841) [as required by Section 807.92(a)(3)] > FDA Question 5 Page 2 K093794 {1}------------------------------------------------ ## DESCRIPTION OF DEVICE: Smart-Read EZTest – steam [as required by Section 807.92 (a)(4)] The biological indicator consists of a self-contained unit which includes: bacterial spores inoculated onto a paper carrier, a small glass ampoule containing sterile culture medium and color indicator, and a plastic vial that serves as the culture tube. This unit complies to the performance characteristics described in USP 31 and ANSI/AAMI/ISO 11138-1. #### Spores Bacterial endospores of Geobacillus stearothermophilus, ATCC 7953 or equivalent, are inoculated onto a 591 Whatman filter paper carrier cut into ¼" x ¾" strips to be placed inside the plastic body of the device. The population of spores on each individual piece of paper will be between 1.0 x 10(x) and 5 x 10(x). The spore concentration will be labeled as a 10- or 10 concentration. The population data is based on the heat shock population assay. The heat shock procedure is 95°C to 100°C for 15 minutes as outlined in USP 31. #### Culture Tube The culture tube is made of a polypropylene. The plastic culture tube is 1.75" in length and 0.34" in OD. The 0.02″ wall allows the culture tube to be flexible so that the culture media glass ampoule can be broken to activate the test. #### Culture Medium The culture medium, consisting of a formulated soybean casein digest medium containing bromocresol purple, this medium is filled into ampoules of a Type I borosilicate glass, flame sealed and sterilized. The sealed ampoules are placed into the plastic body with the spore paper. #### Filter Paper Vent The plastic body is closed with a ½" square autoclavable filter paper vent material. ## Cap A polypropylene plastic cap is used to hold the vent material in place. This cap has three small holes in it to allow steam and air to freely move in and out of the body. FDA Question 5 Page 3 K093794 <sup>-</sup> Reference USP 21 {2}------------------------------------------------ #### INCUBATION: Following exposure to the sterilization process the exposed Smart-Read EZTest is incubated at 60 + 2°C for 10 hours. If spores have survived (sterilization failure) the unit will turn yellow due to the growth of the surviving spores. #### PHYSICAL PERFORMANCE: Reduced incubation time was established by exposing sets of 100 Bls to a steam process that reduced the spore population to approximately 1 spore per Bl. The exposure was acceptable if at least 30 units demonstrated growth and no more than 80 units demonstrated growth. The units were incubated at 60 ± 2°C for 7 days. The reduced incubation time was identified when 97% of the results of the 7 day results were available. This test was performed with four different spore crops and multiple lots of product from each crop. The data appears below. | Fume and 1 1 Hrs. 1 - 8 Hrs. 3 5 9 Hrs. 3 10 Hrs. 2 10 Hrs. 2 Hrs. 2 Hrs. 2 Hrs. 1 1 2 Hrs. | | | | | | | | | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|--------------------|----------|-----------------------------------|----------------------------|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|---------| | Partial Cycle #1 ************************************************************************************************************************************************************* | | | | | | | | | | | Positives | 0 / 44 | 22 / 44 | 36 / 44 | 43 / 44 | 44 / 44 | 44 / 44 | 44 / 44 | | | | Percent Growth | 046 | 50.0% | 81.896 | 97.7% | 100% | 100% | 100% | | | | Partial Cycle a2 2000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 | | | | | | | | | | | Positives | 30 / 45 | 41 / 45 | 44145 | 44 / 45 | 45 / 45 | 45 / 45 | 45 / 45 | | | | Percent Growth | 66.756 | 91.1% | 97.8% | 97.8% | 100% | 100% | 100% | | | | Partlat Cycle #3 20 22 10 12 10 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 | | | | | yr - Sitter Art Start Time | 2017 2017 | 334 - 15 | | | | Positives | 8/73 | 49173 | 65 / 73 | 71 / 75 | 73/73 | 73 / 73 | 73/73 | | | | Percent Growth | 11.096 | 67.1% | 89.0% | 97.3% | 100% | 100% | 100% | | | | Partial Cycle #4 11 12 18 11 12 12 12 12 12 12 12 12 12 12 12 12 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 | | | | | | | | | | | Positives | 58 / 80 | 71 / 80 | 76180 | 78 / 80 | 80 / 80 | 80 / 80 | 80 / 80 | | | | Percent Growth | 72.5% | 88.8% | 95.0% | 97.5% | 100% | 100% | 100% | | | | | | | | | | | | | | | Time 20 1 1 2 3 1 1 7 Hrs 1 8 Hrs : 10 Hrs 1 10 Hrs. 1 - 24 Hrs. 1 2 Hrs. 1 2 Hrs. 1 2 168 Hrs. | | | | | | | | | | | Partlal Cycle #5 1 = 1 = 1 = 1 = 1 = 1 = 1 = = = = = | | | | | | | and and and the same of the comments of the mail of the mail of the | | | | Positives | 32/51 | 51 / 52 | 52 / 52 | 52 / 52 | 52 / 52 | 52 / 52 | 52 / 52 | | | | Percent Growth | 61.5% | 98.1% | 100% | 100% | 100% | 100% | 100% | | | | Partial Cycle #6 " " 130 " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " | | | | | | | 中国有意義 (1) 1977年 1997年 文章書 (1) 文字 (1) 文字 | | | | Positives | 44 / 53 | 53 / 53 | 23 / 53 | 23 / 53 | 23 / 23 | 23 1 23 | 53 / 53 | | | | Percent Growth | 83.096 | 100% | 100% | 100% | 100% | 100% | 100% | | | | Partial Cycle #7 . 3 . 0 . 0 . 0 . 0 . 0 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 1 | | | | | | 2017 - 12:55 | | | | | Positives | 36/73 | 66 / 73 | 72173 | 73 / 73 | 73 / 73 | 73 / 73 | 73 / 73 | | | | Percent Growth | 49.3% | 90.4% | 98.646 | 100% | 100% | 100% | 100% | | | | Time 3 1 1 1 2 0 1 7 Hrs. 1 8 Hrs. 3 10 Hrs. 3 2 24 Hrs. 2 24 Hrs. 1 2 72 Hrs. 1 2 188 Hrs. | | | | | | | | | | | Partial Cycle #8 : - fr. - fr. - fr. - 1. 2 . 1. 1. 1. 1. 1. | | | | | | | 1 * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * | | | | Positives | 26 / 48 | 43 / 48 | 48 / 48 | 48 / 48 | 48 / 48 | 48 / 48 | 48 / 48 | | | | Percent Growth | 52.4% | 89.696 | 100% | 100% | 100% | 100% | 100% | | | | Partial Cycle #9 | - F | . | | : . | | | | | | | Positives | 30 / 46 | 44 / 46 | 45146 | 45 / 46 | વર્ણ | વર્ણ | વર્સ | | | | Percent Growth | 65.2% | 95.7% | 97.8% | 97.8% | 100% | 100% | 100% | | | | Partial Cycle #10: ' | | - પ્રતિ<br>。 | . : : *! | | <br>- - | | 12 - 11 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - | | | | Positives | રવ / 60 | રેક / સ્વિ | 59 / 60 | રત દર | ટવે છે. | રત ભ્વ | સ્ત્ર | | | | Percent Growth | 90.096 | 96.7% | 88.396 | 98.346 | 98.366 | 98.3% | 100% | | | | Partial Cycle #11: | :<br>. | | | . . | | | | | | | | 46 / 61 | -----<br><br>55/61 | | 60 / 61 | λί<br>-<br>50 / 61 | | | | | | Positives | | | 58161 | | | ୧୦ । ୧। | રા । રા | | | | Percent Growth | 75.4% | 90. 296 | 95.196 | 98.4% | 98.4% | 98.496 | 100% | | | | Time Time Time | | | | 7 Hrs ! 8 Hrs. 8 9 Hrs. 1 10 Hrs. | | | 24 Hrs. 72 Hrs. 1 3 168 Hrs. | | | | Partial Cycle 12 " | | | | | 1 - 1983 - 1<br>ייר | | . | DA Question 5 Page 4 | | | Positives | 28 / 80 | 71 / 80 | 76 / 80 | 78 / 80 | 80 / 80 | 80 / 80 | 80 / 80 | | | | Percent Growth | 72.5% | 88.894 | વે 5 જુ | 97.5% | 100% | 100% | 100% | | K093794 | | | | | | | | | | | | Biological Data for Smart-Read Reduced Incubation Time {3}------------------------------------------------ K093794 ρ⁴δ5 #### INTENDED USE OF DEVICE: [as required by Section 807.92(a)(5)] Smart-Read EZTest – Steam self-contained biological indicators (SCBI) are for monitoring the efficacy of pre-vacuum saturated steam sterilization processes. Performance characteristics are established in accordance with USP 31 for the 121°C steam process. Additional pre-vacuum saturated steam sterilization temperatures are also included in the Certificate of Analysis. Smart-Read EZTest – steam SCBIs are also appropriate for use in processes of 132°C, 134°C and 135°C. The Smart-Read EZTest - steam SCBI is placed into a sterilizer load in a horizontal position in a location that is judged to be the most difficult to sterilize. A sterilization cycle appropriate for the particular type of load is run. Following the sterilization cycle, the operator removes the SCBI from the load and waits approximately 10 minutes until the device has cooled so it can be comfortably handled. The operator then activates the self-contained ampoule by placing it into the activation well in the Smart-Well incubator and gently pulling forward. This action mechanically flexes the side wall of the plastic vial to break the glass ampoule containing the culture medium. The activated Smart-Read EZTest – steam SCBI is then placed in an incubator at 60 ± 2°C for 10 hours which provides conditions conducive to growth of any bacterial spores which might have survived the sterilization cycle. Microbial growth of the indicator organisms will produce acid in the microbiological culture medium containing bromcresol purple, thus lowering the pH of the solution and invoking a yellow color change of the media components. If sterilization conditions are met the spores have been killed and the media will remain clear and purple at the conclusion of the 10 hour incubation time. ## SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF Smart-Read EZTest – steam COMPARED TO THE CHARACTERISTICS OF THE PREDICATE DEVICE: [as required by Section 807.92(a)(6)] Smart-Read EZTest – steam, is similar in type to other products presently in the marketplace whose useful intent is the monitoring of sterilizer efficacy, has been determined to be safe and effective. Active components of the indicator have been in routine use more than 30 years. They predate the medical device amendment of the Food and Drug Cosmetic Act and are recognized by the United States Pharmacopoeia 31. The incubation time of 10 hours at 60 + 2°C meets the requirements of the CDRH guidance "Validating Biological Indicator Incubation Time". > FDA Question 5 Page 5 K093794 {4}------------------------------------------------ K093799 p5 of 5 DETERMINATION OF SUBSTANTIALLY EQUIVALENT BASED ON PHYSICAL CHARACTERISTICS: [as required by Section 807.92 (b)(1)] A comparison is provided of the Smart-Read EZTest – steam to the legally marketed predicate device describing similarities and differences such as technology and other important characteristics. - A. EZTest - steam (K930682): ls manufactured in precisely the same manner as the Smart-Read EZTest device. The only difference is the extension of the label claims for an incubation temperature of 60 + 2°C and a reduced incubation time (RIT) of 10 hours. - B. EZTest steam (K963841): This device is exactly the same as the Smart-Read EZTest - steam. The difference is the extension of the label claims for incubation temperature of 60 ± 2°C and the RIT of 10 hours. ### DETERMINATION OF SUBSTANTIALLY EQUIVALENCE BASED ON IN USE PERFORMANCE: [as required by Section 807.92 (b)(2)] The resistance performance of Smart-Read EZTest - steam is exactly the same as EZTest - steam K930682 and K963841. CONCLUSION: fas required by Section 807.92 (b) (3)] Smart-Read EZTest – steam is appropriate for monitoring pre-vacuum steam sterilization processes of 121°C, 132°C, 134°C and 135°C and meets the USP 31, ANSI/AAMI/ISO 11138-1, and ANSI/AAMI ST 59:1999 requirements. > FDA Question 5 Page 6 K093794 {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol resembling three overlapping lines or strokes, possibly representing a stylized human figure or a symbol of health and well-being. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 John R. Gillis, Ph.D. President SGM Biotech, Incorporated 10 Evergreen Drive, Suite E Bozeman, Montana 59715 # MAY 1 4 2010 Re: K093794 Trade/Device Name: Self-Contained Biological Indicator Smart-Read® EZTest® - Steam Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: April 26, 2010 Received: April 27, 2010 Dear Dr. Gillis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2- Dr. Gillis Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. h for. Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ## TAB 9 Indications for Use KO93794 510(k) Number (if known): Smart-Read EZTest - Steam Device Name: Self-contained Biological Indicator Indications for Use: Smart-Read EZTest – Steam self-contained biological indicators (SCBI) are for monitoring the efficacy of saturated steam sterilization processes. Performance characteristics are established in accordance with USP 31 for the pre-vacuum 121°C steam process. Additional pre-vacuum saturated steam sterilization temperatures are also included in the Certificate of Analysis. Smart-Read EZTest – steam SCBIs are also appropriate for use in saturated steam processes of 132°C, 134°C and 135°C. The Smart-Read EZTest – steam SCBI is placed into a sterilizer load in a horizontal position in a location that is judged to be the most difficult to sterilize. A sterilization cycle appropriate for the particular type of load is run. Following the sterilization cycle, the operator removes the SCBI from the load and waits approximately 10 minutes until the device has cooled so it can be comfortably handled: The operator then activates the self-contained ampoule by placing it into the activation well in the Smart-Well incubator and gently pulling forward. This action mechanically flexes the side wall of the plastic vial to break the glass ampoule containing the culture medium. The activated Smart-Read EZTest steam SCBI is then placed in an incubator at 60 ± 2°C for 10 hours which provides conditions conducive to growth of any bacterial spores which might have survived the sterilization cycle. Microbial growth of the indicator organisms will produce acid in the microbiological culture medium containing bromcresol purple, thus lowering the pH of the solution and invoking a yellow color change of the media components. If sterilization conditions are met the spores have been killed and the media will remain clear and purple at the conclusion of the 10 hour incubation time. signature (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 14093794 510(k) Number: > FDA Question 1 Page 2 K093794
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...