K093583 · St Jude Medical · DQK · Dec 15, 2009 · Cardiovascular
Device Facts
Record ID
K093583
Device Name
EP -NURSEMATE, EP -NURSEMATE WITH PHYSIO MODULE
Applicant
St Jude Medical
Product Code
DQK · Cardiovascular
Decision Date
Dec 15, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The EP-NurseMate™ is intended to be used as an extension of the EP-WorkMate™ Recording System for the viewing and measurements of cardiac electrograms, and electronic data entry of procedural events during a clinical electrophysiology procedure. The EP-NurseMate™ with Physio Module is intended to be used as an extension of the EP-WorkMate™ Recording System for the viewing and measurements of cardiac electrograms, receiving vital signs measurements from a patient monitoring unit, and electronic data entry of procedural events during a clinical electrophysiology procedure.
Device Story
EP-NurseMate and EP-NurseMate with Physio Module are remote workstations extending the EP-WorkMate Recording System. Used in clinical electrophysiology (EP) labs by EP team members (nurses/technicians) to perform real-time charting (event titles, medications, comments) and monitor physiologic data. Inputs include cardiac electrograms from the EP-WorkMate system and vital signs from a physiologic monitoring unit. The device displays these inputs on a touchscreen LCD monitor, allowing secondary personnel to view data and enter procedural events independently of the primary recording system. It does not control patient monitors or cardiac stimulation. The system facilitates procedural documentation and monitoring, potentially improving workflow efficiency and data management during EP studies.
Clinical Evidence
Clinical simulation testing was performed to confirm that design changes met requirements and did not adversely affect safety or efficacy. No clinical patient data was required.
Technological Characteristics
Hardware includes PC, touchscreen LCD monitor, keyboard, and cart. The Physio Module version includes a physiologic monitoring device. Software is a modified version of the EP-WorkMate application. Connectivity is via system interconnections to the EP-WorkMate Recording System. No specific materials or sterilization methods are cited.
Indications for Use
Indicated for use during clinical electrophysiology procedures.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for St. Jude Medical. The logo consists of a 3x3 grid of squares on the left, followed by the text "ST. JUDE MEDICAL" in a bold, sans-serif font. Below the company name, there is a tagline that reads "More control Less risk."
K093583
p 1/4
### 510(k) SUMMARY
#### Submitter's Name/Contact Person
Eleanor J. Fox, M.S. Regulatory Affairs Specialist (856) 753-8533
#### Alternative:
Timothy J. Kappers, MBA, RAC Senior Manager, Regulatory Affairs (651) 756-2904
St. Jude Medical Cooper Run Executive Park 575 Route 73 North, Building D West Berlin, NJ 08091-9293 (USA)
#### FDA Registration # 2248049
#### General Information
| Trade Name | EP-NurseMate™ Remote Review and<br>Charting Station<br>(also referred to as EP-NurseMate™)<br><br>EP-NurseMate™ with Physio Remote<br>Review and Charting Station with<br>Physiologic Monitoring Module<br>(also referred to as EP-NurseMate™ with<br>Physio Module) |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name | Computer, diagnostic, programmable |
| Classification Name | DQK, Class II, 21 CFR 870.1425 |
| Predicate Device | NurseMate™<br><br>NurseMate with Physio Module™ |
# Device Description
The EP-NurseMate™ and the EP-NurseMate™ with Physio Module, Version 4.2.0, are modifications to the EP-WorkMate TM Recording System. The EP-NurseMate is connected to the EP-WorkMate Recording System for real-time charting, physiologic monitoring, and data analysis during electrophysiology (EP) studies. Expanding the EP-
DEC 1 5 2009
{1}------------------------------------------------
K093583
p3/4
Image /page/1/Picture/1 description: The image shows the logo for St. Jude Medical. The logo consists of a 3x3 grid of squares on the left, followed by the text "ST. Jude Medical" in a serif font. Below the text is the tagline "MORE CONTROL. LESS RISK."
WorkMate™ Recording System with the optional EP-NurseMate™ or EP-NurseMate™ with Physio Module creates an additional workstation for a member of the EP team to perform charting (e.g., event titles, medications, and comments) and monitor physiologic data from their own display.
# Software Description
The EP-NurseMate™ and the EP-NurseMate™ with Physio Module's functions are controlled by system software. The menu driven software is controlled through a user interface with a touchscreen LCD monitor or keyboard. The EP-NurseMate™ and the EP-NurseMate™ with Physio Module software is a modified version of the EP-WorkMate™ application software. The software was modified using the same Microsoft Visual C++ 6.0 Integrated Development Environment (with Visual Studio Service Pack 6.0A) within which it was originally developed. It is important to note that NurseMate™/EP-NurseMate™ with Physio Module do not control the patient monitor and do not control cardiac stimulation.
#### Hardware Description
The EP-NurseMate™ hardware consists of a PC, a touchscreen LCD monitor, a keyboard, and a cart.
The EP-NurseMate™ with Physio Module consists of a PC, a touchscreen LCD monitor, a keyboard, a physiologic monitoring device, and a cart.
A graphical presentation of the system interconnections is provided in Figures 1 and 2 of for EP-NurseMate™ and EP-NurseMate™ with Physio, respectively.
Image /page/1/Figure/9 description: The image shows a diagram of a medical system. The system includes components such as a patient, Amp/SCU, EP-WorkMate computer, NurseMate computer, and various monitors. Communication lines connect the Amp/SCU to the EP-WorkMate and NurseMate computers, and the EP-WorkMate and NurseMate computers are connected to their respective monitors.
Figure 1 EP-WorkMate™ with NurseMate™ Block Diagram
{2}------------------------------------------------
K092583
p3/4
Image /page/2/Figure/1 description: The image shows the logo for St. Jude Medical. The logo consists of a grid of black squares on the left, followed by the text "ST. JUDE MEDICAL" in a bold, serif font. Below the company name, there is a tagline that reads "MORE CONTROL LESS RISK."
Image /page/2/Figure/2 description: The image shows a diagram of a medical monitoring system. The system includes components such as an Amp/SCU unit connected to a patient, EP-WorkMate and NurseMate computers, and various monitors. The diagram also depicts the communication lines between these components, as well as the roles of medical personnel such as doctors and nurses in utilizing the system.
Figure 2: EP-WorkMate® with NurseMate™ with Physio Module Block Diagram
# Indications for Use
The EP-NurseMate™ is indicated for use during clinical electrophysiology procedures.
The EP-NurseMate™ with Physio is indicated for use during clinical electrophysiology procedures.
#### Intended Use
The EP-NurseMate™ is intended to be used as an extension of the EP-WorkMate™ Recording System for the viewing and measurements of cardiac electrograms, and electronic data entry of procedural events during a clinical electrophysiology procedure.
The EP-NurseMate™ with Physio Module is intended to be used as an extension of the EP-WorkMate™ Recording System for the viewing and measurements of cardiac electrograms, receiving vital signs measurements from a patient monitoring unit, and electronic data entry of procedural events during a clinical electrophysiology procedure.
# Device Comparison to the Cleared Device
The EP-NurseMate and the EP-NurseMate with Physio Modules have the same indications for use and fundamental scientific technology as the predicate devices. All technological characteristics of the devices are substantially equivalent to the predicate devices.
{3}------------------------------------------------
K093583
p 4/4
Image /page/3/Picture/1 description: The image shows the logo for St. Jude Medical. The logo consists of a grid of nine squares on the left and the words "ST. JUDE MEDICAL" in a stylized font on the right. Below the company name is the tagline "MORE CONTROL LESS RISK."
#### Summary of Non-Clinical Testing
Clinical simulation testing was performed to confirm that the changes met design requirements and did not adversely affect the safety or efficacy of the devices.
### Summary of Design Control Activities
The development of the EP-NurseMate™ and the EP-NurseMate™ with Physio Module were performed in accordance with St. Jude Medical's Quality System requirements, and in compliance with Quality System Regulation design control requirements documented in 21 CFR 820.30. A Declaration of Conformity with Design Controls is provided in Attachment J.
#### Conclusion
The EP-NurseMate™ and the EP-NurseMate™ with Physio Module have the same indications for use and fundamental scientific technology as the predicate devices. All technological characteristics of the EP-WorkMate are substantially equivalent to the predicate device.
Where operational and performance differences exist between the proposed device and the predicate device, performance testing demonstrated that these differences do not adversely affect the device's safety and effectiveness.
Therefore, St. Jude Medical considers the EP-NurseMate™ and the EP-NurseMate with Physio Module to be substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a seal for the Department of Health & Human Services (HHS). The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is a stylized image of a caduceus, a symbol often associated with medicine and healthcare.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
DEC 1 5 2009
St. Jude Medical, Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo MN 53313
Re: K093583
Trade/Device Name: EP-NurseMate and EP-NurseMate with Physio Module version 4.2 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: December 5, 2009 Received: December 9, 2009
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{5}------------------------------------------------
Page 2 – Mr. Mark Job
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for St. Jude Medical. The logo consists of a grid of squares on the left and the words "ST. Jude Medical" on the right. Below the company name is the tagline "More control. Less risk."
# Indications for Use
# 510(k) Number (if known)_ K093583
#### Device Name:
EP-NurseMate™ (EP-NurseMate™ Remote Review and Charting Station) EP-NurseMate™ with Physio Module (EP-NurseMate™ Remote Review and Charting Station with Physiologic Monitoring Module)
### Indications for Use:
The EP-NurseMate™ is indicated for use during clinical electrophysiology procedures.
The EP-NurseMate™ with Physio Module is indicated for use during clinical electrophysiology procedures.
| Prescription Use X<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
|---------------------------------------------------|--------|------------------------------------------------|
|---------------------------------------------------|--------|------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|------------------------------------|
| Division of Cardiovascular Devices |
| 510(k) Number | 4093583 |
|---------------|---------|
|---------------|---------|
Page 1 of 1
EP-NurseMate Version 4.2 - St. Jude Medical - November 5, 2009 - Special 510(k) - Page 27 of 2005
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.