K093547 · Pinpointe USA, Inc. · PDZ · Oct 15, 2010 · General, Plastic Surgery
Device Facts
Record ID
K093547
Device Name
PINPOINTE FOOTLASER
Applicant
Pinpointe USA, Inc.
Product Code
PDZ · General, Plastic Surgery
Decision Date
Oct 15, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PinPointe™ FootLaser™ and the delivery accessories that are used with them are intended for use in surgical procedures involving open, laparoscopic and endoscopic ablation, vaporization, excision, incision, and coagulation of soft tissue in the medical specialties of general and cosmetic dentistry, otolaryngology/ENT surgery, and dermatology & plastic surgery including intraoral soft tissue dental surgery, oral maxillofacial and cosmetic surgery, general surgery, E.N.T. surgery, podiatry, and dermatology and plastic surgery. The PinPointe™ FootLaser™ is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
Device Story
PinPointe FootLaser is a flashlamp-pumped solid-state Nd:YAG laser system (1064 nm) used for soft tissue surgery and onychomycosis treatment. System includes main console with control panel, footswitch, aiming beam (630-680 nm), and various delivery devices (fiber-optic handpieces, guide tips). Physician operates system via footswitch to deliver pulsed laser energy to target tissue or nail. Aiming beam assists in positioning. Device enables precise tissue ablation, vaporization, and coagulation. In onychomycosis, laser energy targets nail bed/plate to promote clear nail growth. Used in clinical settings (dermatology, podiatry, surgery). Output allows clinicians to perform minimally invasive procedures or treat nail fungal infections, potentially improving patient nail appearance and clinical outcomes.
Clinical Evidence
Clinical study demonstrated safety and effectiveness for temporary increase of clear nail in patients with onychomycosis (dermatophytes/yeasts) at 6 and 12 months post-treatment.
Technological Characteristics
Flashlamp-pumped solid-state Nd:YAG laser; 1064 nm wavelength; 630-680 nm aiming beam; pulsed output mode (5-100 Hz); 20-200 mJ pulse energy; ≤ 6W power; 100-700 μsec pulse duration. System includes console, footswitch, and reusable/sterilizable fiber-optic delivery devices. Electrical requirements: 90-130 VAC or 200-240 VAC, 50/60 Hz. Dimensions: 32" x 13" x 14"; Weight: 17.2 kg.
Indications for Use
Indicated for patients with onychomycosis (dermatophytes/yeasts) for temporary increase of clear nail; also indicated for soft tissue surgical procedures (ablation, vaporization, excision, incision, coagulation) in dentistry, ENT, dermatology, plastic surgery, and general surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers (K083616)
Incisive Family of InPulse Nd:YAG Lasers (K083215)
Submission Summary (Full Text)
{0}------------------------------------------------
693547
CONFIDENTIAL
Section 5 - 510(k) Summary
DCT1 5 2010
### I. General Information
Submitter:
PinPointe USA. Inc. 275 Airpark Boulevard Suite 100 Chico, CA 95973 530-809-3800 Tel: Fax: 866-235-1386
USA. Inc.
John Strisower Chief Executive Officer, PinPointe USA, Inc
Summary Preparation Date:
October 13, 2010
II. Names
Trade Name(s):
Contact Person:
PinPointe™ FootLaser™ (and delivery device accessories)
Common Name(s):
Laser Powered Surgical Instruments
Primary Classification Names: Laser Powered Surgical Instruments (and Accessories) (21 CFR Part 878.4810, Product Code GEX)
### III. Predicate Devices
- PathoLase Family of PinPointe™ and PinPointe™ FootLaser™ Nd:YAG Lasers ● (K083616)
- Incisive Family of InPulse Nd: Y AG Lasers (K083215). .
### Product Description IV.
:
.
The PinPointe™ FootLaser™ is comprised of the following main components:
- Main console containing the major electrical components, including:
- > Control/ Display Panel with the:
- Kevswitch (that controls authorized access to the laser system); ●
- emergency Laser Stop button; .
- Displays (laser emission indicator, average power, pulse energy, . repetition rate)
- Standby button (default mode when laser system turned on places . system into the Standby mode preventing laser emission).
- Ready button (places system into the Ready mode allowing laser ● cmission when the footswitch is depressed and a fiber optic is properly attached);
- 1064 nm treatment laser (solid state Nd: YAG laser rod) with flashlamp and > associated light regulation components and electronics;
- A 630 -680 nm (red) aiming beam diode laser;
- Delivery device fiber-optic connector port; A
- Remote interlock connector (External door interlock connector); A
{1}------------------------------------------------
K093547
- Connector ports for the footswitch and power cord; A
- Accessory holder (attached to the rear of the main console); A
- . Footswitch;
A
- Medical grade power cord; .
- Delivery Devices for Non-Contact and Contact with Intact Skin/Tissue: .
- No Standoff: Reusable, cleanable, tip is provided for non-Guide Tip contact use to direct and control the placement of the laser beam (free beam) at the treatment location. The Guide tip attaches to the end of the handpiece. The optical fiber is threaded through the handpiece and fits securely into the bore of the Guide tip;
- > Guide Tip -With Standoff: Reusable, cleanable, tip is provided for minimal-contact with intact skin/ tissue to direct and control the placement of the laser beam at the treatment The Guide tip attaches to the end of the location. The optical fiber is threaded through the handpiece. handpiece and fits securely into the bore of the Guide tip;
- Delivery Devices for Contact with Breached Surfaces: ●
- Reusable, cleanable, sterilizable optical fibers (range of A Optical Fibers . 200 - 1000 um diameter) provided non-sterile, clean and ready for sterilization (steam autoclave).
- Handpieces -Reusable, cleanable, sterilizable handpieces (large and A small diameter shafts) provided non-sterile, clean and ready for sterilization (steam autoclave). The optical fiber is threaded through the handpiece and secured and held in place with the handpiece locking cap;
- Handpiece Tips Disposable single-use tips are provided in straight and A curved configurations and are used to direct and control the placement of the optical fiber tip at the treatment The handpiece tips attach to the end of the location. handpiece. The optical fiber is threaded through both the handpiece and the handpiece tip;
- . Accessories:
- A Safety Glasses
- . Tools:
- Optical Fiber Striper; A
- Optical Fiber Cleaver (carbide wedge, ceramic, or equivalent scribe for cleaving A the optical fibers).
#### V. Indications for Use
The PinPointe™ FootLaser™ and the delivery accessories that are used with them are intended for use in surgical procedures involving open, laparoscopic and endoscopic ablation, vaporization, excision, incision, and coagulation of soft tissue in the medical specialties of general and cosmetic dentistry, otolaryngology/ENT surgery, and dermatology & plastic surgery including intraoral soft tissue dental surgery, oral maxillofacial and cosmetic surgery, general surgery, E.N.T. surgery, podiatry, and dermatology and plastic surgery.
{2}------------------------------------------------
### Podiatry
Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:
- Matrixectomy
- . Radical nail excision
- . Periungual and subungual warts
- Plantar warts .
- . Neuromas
Plantar warts
Removal of tattoos
Treatment of keloids
The PinPointe™ FootLaser™ is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
D
.
.
.
.
### Dermatology and Plastic Surgery
Dermatology and plastic surgery (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:
- Lesions of skin and subcutaneous . ● tissue .
- . Telangiectasia :
- Port wine lesions .
- Spider veins .
- Hemangiomas ●
## Oropharangeal / Dental Surgery
Indicated for:
- Abscess incision and drainage .
- Aphthous ulcers treatment .
- . Biopsies, excisional and incisional
- Crown lengthening ●
- Exposure of unerupted / partially . erupted teeth
- . Fibroma removal
- Frenectomy .
- Frenotomy ●
- Gingival incision and excision ●
- Gingivectomy .�
- . Gingivoplasty
- Hemostasis ●
- Implant recovery .
- Lesion (tumor) removal .
- . Leukoplakia
- Operculectomy ●
- Oral papillectomy .
## General Surgery
Indicated for:
.
-
- Pulpotomy ● Pulpotomy as adjunct to root canal � therapy
Periungual and subungual warts
Debridement of decubitus ulcer
- . Removal of filling material such as gutta percha or resin as adjunct treatment during root canal retreatment
- . Selective ablation of enamel (first degree) caries removal
- Sulcular debridement (removal of . diseased or inflamed soft tissue in the periodontal pocket) to improve clinical indices including gingival index, gingival bleeding index. probe depth, attachment loss, and tooth mobility
- Tissue retraction for impressions .
- Vestibuloplasty �
- Open, laparoscopic, and endoscopic general surgery (ablation, vaporization, . incision, excision, and coagulation of soft tissue) including:
-
1
.
- Cholecystectomy .
- Lymphadenectomy -
- Mastectomy -
- Partial nephrectomy
- Hepatectomy .
- -Parathyroidectomy -Removal of tumors
Removal of lesions
Removal of polyps
Thyroidectomy
Premarket Notification, 510(k) Submission for: PinPointe™ FootLaser™
{3}------------------------------------------------
# Kog3547
### CONFIDENTIAL
- Pilonidal cystectomy .
- Pancreatectomy -
- Resection of lipoma -
- -Splenectomy
- Pelvic adhesiolysis -
- Hemorrhoidectomy ↓
- Herniorrhaphy .
- Tumor biopsy .
- Tonsillectomy .
- Debridement of decubitus ulcers .
- Appendectomy .
### Endonasal Surgery
Endonasal surgery (ablation, vaporization, incision, and coagulation of soft tissue) including:
- � Lesions or tumors of the oral, nasal, glossal, pharyngeal & laryngeal tissues
- Tonsillectomy .
- Adenoidectomy .
The safety and effectiveness of the PinPointe™ FootLaser™ for use for the temporary increase of clear nail in patients with onychomycosis was demonstrated in clinical studies.
### VI. Summary of Technological Characteristics
The technological characteristics of the PinPointe™ FootLaser™ are identical to those of the predicate devices.
| Characteristic | K093545 - PinPointe™<br>FootLaser™ and Delivery<br>Devices; Incisive, Inc. | | | K083616 - PathoLase Family of<br>PinPointe™ and PinPointe™<br>FootLaser™ Lasers; PathoLase, Inc. | | | K0832153 - Incisive Family of<br>InPulse Nd:YAG Lasers;<br>Incisive, LLC | | |
|------------------------------------|----------------------------------------------------------------------------|------------|------------|--------------------------------------------------------------------------------------------------|------------|-----------|--------------------------------------------------------------------------|------------|------------|
| Wavelength | 1064 nm | | | 1064 nm | | | 1064 nm | | |
| Aiming Beam | 630-680 nm (≤ 2.5 mW) | | | 630-680 nm (≤ 2.5 mW) | | | 630-680 nm (≤ 2.5 mW) | | |
| Model | 6 W | 30 W | 100.W | 6 W | 30 W | 100 W | 6 W | 30 W | 100 W |
| Energy Per Pulse<br>(mJ) | 20-200 | 20-1000 | 20 - 3500 | 20-200 | 20-1000 | 20 - 3500 | 20-200 | 20-1000 | 20 - 3500 |
| Power (Watts) | ≤ 6 | ≤ 30 | ≤ 100 | ≤ 6 | ≤ 30 | ≤ 100 | ≤ 6 | ≤ 30 | ≤ 100 |
| Pulse Duration<br>(μsec) | 100 - 700 | 350 - 3000 | 350 - 3000 | | 100 - 3000 | | 100 - 700 | 350 - 3000 | 350 - 3000 |
| Output Mode | Pulsed, Multi-Mode | | | Pulsed, Multi-Mode | | | Pulsed, Multi-Mode | | |
| Repetition Rate | 5 - 100 Hz | | | 5 - 100 Hz | | | 5-100 Hz | | |
| Laser Media | Flashlamp-Pumped Solid State<br>Laser Rod | | | Flashlamp-Pumped Solid State<br>Laser Rod | | | Flashlamp-Pumped Solid State<br>Laser Rod | | |
| User Interface | Push-button control panel | | | LCD Color touch screen | | | Push-button control panel | | |
| Laser Activation | Footswitch | | | Footswitch | | | Footswitch | | |
| Delivery Devices<br>• How supplied | • Non-sterile, reusable, cleanable, sterilizable | | | • Non-sterile, reusable, cleanable, sterilizable | | | • Non-sterile, reusable, cleanable, sterilizable | | |
| System<br>Dimensions | 32" x 13" x 14" (H x W x D) | | | 32" x 13" x 14" (H x W x D) | | | 32" x 13" x 14" (H x W x D) | | |
| System Weight | 17.2 kg (38 lbs) | | | 17.2 kg (38 lbs) | | | 17.2 kg (38 lbs) | | |
| Electrical Reqs. | 90-130 VAC, 50/60 Hz<br>200-240 VAC, 50/60 Hz | | | 90-130 VAC, 50/60 Hz<br>200-240 VAC, 50/60 Hz | | | 90-130 VAC, 50/60 Hz<br>200-240 VAC, 50/60 Hz | | |
{4}------------------------------------------------
Clinical study demonstrated the safety and effectiveness of the PinPointe™ FootLaser™ for use for the temporary increase of clear nail at 6 and 12 months following treatment in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
### VIII. Safety and Effectiveness Information
The review of the indications for use, technical characteristics, and clinical study results provided demonstrates that the PinPointe™ FootLaser™ is substantially equivalent to the predicate devices and is safe and effective for use for the temporary increase of clear nail at 6 and 12 months following treatment in patients with onychomycosis.
### IX. Conclusion
The PinPointe™ FootLaser™ was found to be substantially equivalent to the predicate devices.
The PinPointe™ FootLaser™ share identical indications for use, similar design features, and functional features with, and thus are substantially equivalent to, the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with outstretched wings, positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The logo is circular in shape.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 13, 2013
PinPointe USA. Inc. % Ms. Anne Worden Regulatory Consultant 3637 Bernal Avenue Pleasanton, California 94566
Re: K093547
Trade/Device Name: PinPointe FootLaser and Accessories Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: PDZ, GEX Dated: October 13, 2010 Received: October 14, 2010
Dear Ms. Worden:
This letter corrects our substantially equivalent letter of October 15, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{6}------------------------------------------------
Page 2 - Ms. Anne Worden
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely vours. FOR
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use Statement
OCT 1 5 2010
| 510(k) Number (if known): | K093547 |
|---------------------------|-----------------------|
| Device Name: | PinPointe™ FootLaser™ |
| Indications for Use: | |
The PinPointe™ FootLaser™ and the delivery accessories that are used with them are intended for use in surgical procedures involving open, laparoscopic and endoscopic ablation, vaporization, excision, incision, and coagulation of soft tissue in the medical specialites of general and cosmetic dentistry, otolaryngology/ENT surgery, and dermatology & plastic surgery including intraoral soft tissue dental surgery, oral maxillo-facial and cosmetic surgery, general surgery, E.N.T. surgery, podiatry, and dermatology and plastic surgery.
### Podiatry
Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:
- Matrixectomy
- . Periungual and subungual warts
- Plantar warts ●
- Radical nail excision ●
- Neuromas .
The PinPointe™ FootLaser™ is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
*** Page 1 of 4 ***
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Ogden for mxm
Division Size 500
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K093547
Page 1 of 4
Premarket Notification, 510(k) Submission for: PinPointe™ FootLaser™
Section 4: Indications for Use
{8}------------------------------------------------
## Indications for Use Statement - Continued
510(k) Number (if known): K093547
| Device Name: | PinPointe™ FootLaser™ |
|--------------|-----------------------|
|--------------|-----------------------|
Indications for Use - Continued:
### Dermatology and Plastic Surgery
Dermatology and plastic surgery (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:
- Lesions of skin and subcutaneous tissue .
- Telangiectasia .
- Port wine lesions ●
- . Spider veins
- Hemangiomas ●
- Plantar warts ●
- . Periungual and subungual warts
- Removal of tattoos .
- Debridement of decubitus ulcer ●
- . Treatment of keloids
## Oropharangeal / Dental Surgery
Indicated for:
- Abscess incision and drainage .
- Aphthous ulcers treatment .
- Biopsies, excisional and incisional .
- Crown lengthening ●
- Exposure of unerupted / partially erupted teeth
- . Fibroma removal
- . Frenectomy
- Frenotomy .
- Gingival incision and excision ●
>
- Gingivectomy ●
- . Gingivoplasty
Prescription Use
(Part 21 CFR 801 Subpart D)
*** Page 2 of 4 ***
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number: K093547
**Page 2 of 4**
Premarket Notification, 510(k) Submission for: PinPointe™ FootLaserTM
Section 4: Indications for Use
{9}------------------------------------------------
## Indications for Use Statement - Continued
510(k) Number (if known): K093547
Device Name: PinPointe™ FootLaser™
Indications for Use - Continued:
### Oropharangeal / Dental Surgery - Continued
Indicated for:
- Hemostasis .
- Implant recovery
- . Lesion (tumor) removal
- Leukoplakia .
- . Operculectomy
- Oral papillectomy ●
- Pulpotomy ●
- Pulpotomy as adjunct to root canal therapy .
- Removal of filling material such as gutta percha or resin as adjunct treatment during root . canal re-treatment
- . Selective ablation of cnamel (first degree) caries removal
- Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal ● pocket) to improve clinical indices including gingival bleeding index, probe depth, attachment loss, and tooth mobility
- Tissue retraction for impressions
- Vestibuloplasty ●
## General Surgery
Indicated for:
- Open, laparoscopic, and endoscopic general surgery (ablation, vaporization, incision, . excision, and coagulation of soft tissue) including:
- Cholecystectomy -
- Lymphadenectomy .
- Mastectomy .
- Partial nephrectomy .
- Hepatectomy .
Prescription Use > (Part 21 CFR 801 Subpart D) *** Page 3 of 4 *** AND/OR
Over-The-Counter Use . (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
\$\frac{\text{Nid R.P.Q.den}}{\text{(Division Sigh-Off)}}\$ for MXM
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K093547
**Page 3 of 4**
Premarket Notification, 510(k) Submission for: PinPointe™ FootLaser™
Section 4: Indications for Use
{10}------------------------------------------------
### Indications for Use Statement - Continued
510(k) Number (if known): K093547
## Device Name: PinPointe™ FootLaser™
### Indications for Use - Continued:
## General Surgery - Continued
Indicated for:
- Pilonidal cystectomy -
- Pancreatectomy .
- Resection of lipoma .
- Splenectomy -
- Pelvic adhesiolysis -
- Hemorrhoidectomy -
- Removal of lesions -
- Thyroidectomy -
- Removal of polyps -
- Parathyroidectomy -
- Removal of tumors -
- Hemiorrhaphy .
- Tumor biopsy .
- Tonsillectomy .
- Debridement of decubitus ulcers .
- Appendectomy
### Endonasal Surgery
Endonasal surgery (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:
- Lesions or tumors of the oral, nasal, glossal, pharyngeal & laryngeal tissues ●
- . Tonsillectomy
- Adenoidectomy .
*** Page 4 of 4 ***
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | ✓ |
|-------------------------------------------------|---|
|-------------------------------------------------|---|
AND/OR
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
|------------------------------------------------|--|
|------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Ogden for mam
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K093547
**Page 4 of 4**
Premarket Notification, 510(k) Submission for: PinPointe™ FootLaser™
Section 4: Indications for Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.