K093499 · Lexington International, LLC · OAP · Nov 23, 2009 · Physical Medicine
Device Facts
Record ID
K093499
Device Name
HAIRMAX LASERCOMB
Applicant
Lexington International, LLC
Product Code
OAP · Physical Medicine
Decision Date
Nov 23, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HairMax LaserComb is indicated to promote hair growth in males, 30-60 years of age with androgenetic alopecia who have Norwood Hamilton Classifications of IIa to V patterns of hair loss and Fitzpatrick Skin Types I to IV.
Device Story
Hand-held low-level laser device; emits laser light to scalp to promote hair growth. Comb teeth part hair to ensure light reaches scalp. Operated by user at home. Modified version adds Lithium-ion battery power supply option to original wall-transformer-only design; includes cosmetic changes to housing color and dimensions. Device includes audible timer. User applies device to scalp; laser light exposure intended to stimulate hair follicles. Benefits include potential promotion of hair growth in specified male population.
Clinical Evidence
No new clinical data provided. Substantial equivalence based on identical technological characteristics and performance testing confirming compliance to design specifications.
Technological Characteristics
Low-level laser device; Class II infrared lamp (21 CFR 890.5500). Power source: wall transformer or Lithium-ion battery. Features comb teeth for scalp parting and audible timer. Plastic housing. No software-based algorithms.
Indications for Use
Indicated for males, 30-60 years old, with androgenetic alopecia (Norwood-Hamilton IIa-V) and Fitzpatrick Skin Types I-IV to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Lexington International, LLC. The word "Lexington" is in a bold, sans-serif font, with a curved line above it. Below "Lexington" is the text "International, LLC" in a smaller font. To the right of the text is a star.
K093499
Lexington International, LLC 777 Yamato Road, Suite 105 Boca Raton, FL. 33431
# 6. 510(k) Summary
Lexington International, LLC
HairMax LaserComb ver. 5.2
# NOV 2 3 2009
# Submitter's Contact Information
| Name: | David Michaels, Managing Director |
|---------------|-----------------------------------|
| Address: | Lexington International, LLC |
| | 777 Yamato Rd. Suite 105 |
| | Boca Raton, FL 33431 |
| Telephone: | (561) 314-2430 |
| Facsimile: | (561) 892-0747 |
| E-mail: | DM@HairMax.com |
| Manufacturer: | Lexington International, LLC |
Establishment Registration Number: 3006182775
## Official Correspondent
| Name: | Cherita James, Regulatory Consultant |
|------------|--------------------------------------|
| Address: | M Squared Associates, Inc. |
| | 901 King Street, Suite 200 |
| | Alexandria, VA 22314 |
| Telephone: | (703)562-9800 |
| Facsimile: | (703) 562-9797 |
| E-mail: | CJames@MSquaredAssociates.com |
Trade Name: HairMax LaserComb Common or Usual Name: Lamp, nonheating, for promotion of hair growth. Product Code: OAP
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for Lexington International, LLC. The word "Lexington" is in a bold, sans-serif font, with a horizontal line running through the middle of the word. Above the word "Lexington" is a curved line. Below the word "Lexington" is the text "International, LLC" in a smaller, sans-serif font. To the right of the word "Lexington" is a star.
ton International. LLC Yamato Road, Suite 105 Boca Raton, FL. 33431
Classification Name: 21 CFR 890.5500 Infrared lamp Predicate Device: HairMax LaserComb K060305
### November 9, 2009
### Device Description:
Similar to the HairMax LaserComb K060305, the modified HairMax LaserComb consists of a hand-held low level laser device that emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp.
The predicate unit operates from solely on a class 2, UL and CUL rated wall The modified unit adds a Lithium ion battery power supply option. In transformer. addition, cosmetic changes are being made to the device color and exterior plastic housing. Changes do not alter form fit or function as compared to the predicate device.
#### Intended Use / Indications for Use
The HairMax LaserComb is indicated to promote hair growth in males, 30-60 years of age with androgenetic alopecia who have Norwood Hamilton Classification IIa -V patterns of hair loss and Fitzpatrick Skin Types I to IV.
#### Technological Characteristics
The modifications to the HairMax LaserComb since its previous clearance in K060305 include changes to available power supply, and consequently the device casing dimensions and color. These minor differences do not affect the safety or performance of the device and do not change the intended use of the HairMax LaserComb.
#### Nonclinical Testing
Based on the Risk Analysis, the verification and validation tests that were performed and the acceptance criteria applied for each are listed in Section 10. The modified HairMax LaserComb was subject to the same preclinical requirements as the
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Lexington International, LLC. The word "Lexington" is in a bold, sans-serif font, with a line underneath it. Below that is the word "International, LLC" in a smaller, sans-serif font. There is a curved line above the word "Lexington" and a star to the right of the word.
aton International. LLC 777 Yamato Road, Suite 105 Boca Raton, FL. 33431
predicate device. Performance testing was conducted to confirm compliance to design specifications; all functions were verified to operate as designed.
## Substantial Equivalence
The modified HairMax LaserComb is as safe and effective as the predicate device. The modified device has the same intended use of affecting hair growth as the predicate device. The modified HairMax LaserComb has the same indications, i.e., treating androgenetic alopecia, and the same specific indication of promoting hair growth in males, 30-60 years of age with androgenetic alopecia who have Norwood-Hamilton Classifications of I to V patterns of hair loss and Fitzpatrick Skin types of I-IV as the predicate device.
The modified HairMax LaserComb is identical in technological characteristics as the device cleared in K060305, including its laser power, wavelength, laser delivery method, its comb component, its instructions for use and its audible timer. The cosmetic change to device color and dimensions of the exterior plastic housing and addition of a battery does not affect the HairMax LaserComb's fundamental scientific technology.
### Conclusion:
The modified HairMax LaserComb has the following similarities to the HairMax LaserComb:
- . has the same indicated use,
- uses the same operating principle, .
- incorporates the same basic device design and physical properties, .
- . incorporates the same materials.
Therefore the modification to the modified HairMax LaserComb can be found substantially equivalent to the HairMax LaserComb cleared in K060305.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Lexington International, LLC % M Squared Associates, Inc. Ms. Cherita James Regulatory Consultant 901 King Street, Suite 200 Alexandria, Virginia 22314
NOV 2 3 2009
Re: K093499
Trade/Device Name: HairMax LaserComb Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP, GEX Dated: November 10, 2009 Received: November 12, 2009
Dear Ms. James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it . may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Cherita James
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M Millerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Lexington International, LLC. The word "Lexington" is in a larger, bold font, with a line underneath it. Below that, the words "International, LLC" are in a smaller font. There is a curved line above the word "Lexington" and a star to the right of the logo.
Lexington International, LLC 777 Yamato Road, Suite 105 Boca Raton, FL. 33431
# 5. Indications for Use
510(k) Number (if known): Ko 9 3 499
Device Name: HairMax LaserComb
Indications for Use:
The HairMax LaserComb is indicated to promote hair growth in males, 30-60 years of age with androgenetic alopecia who have Norwood Hamilton Classifications of IIa to V patterns of hair loss and Fitzpatrick Skin Types I to IV.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark N. Mulhausen
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K093499
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.