COLLAWOUND ART

K093351 · Collamatrix, Inc. · FRO · Jun 16, 2010 · SU

Device Facts

Record IDK093351
Device NameCOLLAWOUND ART
ApplicantCollamatrix, Inc.
Product CodeFRO · SU
Decision DateJun 16, 2010
DecisionSESE
Submission TypeAbbreviated
Device ClassClass U
AttributesTherapeutic

Indications for Use

CollaWound ART is a wound dressing intended for use as a barrier to resist microbial colonization within the dressing and reduce microbial penetration through the dressing. CollaWound ART will be used for the management of partial and full thickness wounds, chronic vascular ulcers, ulcers, diabetic ulcers, ulcers, venous pressure tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, dehiscence), trauma wounds (abrasions, lacerations, post-laser surgery, wound second-degree bums, and skin tears) and draining wounds.

Device Story

CollaWound ART is an antimicrobial wound dressing composed of porcine-derived collagen and polyhexamethylene biguanide (PHMB). The device is applied topically to the wound bed by a healthcare provider. It functions by absorbing wound exudates to form a gel-like structure, maintaining a moist microenvironment that supports granulation tissue formation and re-epithelialization. The collagen matrix serves as a physical barrier, while the incorporated PHMB acts to reduce or prevent microbial colonization within the dressing and inhibit microbial penetration through the dressing. It is intended for various wound types, including chronic ulcers, surgical sites, and trauma wounds. The device provides a protective environment to facilitate healing and manage bioburden.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per FDA Blue Book G95-1 and ISO 10993, including cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, hemocompatibility, and pyrogenicity. Additional bench studies included release kinetics and antimicrobial effectiveness testing.

Technological Characteristics

Wound dressing composed of porcine-derived collagen and polyhexamethylene biguanide (PHMB). Functions as a physical barrier and antimicrobial agent. Biocompatibility verified per ISO 10993. No electronic components, software, or energy sources.

Indications for Use

Indicated for management of partial/full thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, skin tears), and draining wounds. Acts as a barrier to resist microbial colonization and penetration.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # COLLAMATRIX 510(k) summary Summary information #### 1. Date Prepared August 1, 2009 #### Submitter name and address 2. Collamatrix Inc. 1F, No.50-1, Keyan Road, Jhunan Science Park, Miaoli County, 350, Taiwan #### 3. Contact person | Name: | Dennis J. N. Seah | |-------|-------------------| | Tel: | + 886 2 7711 3699 | | Fax: | + 886 2 7711 3599 | #### Device names 4. | Propriety name: | CollaWound ART | |----------------------|------------------------------| | Common name: | Antimicrobial wound dressing | | Classification name: | Dressing, Wound | #### 5. Device classification | Regulatory class: | Unclassified | |-------------------|--------------| | Product code: | FRO | #### 6. Device description CollaWound ART is a wound dressing comprised of collagen and polyhexamethylene biguanide (PHMB). The collagen is prepared from pig skin. CollaWound ART absorbs wound exudates and forms gel-like structure that maintains a moist microenvironment at > 1F, No. 50-1, Keyan Road, Jhunan Science Park, Miaoli Country, 350, Taiwan Page 1 of 3 Tel: +886 2 7711 3699 Fax: +886 2 7711 3599 ### JUN 1 & 2010 K09335) page 1/3 {1}------------------------------------------------ K093351 page 2/3 ### COLLAMATRIX the wound bed and aids in granulation tissue formation and re-epithelialization. The dressing acts as a protective barrier to the invasion of foreign microorganisms. The PHMB content is intended to reduce or prevent microbial colonization or growth on the device. #### 7. Intended use CollaWound ART is a wound dressing intended for use as a barrier to resist microbial colonization within the dressing and reduce microbial penetration through the dressing. CollaWound ART will be used for the management of partial and full thickness wounds, chronic vascular ulcers, ulcers, diabetic ulcers, ulcers, venous pressure tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, dehiscence), trauma wounds (abrasions, lacerations, post-laser surgery, wound second-degree bums, and skin tears) and draining wounds. #### Statement of Substantial equivalence 8. CollaWound ART is a PHMB-containing wound dressing device similar to predicate PHMB-containing devices that are previously approved by the agency and allowed for marketing towards the management of wounds. ### Predicate devices are listed below | Trade name: | FortaDermTM Antimicrobial PHMB Wound Dressing coated with 0.1% | |-------------|----------------------------------------------------------------| | | Polyhexamethylene Biguanide Hydrochloride (PHMB) (K051647) | | Company: | Organogenesis Inc. | XCell Antimicrobial Cellulose Wound Dressing (K024054) Trade name: Xylos® Corporation Company: The proposed device, CollaWound ART, is another PHMB-containing wound dressing that is quite similar with respect to the indication for use, technological characteristics and material to the above devices in terms of the substantial equivalency under the 510(k) regulations. #### 9. Safety and effectiveness Biocompatibility tests have confirmed that CollaWound ART meets the requirements 1F, No. 50-1, Keyan Road, Jhunan Science Park, Miaoli Country, 350, Taiwan Page 2 of 3 Tel: +886 2 7711 3699 Fax: +886 2 7711 3599 {2}------------------------------------------------ K093351 page 3/3 ## COLLAMATRIX stated in the FDA Blue book memorandum G95-1 and ISO 10993. The following studies were conducted: - (a) Cytotoxicity - (b) Irritation test - (c) Sensitization - (d) Systemic toxicity test - (e) Genotoxicity - (f) Hemocompatibility/hemolysis - (g) Pyrogenic test - (h) Release kinetic - (i) Antimicrobial effectiveness study ### 10. Conclusion CollaWound ART is essentially equivalent in indication for use, technological characteristics, material, safety and effectiveness to the commercially available predicate device, and therefore meets the requirements as defined in 21 CFR § 807. Page 3 of 3 {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle or bird symbol, represented by a series of curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 ## JUN 1 8 2010 Collamatrix, Inc. % Mr. Dennis J.N. Seah 1F, No. 50-1, Keyan Road, Jhunan Science Park Miaoli County, 350 Taiwan Re: K093351 Trade/Device Name: CollaWound ART Regulatory Class: Unclassified Product Code: FRO Dated: May 10, 2010 Received: May 14, 2010 Dear Mr. Seah: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Mr. Dennis J.N. Seah If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Statement of indications for use 510(K) Number (if known): K093351 Device Name: CollaWound ART Indications for Use: CollaWound ART is a wound dressing intended for use as a barrier to resist microbial colonization within the dressing and reduce microbial penetration through the dressing. CollaWound ART is intended for the management of wounds including: - partial and full thickness wounds . - pressure ulcers . - venous ulcers . - diabetic ulcers . - . chronic vascular ulcers - tunneled/undermined wounds . - surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, . wound dehiscence) - trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) . - draining wounds . Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Keene for MXM (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K093351
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...