FORTADERM ANTIMICROBIAL WOUND DRESSING

K051647 · Organogenesis, Inc. · FRO · Nov 8, 2005 · SU

Device Facts

Record IDK051647
Device NameFORTADERM ANTIMICROBIAL WOUND DRESSING
ApplicantOrganogenesis, Inc.
Product CodeFRO · SU
Decision DateNov 8, 2005
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The FortaDerm™ Antimicrobial PHMB Wound Dressing is intended for the management of wounds and as an effective barrier to resist microbial colonization within the dressing and reduce microbes penetrating through the dressing. FortaDerm™ Antimicrobial PHMB Wound Dressing may be used for the management of: partial and full thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, wound dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) and draining wounds. The device is intended for one-time use.

Device Story

FortaDerm™ Antimicrobial PHMB Wound Dressing is a two-layer, fenestrated, crosslinked porcine intestinal collagen sheet coated with 0.1% Polyhexamethylene Biguanide hydrochloride (PHMB). Supplied dry in various sizes (4x4 cm to 12x36 cm) in sterile, single-use pouches. Used by clinicians for wound management; acts as a physical barrier to resist microbial colonization within the dressing and reduce microbial penetration. Benefits patients by providing a protective environment for wound healing while managing bioburden.

Clinical Evidence

Bench testing only. Device passed all biocompatibility and performance requirements.

Technological Characteristics

Two-layer, fenestrated, crosslinked porcine intestinal collagen sheet coated with 0.1% Polyhexamethylene Biguanide hydrochloride (PHMB). Supplied dry in sheet form. Sterile, single-use.

Indications for Use

Indicated for management of partial and full thickness wounds, including pressure, venous, diabetic, and chronic vascular ulcers; tunneled/undermined wounds; surgical wounds (donor sites/grafts, post-Moh's, post-laser, dehiscence); trauma wounds (abrasions, lacerations, second-degree burns, skin tears); and draining wounds.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K051647 page NOV - 8 2005 ### 510(K) SUMMARY ### Submitted by: Organogenesis Inc. 150 Dan Road Canton, Massachusetts 02021 ### Contact Patrick R. Bilbo Telephone: (781) 401-1155 Facsimile: --------------------------------------------------------------------------------------------(781) 575-1570 October 21, 2005 Date: ### Device: | Trade Name: | FortaDerm™ Antimicrobial PHMB Wound<br>Dressing coated with 0.1% Polyhexamethylene<br>Biguanide Hydrochloride (PHMB) | |----------------------|----------------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Topical Wound Dressing, Wound Management<br>Biomaterial | | Classification Name: | Dressing, Wound (KMF) | | Classification: | Unclassified | ### Predicate Devices: The relevant predicate devices are FortaDerm™ (PuraPly™) Wound Dressing (K011026) manufactured by Organogenesis, Inc. and Xcell Antimicrobial Cellulose Wound Dressing (K024054) manufactured by Xylos Corporation. ### Statement of Substantial Equivalence: The FortaDerm™ Antimicrobial PHMB Wound Dressing is similar with respect to intended use, technological characteristics, materials and physical construction to the predicate devices in terms of section 510(k) equivalency. ### Intended Use: The FortaDerm™ Antimicrobial PHMB Wound Dressing is intended for the management of wounds and as an effective barrier to resist microbial colonization within the dressing and reduce microbes penetrating through the dressing. FortaDerm™ Antimicrobial PHMB Wound Dressing may be used for the management of: partial and full thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, wound {2}------------------------------------------------ K051647 page 2/2 dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) and draining wounds. ### The device is intended for one-time use. ### Device Description: FortaDerm™ Antimicrobial PHMB Wound Dressing consists of two layers, crosslinked sheet of fenestrated sheet of porcine intestinal collagen coated with 0.1 % Polyhexamethylene Biguanide hydrochloride (PHMB). FortaDerm Wound Dressing is supplied dry in sheet form in sizes ranging from 4 x 4 cm to 12 x 36 cm. The device is packaged in sterile, sealed single pouches. ### Performance Data: FortaDerm™ Antimicrobial PHMB Wound Dressing was subjected to a number of tests to assess biocompatibility and performance. The device passed the requirements of all tests. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV - 8 2005 Patrick R. Bilbo, Vice President Regulatory and Clinical Affairs Organogenesis, Inc. 150 Dan Road Canton, Massachusetts 02021 Re: K051647 Trade/Device Name: FortaDerm™ Antimicrobial PHMB Wound Dressing Regulatory Class: Unclassified Product Code: FRO Dated: October 21, 2005 Received: October 25, 2005 Dear Mr. Bilbo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ ### Page 2- Patrick R. Bilbo, Vice President This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Barbara boehup for Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use # 510(k) Number (if known): KOSIC47 Device Name: FortaDerm™ Antimicrobial PHMB Wound Dressing Indications for Use: The FortaDerm™ Antimicrobial PHMB Wound Dressing is intended for the management of wounds and as an effective barrier to resist microbial colonization within FortaDerm™ the dressing and reduce microbes penetrating through the dressing. Antimicrobial PHMB Wound Dressing may be used for the management of: partial and full thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, wound dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) and draining wounds. The device is intended for one-time use. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) bare buem Page**of** (Division St Division of General, Restorative, and Neurological Devices 510(k) Number K051647 (Posted November 13, 2003)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...