Syntec Osteo-plate and screw Fixation is intended to be used on small bones or large bones for fracture fixation, including the fractures of clavicle, scapula, pelvis, long bones (humerus, ulna, radius, femur, tibia, and fibula), and small bones (metacarpals, metatarsals, and phalanges).
Device Story
Syntec Osteo-plate and screw Fixation consists of metallic bone plates and screws for fracture fixation. Plates range 3.8-17.5mm width, 15-319mm length, 2-22 holes. Screws range 2.4-7.3mm diameter, 6-180mm length. Includes self-tapping/self-drilling options and locking head screw/plate designs. Used by surgeons in clinical settings to fasten plates to bone or as lag screws to hold bone fragments together. Provides mechanical stabilization of fractures to facilitate healing.
Clinical Evidence
Bench testing only. Comparison of compression strength, pull-out strength, and torsional strength against predicate device demonstrated substantial equivalence.
Technological Characteristics
Materials: Stainless steel (ASTM F138) and Titanium alloy Ti-6Al-4V (ASTM F136). Mechanical fixation via plates and screws. Various sizes and configurations (locking/non-locking, self-tapping/self-drilling).
Indications for Use
Indicated for fracture fixation of small and large bones, including clavicle, scapula, pelvis, long bones (humerus, ulna, radius, femur, tibia, fibula), and small bones (metacarpals, metatarsals, phalanges).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Syntec-Taichung Non-sterile Bone Plate and Screw Implants (K983495)
Submission Summary (Full Text)
{0}------------------------------------------------
K092889
# '/ 3
JUN 2 5 2010
## 510(K) SUMMARY
Submitter's name : Address :
Contact person :
Common name:
Produce Code :
Predicate device:
Name of the device :
Classification name :
Regulation Number :
Trade or proprietary name :
Syntec Scientific Corporation 2, Kung San Rd, Chuan Shing Industrial Zone, Shen Kang, Chang Hua, Taiwan Tel : 886-4-7987099 Fax: 886-4-7987077 Carol Chang Syntec Osteo-plate and screw Fixation Syntec Osteo-plate and screw Fixation bone plate and bone screw Smooth or threaded metallic bone fixation fastener. HWC, HRS 888.3030, 888.3040 II Syntec-Taichung Non-sterile Bone Plate and Screw Implants (K983495) August 01, 2009
Preparation of the date:
#### Material:
Class :
The Syntec Osteo-plate and screw Fixation provides two materials, Stainless Steel and Titanium Alloy (Ti-6Al-4V) and commercially followed by ASTM F138, and ASTM F136.
#### Description of the Device:
Those are designed with various sizes of partially and fully threaded to satisfy with different kinds of fractures as using on small bones or large bones. The subject components are plates available in width from 3.8mm to 17.5mm, in total length from 15mm to 319mm and the hole numbers are 2 to 22 holes. The threads of the screws are available in diameters from 2.4mm to 7.3mm, and in length from 6mm to 180mm. Those
{1}------------------------------------------------
K092889 #2/3
screws have been designed for fitting on different symptoms of fractures and the plates have been designed to distribute for local anatomies and should be necessary to be used with its intended screws. Basically, the screws used to fasten plates onto bone, or, as lag screws, to hold together fragments of bone. We have totally the same kinds of the screws and plates as the predicate devices' functions, but with self-tapping or self-drilling to promote the operation efficiency. Besides this, we have another alternative screws and plates which are designed with locking head of the screws taken place with the traditional screw head, and threaded aside locking hole of the plates taken place with the aside plain hole of the plates.
#### Indications for Use:
Syntec Osteo-plate and screw Fixation is intended to be used on small bones or large bones for fracture fixation, including the fractures of clavicle, scapula, pelvis, long bones (humerus, ulna, radius, femur, tibia, and fibula), and small bones (metacarpals, metatarsals, and phalanges).
#### Technology Characteristics:
The design, materials and indications for use of the Osteo-plate and screw are equivalent to device previously approved for market in the United States.
| | Predicate Device | Applicant |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Sponsor | Syntec Scientific Corporation | Syntec Scientific Corporation |
| Device name | Syntec-Taichung Non-sterile<br>Bone Plate and Screw implants | Syntec Osteo-plate and screw<br>Fixation |
| 510(k) number | K983495 | - |
| Regulation number/<br>name/ class/ product<br>code | Class II §888.3030, 888.3040<br>Single/ Multiple component<br>metallic bone fixation<br>appliances and accessories/<br>Smooth or threaded metallic<br>bone fixation fastener<br>HRS, HWC | Class II §888.3030, 888.3040<br>Smooth or threaded metallic bone<br>fixation fastener<br>HRS, HWC |
| Intended Use | The bone plates and screws are<br>provided non-sterile. The | Syntec Osteo-plate and screw<br>Fixation is intended to be used on |
Device Comparison Table:
{2}------------------------------------------------
K092889 #3/3
| | devices are intended to treat<br>fractures of various bones,<br>including the clavicle, scapula,<br>pelvis, long bone (humerus,<br>ulna, radius, femur, tibia, and<br>fibula), and small bone<br>(metacarpals, metatarsals, and<br>phalanges). | mini, small bones or large bones<br>for fracture fixation, including the<br>fractures of clavicle, scapula,<br>pelvis, long bone (humerus, ulna,<br>radius, femur, tibia, and fibula),<br>and small bone (metacarpals,<br>metatarsals, and phalanges). |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Diameters of screws | From 1.5mm to 7.3mm | From 2.4mm to 7.3mm |
| Lengths of plates | From 17mm to 262mm | From 15mm to 319mm |
| Materials | Stainless steel and<br>titanium alloy (Ti-6Al-4V) | Stainless steel and<br>titanium alloy (Ti-6Al-4V) |
#### Performance Data:
The test results of the Compression strength of plates and screws, Pull-out strength of screws, and Torsional strength of screws were compared to the results of the Syntec-Taichung Non-sterile Bone Plate and Screw implants and demonstrated substantial equivalence.
### Conclusion:
The Syntec Osteo-plate and screw Fixation is no new issue of safety or effectiveness. The device does not additional concerns regarding safety and effectiveness and may, therefore, be considered substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of flowing lines.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
## JUN 2 6 2010
Syntec Scientific Corporation % Ms. Carol Chang 3F1. 96 Chung Hsiao E. Rd. Sec. 3 Taipei China (Taiwan) 106
Re: K092889
Trade/Device Name: Syntec Osteo-plate and screw Fixation Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: June 10, 2010 Received: June 15, 2010
Dear Ms. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medica! device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Ms. Carol Chang
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melke Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATION FOR USE
# Indications for Use
## 510(k) Number (if known): K092889 Device Name: Syntec Osteo-plate and screw Fixation Indications for Use:
Syntec Osteo-plate and screw Fixation is intended to be used on small bones or large bones for fracture fixation, including the fractures of clavicle, scapula, pelvis, long bones (humerus, ulna, radius, femur, tibia, and fibula), and small bones (metacarpals, metatarsals, and phalanges).
Prescription Use _____________________________ AND/OR Over-The-Counter Use_ ( Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Smuta for mxin
Division Sign-Q Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K092889
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.