The Gyrus ACMI® Flexible Endoscope Storage-Sterilization Trays are intended to be used to enclose and protect Gyrus ACMI flexible endoscopes during sterilization. The trays are to be used in conjunction with an FDA cleared sterilization wrap. The trays are optional accessories to the Gyrus ACMI endoscopes for which they are designed. Maintenance of sterility depends on the sterilization wrap, not the tray. The trays are indicated for ETO sterilization of only of the following : Gyrus ACMI Flexible Endoscopes ICN-0564, ICN-0565 ICN DUR-D DUR-D, DUR-DBA ACN ACN-2T, ACN-2TBA DUR-8 DUR-8, DUR-8-BA, DUR-8E, DUR-8E-BA,DUR-8 Ultra AUR AUR AUR-7. AUR-735
Device Story
Plastic storage-sterilization trays designed to house and protect Gyrus ACMI flexible endoscopes during ethylene oxide (EtO) sterilization and subsequent storage. Trays feature large holes (approx. 7mm) to facilitate gas ingress/egress. Used in clinical settings by healthcare staff; trays are wrapped in FDA-cleared sterilization wrap prior to processing. Trays do not contact patients. Protects endoscopes from physical damage; sterility maintenance relies on the external wrap. Validated for 100% EtO sterilization cycles.
Clinical Evidence
Bench testing only. PCDs inoculated with biological indicators placed in trays, wrapped, and processed in 30-minute half-cycles. Demonstrated 6-log reduction (SAL 10^-6). EtO residual testing performed after 60-minute full cycles with 12-hour aeration; results within ISO 10993-7 limits.
Technological Characteristics
Constructed of biocompatible RADEL-R (polyphenylsulfone). Dimensions feature large 7mm diameter holes for gas exchange. Compatible with 100% EtO sterilization. Biocompatibility per ISO 10993. Non-patient contacting.
Indications for Use
Indicated for ETO sterilization of specific Gyrus ACMI flexible endoscopes (ICN, DUR-D, ACN, DUR-8, and AUR series) to provide physical protection during sterilization and storage. Not for patient contact.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Symmetry Medical Poly Vac Instrument Delivery System (K040223)
{0}------------------------------------------------
Kogalasa
# DEC 1 1 2009
- --
### 510(k) Summary of Safety and Effectiveness Gyrus ACMI Inc. Gyrus ACMI Flexible Endoscope Storage-Sterilization Trays
### General Information
Manufacturer:
Contact Person:
Gyrus ACMI Inc. 136 Turnpike Rd. Southborough, MA 01772-2104
Lorraine Calzetta Regulatory Affairs Tel. #: 508-804-2752 Fax #: 508-804-2624
December 3, 2009,
Date Prepared:
### Device Description
Classification Name: Sterilization, wrap, containers and trays (21CFR 880.6850), Class II
or Accessories to Endoscope (21CFR876.1500) Class II
Trade Name:
Gyrus ACMI Flexible Endoscope Storage-Sterilization Trays
Generic/Common Name:
Endoscope and accessories
### Predicate Devices
Symmetry Medical Poly Vac Instrument Delivery System Olympus Sterilization Trays
K040223 K033222
### Intended Uses
The Gyrus ACMI® Flexible Endoscope Storage-Sterilization Trays are intended to be used to enclose and protect Gyrus ACMI flexible endoscopes during sterilization. The trays are to be used in conjunction with an FDA cleared sterilization wrap. The trays are optional accessories to the Gyrus ACMI endoscopes for which they are designed. Maintenance of sterility depends on the sterilization wrap, not the tray.
The trays are indicated for ETO sterilization of only of the following :
Gyrus ACMI Flexible Endoscopes ICN-0564, ICN-0565 ICN DUR-D DUR-D, DUR-DBA
{1}------------------------------------------------
7
#### ACN ACN-2T, ACN-2TBA DUR-8 DUR-8, DUR-8-BA, DUR-8E, DUR-8E-BA,DUR-8 Ultra AUR AUR AUR-7. AUR-735
EtO Sterilization parameters:
EtO Sterilize using 100% ethylene oxide
Temp: 55°C, Vacuum 97mmHgV, Preconditioning time 60 minutes, EtO concentration 725-750 mg/L, Exposure time 60 minutes, Humidity 35%-80%, Aeration: 12 hours @55 ℃
### Product Description
The Gyrus ACMI® Flexible Endoscope Storage-Sterilization trays are comprised of plastic lids and bottoms that contain numerous large holes (approximately 7mm in diameter) that permit ready ingress and egress of sterilization gases. The trays are designed to provide protection from physical damage to the flexible endoscope during sterilization and storage. The trays are constructed of biocompatible RADEL-R. The trays do not contact the patient. Radel-R is a polyphenylsulfone plastic that is widely used in medical devices. Radel-R meets the requirements for biocompatibility pursuant to ISO-10993 and is compatible with EtO sterilization.
### Performance Data
The PCDs (process challenge devices) were placed in the trays and inoculated with FDA cleared biological indicator organisms, and chemical indicators were placed. The trays were wrapped with two layers of FDA approved sterilization wrap and placed into ethylene oxide sterilizer for processing. The system was sterilized successfully in. a 30 minute half cycle demonstrating 6 log reduction capability ( SAL of 10-6). Test systems were exposed to 60 minute full cycles and ethylene oxide residual testing was performed, with a 12 hour aeration time. Pursuant to ISO10993-7, residual concentrations were within acceptable limits.
### Technological Characteristics and Substantial Equivalence
The Gyrus ACMI® Endoscope Storage-Sterilization Trays are composed of the same naterials and utilize similar features as that of the predicates.
In summary, the Gyrus ACMI Endoscope Storage Sterilization Trays are substantially equivalent to the predicate device and presents no new questions of safety or efficacy.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
## DEC 1 1 2009
Ms. Lorraine Calzetta Regulatory Affairs Gyrus ACMI, Incorporated 136 Turnpike Road Southborough, Massachusetts 01772
Re: K092682
Trade/Device Name: Gyrus ACM1® Flexible Endoscope Storage-Sterilization Trays Regulation Number: 21CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: October 21, 2009 Received: November 6, 2009
Dear Ms. Calzetta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Ms. Calzetta
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours
Susan Tanner
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Traditional 510(k) Notitication K092682
Device Name: Gyrus ACMI® Flexible Endoscope Storage-Sterilization Trays
510(k) Number: K O 9 2682
Device Name: Gyrus ACMI® Flexible Endoscope Storage-Sterilization Trays
Indications for use:
The Gyrus ACMI® Flexible Endoscope Storage-Sterilization Trays are intended to be used to encloseand protect Gyrus ACMI flexible endoscopes during sterilization. The trays are to be used in conjunction with an FDA cleared sterilization wrap. The trays are optional accessories to the Gyrus ACMI endoscopes for which they are designed. Maintenance of sterility depends on the sterilization wrap, not the tray.
The trays are indicated for ETO sterilization of only the following :
| | Gyrus ACMI Flexible Endoscopes | |
|-------|-------------------------------------------------|--|
| ICN | ICN-0564, ICN-0565 | |
| DUR-D | DUR-D, DUR-DBA | |
| ACN | ACN-2T, ACN-2TBA | |
| DUR-8 | DUR-8, DUR-8-BA, DUR-8E, DUR-8E-BA, DUR-8 Ultra | |
| AUR | AUR-7, AUR-735 | |
EtO Sterilization parameters:
EtO Sterilize using 100% ethylene oxide,
Temp: 55°C, Vacuum 97mmHgV, Preconditioning time 60 minutes, EtO concentration 725-750 mg/L, Exposure time 60 minutes, Humidity 35%-80%, Aeration: 12 hours @55 ℃
Prescription Use: (Per 21 CFR 801.109)
Over-the-Counter Use: OR
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth F. (Laurie-Wells)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
PAGE. I OF 1
4
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.