K033222 · Olympus Winter & Ibe GmbH · KCT · Aug 2, 2004 · General Hospital
Device Facts
Record ID
K033222
Device Name
OLYMPUS STERILIZATION TRAYS
Applicant
Olympus Winter & Ibe GmbH
Product Code
KCT · General Hospital
Decision Date
Aug 2, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Olympus sterilization trays are intended to be used to enclose Olympus medical devices including hand instruments, trocars, camera heads, adapter, and endoscopes to be sterilized by a health care provider. It is intended to allow steam sterilization of the enclosed medical device. The qualification for sterilizing endoscopic instruments were validated with a process challenge device (PCD) by a GLP approved laboratory. The PCD has a diameter of 2 mm and a length of 1500 mm. The Olympus sterilization trays are to be sterilized in the following cycle: Prevacuum Steam: 132°C - 134°C - 5 minutes minimum. Drying Time: 10 – 20 minutes as needed.
Device Story
Sterilization trays designed to house Olympus medical instruments during steam sterilization; perforated thermoplastic base and lid with stainless steel locking tabs; includes silicone mats for instrument fixation. Used in clinical settings by healthcare providers. Trays allow steam penetration to enclosed instruments; must be used with approved sterilization wrap to maintain microbial barrier. Validated for prevacuum steam sterilization cycles (132°C–134°C). Benefits include organized instrument handling and protection during the sterilization process.
Clinical Evidence
Bench testing only. Validation performed using a process challenge device (PCD) with 2 mm diameter and 1500 mm length in a GLP-approved laboratory to confirm steam sterilization efficacy. Material biocompatibility and compatibility with prevacuum steam sterilization cycles (132°C-134°C) were verified.
Technological Characteristics
Materials: Thermoplastic RADEL, medical grade stainless steel (handles/locking tabs), silicone (mats). Design: Perforated base and lid for aeration. Sterilization: Prevacuum steam (132°C-134°C). Connectivity: None. Form factor: Various tray dimensions for specific instrument sets.
Indications for Use
Indicated for use by healthcare providers to enclose Olympus medical devices (hand instruments, trocars, camera heads, adapters, and endoscopes) during steam sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
## 033222
## AUG - 2 2004
## 510(k) Summary Olympus Sterilization Trays
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92
A. Submitter's Name, Address, Phone, Contact Person and Summary Date
| Manufacturer of subject device: | Olympus Winter & Ibe<br>Kuehnstr. 61<br>D-22045 Hamburg<br>Germany |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration number: | 8010313 |
| Contact Person: | Tina Steffanie-Oak<br>Olympus America Inc.<br>Two Corporate Center Drive<br>Melville NY 11747-3157<br>Registration Number: 2429304<br>Telephone: (631) 844-5477 |
| B. Device Name | |
| Device Name: | Olympus sterilization trays |
| Common/Usual Name: | Sterilization container |
| Classification Name: | Sterilization container |
| Classification: | 21 CFR 882.1480<br>Class II |
## C. Predicate Devices:
PolyVac Delivery Systems (K012105)
D. Device Description
The Sterilization trays, which are the subject of this premarket notification, are as follows:
{1}------------------------------------------------
| Item no. | Outer dimensions<br>WxHxD[mm] | Detailed product description |
|-----------|-------------------------------|----------------------------------------------------------------------------|
| A5938* | 595 x 61 x 295 | Instrument tray, with lid, for video telescope |
| A5951* | 478 x 41 x 224 | Insert tray, for 2.7 and 4.0 mm TruView arthroscopes |
| A5952* | 478 x 41 x 224 | Insert tray, for Diver-Line hand instruments |
| A5961* | 665 x 52 x 204 | Insert tray, for ureteroscopes, with lid |
| A5975* | 532 x 63 x 128 | Instrument tray, small, with lid |
| A5976* | 478 x 68 x 224 | Insert tray, for urology |
| A5977* | 478 x 68 x 224 | Insert tray, for 4 TroQ trocars |
| A5984* | Set: A5994 and WA05974 | Instrument tray, for camera head and adapter |
| A5992* | 478 x 54 x 224 | Instrument tray, for A3336A, A2027A and A5226A |
| A5994* | 478 x 68 x 224 | Insert tray, for camera head and adapter |
| A05951A* | 478 x 41 x 224 | Insert tray, for TrueView II and mini TrueView telescopes |
| A05952A* | 665 x 65 x 204 | Instrument tray, for SonoSurg instruments, with lid |
| A05961A* | 405 x 95 x 200 | Instrument tray, for compact cystoscope and compact hysteroscope, with lid |
| WA05970A | 537 x 139 x 268 | Instrument tray, with lid and silicone mat |
| WA05971A | 482 x 60 x 224 | Insert tray, for the upper part of WA05970A, with silicone mat |
| WA05972A | 480 x 47 x 224 | Insert tray, for 5 HiQ shafts and jaw inserts |
| WA05973A | 460 x 39 x 207 | Insert tray, for the bottom of WA05970A, with silicone mat |
| WA05974A | 488 x 31 x 235 | Lid, for insert trays |
| WA05980A | 438 x 176 | Spare mat, nubbly, silicone, for WA05970A and WA05971A |
| WA05981A | 400 x 175 | Spare mat, nubbly, silicone, for WA05973A |
| WA05990A | 446 x 88 x 49 | Instrument tray, for two telescopes, with lid |
| WA05994A* | 478 x 68 x 224 | Insert tray, for camera head and adapter |
| WA05995A* | 478 x 68 x 224 | Insert tray, pediatric urology |
| WA05980A* | 595 x 61 x 295 | Instrument tray, for video telescope |
The Olympus sterilization trays consist of perforated bottoms with or without perforated lids. The material of the bottom is thermoplastic, the material of the lid is thermoplastic with handles and locking tabs made from medical grade stainless steel. The silicone mats for fixture of the instruments are perforated for aeration.
The stripes are designed for minimal contact area with instruments and secure fixing.
{2}------------------------------------------------
Olympus sterilization trays will be sold non-sterile and can be reused after proper cleaning and sterilization.
E. Intended Use of Device
The Olympus sterilization trays are intended to be used to enclose Olympus medical devices including hand instruments, trocars, camera heads, adapter, and endoscopes to be sterilized by a health care provider. It is intended to allow steam sterilization of the enclosed medical device.
The qualification for sterilizing endoscopic instruments were validated with a process challenge device (PCD) by a GLP approved laboratory. The PCD has a diameter of 2 mm and a length of 1500 mm.
The Olympus sterilization trays are to be sterilized in the following cycle:
Prevacuum Steam:
Drying Time:
Comparison to Predicate Devices
| | Subject | Predicate |
|-------------------------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------|
| | Olympus Sterilization<br>Trays | Poly Vac's Delivery<br>Systems |
| ELEMENT | | |
| Intended Use: to enclose<br>medical devices during steam sterilization. | Yes | Yes |
| Intended to be reused | Yes | Yes |
| Material Composition | Thermoplastic RADEL® | Thermoplastic RADEL® |
| Consist of base with lid, lid can be fastened<br>to the base by a locking tab | Yes | Yes |
| Usable with silicone mats or special silicone<br>holders | Yes | Yes |
| PERFORMANCE | | |
| Sterilant Pentration for Steam with<br>Prevacuum validated | Yes | Yes |
| Microbial Barrier Properties | To be used with approved<br>sterilization wrap | To be used with approved<br>sterilization wrap |
| Material Compatibility | Compatible to Steam<br>Sterilization at 134°C | Compatible to Steam<br>Sterilization at 132°C |
| Toxicological Properties | Material biocompatible | Material biocompatible |
| Shelf Life | depends on wrapping<br>material and storage<br>conditions. | n.a. |
| Drying time | 10 - 20 minutes | 20 - 40 minutes |
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 2 2004
Olympus Winter & Ibe GmbH C/O Ms. Tina Steffanie-Oak Olympus America, Incorporated Two Corporate Center Drive Melville, New York 11747-3157
Re: K033222
Trade/Device Name: Olympus Sterilization Trays Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: May 20, 2004 Received: May 21, 2004
Dear Ms. Steffanie-Oak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Steffanie-Oak
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements incan that I DAT has mace a development regulations administered by other Federal agencies. 01 the Act of any 1 ederal states and equirements, including, but not limited to: registration 1 ou inust comply was 807); labeling (21 CFR Part 801); good manufacturing practice alle listing (21 CF Part 801), and 1001), and many (QS) regulation (21 CFR Part 820); and if requirements as set form in the quality significal (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anon you to objan middle substantial equivalence of your device to a premarket notification. - The PDF results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), if you desire specific acroof Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may overn of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Clus
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K033222 Olympus Sterilization Trays Device Name: Indications for Use:
The Olympus sterilization trays are intended to be used to enclose Olympus medical devices including hand instruments, trocars, camera heads, adapter, and endoscopes to be sterilized by a health care provider. It is intended to allow steam sterilization of the enclosed medical device.
The qualification for sterilizing endoscopic instruments were validated with a process challenge device (PCD) by a GLP approved laboratory. The PCD has a diameter of 2 mm and a length of 1500 mm.
The Olympus sterilization trays are to be sterilized in the following cycle:
| Prevacuum Steam: | 132°C - 134°C - 5 minutes minimum. |
|------------------|------------------------------------|
| Drying Time: | 10 – 20 minutes as needed. |
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Re Melby
esthesiology, General Hospital,
510(k) Number: K033222
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.