K092414 · Conmed Corporation · KCT · Mar 10, 2010 · General Hospital
Device Facts
Record ID
K092414
Device Name
CONMED DETACHATIP INSTRUMENT TRAY, 33CM AND 43CM
Applicant
Conmed Corporation
Product Code
KCT · General Hospital
Decision Date
Mar 10, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The ConMed™ DetachaTip® Instrument Trays are perforated containment devices for medical device sterilization. ConMed" DetachaTip® Instrument Trays are a family of containment devices used to conveniently organize, sterilize, and transport Detachatip instruments between uses. The ConMed instrument trays are not intended to maintain sterility. We do not recommend that the trays be used to store sterilized contents. The 33cm DetachaTip Instrument Tray (1-1027) is designed for use with ConMed's 33cm DetachaTip multiuse scissors, DetachaTip multiuse graspers, DetachaTip multiuse dissectors and DetachaTip handles. The trays are intended to be used specifically with the following: [List of 33cm instruments]. The 43cm DetachaTip Instrument Tray (1-4327) are designed for use only with The 15cm DetachaTip multiuse scissors, DetachaTip multiuse graspers, Convice 3 +20m Detacha is is and DetachaTip handles. The trays are intended to be used specifically with the following: [List of 43cm instruments].
Device Story
Perforated containment trays for organizing, sterilizing, and transporting ConMed DetachaTip surgical instruments; used in clinical settings by healthcare staff. Trays feature Radel R Polyphenylsulfone base/lid, silicone mat, and stainless steel handles. Instruments are placed in tray; lid locked; tray subjected to prevacuum steam sterilization (132°C, 4 min exposure, 20 min dry). Perforations facilitate steam penetration. Device is reusable; not intended for long-term storage of sterile contents. Benefits include organized instrument handling and effective sterilization processing.
Clinical Evidence
Bench testing only. Validation performed using biological indicators and thermocouples to confirm sterilization efficacy and drying performance during prevacuum steam sterilization cycles (132°C, 4-minute exposure, 20-minute dry time).
Technological Characteristics
Materials: Radel R Polyphenylsulfone (base, tray, lid), silicone rubber (mat), stainless steel (handles). Design: Perforated lid/base for steam penetration; locking lid mechanism. Dimensions: 33cm and 43cm tray variants. Sterilization: Prevacuum steam (132°C).
Indications for Use
Indicated for use as a containment device to organize, sterilize, and transport specific ConMed DetachaTip multiuse surgical instruments (scissors, graspers, dissectors, handles) in 33cm and 43cm lengths. Not intended for maintaining sterility or storage of sterilized contents.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Paragon Medical Surgical Instrument Delivery Tray (K032119)
Submission Summary (Full Text)
{0}------------------------------------------------
K09241.14
Image /page/0/Picture/1 description: The image shows the logo for Conmed Corporation. The logo consists of a stylized "C" shape in a black square on the left, followed by the word "CONMED" in large, bold letters. Below the word "CONMED" is the word "CORPORATION" in smaller letters. There is a registered trademark symbol next to the "D" in "CONMED".
MAR 1 0 2010
# 510(k) Summary of Safety and Effectiveness
# ConMed™ Detachatip® Instrument Trays
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for the 510(k) Number
# A. Submitter
ConMed Corporation 525 French Road Utica, NY 13502 Registration Number: 1320894 Date: August 6, 2009
# B. Company Contact
Sandy Coveleski Regulatory Affairs Specialist ConMed Corporation 525 French Road Utica, NY 13502
Phone: 315-624-3435 Fax: 315-624-3225 e-mail: sandy_coveleski@mail.conmed.com
### C. Device Name
| Trade Name: | ConMed™ Detachatip® Instrument Trays |
|------------------------|---------------------------------------------------------------------------|
| Common Name: | Instrument Sterilization Tray |
| Classification Name: | Sterilization Wrap Containers, Trays, Cassettes,<br>and other Accessories |
| Proposed Class/Device: | Class II |
| Product Code: | KCT |
| Regulation Number: | 21 CFR 880.6850 |
| Panel: | 880 General Hospital |
{1}------------------------------------------------
### D. Predicate Device
Paragon Medical Surgical Instrument Delivery Tray Paragon Medical 510(K) # K032119
## E. Intended Use
The ConMed™ DetachaTip® Instrument Trays are perforated containment devices for medical device sterilization. ConMed" DetachaTip® Instrument Trays are a family of containment devices used to conveniently organize, sterilize, and transport Detachatip instruments between uses. The ConMed instrument trays are not intended to maintain sterility. We do not recommend that the trays be used to store sterilized contents.
The 33cm DetachaTip Instrument Tray (1-1027) is designed for use with ConMed's 33cm DetachaTip multiuse scissors, DetachaTip multiuse graspers, DetachaTip multiuse dissectors and DetachaTip handles. The trays are intended to be used specifically with the following:
| CAT. NO. | DESCRIPTION |
|----------|-----------------------------------|
| 2-1003 | METZENBAUM, 33CM LENGTH |
| 2-1004 | MINI-METZENBAUM, 33CM LENGTH |
| 2-1005 | 5MM BABCOCK GRASPER, 33CM LENGTH |
| 2-1008 | FENESTRATED GRASPER, 33CM LENGTH |
| 2-1009 | CURVED DISSECTOR, 33CM LENGTH |
| 1-1010 | STANDARD HANDLE |
| 2-1013 | HOOK, 33CM LENGTH |
| 2-1014 | 10MM BABCOCK, 33CM LENGTH |
| 1-1015 | COMPACT HANDLE |
| 2-1017 | RIGHT ANGLE MEEKER DISSECTOR 33CM |
| 2-1018 | TAPERED DISSECTOR, 33CM LENGTH |
| 2-1019 | ALLIS GRASPER, 33CM LENGTH |
| 1-1024 | IN-LINE HANDLE |
| 1-1028 | ENDOWEAVE GRASPER, 33 CM LENGTH |
The 43cm DetachaTip Instrument Tray (1-4327) are designed for use only with The 15cm DetachaTip multiuse scissors, DetachaTip multiuse graspers, Convice 3 +20m Detacha is is and DetachaTip handles. The trays are intended to be used specifically with the following:
{2}------------------------------------------------
| CAT. NO. | DESCRIPTION |
|----------|-----------------------------------|
| 2-4301 | METZENBAUM, 43CM LENGTH |
| 2-4304 | MINI-METZENBAUM, 43CM LENGTH |
| 2-4305 | 5MM BABCOCK GRASPER, 43CM |
| 2-4308 | CURVED DISSECTOR, 43CM LENGTH |
| 2-4307 | FENESTRATED GRASPER, 43CM LENGTH |
| 2-4314 | 10MM BABCOCK, 43CM |
| 2-4317 | RIGHT ANGLE MEEKER DISSECTOR 43CM |
| 2-4318 | TAPERED DISSECTOR, 43CM |
| 2-4319 | ALLIS GRASPER, 43CM |
| 2-4328 | ENDOWEAVE GRASPER, 43CM |
Materials The ConMed™ DETACHATIP® INSTRUMENT TRAYS consist of a Radel R Polyphenylsulfone base, a Radel R Polyphenylsulfone tray, a silicone rubber mat, a Radel R Polyphenylsulfone lid with tray lid clips with perforations to facilitate steam penetration, and stainless steel carrying handles. The tray holds the ConMed™ DETACHATIP® surgical instruments before, during, and after the sterilization process. The tray set has a locking lid to contain the instruments.
Sterilant Penetration The the ConMed™ DETACHATIP® INSTRUMENT TRAYS have been validated to perform effectively during prevacuum steam sterilization and drying cycles, using biological indicators and thermocouples to support sterilization and drying processes.
#### Pre-vacuum
Min temperature =132°C Min. exposure = 4 minutes Min. dry time = 20 minutes
CAUTION: Testing demonstrates that a minimum dry time of 20 minutes is required to prevent wet packs when using the prevacuum cycle.
Note: Validation was conducted using wrapped trays. The device should Note: Vanadion was other with FDA cleared wrap indicated for these sterilization cycles.
SheIf Life The sterilization tray is reusable and will not be serviced or repaired.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR 1 0 2010
Ms. Sandy Coveleski Regulatory Affairs Specialist ConMed Corporation 525 French Road Utica, New York 13502
Re: K092414
Trade/Device Name: ConMedTM DetachaTip® Instrument Trays 33cm DetachaTip Instrument Tray (1-1027) 43cm Detachatip Instrument Tray (1-4327) Regulation Number: 21CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: February 16, 2010 Received: February 17, 2010
Dear Ms. Coveleski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2- Ms. Coveleski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm ! 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours
Susan Runser
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
| 510(k) Number (if known): | | 092414 | | | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--------------------------------------------------------------------------------------------------|--|--|--|--|--|--|
| | | | | | | | | |
| A Property Company Company Company Company Company Company Company Comments of Children Comments of Children Comments of Children Comments of Children Comments of Children Co | | C 11 1 M . 1 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . | | | | | | |
| Family Name: | ConMed" DetachaTip® Instrument Trays |
|--------------|------------------------------------------|
| Trade Name: | 33cm DetachaTip Instrument Tray (1-1027) |
| | 43cm DetachaTip Instrument Tray (1-4327) |
### Indications for Use
The ConMed" DetachaTip® Instrument Trays are perforated containment devices for medical device sterilization. ConMed" DetachaTip® Instrument Trays are a family of containment devices used to conveniently organize, sterilize, and transport Detachatip instruments between uses. The ConMed instrument trays are not intended to maintain sterility. We do not recommend that the trays be used to store sterilized contents.
The 33cm DetachaTip Instrument Tray (1-1027) is designed for use with ConMed's 33cm DetachaTip multiuse scissors, DetachaTip multiuse graspers, DetachaTip multiuse dissectors and DetachaTip handles. The trays are intended to be used specifically with the following:
| Cat. No. | Description |
|----------|-----------------------------------|
| 2-1003 | Metzenbaum, 33cm length |
| 2-1004 | Mini- Metzenbaum, 33cm length |
| 2-1005 | 5mm Babcock Grasper, 33cm length |
| 2-1008 | Fenestrated Grasper, 33cm length |
| 2-1009 | Curved Dissector, 33cm length |
| 1-1010 | Standard Handle |
| 2-1013 | Hook, 33cm length |
| 2-1014 | 10mm Babcock, 33cm length |
| 1-1015 | Compact Handle |
| 2-1017 | Right Angle Meeker dissector 33cm |
| 2-1018 | Tapered Dissector, 33cm length |
| 2-1019 | Allis Grasper, 33cm length |
| 1-1024 | In-line Handle |
| 1-1028 | Endoweave Grasper, 33 cm length |
The 43cm DetachaTip Instrument Tray (1-4327) are designed for use only with ConMed's 43cm DetachaTip multiuse scissors, DetachaTip multiuse graspers, DetachaTip multiuse dissectors and DetachaTip handles. The trays are intended to be used specifically with the following:
{6}------------------------------------------------
## INDICATIONS FOR USE
| Cat. No. | Description |
|----------|-----------------------------------|
| 2-4301 | Metzenbaum, 43cm length |
| 2-4304 | Mini- Metzenbaum, 43cm length |
| 2-4305 | 5mm Babcock Grasper, 43cm |
| 2-4308 | Curved Dissector, 43cm length |
| 2-4307 | Fenestrated Grasper, 43cm length |
| 2-4314 | 10mm Babcock, 43cm |
| 2-4317 | Right Angle Meeker dissector 43cm |
| 2-4318 | Tapered Dissector, 43cm |
| 2-4319 | Allis Grasper, 43cm |
| 2-4328 | Endoweave Grasper, 43cm |
Sterilize the 33cm DetachaTip Instrument Tray (1-1027) and the 43cm DetachaTip Instrument Tray (1-4327) using the following parameters:
| Method | Cycle | Temperature | Exposure Time | Dry Cycle Time |
|-----------------|------------|--------------|---------------|----------------|
| Steam (wrapped) | Pre-vacuum | 270°F(132°C) | 4 minutes | 20 minutes |
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-the-Counter Use__X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE If NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Eshlata T. Queda-Wells
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K092414
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.