TTS100 PORTABLE HYPERTHERMIC PERFUSION DEVICE

K092366 · Thermal Therapeutic Systems, Inc. · LGZ · Mar 4, 2010 · General Hospital

Device Facts

Record IDK092366
Device NameTTS100 PORTABLE HYPERTHERMIC PERFUSION DEVICE
ApplicantThermal Therapeutic Systems, Inc.
Product CodeLGZ · General Hospital
Decision DateMar 4, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the TTS100 Portable Hyperthermic Perfusion Device is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating, warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician.

Device Story

The TTS100 is a prescription-use portable hyperthermic perfusion device consisting of a durable console and a sterile, single-use Hyperthermic Lavage Tubing Set. The console features a heater, roller-type fluid pump, microprocessor, and touchscreen interface. The disposable set includes a fluid reservoir, inlet/outlet tubing, and auxiliary temperature sensors; portions of the accessory kit may be implanted during lavage. The device circulates sterile fluid through the cavity; fluid is heated in the reservoir and pumped to the patient, returning via gravity. The system monitors fluid temperature and pressure, providing real-time data and alarms on the touchscreen. Software automatically responds to unsafe conditions or sensor failures, while independent protection circuits ensure safety during software faults. Used in clinical settings, the device allows physicians to control cavity temperature, potentially benefiting patients by maintaining therapeutic hyperthermia. The closed-loop disposable design prevents fluid contact with console components, facilitating rapid post-procedure processing.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including software verification/validation, biocompatibility testing (ISO 10993), electrical safety/EMC testing (IEC 60601), and human factors studies to mitigate user-related risks.

Technological Characteristics

Durable console with heater, roller pump, and touchscreen; sterile single-use tubing set. Materials comply with ISO 10993. Connectivity includes sensor monitoring of fluid path. Sterilization via ethylene oxide (ISO 11135). Software-controlled monitoring of temperature and pressure with independent hardware protection circuits. Complies with IEC 60601-1 and IEC 60601-1-2.

Indications for Use

Indicated for patients requiring thoracic or peritoneal cavity lavage with warmed, physiologically compatible sterile solution to achieve a target temperature, as determined by a physician-selected protocol.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## MAR - 4 2010 ThermalTherapeuticSystems, Inc. POB 82670 Pittsburgh, PA 15218 Phone: 412.571.2622 www.thermaltherapeutics.com ## 510(k) Summary ## Model TTS100 Portable Hyperthermic Perfusion Device #### 510(k) Number: K092366 Submitted in accordance with the requirements of SMDA 1990 and 21CFR807.92. ## 1.0 21CFR 807.92(A)(1) 1.1 Applicant's Information J. Michael Fausset BS CP Chief Technical Officer ThermalTherapeutic Systems, Inc. Mobile: 765 427.3308 Fax: 412 291.1951 E: imf@thermaltheraneutics.com I: http://www.thermaltherapeutics.com Medical Establishment Registration No .: pending 1.2 Submitter's Information James Jochen Rogers General Manager Coastal Consulting Group, Ltd. P.O. Box 961 Moon Township, PA 15108 Mobile: 724 713.2298 Fax: 412 291.1951 E: ir@coastalco.com I: http://www.coastakcg.com - 1.3 Date March 1, 2010 ## 2.0 21CFR 807.92(A)(2) 2.1 Device Information Trade/Proprietary Name: TTS100 Portable Hyperthermic Perfusion Device Common Name: TTS100 Portable Hyperthermic Perfusion Device Model Number: TTS100 ·Regulation Number: None Regulation Name: None Regulatory Class: Unclassified, pre-amendment device; non-exempt from 510(k) Device Classification Name: Warmer, thermal, Infusion fluid Classification Panel: General Hospital Classification Product Code(s): LGZ 2.2 Device Classification Infusion fluid thermal warmers and their accessories are unclassified, Pre-Amendment devices. 510(k) premarket notification submissions for these devices are reviewed by the General Hospital Devices Branch, Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices, {1}------------------------------------------------ ## ThermalTherapeuticSystems, Inc. ## 3.0 21CFR 807.92(A)(3) 3.1 Predicate Device Information | PREDICATE DEVICES | | | | |-------------------|---------------------------|--------------------------|--------------------------| | 510(k) # | DEVICE | 510(k) SPONSOR | 510(k)<br>CLEARANCE DATE | | K993330 | ThermoChem-HT System | Hemocleanse, Inc. | 12/30/1999 | | K070654 | Belmont Hyperthermia Pump | Belmont Instrument, Inc. | 6/8/2007 | ## 4.0 21CFR 807.92(A)(4) #### 4.1 Device Description The TTS100 Portable Hyperthermic Perfusion Device is a prescription device comprised of two components - one durable and one disposable. The TTS L00 Console is the durable component - a portable, unit containing heater, pump, user touchscreen, microprocessor, and interface electronics. The primary user interface of the TTS100 is a large, color touchscreen display that receives operational commands from the clinical user, and presents normal operating conditions, flow rate, output fluid temperature, patient temperature, target temperature, line pressure, alarm and status messages, various timers and auxiliary temperatures. The TTS100 software monitors various sensors in the fluid path to ensure safe operation, automatically reacts to unsafe conditions, and alerts the clinical user to those conditions for resolution. Independent protection circuits prevent unsafe operation in the event of system software fault. The disposable, single-use component is the Hyperthermic Lavage Tubing Set, which has two subcomponents - a Circulation Loop and an Accessory Kit. The Circulation Loop contains the filtered soft-shell fluid reservoir, inlet ports, vent lines, integrated distributed sensors, patient inlet circulation loop tubing, patient outlet circulation loop tubing, and drain line. The Accessory Kit - an integral component of the Hyperthermic Lavage Tubing Set - contains patient inlet and patient outlet tubing, tubing connectors, and auxiliary temperature sensors. Portions of the Accessory Kit can be implanted into the patient for use during the lavage procedure. The Circulation Loop does not have direct patient contact - only the circulating fluid within the loop contacts the patient. The Hyperthermic Lavage Tubing set is designed such that fluid travels through the Disposable Set without wetting the heater, pump or other TTS system components (e.g., only the internal surfaces of the Disposable Set are wetted), enabling rapid and straightforward post-processing of the TTS100 Console between patients. The Hyperthermic Lavage Tubing Set is supplied sterile, packaged in a proprietary-design container, and is only intended to be used in conjunction with the TTS100 Portable Hyperthermic Perfusion Device. ### 5.0 21CFR 807.92(A)(5) - 5.1 Indications for Use The intended use of the TTS100 Portable Hyperthermic Perfusion Device is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating, warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician. #### 6.0 21CFR 807.92(A)(6) #### 6.1 Technological Characteristics Comparison to Predicate Device(s) As compared to the predicate devices, the TTS100 Portable Hyperthermic Perfusion Device has the same operating principle, energy type, environmental specifications, and performance specifications. All devices use a roller-type fluid pump, touchscreen-based clinical user Interface for machine set-up and control, sensor monitoring of various fluid temperatures, and a proprietary-design disposable set including large fluid reservoir to circulate sterile fluid into and out of the body cavity. For the TTS100, fluid in the reservoir is heated via direct heat to the desired temperature, and the heater drain tubing passes through a roller pump and out to the patient inlet. Fluid returns from the patient outlet wia gravity drain back to the fluid reservoir. The TTS100 monitors circulating fluid temperature and pressure, and automatically responds to ineffective or unsafe operating conditions. {2}------------------------------------------------ ## ThermalTherapeuticSystems, Inc. ## 7.0 21CFR 807.92(B)(1) #### 7.1 Performance Data - Discussion of Non-Clinical Tests - 7.1.1 FDA Recognized Consensus Standards The TTS100 is designed to, complies with, and has been tested as part of verification and validation activities to, the FDA Recognized Consensus Standards listed in the table below, as applicable to device features and components: | REQUIREMENT | REFERENCE # | TITLE | |---------------------------|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Biocompatibility | AAMI/ANSI/ISO 10993-1:2003(E) | Biological evaluation of medical devices - Part<br>1: Evaluation and testing | | | AAMI/ANSI/ISO 10993-5:1999 | Biological evaluation of medical devices - Part<br>5: Tests for in vitro cytotoxicity | | | AAMI / ANSI / ISO 10993-<br>7:1995(R)2001 | Biological Evaluation of Medical Devices - Part<br>7: Ethylene Oxide Sterilization Residuals. | | | AAMI/ANSI/ISO 10993-10:2002 | Biological evaluation of medical devices -<br>Part 10: tests for irritation and delayed-type<br>hypersensitization | | | AAMI/ANSI/ISO 10993-11:2006 | Biological evaluation of medical devices - Part<br>11: Tests for systematic toxicity | | Electrical<br>Safety, EMC | IEC<br>60601-1:1988+A1:1991+A2:1995<br>(IEC 60601-1+A1+A2) | Medical electrical equipment - Part 1: General<br>Requirements for Safety, 1988; Amendment 1,<br>1991-11, Amendment 2, 1995 | | | IEC 60601-1-2:2001+A1:2004<br>(IEC 60601-1-2+A1) | Medical Electrical Equipment - Part 1-2: General<br>Requirements for Safety - Collateral standard:<br>Electromagnetic Compatibility - Requirements<br>and Tests (Ed. 2.1) | | Sterility | AAMI / ANSI / ISO 11135-1:2007 | Sterilization of health care products -<br>Ethylene oxide - Part 1: Requirements for the<br>development, validation, and routine control of<br>a sterilization process for medical devices. | | Packaging | ASTM D4169-05:2005 | Standard Practice for Performance Testing of<br>Shipping Containers and Systems | #### 7.1.2 Software Verification/Validation Testing As part of software verification/validation activities, and in support of substantial equivalence, the following tests were carried out the TTS100 to evaluate its ability to meet product performance specifications. Specifically, the TTS100 was tested to assess ... - I The ability of the TTS100 to heat fluids over the full range of fluid flowrates. - The ability of the TTS100 to maintain fluid temperature over the full range of fluid flowrates. E - E The ability of the TTS100 to detect and automatically respond to unsafe or ineffective operating conditions, as caused by the failure of the system sensors, excessive fluid temperature, excessive fluid pressure, or system-internal component failures. - The ability of the TTS100 to inform the clinical user of those unsafe or ineffective operating conditions via notifications, alerts, alarms, and/or system faults. - D The ability of the TTS100 to mitigate against known or predictable operator errors. - The ability of the TTS100 to store treatment parameters and data in non-volatile memory. D - ם Measured temperature accuracy over the full range of fluid flowrates within the operating pressure range. - D Measured pressure accuracy over the full range of fluid flowrates within the operating temperature range. - Fluid flowrate accuracy over the full range of fluid flowrates within the operating temperature and pressure range. The TTS100 performed within specification in all of the above tests. {3}------------------------------------------------ # K092366 ## ThermalTherapeuticSystems, Inc. #### 7.1.3 Human Factors A human factors non-cilnical study evaluated user- and patient-safety risks associated with the design of the TTS100, and demonstrated that the task-risks have been identified, addressed and mitigated. ## 8.0 21CFR 807.92(s)(2) - 8.1 Performance Data Discussion of Clinical Tests N/A - 9.0 21CFR 807.92(B)(3) #### 9.1 Conclusions Based upon safety and performance testing and compliance with voluntary standards, ThermalTherapeutic Systems, Inc. believes that the TTS100 Portable Hyperthermic Perfusion Device is substantially equivalent to the predicate devices, and does not raise any new questions of safety or effectiveness. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three curved lines. DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Thermal Therapeutic Systems, Incorporated C/O Mr. James Jochen Rogers General Manager Coastal Consulting Group, Limited P.O. Box 961 Moon Township, Pennsylvania 15108 JAN 1 0 2017 Re: K092366 Trade/Device Name: TTS100 Portable Hyperthermic Perfusion Device Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LGZ Dated: March 3, 2010 Received: March 4, 2010 Dear Mr. Rogers: This letter corrects our substantially equivalent letter of March 4, 2010. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {5}------------------------------------------------ Page 2 - Mr. James Jochen Rogers CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Erin I. Keith -S Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K092366 Device Name: TTS100 Portable Hyperthermic Perfusion Device Indications For Use: The intended use of the TTS100 Portable Hyperthermic Perfusion Device is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating, warmed, physiologically compatible sterile solution, according lo a protocol to be selected by the physician. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE-ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Charles Schwab (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K092366 Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...