K070654 · Belmont Instrument Corp. · LGZ · Jun 8, 2007 · General Hospital
Device Facts
Record ID
K070654
Device Name
HYPERTHERMIA PUMP
Applicant
Belmont Instrument Corp.
Product Code
LGZ · General Hospital
Decision Date
Jun 8, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Belmont Hyperthermia Pump is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.
Device Story
The Belmont Hyperthermia Pump is a system for rapid perfusion of warmed sterile solution into thoracic or peritoneal cavities. It utilizes a roller-type fluid pump and a heat exchanger to circulate and warm fluid. The system includes a disposable set with a large fluid reservoir; fluid is drained from the patient, pumped through the heat exchanger, and returned to the body cavity. A microprocessor-based system monitors fluid temperature, patient temperature, line pressure, and air in the fluid path. A touch screen displays flow rate, fluid volume, temperatures, and pressure, while providing procedural guidance. The device features a hardware override circuit to prevent unsafe operation in the event of computer failure. It is intended for clinical use under physician supervision. By maintaining precise cavity temperatures, the device supports hyperthermia treatment protocols, potentially benefiting patients by facilitating controlled thermal therapy.
Clinical Evidence
Bench testing only. Testing verified system performance regarding fluid pumping accuracy across various flow rates, back pressures, and ambient temperatures; the ability to warm fluid to target temperatures; and the efficacy of safety sensors and alarm systems for detecting operator errors, sensor failures, and internal malfunctions.
Technological Characteristics
Microprocessor-controlled roller-type fluid pump; heat exchanger for fluid warming; touch screen interface; disposable fluid circuit with reservoir. Monitors fluid/patient temperature, line pressure, and air in path. Includes hardware override circuit for safety. Connectivity/software details not specified beyond microprocessor control.
Indications for Use
Indicated for patients requiring thoracic or peritoneal cavity lavage with warmed sterile solution to achieve a target temperature, as determined by a physician-selected protocol.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's seal, which includes an emblem with three stylized human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### JAN 1 0 2017
Mr. Uraiwan P. Labadini Ouality Assurance/Regulatory Affairs Manager Belmont Instrument Corporation 780 Boston Road Billerica, Massachusetts 01821
Re: K070654
Trade/Device Name: Belmont Hyperthermia Pump Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LGZ Dated: March 9, 2007 Received: March 9, 2007
Dear Mr. Labadini:
This letter corrects our substantially equivalent letter of June 8, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting (reporting of
{1}------------------------------------------------
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 8091 ), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Belmont Hyperthermia Pump
Indications For Use:
The intended use of the Belmont Hyperthermia Pump is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
A
(Oblision Sign-Off) Civision of Anesthestology, General Hoapttal, infection Control, Dental Devices
Page 1 of _1
510(k) Number:_
{3}------------------------------------------------
K070654
Image /page/3/Picture/1 description: The image shows the logo for Belmont Instrument Corporation. The logo features the word "Belmont" in a stylized font, with the words "Instrument Corporation" in a smaller font underneath. Below the company name is the tagline "Creating a New Standard of Care".
JUN - 8 2007
Registered in Accordance with ISO 13485
### Premarket Notification 510(k) Summary [As Required by 21 CFR 807.92(a)]
Date prepared: March 9, 2007
- Submitter & 1. Belmont Instrument Corporation · Manufacturing Site: 780 Boston Road Billerica, MA 01821
Establishment Registration Number: 1219702
- 2. Contact Person: Uralwan P. Labadini, Quality Assurance/Regulatory Affairs Manager
Telephone: (978) 663-0212 Ext. 28 Fax: (978) 663-0214
- 3. Trade Name: The Belmont Hyperthermia Pump
- 4. Common name: Warmer, Thermal, Infusion Fluid
- 5. Classification name: Warmer, Thermal, Infusion Fluid
- હે. Product Code: 80 LGZ Device Class: Class II
7.
- Performance Standards: No performance standards have been officially adopted by the F.D.A.
- The Belmont Hyperthermia Pump is substantially equivalent to the 8. ThermoChem-HT System, which was the subject of Premarket Notification #K993330 and received F.D.A 510(k) concurrence to market on December 30, 1999.
780 Boston Road, Billerica, MA 01821 tel 978.663.0212 fax 978.663.0214 www.belmontinstrument.com
{4}------------------------------------------------
#### 9. Brief Description:
The Belmont Hyperthermia Pump combines advanced microprocessor technology with an efficient mechanical system to provide a high speed, simple and safe system for rapid perfusion of warmed fluid. It can raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.
The system monitors fluid temperature, patient temperature, line pressure, and air in the fluid path to ensure safe operation and alarms at all unsafe conditions. A hardware override circuit prevents unsafe operation in case of system computer A touch screen displays flow rate, total fluid pumped, output fluid failure. femperature, targef temperature, patient temperature, line pressure, alarm and status messages and proper procedures to proceed safely after an alarm situation.
#### 10. Intended Use
The intended use of the Belmont Hyperthermia Pump is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.
- 11. Summary of the technological characteristics of the Belmont Hyperthermia Pump compared to the predicate device. ThermoChem-HT System.
These two devices have the same operating principle, energy type, environmental specifications, or performance specifications. These two devices use roller-type fluid pump, touch screen to direct the user through set-up and use, a disposable set including large fluid reservoir to circulate sterile fluid into and out of the body cavity. Flow from the patient outlet is drained into the large reservoir, then through a roller pump and then to a heat exchanger. The heat exchanger warms fluid to the desired target temperature and then passes to the body cavity through the patient line/return line. These systems also monitor the circulating sterile fluid temperature. The Hyperthermia Pump sounds an audible alarm, stops heating, and pumping at all unsafe conditions.
- 12. Summary of Nonclinical Tests and Results
In order to verify performance of the Belmont Hyperthermia Pump in support of substantial equivalence, the following tests were carried out:
- The ability of the system to pump fluids accurately over the full range of flow a. rate and operating conditions including different input fluid temperatures, different back pressure, and change in ambient temperature.
{5}------------------------------------------------
- The ability of the system to warm fluid to the desired target temperature over b. the full range of flow rate and operating conditions.
- The ability of the system to detect and alarm at unsafe or ineffective C. operating conditions including operator errors, the failure of the system sensors, and other internal system malfunctions.
The Belmont Hyperthermia Pump performed within specification in all of the above tests.
- Conclusion: The Belmont Hyperthermia Pump is substantially equivalent to the 13. ThermoChem-HT System, a legally marketed devices intended to circulate warmed sterile fluid through the body cavity.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.