K092224 · Novus Scientific Pte, Ltd. · FTL · Jan 25, 2010 · General, Plastic Surgery
Device Facts
Record ID
K092224
Device Name
TIGR SURGICAL MESH , MODEL WK-6
Applicant
Novus Scientific Pte, Ltd.
Product Code
FTL · General, Plastic Surgery
Decision Date
Jan 25, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TIGR Matrix Surgical Mesh is intended for use in reinforcement of soft tissue where weakness exists.
Device Story
TIGR Matrix Surgical Mesh is a knitted, synthetic, resorbable surgical mesh. It consists of two fiber types with different degradation profiles: 40% by weight is a copolymer of polyglycolide, polylactide, and polytrimethylene carbonate (functional life ~2 weeks; fully absorbed ~4 months); 60% by weight is a copolymer of polylactide and polytrimethylene carbonate (functional life ~9 months; fully absorbed ~36 months). Both fibers degrade via bulk hydrolysis. The device is used by surgeons for soft tissue reinforcement. It provides mechanical support during the healing process and is gradually absorbed by the body, eliminating the long-term presence of permanent foreign material.
Clinical Evidence
No human clinical data provided. Effectiveness was demonstrated via an in vivo study in a sheep hernia repair model comparing the subject device to Prolene Mesh. Results indicated equivalent performance regarding tissue integration and local tolerance. Biocompatibility and safety were evaluated in accordance with ISO 10993 and 21 CFR Part 58 (GLP), confirming the device is non-toxic, non-mutagenic, non-sensitizing, and biocompatible.
Technological Characteristics
Knitted synthetic resorbable mesh. Materials: Copolymer of polyglycolide, polylactide, and polytrimethylene carbonate (40%); copolymer of polylactide and polytrimethylene carbonate (60%). Degradation mechanism: Bulk hydrolysis. Performance metrics: Mesh thickness, density, pore diameter, suture retention strength, tear strength, and burst strength. Biocompatibility testing per ISO 10993.
Indications for Use
Indicated for reinforcement of soft tissue where weakness exists in patients requiring surgical mesh repair.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
K092224
Page 1 of 2
## 510(k) Summary
#### Submitter's Information:
Name: Address:
Phone: Contact Person:
## Date of Preparation:
Device Name:
Trade Name: Common Name: Classification Name: Classification Product Code: Regulatory number:
Predicate Device Names:
Novus Scientific Pte Ltd Nordic European Centre, 3 International Business Park #01-20 (S) 609927 +65 68900360 Kelvin Koh
17 July 2009
TIGR Matrix Surgical Mesh Surgical Mesh Mesh. Surgical, Polymeric FTL \$878.3300
Prolene Mesh (K001122) Mersilene Mesh (K851086) Ultrapro Mesh (K033337)
#### Device Description:
TIGR Matrix Surgical Mesh is knitted from two different synthetic resorbable fibers, possessing different degradation characteristics. The first fiber, making up 40% of the matrix by weight, is a copolymer of polyglycolide, polylactide, and polytrimethylene carbonate.
The second fiber, making up 60% of the matrix by weight, is a copolymer of polylactide, and polytrimethylene carbonate. Both fibers degrade by bulk hydrolysis once implanted, resulting in a decreasing strength retention followed by mass loss of the fibers.
Based on the product's absorption characteristics, in vitro testing showed that the first fiber (polyglycolide, polylactide, and polytrimethylene carbonate) loses its functional capabilities after 2 weeks and in vivo studies in the abdominal wall of sheep showed that the first fiber was fully absorbed after 4 months. The same in vitro testing showed that the second fiber (polylactide, and polytrimethylene carbonate) loses its functional capabilities after 9 months and in vivo studies in the abdominal wall of sheep indicated that the second fiber should be absorbed after approximately 36 months.
JAN 9 5 2010
{1}------------------------------------------------
K092224
Page 2 of 2
#### Intended Use:
TIGR Matrix Surgical Mesh is intended for use in reinforcement of soft tissue where weakness exists.
#### Technological Characteristics:
The physical and mechanical properties of the TIGR Matrix Surgical Mesh, such as mesh thickness, density, pore diameter, mesh knit characteristics, suture retention strength, tear strength, and burst strength, has similar performance characteristics to the currently marketed predicate devices.
#### Performance data:
The biocompatibility and safety tests conducted for TIGR Matrix Surgical Mesh were selected in accordance with "ISO 10993, Biological Evaluation of Medical Devices Part I: Evaluation and Testing." All studies were conducted in accordance to 21 CFR, Part 58, Good Laboratory Practices. Based on the results from these studies, TIGR Matrix Surgical Mesh is considered to be non-toxic, nonmutagenic, non-sensitizing, biocompatible and safe for its intended use.
The effectiveness of TIGR Matrix Surgical Mesh was compared in vivo in a Sheep hernia repair model to the Prolene Mesh. The overall performance of TIGR Matrix Surgical Mesh, including tissue integration, local tolerance was equivalent to its predicate device.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with outstretched wings, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Novus Scientific, Pte, Ltd. % Mr. Kelvin Koh Nordic European Centre 3 International Business Park #01-20 Singapore 609927
JAN 2 5 2010
Rc: K092224
Trade/Device Name: TIGR Matrix Surgical Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh, Polymeric Regulatory Class: Class II Product Code: FTL Dated: November 18, 2009 Received: November 23, 2009
Dear Mr. Koh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Kelvin Koh
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A. Millikenn
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K09 2224
# Indications for Use
510(k) Number:
K092224
Device Name:
TIGR Matrix Surgical Mesh
## Indications for Use:
TIGR™ Matrix Surgical Mesh is indicated for use in reinforcement of soft tissue where weakness exists.
X Prescription Use_
(Per 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David Kramer for MXM
(Division Sign (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K092224
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.