WEDGELOC 180X STANDARD SUTURE ANCHOR WITH OPTI FIBER SUTURES,MODEL 1000-01 TAB VARYING DIAMETERS,
Applicant
Medshape Solutions
Product Code
HWC · Orthopedic
Decision Date
Sep 16, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MedShape Solutions, Inc., WedgeLoc™ 180X Suture Anchor with Opti-Fiber™ Sutures are intended for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist and elbow in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial collateral Elbow: Ligament Reconstruction
Device Story
WedgeLoc™ 180X Suture Anchor with Opti-Fiber™ Sutures; used for soft tissue-to-bone fixation in orthopedic procedures. Device functions as mechanical fastener; anchors suture into bone to facilitate tissue repair. Used by surgeons in clinical/OR settings. Provides secure attachment point for sutures; enables reattachment of tendons/ligaments to bone. Benefits patient by restoring structural integrity of damaged joints/tissues.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use.
Technological Characteristics
Suture anchor system for bone fixation. Materials include non-degradable synthetic components. Mechanical fixation principle. Class II device.
Indications for Use
Indicated for fixation of suture to bone in shoulder, foot/ankle, knee, hand/wrist, and elbow procedures, including rotator cuff, Bankart, SLAP, biceps, acromio-clavicular, deltoid, capsular, ligament, tendon, and hallux valgus repairs/reconstructions.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
WedgeLoc™ Suture Anchor with Opti-Fiber™ Suture (K083792)
{0}------------------------------------------------
K091202 (pg 1/2)
Image /page/0/Picture/1 description: The image shows the logo for Medshape Solutions. The word "MEDSHAPE" is in large, bold, black letters on the top line. Underneath the first word is a horizontal line, and below that is the word "SOLUTIONS" in smaller, black letters.
SEP 16 2009
Special 510(k) Page 10 of 140
## Special 510(k) Summary
Special 510(k) Number:
April 23rd, 2009 Date Prepared:
This Special 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
- Submitter: A. MedShape Solutions, Inc. (MSS) 1575 Northside Drive. Suite 440 Atlanta, Georgia 30318
- B. Company Contact: Jack Griffis Vice President, Research & Development (404) 249-9156 x11 (direct) (404) 249-9158 (fax) Jack.Griffis@MedShapeSolutions.com
- C. Device Information: WedgeLoc™ 180X Suture Anchor with Opti-Fiber™ Sutures Trade Name: Common Name: Suture Anchor
- Fastener, Fixation, Non-degradable, Soft Tissue Classification Name: D. HWC/MBI 21 CFR 888.3040 Suture, Nonabsorbable, Synthetic, Polyethylene GAT 21 CFR 878.500
- E. Predicate Device(s): MSS, WedgeLoc™ Suture Anchor with Opti-Fiber™ Suture, K083792 Arthrex PushLock™ Suture Anchor, K051219 Arthrex FiberWIRE™ Polvblend Suture, K021434
- F. Labeling and Intended Use:
NOTE: Draft labels and instructions for use can be found in Attachment D. However, no changes to the labeling or Instructions for Use have been made to the original submitted information per K083792.
The proposed WedgeLoc™ 180X Suture Anchor and Opti-Fiber™ Suture has the same intended uses as the previously cleared device (K083792). In particular, both devices are indicated for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist and elbow in the following procedures:
{1}------------------------------------------------
1,091202 Lpg 212)
Special 510(k) Page L of 10
- Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Tenodesis. Acromio-Clavicular Separation Repair. Biceps Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Lateral Stabilization, Medial Stabilization, Achilles Tendon Foot/Ankle: Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair
- Medial Collateral Ligament Repair, Lateral Collateral Ligament Knee: Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Collateral Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Ligament Ligament Reconstruction, Radial Collateral Reconstruction
- Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Elbow: Radial collateral Ligament Reconstruction
- Substantial Equivalence Summary: G.
The proposed WedgeLoc™ 180X Suture Anchor is is substantially equivalent to the predicate WedgeLoc™ Suture Anchor with Opti-Fiber™ Suture (K08379) and the Arthrex PushLock™ Suture Anchor (K051219) in which the basic features and intended uses are the same. In addition, the technological characteristics of the WedgeLoc™ 180X and the WedgeLoc™ Suture Anchors are equivalent. Any differences between the WedgeLoc™ 180X Suture Anchor and the predicate WedgeLoc™ or Arthrex PushLock™ Suture Anchors are considered minor and do not raise questions concerning safety and effectiveness. Based on the information submitted, MedShape Solutions, Inc. has determined that the new WedgeLoc™ 180X Suture Anchor is substantially equivalent to the currently marketed predicate device.
04/23/2009
Jack Griffes Vice President, Research & Development
Image /page/1/Picture/11 description: The image shows the logo for MEDSHAPE SOLUTIONS. The word "MEDSHAPE" is in large, bold, black letters, with a line underneath it. Below the line, the word "SOLUTIONS" is written in smaller, black letters.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing the bird's body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all capital letters and is evenly spaced around the circle.
SEP 1 6 2009
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
MedShape Solutions % Jack Griffis · 1575 Northside Drive. Suite 440 Atlanta, Georgia 30318
#### Re: K091202
Trade/Device Name: WedgeLoc™ 180x Suture Anchor and Opti-Fiber™ Sutures Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth or threaded metallic bone fixation fastener
Regulatory Class: Class II
Product Code: HWC, MBI, GAT
Dated: September 2, 2009
Received: September 4, 2009
### Dear Mr. Griffis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Jack Griffis
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Boubeau Buchus
for
Mark N. Melk Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
#### 510(k) Number: K091202
Device Name: WedgeLoc™ 180X Suture Anchor and Opti-Fiber™ Sutures
#### Indications for Use:
The MedShape Solutions, Inc., WedgeLoc™ 180X Suture Anchor with Opti-Fiber™ Sutures are intended for fixation of suture to bone in the shoulder, foot/ankle, knee, hand/wrist and elbow in the following procedures:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction
Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial collateral Elbow: Ligament Reconstruction
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Onetu
(Division Sign-Off) for mxn
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091202
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.