ZEUS (ZIO ECG UTILIZATION SERVICE) SYSTEM

K091075 · iRhythm Technologies, Inc. · DQK · Jul 21, 2009 · Cardiovascular

Device Facts

Record IDK091075
Device NameZEUS (ZIO ECG UTILIZATION SERVICE) SYSTEM
ApplicantiRhythm Technologies, Inc.
Product CodeDQK · Cardiovascular
Decision DateJul 21, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The Zio ECG Analysis System is intended for use by qualified medical professionals for assessment of a patients recorded ambulatory ECG data. The system is intended to be marketed as a service that downloads and analyzes up to 14 days of ECG data. Analysis metrics are provided in a report which is available for clinician review and printing. The reported ECG metrics include single-lead analysis on a beat by beat basis heart rate measurement and rhythm analysis. The report does not contain diagnostic interpretation the reported analysis is provided for review by the intended user to render a diagnosis based on clinical judgment and experience.

Device Story

ZEUS System processes/analyzes up to 14 days of continuous single-lead ambulatory ECG data; enables iRhythm to provide ECG analysis service. System downloads data, performs automated analysis of beats, runs, rhythms, and heart rate; generates reports for clinician review. Used by medical professionals to render diagnoses based on clinical judgment. Benefits include extended recording duration (up to 14 days) to capture transient arrhythmias, reducing need for multiple tests and improving patient experience. System combines end-to-end processing with automated analysis.

Clinical Evidence

Bench testing only. Automated analysis algorithm performance verified against industry standards ANSI/AAMI EC57 and IEC60601-2 51. Equivalence to predicates demonstrated through comparative analysis of 48-hour continuous ECG recordings and quantification of sensitivity and positive predictivity metrics.

Technological Characteristics

Software-based ECG processing system. Performs automated beat, run, rhythm, and heart rate analysis on single-lead ECG data. Connectivity via website for report delivery. Compliant with ANSI/AAMI EC57 and IEC60601-2 51 standards.

Indications for Use

Indicated for qualified medical professionals to assess recorded ambulatory ECG data. Not for pediatric patients or patients with paced beats.

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for iRhythm Technologies, Inc. The logo features the word "iRhythm" in a stylized font, with a series of curved lines above the word. The lines are arranged in an ascending pattern, resembling a sound wave or a heartbeat rhythm. K 091075 Page 1 of 3 ### 510(k) PREMARKET NOTIFICATION JUL 21 2009 ## Executive Summary #### ﮩ INTRODUCTION The Zio ECG Utilization Service System ( ZEUS System ) is an electrocardiogram (ECG) processing and analysis system designed to handle continuously recorded single lead ECG data It downloads stores analyzes and sorts the ECG data to generate a report of the findings contained within the data thereby enabling the provision of a complete ECG processing and analysis service The ZEUS System is substantially equivalent to two predicate systems both of which have similar intended uses The scope of functionality of the ZEUS system is encompassed by a combination of the two predicate systems Automated ECG analysis performance was quantified for any claimed analysis metrics The resulting statistics demonstrate sensitivity and positive predictivity levels which satisfy requirements and minimize safety or efficacy concerns ## 2 BACKGROUND The ZEUS System is intended to be used with the Zio family of ECG monitoring devices The Zio devices enable continuous ambulatory ECG data collection over a span of up to 14 days This long recording period increases the likelihood that any transient arrhythmias a patient may be experiencing will be captured Despite the clinical usefulness of recording ECG data for up to 14 days analysis systems currently available on the market are either unable to process recordings of this length or the intensive level of human effort required to do so makes it impractical This often results in the need for patients to receive multiple monitoring tests which results in higher costs and a less than optimal patient experience To provide our customers with a complete suite of products resulting in a simplified experience with improved diagnostic function at lower cost it was necessary to design a system that was capable of analyzing up to 14 days of continuously recorded single-lead ECG data The ZEUS System meets that requirement and enables iRhythm Technologies Inc ((Rhythm) to provide a complete ECG processing and analysis service to our customers #### 3 | INTENDED USE The Zio ECG Analysis System is intended for use by qualified medical professionals for assessment of a patients recorded ambulatory ECG data The system is intended to be marketed as a service that downloads and analyzes up to 14 days of ECG data Analysis metrics are provided in a report which is available for clinician review and printing The reported ECG metrics include single-lead analysis on a beat by beat basis heart rate measurement and rhythm analysis The report does not contain diagnostic interpretation the reported analysis is provided for review by the intended user to render a diagnosis based on clinical judgment and experience #### DEVICE DESCRIPTION ঞ্চ The ZEUS System is a software system designed to be used internally at IRhythm to provide a complete ECG processing and analysis service It includes software that enables ECG data download automated ECG analysis automated report generation and report delivery 650 Townsend Street Sule 380 | San Francısco CA 94103 | (415) 632 5701 F | (415) 632 5701 F | www.irhythmlech.com Rhythm Technologies Inc Page 7 1 {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for iRhythm. The logo consists of the word "iRhythm" in a stylized font, with a small graphic above it. The graphic appears to be a series of curved lines, possibly representing a heartbeat or rhythm. (- 1 via a website It is capable of automated ECG analysis on continuously recorded singlelead ECG data up to 14 days in duration # 5 SUBSTANTIAL EQUIVALENCE Legally marketed predicate systems | 510(k) Number | Product | Manufacturer | |---------------|-------------------------------------------------------|-------------------------| | K042463 | Ambulatory ECG Analysis<br>System | Rozinn Electronics Inc | | K062282 | Automated ECG Analysis and<br>Interpretation Software | Monebo Technologies Inc | The ZEUS System is substantially equivalent to the Rozinn Holter Plus System and the Monebo Analysis Software Library The Rozınn Holter Plus System is an end to end ECG processing system implemented in the form of software used by the clinic to download ECG data and process a complete report output It enables human-assisted ECG analysis The Monebo Software Library is a software library that enables automated ECG analysis of beats runs rhythms and heart rate The ZEUS System combines the end to-end ECG processing capabilities of the Rozinn system with the automated ECG analysis functionality of the Monebo software Ibrary The table below summanzes key similarities of the ZEUS system with the predicate systems | | ZEUS System | Rozinn Holter Plus | Monebo SW Library | |------------------------|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------| | Intended Use | Recorded ECG data<br>processing and<br>analysis | Recorded ECG data<br>processing and<br>analysis | Recorded ECG data<br>analysis | | System<br>capabilities | ECG data download<br>analysis and reporting | ECG data download<br>analysis and reporting | ECG data analysis only | | Data<br>Analysis | Automated beat run<br>rhythm and heart rate<br>analysis | Preliminary automated<br>analysis of beat run<br>and heart rate analysis<br>with human intervention<br>required to identify<br>certain rhythms | Automated beat run<br>rhythm and heart rate<br>analysis | Extensive performance testing was conducted on the automated analysis algorithm portions of the ZEUS system to ensure that automated performance was substantially equivalent to the Monebo automated analysis metrics Equivalence with the Rozinn Holter Plus System was demonstrated through a study that compared analysis reports for a 48 hour continuous ECG recording 650 Townsend Street Suite 380 | San Francisco CA 94103 | (415) 632 5701 F | (415) 632 5701 F | www.rhythmlech.com l {2}------------------------------------------------ ![iRhythm](https://i.imgur.com/9999999.png) #### 6 COMPLIANCE TESTING Compliance testing was performed to two industry standards - . ANSI/AAMI EC57 - . IEC60601-2 51 These two standards define the required testing and reporting methods used to assess the performance of an automated analysis algonthm All metrics reported in this submission were obtained using the methods required by these two standards Further details on compliance to these standards can be found in Section 10 of this submission These standards do not provide minimum performance requirements Minimum performance standards were developed internally at iRhythm in consultation with our clinical team and are documented in Section 13 of this submission Section 17 describes in greater detail the methods and test results used to venfy that the automated analysis algorithm portions of the ZEUS system meet the requirements when tested in compliance with the ındustry standards above #### PRECAUTIONS 7 The ZEUS system is not designed for analysis of ECG recordings from pediatric patients or those patients with paced beats #### CONCLUSION 8 - Based on thorough verfication testing quantification of algorithmic performance and substantial equivalence companson with predicate systems it is our determination that the ZEUS system is substantially equivalent to the legally marketed predicates {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUL 21 2009 Mr. Michael Righter Regulatory and Quality Assurance Engineer iRhythm Technologies, Inc. 650 Townsend Street, Suite 380 San Francisco, CA 94103 Re: K091075 Trade/Device Name: ZEUS System Model S100 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Codes: DOK Dated: July 13, 2009 Received: July 15, 2009 Dear Mr. Righter: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Michael Righter Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-315.0 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Bram D. Žuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for iRhythm. The logo consists of the word "iRhythm" in a serif font, with the "i" in lowercase and the "Rhythm" in uppercase. Above the word "iRhythm" are five diagonal lines that are progressively shorter from left to right. The lines are filled with a pattern of small dots. # INDICATIONS FOR USE STATEMENT 510(k) Number: Kogin75 ZEUS System __ Indications for Use: Device Names: The ZEUS System is intended for use by qualified medical professionals for assessment of a patient's recorded ambulatory ECG data. The system is intended to be marketed as a service that downloads and analyzes up to 14 days of ECG data. Analysis metrics are provided in a report which is available for clinician review and printing. The reported ECG metrics include single-lead analysis on a beat by beat basis, heart rate measurement and rhythm analysis. The report does not contain diagnostic interpretation; the reported analysis is provided for review by the intended user to render a diagnosis based on clinical judgment and experience. AND/OR Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Tittillsk-lon.B.Zuckerman (Division Sign-Off) 7/21/09 Division of Cardiovascular Devices 510(k) Number K09/075 650 Townsend Street, Suite 380 | San Francisco, CA 94103 | (415) 632-5701 | | (415) 632-5701 | | www.irhythmtech.com
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...