The Holter Plus™ Ambulatory ECG Analysis System is intended for general hospital or clinical use by medical professionals whenever it is required to assess a patient's ambulatory ECG data. This product allows a trained physician or health care professional, after having performed a long-term continuous electrocardiograph (ECG) recording on a digital flash memory Holter recorder, to download and analyze the data from the recorder, review it and produce printed reports. This will enable the evaluation of arrhythmias, ischemic attacks, reporting of PQRST intervals, clinical and enidemiological research studies, evaluation of a patients response after resuming occupational or recreational activities(e.g., after M.I. or cardiac surgery, and/or evaluation of patients with pacemakers. The need to record, review, edit and archive these ambulatory ECG data is most commonly encountered in the Cardiology areas of the hospital during cardiac patient assessment. This device is available for sale only upon the order of a physician or other related licensed medical professional, and not intended for any home use applications.
Device Story
Holter Plus™ is a PC-based ambulatory ECG analysis system. It ingests digital ECG data from a Braemar RZ153+ flash memory recorder. The system performs high-speed analysis to classify ECG morphologies and calculate parameters like heart rate trends and RR variability. It is used in hospital cardiology departments by physicians or trained healthcare professionals. The system provides a graphical interface for clinicians to review, edit, and print reports. It does not perform automated diagnosis; it serves as a tool for clinical evaluation of arrhythmias, ischemia, and pacemaker function. The device facilitates patient management by enabling retrospective review and archiving of long-term cardiac recordings.
Clinical Evidence
Bench testing only. Performance testing confirmed equivalence to predicate devices through system tests, design reviews, and product verification testing.
Technological Characteristics
PC-based system; Windows-based analysis software; interfaces with digital flash memory Holter recorders (Braemar RZ153+). Provides graphical display of ECG waveforms, heart rate trends, and RR variability. No specific materials or sterilization methods applicable as it is a software/computer-based analysis system.
Indications for Use
Indicated for clinical assessment of ambulatory ECG data in patients requiring long-term continuous ECG monitoring, including evaluation of arrhythmias, ischemic attacks, PQRST intervals, post-MI or post-cardiac surgery recovery, and pacemaker function. For use by medical professionals in hospital/clinical settings; prescription only.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Northeast Monitoring Inc. Holter for Windows System (K930564)
Advanced BioSensor ABS Holter (formerly Micro SI Ambulatory ECG Monitor) System (K881905)
Oxford Scientific Medilog Excel 2 System (K002544)
Biomed Systems Inc. WinPV Holter System (K8841524)
Submission Summary (Full Text)
{0}------------------------------------------------
## SEP 2 2 2004
## 510(k) Summary of Safety and Effectiveness
#### Submitter Information:
Rozinn Electronics Inc. 71-22 Myrtle Ave. Glenside. NY 11385-7254 Phone: 718-386-5526 Contact: Mark Rosoff, President
#### Product Name:
Holter Plus™ Ambulatory ECG Analysis System Proprietary: Common: Programmable Diagnostic Computer and Software. Classification: Class II (see 21 CFR - 870.1425)
#### Predicate Devices:
The predicate device(s) are the following:
- 1. Northeast Monitoring Inc. Holter for Windows System (510(k) K930564).
- 2. Caird Technology Holter 2000 System. (510(k) K012686).
- Advanced BioSensor ABS Holter (formerly Micro SI Ambulatory ECG Monitor) 3. System (510(k) K881905).
- 4. Oxford Scientific Medilog Excel 2 System. (510(k) K002544).
- 5. Biomed Systems Inc. WinPV Holter System. (510(k) K8841524).
These predicate devices have the same performance specifications as the Holter Plus™ Ambulatory ECG Analysis System
#### Device Description:
The Holter Plus™ Ambulatory ECG Analysis System is an ambulatory ECG analysis system comprised of a standard computer display, a standard Personal Computer (PC) based system, and hardware used to acquire the ambulatory ECG data (specifically the Braemar RZ153+ data recorder - 510(k) K993618). The Holter Plus™ Ambulatory ECG Analysis System System can provide the retrospective display, storage and recording (or printing) of ambulatory ECG waveform and annotated data that is provided by the Braemar RZ153+ recorder.
The Holter Plus™ Ambulatory ECG Analysis System allows the assessment and review of a patient's recorded ambulatory ECG records, and presents an analysis for the clinician's review and editing. This product performs a high-speed analysis of the data, classify the data into appropriate ECG morphologies, and allow the results to be printed if desired. The Holter Plus TM does not perform any diagnosis of data, but only displays the
{1}------------------------------------------------
ECG morphologies, and associated curves calculated form the data such as heart rate trends and RR variability in graphical form. The physician will be able to review, edit, and print the data collected. The Holter Plus™ will provide information such as when patient-activated events occurred.
The Holter Plus™ Ambulatory ECG Analysis System provides storage of patient data for archiving and printing various reports. This includes waveform and patient parameter data for the ambulatory ECG data and assessment. The waveform and parameter data that has been stored in a patient file can be retrieved and reviewed on the display or printed out on a laser printer.
## Intended Uses:
The Holter Plus™ Ambulatory ECG Analysis System is intended for general hospital or clinical use by medical professionals whenever it is required to assess a patient's ambulatory ECG data. This product allows a trained physician or health care professional, after having performed a long-term continuous electrocardiograph (ECG) recording on a digital flash memory Holter recorder, to download and analyze the data from the recorder, review it and produce printed reports. This will enable the evaluation of arrhythmias, ischemic attacks, reporting of PQRST intervals, clinical and enidemiological research studies, evaluation of a patients response after resuming occupational or recreational activities(e.g., after M.I. or cardiac surgery, and/or evaluation of patients with pacemakers.
The need to record, review, edit and archive these ambulatory ECG data is most commonly encountered in the Cardiology areas of the hospital during cardiac patient assessment. This device is available for sale only upon the order of a physician or other related licensed medical professional, and not intended for any home use applications.
## Technological Comparison to Predicate Device(s):
The Holter Plus™ Ambulatory ECG Analysis System uses the same type of technology (i.e. personal computers and Microsoft Windows-based data analysis software) that is found in the predicate devices listed above.
## Summary of Performance Testing:
Tests demonstrating ambulatory ECG testing equivalence with the predicate devices listed above has been performed, and these results show similar performance with these devices.
Conformance to the product development procedures and plans have been assured by the application of the system tests, design reviews, and product verification testing performed prior to product release.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
SEP 2 2 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Rozinn Electronics, Inc. c/o Mr. Robert Mosenkis President Citech 5200 Butler Pike Plymouth Meeting, PA 19462-1298
Re: K042463 K042405
Trade Name: Holter Plus™ Ambulatory ECG Analysis System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: September 10, 2004 Received: September 13, 2004
Dear Mr. Mosenkis:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becternined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated to togals actment date of the Medical Device Amendments, or to Conninered processified in accordance with the provisions of the Federal Food, Drug, devices that have been recuire in asse equire approval of a premarket approval application (PMA). allu Cosmette Act (71ct) market the device, subject to the general controls provisions of the Act. The r ou may, dicierore, market the Act include requirements for annual registration, listing of general voltable provider ing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may be subject to basil adalarians, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Robert Mosenkis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
B.Zimmerman for
B. Z. Zimmerman, M.D.
Bram D. Zuckerman, M.D Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K042463
Ind, fort
# Indications for Use
510(k) Number (if known):
Device Name: Holter Plus
Indications for Use:
The Holter Plus ™ Ambulatory ECG Analysis System is intended for general hospital or clinical use by medical professionals whenever it is required to assess a patient's ambulatory ECG data. This product allows a trained physician or health care professional, after having performed a long-term continuous electrocardiograph (ECG) recording on a digital flash memory Holter recorder, to download and analyze the data from the recorder, review it and produce printed reports. This will enable the evaluation of arrhythmias, ischemic attacks, reporting of PQRST intervals, clinical and epidemiological research studies, evaluation of a patients response after resuming occupational or recreational activities (e.g., after M.I. or cardiac surgery, and/or evaluation of patients with pacemakers.
The need to record, review, edit and archive these ambulatory ECG data is most commonly encountered in the Cardiology areas of the hospital during cardiac patient assessment. This device is available for sale only upon the order of a physician or other related licensed medical professional, and not intended for any home use applications.
Prescription Use (Per 21 CFR 801 Subpart D)
AND/OR
Over-The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummenstiel
510(k) Number
Page of of
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.