All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion. Cemented femoral stems and cemented acetabular cups are intended for cemented fixation only. Press-fit femoral stems and acetabular cups are intended for press-fit fixation. Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon. Femoral heads and endoprostheses are intended for use in cemented and press-fit applications.
Device Story
The Novation CFS Press-Fit and Cemented Femoral Stems are orthopedic implants used in hip arthroplasty. The press-fit version features a calcar collar, grit-blast body, and satin-finish neck. The cemented version is cast cobalt chrome alloy with a satin-finish neck. Both utilize a modified 12/14 femoral head taper connection. These devices are implanted by orthopedic surgeons in a clinical/OR setting to replace the femoral component of the hip joint. They interface with previously cleared femoral heads and unipolar sleeves. The device restores joint function and mobility in patients with degenerative hip conditions or fractures. Clinical benefit is derived from the mechanical replacement of the diseased or damaged femoral head and neck, facilitating joint articulation and weight-bearing.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Cobalt chrome alloy. Design: Femoral stems with 12/14 taper connection; press-fit version includes calcar collar and grit-blast body; cemented version is cast. Sterilization: Same as predicate devices. Connectivity: None (mechanical implant).
Indications for Use
Indicated for skeletally mature individuals undergoing primary hip replacement for osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative hip problems, or proximal femoral fractures. Also indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed reconstructions with sufficient bone stock, and restoration of mobility after fusion.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
{0}------------------------------------------------
# Exactech® Novation® CFS™ Press-Fit and Cemented Femoral Stems Special 510(k) - 510(k) Summary of Safety and Effectiveness
### APR 2 1 2009
Exactech® İnc. Sponsor: 2320 N.W. 66th Court Gainesville, FL 32653
> Phone: (352) 377-1140 Fax: (352) 378-2617
FDA Establishment Number 1038671
- Tara R. Patterson Contact: Associate Regulatory Affairs Specialist
March 18, 2008 Date:
Trade or Proprietary or Model Name(s): Exactech® Novation® CFS™ Press-Fit and Cemented Femoral Stems
Common Name: Cemented Femoral Hip Prosthesis
Press-Fit Femoral Hip Prosthesis
### Classification Name:
Prosthesis, hip, semi-constrained, metal/polymer, uncemented
Prosthesis, hip, semi-constrained, metal/polymer, cemented
Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented
Information on devices to which Substantial equivalence is claimed:
| 510(k) Number | Trade or Proprietary or Model Name | Manufacturer |
|---------------|-----------------------------------------|----------------|
| K042842 | Novation® 12/14 Press-Fit Femoral Stems | Exactech, Inc. |
| K083392 | Novation® Cemented Plus Femoral Stems | Exactech, Inc. |
#### Indications for Use:
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures
{1}------------------------------------------------
K090764 (pg 2/2)
### Exactech® Novation® CFSTM Press-Fit and Cemented Femoral Stems Special 510(k) - 510(k) Summary of Safety and Effectiveness
where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.
- . Cemented femoral stems and cemented acetabular cups are intended for cemented fixation only.
- Press-fit femoral stems and acetabular cups are intended for press-fit fixation. Press-fit . components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon.
- . Femoral heads and endoprostheses are intended for use in cemented and press-fit applications.
### Device Description:
The proposed Novation CFS Press-Fit and Cemented Femoral Stems are modifications to the existing Novation Press-Fit and Novation Cemented Plus femoral stem devices previously cleared in K042842 and K083392, respectively. The proposed Novation CFS Press-Fit stem has the same general design features as the predicate device. The modifications include the addition of a calcar collar, grit-blast body, and a satin finish neck. The proposed CFS Cemented stem has the same general design features as the Novation Cemented Plus stem, but the proposed stems are cast instead of forged cobalt chrome alloy and the neck is a satin finish. Both proposed stems contain a modified 12/14 femoral head taper connection.
The proposed stems mate with previously cleared 12/14 Co-28Cr-6Mo femoral heads (K041906), Zirconia ceramic femoral heads (K050398 and K060107) and 12/14 Unipolar Sleeves (K010081). The proposed stems are not compatible with alumina ceramic femoral heads (K032964 and K051682).
The predicate and proposed devices have the same intended use and basic fundamental scientific technology and share the following similarities:
- · the same indications for use
- · similar design features
- · incorporate the same materials
- · the same shelf life
- · are packaged and sterilized using the same materials and processes.
### Substantial Equivalence Conclusion:
Results of engineering studies referenced in this 510(k) submission demonstrate that the proposed Novation CFS Press-Fit and Cemented Femoral Stems are substantially equivalent to the cleared predicate devices.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its back, representing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## APR 2 1 2009
Exactech® Inc. % Ms. Tara R. Patterson Associate Regulatory Affairs Specialist 2320 N.W. 66th Court Gainesville, FL 32653
Re: K090764
Trade/Device Name: Novation CFS Press-Fit and Cemented Stems Regulation Number: 21 CFR 888.3360 Regulation Name: Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis Regulatory Class: Class II Product Code: LWJ, LZO, JDI Dated: March 19, 2009 Received: March 23, 2009
Dear Ms. Patterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 16 ass starterior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, attrefy missions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your dovied to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be actived that i Driver in that your device complies with other requirements of the Act that + Dr has intatutes and regulations administered by other Federal agencies. You must or any I occur all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Of K F art 607), accember (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
f.o.L Peter D. R.
Mark N. Melkerson Dep DIRECTOR
Director
MD 09/21/15
Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Exactech® Novation® CFSTM Press-Fit and Cemented Femoral Stems Special 510(k) - Indications for Use
510(k) Number:
K090764(pg1/1)
## Device Name: Exactech® Novation® CFS™ Press-Fit and Cemented Femoral Stems
#### INDICATIONS
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, theumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.
- t Cemented femoral stems and cemented acetabular cups are intended for cemented fixation only.
- Press-fit femoral stems and acetabular cups are intended for press-fit . fixation. Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon.
- Femoral heads and endoprostheses are intended for use in cemented and . press-fit applications.
Prescription Use X (Part 21 CFR 801 Subpart D) and/or
Over-The-Counter Use (21 CFR 807 Subpart C)
Please do not write below this line - use another page if needed.
Image /page/4/Picture/13 description: The image shows a document with the title "Concurrence of CDRH, Office of Device Evaluation (ODE)". There is a signature above the text "(Division Sign-Off)". Below this is the text "Division of General, Restorative, and Neurological Devices". The 510(k) number is K090764.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.