FOTONA XP ND: YAG LASER SYSTEM FAMILY, FOTONA XP, XP-2 FOCUS, FOTONA XP CHANGER,MODEL MOO1-14F, MOO1-14F, MOO1-14F/2,
K090126 · Fotona D.D. · GEX · Jun 29, 2009 · General, Plastic Surgery
Device Facts
Record ID
K090126
Device Name
FOTONA XP ND: YAG LASER SYSTEM FAMILY, FOTONA XP, XP-2 FOCUS, FOTONA XP CHANGER,MODEL MOO1-14F, MOO1-14F, MOO1-14F/2,
Applicant
Fotona D.D.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 29, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Nd: YAG laser (1064 nm wavelength) in dermatology and other surgical areas: - Surgical incison, excision, vaporization, ablation and coagulation of soft tissue. All soft tissue is included, striated and smooth tissue, muscle, cartilage, mucous membrane, lymph vessels and nodes, organs and glands. It is further indicated for laser assisted lipolysis - Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of -PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin - Treatment of wrinkles - - Treatment of mild to moderate inflammatory acne vulgaris - - Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins - Endo Venous Laser Therapy of superficial incompetent tributary veins associated with varicose veins and varicosities Nd:YAG laser (1064 nm wavelength) in dentistry: - Excisional and incisional biopsies - Excision and vaporization of herpes simplex I and II - Exposure of unerupted teeth - Fibroma removal - Frenectomy and frenotomy - - Gingival troughing for crown impressions - -Gingivectomy - Gingivoplasty - - Gingival incision and excision - - Hemostasis - - Implant recovery - - Incision and drainage of abscess - - Laser assisted uvulopaletoplasty (LAUP) - - Operculectomy - - Oral papillectomies - - Pulpotomy and pulpotomy as an adjunct to root canal therapy - - Reduction of denture hyperplasia - - Reduction of gingival hypertrophy - - Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy - - Removal of post-surgical granulations - - Soft tissue crown lengthening - - Sulcular debridement or soft tissue curettage (removal of diseased or imflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility) - - Tissue retraction for impression - - Treatment of aphtous ulcers - - Vestibuloplasty
Device Story
Fotona XP Nd:YAG (1064 nm) laser system; uses flashlamp-pumped optical cavity; red diode aiming beam; fiber optic delivery to handpiece. Operated via footswitch. Two modes: PULSE (selective photothermolysis; larger spots; hair removal, lesions, acne, wrinkles) and QCW (Quasi-Continuous Wave; heating/incision; smaller spots; surgery, dentistry, lipolysis, endovenous therapy). User selects parameters (fluence, pulse duration, repetition rate, power) via keyboard interface. Used by clinicians in office/clinic settings. Output is laser energy for tissue interaction; enables precise surgical cutting, coagulation, and thermal tissue modification. Benefits include minimally invasive treatment options for various dermatological and dental conditions.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and established clinical indications for Nd:YAG laser systems.
Technological Characteristics
Nd:YAG laser (1064 nm); flashlamp-pumped; fiber optic delivery; red diode aiming beam; variable pulse duration (0.1-50 msec); repetition rate up to 100 Hz; max pulse energy 10 J. Modes: PULSE and QCW. Standalone system.
Indications for Use
Indicated for patients requiring soft tissue surgical procedures (incision, excision, ablation, coagulation, vaporization) in dermatology, general surgery, and dentistry; hair reduction (all skin types I-VI); treatment of wrinkles, acne vulgaris, vascular/pigmented lesions, and varicose veins. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Fotona Fidelis III Er: YAG/ Nd: YAG Laser System (K070355)
Millennium Periolase Nd: YAG Laser System (K030290)
{0}------------------------------------------------
# 5. 510(k) Summary
K090126
Fotona d.d. Submitter's Name: Stegne 7 SI-1210 Ljubljana, Slovenia Phone: +386 15009100 + 386 5009 200 Fax: Contact Person: Stojan Trošt, QA&RA Manager Phone: + 386 1 5009 299 E-mail: stojan.trost@fotona.com
JUN 2 9 2009
Date: January 14, 2009
Device Name: Trade name: Fotona XP Nd:YAG Laser System Family Nd:YAG Surgical Laser Common name: Classification name: Instruments, Surgical, Powered, Laser 79-GEX
## DEVICE DESCRIPTION
The Fotona XP Laser System Family is based on the Nd: YAG (1064 nm) laser technology. There is one optical cavity containing the Nd: YAG crystal. The Nd: YAG laser is activated by means of the use of flashlamps. After the cavity, a red diode aiming beam is reflected onto a coaxial beam path, using a beamsplitter assembly. The combined therapeutic and aiming beams are guided by an optical fiber delivery system to an optical handpiece. Optionally, the laser beams can be guided through fiber having SMA905 connector on the proximal end bare fiber on distal end. The user activates laser emission by means of a footswitch.
The Nd:YAG (1064 nm) laser can be operated with variable pulse durations in the range of 0.1-50 msec, with repetition rates up to 100 Hz and laser pulse energies up to 10 J. Depending on the type of treatment, the user can choose between a PULSE mode user interface, and a OCW (Ouasi-Continuous Wave) mode user interface. The laser operation is the same for both modes, the difference is only in the laser parameters that can be directly selected from the keyboard.
In the PULSE user interface mode the user can directly select the handpiece treatment spot size, laser fluence, pulse duration and pulse repetition rate. This mode is intended primarily for treatments at relatively small repetition rates, or in individual pulses, and at relatively high pulse energies. In this mode, a major expected treatment effect is the selective photothermolysis of target chromophores in soft tissue. For this reason it is more effective and safer that the energy is delivered to target chromophores in a relatively short time, in order not to cause damage to the surrounding tissue. Typically, the illuminated areas, or spotsizes are larger than 1 mm. A spotsize is determined by choosing and adjusting the handpiece. In this mode, the relevant and sufficient treatment parameters that have been accepted by the medical community are laser pulse energy, laser pulse fluence (energy over the spot-size area), pulse duration and pulse repetition rate. Recommended indications for use in this mode are:
- Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of । PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin
{1}------------------------------------------------
- Treatment of wrinkles
- Treatment of mild to moderate inflammatory acne vulgaris ।
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemaongiomae, warts, telangiectasiae, venus lake, leg veins and spider veins
In the QCW user interface mode the user can directly select the laser pulse repetition rate, pulse width and the average laser power. This mode is intended primarily for treatments at relatively high repetition rates and small pulse energies. In this mode, a major expected treatment effect is incision, excision and coagulation of tissue via heating of the tissue in close contact with the fiber delivery of a known diameter (typically smaller than 1.0 mm), and over a time typically longer than 0.5 sec. For this reason, in this mode, average power, repetition rate (frequency) and pulse-width are the relevant and sufficient treatment parameters, accepted by the medical community, for performing various dental or surgical procedures. Recommended indications for use in this mode are:
- Surgical incison, excision, vaporization, ablation and coagulation of soft tissue. All soft tissue is included, striated and smooth tissue, muscle, cartilage, meniscus, mucous membrane, lymph vessels and nodes, organs and glands. It is further indicated for laser assisted lipolysis
- Endo Venous Laser Therapy of superficial incompetent tributary veins associated with varicose veins and varicosities
- Indications for use in dentistry: -
- Excisional and incisional biopsies -
- Excision and vaporization of herpes simplex I and II ー
- Exposure of unerupted teeth —
- Fibroma removal -
- Frenectomy and frenotomy -
- Gingival troughing for crown impressions -
- Gingivectomy -
- Gingivoplasty -
- Gingival incision and excision -
- Hemostasis -
- Implant recovery -
- Incision and drainage of abscess -
- Laser assisted uvulopaletoplasty (LAUP) -
- -Operculectomy
- Oral papillectomies -
- Pulpotomy and pulpotomy as an adjunct to root canal therapy -
- Reduction of denture hyperplasia -
- Reduction of gingival hypertrophy -
- Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy
- Removal of post-surgical granulations -
- Soft tissue crown lengthening ー
- Sulcular debridement or soft tissue curettage (removal of diseased or imflamed soft tissue in । the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility)
- -Tissue retraction for impression
- Treatment of aphtous ulcers ।
- Vestibuloplasty -
{2}------------------------------------------------
## STATEMENT OF SUBSTANTIAL EQUIVALENCE
Fotona believes that its Fotona XP Laser System Family is substantially equivalent to other legallymarketed predicate devices. The Fotona XP laser system family is comparable to the following predicate devices in terms of its indications for use, technical specifications, operating performance features, and general design features:
## a) Fotona Fidelis III Er: YAG/ Nd: YAG Laser System (K070355), previously cleared for:
Nd: YAG laser (1064 nm wavelength) in dermatology and other surgical areas:
- Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of । PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin
- । Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins
- Treatment of wrinkles l
- Treatment of mild to moderate inflammatory acne vulgaris —
Nd:YAG laser (1064 nm wavelength) in dentistry:
- Excisional and incisional biopsies l
- Excision and vaporization of herpes simplex I and II ー
- Exposure of unerupted teeth -
- Fibroma removal -
- Frenectomy and frenotomy -
- Gingival troughing for crown impressions -
- Gingivectomy -
- -Gingivoplasty
- Gingival incision and excision -
- -Hemostasis
- Implant recovery ー
- Incision and drainage of abscess
- Laser assisted uvulopaletoplasty (LAUP) -
- -Operculectomy
- Oral papillectomies -
- -Pulpotomy and pulpotomy as an adjunct to root canal therapy
- Reduction of denture hyperplasia -
- Reduction of gingival hypertrophy ﺴﻤ
- Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy
- -Removal of post-surgical granulations
- Soft tissue crown lengthening -
- Sulcular debridement or soft tissue curettage (removal of diseased or imflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility)
- Tissue retraction for impression ।
- Treatment of aphtous ulcers ।
- Vestibuloplasty -
#### b) Millennium Periolase Nd: YAG Laser System (K030290) previously cleared for:
- Ablation, incision, excision, vaporization and coagulation of soft tissues using a contact fiber optic delivery system
{3}------------------------------------------------
- . c) Cooltouch Nd: YAG Laser System (K061618), Dornier Medilas D family (K073427) and Diomed Delta family of laser systems (K051996) previously cleared for:
- Endo Venous Laser Therapy of superficial incompetent tributary veins associated with varicose । veins and varicosities
- d) Cynosure SmartLipo and SmartLipo MPX Nd: YAG laser system (K062321, K080121), and Syneron LipoLite (eLipo) (K073715) previously cleared for:
- Surgical incison, excision, vaporization, ablation and coagulation of soft tissue. All soft tissue is included, striated and smooth tissue, muscle, cartilage, meniscus, mucous membrane, lymph vessels and nodes, organs and glands. It is further indicated for laser assisted lipolysis
- e) Candela GentleY AG Family of Laser Systems (K033172) previously cleared for:
- Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of — PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemaongiomae, warts, telangiectasiae, venus lake, leg veins and spider veins
- Coagulation and hemostasis of soft tissue l
- Treatment of wrinkles —
## f) Cooltouch Nd: YAG Laser System (K040131) and Laserscope Gemini Laser System (K034011) previously cleared for:
- Treatment of mild to moderate inflammatory acne vulgaris -
Details are provided in the Substantial Equivalence Discussion Section of this submission.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol on the right and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion on the left. The caduceus symbol is composed of three stylized lines that resemble a bird or a flame.
Public Health Service
JUN 2 9, 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Fotona D.D. % Stojan Trošt OA/RA Manager Stegne 7, 1210 Ljubljana Slovenia
Re: K090126
Trade/Device Name: Fotona XP Nd: YAG Laser System Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery
and in dermatology
Regulatory Class: II Product Code: GEX Dated: June 16, 2009 Received: June 19, 2009
Dear Stojan Trošt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Stojan Trošt
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of-Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
hereby yours,
Fofh Noto de D.i
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 4. Indications for Use Statement
510(k) Number (if known):
#### Device Name: Fotona XP Nd: YAG Laser System Family
Indications For Use:
Nd: YAG laser (1064 nm wavelength) in dermatology and other surgical areas:
- Surgical incison, excision, vaporization, ablation and coagulation of soft tissue. All soft tissue is included, striated and smooth tissue, muscle, cartilage, mucous membrane, lymph vessels and nodes, organs and glands. It is further indicated for laser assisted lipolysis
- Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of -PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin
- Treatment of wrinkles -
- Treatment of mild to moderate inflammatory acne vulgaris -
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins
- Endo Venous Laser Therapy of superficial incompetent tributary veins associated with varicose veins and varicosities
Nd:YAG laser (1064 nm wavelength) in dentistry:
- Excisional and incisional biopsies
- Excision and vaporization of herpes simplex I and II
- Exposure of unerupted teeth
- Fibroma removal
- Frenectomy and frenotomy -
- Gingival troughing for crown impressions -
- -Gingivectomy
- Gingivoplasty -
- Gingival incision and excision -
- Hemostasis -
- Implant recovery -
- Incision and drainage of abscess -
- Laser assisted uvulopaletoplasty (LAUP) -
- Operculectomy -
- Oral papillectomies -
- Pulpotomy and pulpotomy as an adjunct to root canal therapy -
- Reduction of denture hyperplasia -
- Reduction of gingival hypertrophy -
- Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy
- Removal of post-surgical granulations -
- Soft tissue crown lengthening
- Sulcular debridement or soft tissue curettage (removal of diseased or imflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index,
Neil Ree Qulm for mkm
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
5.10(k) Number K 090126
13/70
{7}------------------------------------------------
probe depth, attachment loss, and tooth mobility)
- Tissue retraction for impression
- Treatment of aphtous ulcers
- Vestibuloplasty
Prescription Use: __ X_ (21 CFR 801 Subpart D)
#### AND/OR
Over-The-Counter Use: (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Niel K.S. Oshun
Sec. mtm
(Division Sign-Off (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K090126
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.