INTELLIGENT ARTICULATING ENDOSCOPIC LINEAR CUTTER REVERSE WITH RELOADS

K083527 · Power Medical Interventions, Inc. · GDW · Dec 18, 2008 · General, Plastic Surgery

Device Facts

Record IDK083527
Device NameINTELLIGENT ARTICULATING ENDOSCOPIC LINEAR CUTTER REVERSE WITH RELOADS
ApplicantPower Medical Interventions, Inc.
Product CodeGDW · General, Plastic Surgery
Decision DateDec 18, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4750
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Intelligent Articulating Endoscopic Linear Cutters Reverse with Reloads have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses.

Device Story

Endoscopic linear cutter; used in general, gastrointestinal, gynecological, and thoracic surgery; operated by surgeons. Device delivers stapling and cutting action to create anastomoses. Powered mechanically by rotational energy; clamp shaft drives clamp screw to close anvil and compress tissue. Modification from predicate involves distal translation of jaw pivot point to enable reverse opening. Device features rigid shaft; electronics, power configuration, and internal gearing/transmission identical to predicate. Used in OR setting.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological comparison and design modification analysis.

Technological Characteristics

Rigid shaft endoscopic linear cutter; powered by rotational energy. Features include clamp shaft, clamp screw, and anvil mechanism for tissue compression. Modification: distally translated jaw pivot point for reverse opening. Electronics, power configuration, and internal gearing/transmission identical to predicate.

Indications for Use

Indicated for patients undergoing general or endoscopic surgery in gastrointestinal, gynecological, general abdominal, or thoracic procedures requiring resection, transection, or creation of anastomoses.

Regulatory Classification

Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K083527 # DEC 1 8 2008 Page 1/2 ### Section E ## Special 510(k) - Summary In Accordance with 21 CFR Section 807.92 Power Medical Interventions, Inc., is submitting the following 510(k) Summary: ## 1) Submitter Information: Power Medical Interventions, Inc. 2021 Cabot Blvd. Langhorne, PA 19047 Ph: 267-775-8151 Fax: 267-775-8123 Applicant: Barbara J. Whitman Date of Notification: November 26, 2008 Name of Device: 2) Trade Name: Intelligent Articulating Endoscopic Linear Cutters Reverse with Reloads Linear Staplers with Implantable Staples Common Name: - Staple, Implantable, GDW Classification Name: 3) Predicate Device: Intelligent Articulating Endoscopic Linear Cutters with Reloads, Power Medical Interventions, Inc., Langhorne, PA. REF i45, i455, i60, i6oS (Ko71708). Device Description: {1}------------------------------------------------ #### Device Modification 5) The Intelligent Articulating Endoscopic Linear Cutter Reverse with Reloads has identical technological features as the predicate device, i45/160 (K071708). The modification made to the device was that the pivot point for the jaws was translated distally, which enables the device to open in Reverse. Both the subject device and the predicate device deliver stapling/cutting action to create anastomoses. 0012 2 - Indications For Use 6) The Intelligent Articulating Endoscopic Linear Cutters Reverse with Reloads have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal and thoracic surgical procedures for resection, transection, and creation of anastomoses. - Comparison to Predicate Devices The Intelligent Articulating Endoscopic Linear Cutters Reverse functions the same as the predicate device, i45/160 (K071708). The electronics, power configuration, and internal gearing & transmission are common to both the Intelligent Articulating Endoscopic Linear Cutters Reverse and 160/145. The difference is that the pivot point for the stapling jaws was translated distally in order to enable the device to open in reverse. The shafts on both the subject device, Intelligent Articulating Endoscopic Linear Cutters Reverse, and the predicate device, i45/i6o, are both rigid. Both devices are powered mechanically by rotational energy. In both the predicate and the subject device, a clamp shaft drives a clamp screw, which causes the anvil to close and compress tissue. The same is true for firing: {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable. ### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## DEC 1 8 2008 Power Medical Interventions, Inc. % Ms. Barbara J. Whitman Director, Regulatory Affairs 2021 Cabot Boulevard West Langhorne, Pennsylvania 19047 Re: K083527 Trade/Device Name: Intelligent Articulating Endoscopic Lincar Cutter Reverse with Reloads Regulation Number: 21 CFR 878.4750 Regulatory Class: II Product Code: GDW Dated: November 26, 2008 Received: November 28, 2008 Dear Ms. Whitman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ## Page 2 - Ms. Barbara J. Whitman This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Mark M. Mulhern Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Section D ## Indications for Use 510(k) Number (if known): K083527 Device Name: Intelligent Articulating Endoscopic Linear Cutter Reverse with Reloads Indications For Use: The Intelligent Articulating Endoscopic Linear Cutters Reverse with Reloads have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdomihal and thoracic surgical procedures for resection, transection, and creation of anastomoses. Prescription Use (Part 21 CFR 801 Subpart D) 家庭的多少家 出家 ANDIOR ver-The-Counter-Use (21 CFR 807 Subpart C) elease do not write below this line-continue onland ther page NEEDED's > ffice of Device Evaluation (ODE) Concurrence of > > (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** 1083527-
Innolitics
510(k) Summary
Decision Summary
Classification Order