The DSA 2000ex device is used in vascular imaging applications. During X-ray exposures, the DSA 2000ex is used to acquire video images from the video display chain provided by the X-ray manufacturer's system. The images are stored in the DSA 2000ex solid state memory, and written to the hard disk medium. Images are processed in real-time to provide increased image usability. The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, image rotation and pixel shifting. The Eigen DSA 2000ex device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists, and cardiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions. Additional functions include allowing measurements to be made for quantizing stenosis and guidance of catheters in the Roadmapping mode.
Device Story
DSA 2000ex is a digital subtraction angiography system for vascular imaging. It acquires video signals from X-ray manufacturer systems during exposure; stores data in solid-state memory and hard disk. Real-time processing includes subtraction, window/level adjustment, noise reduction, landscaping, pixel shifting, and image rotation. Used in cardiology and radiology labs by clinicians to enhance diagnostic images, perform quantitative coronary angiography (QCA), and guide catheters via roadmapping. Output is displayed for clinician review to assist in clinical decision-making and procedural guidance.
Clinical Evidence
Bench testing only. Performance verified through system verification, validation, and benchmarking using test images and phantoms incorporating simulated stenosis.
Technological Characteristics
System for radiological image processing. Features solid-state memory and hard disk storage. Functions include real-time subtraction, window/level adjustment, noise reduction, landscaping, pixel shifting, and image rotation. Connectivity via X-ray lab video interface. Software-based processing.
Indications for Use
Indicated for use in vascular imaging applications in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists and cardiologists. Used for image capture, playback, archiving, stenosis quantification, and catheter guidance via roadmapping.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
## 510(k) Summary
# AUG 2 9 2008
1082341
| Submitter's Name: | Eigen Inc. | |
|----------------------------|-----------------------------------------------------------------------------------------------------|---------------|
| Submitter's Address: | 13366 Grass Valley Avenue, Grass Valley, CA 95945 | |
| Submitter's Telephone: | 530-274-1240 | |
| Contact Name: | Animesh Khemka | |
| Date Summary was Prepared: | July 23, 2008 | |
| Trade or Proprietary Name: | DSA 2000ex | |
| Common or Usual Name: | Digital Subtraction Angiography | |
| Classification Name: | System, Image Processing, Radiological, LLZ<br>Picture Archiving and Communications, 21CFR 892.2050 | |
| Predicate Devices: | Device Name | 510(k) Number |
| | Digital 8 | K901956 |
| | DSA 2000 | K063846 |
### Description of the Device and Summary of the Technological Characteristics
The DSA 2000ex device is used in vascular imaging applications. During X-ray exposures, the DSA 2000ex is used to acquire video images from the video display chain provided by the X-ray manufacturer's system. The images are stored in the DSA 2000ex solid state memory, and written to the hard disk medium. Images are processed in real-time to provide increased image usability. The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, image rotation and pixel shifting. The Eigen DSA 2000ex device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions. Additional functions include allowing measurements to be made for quantizing stenosis and guidance of catheters in the Roadmapping mode.
#### Substantial Equivalence
The Eigen DSA 2000ex is substantially equivalent to other Eigen imaging devices: the Digital 8 and DSA 2000. The DSA 2000ex is a modification of Eigen's DSA 2000 cleared under special 510(k) # K063846 and Eigen's Digital 8, which was cleared under 510(k) # K901956. The modifications made to the DSA 2000 and Digital 8 do not alter the intended use or the fundamental scientific technology of the device.
{2}------------------------------------------------
The DSA 2000ex and its predicates (the Digital 8 and DSA 2000) enhance and process images by performing the following functions:
- E Noise Reduction
- I Window/Level Feature
- 트 Edge Enhancement
- Image Subtraction l
- 트 Still Image Display
- X-ray Lab Command Interface I
- 트 X-ray Lab Video Interface
- Data Archiving l
- Data and Space Management l
- Study Information (e.g., patient, site, and date) 트
- L Configuration Menu
- Still Image Acquisition and Display I
- I Sequence Image Acquisition and Display
- i Automated Contrast
- 트 Landscaping
- I Pixel Shifting
- Modality Work-List Support 트
- I Mix (also called Roadmapping)
- Quantification (also called Quantitative Coronary Angiography 트 (QCA))
The DSA 2000ex device also offers the additional feature of image rotation which is not in the predicate devices.
### Testing and Performance Data
All product and engineering specifications were verified and validated. Test images as well as test phantoms incorporating simulated stenosis were developed and used to verify system performance through verification, validation and benchmarking.
### Conclusion
6
The results of comparing the intended use, function, technological characteristics, mode of
{3}------------------------------------------------
operation and specifications of the DSA 2000ex with those of the predicate devices demonstrate that the DSA 2000ex is substantially equivalent to existing products in the market today.
6
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circular fashion around the emblem.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 9 2008
Eigen, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K082341
Trade/Device Name: DSA 2000ex Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: II Product Code: LLZ Dated: August 14, 2008 Received: August 15, 2008
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR | | |
|-----------------|----------------------------------|--------------|
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K082341
Indications for Use
510(k) Number (if known): pending
Device Name: "DSA 2000ex"
Indications for Use:
The DSA 2000ex device is used in vascular imaging applications. During X-ray exposures, the DSA 2000ex is used to acquire video images from the video display chain provided by the X-ray manufacturer's system. The images are stored in the DSA 2000ex solid state memory, and written to the hard disk medium. Images are processed in real-time to provide increased image usability. The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, image rotation and pixel shifting. The Eigen DSA 2000ex device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists, and cardiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions. Additional functions include allowing measurements to be made for quantizing stenosis and guidance of catheters in the Roadmapping mode.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign Off) | |
|-----------------------------------------|-----------------------------|
| Division of Reproductive, Abdominal and | |
| Radiological Devices EIGEN INC. | K082341 |
| 510(k) Number | SPECIAL 510(K) - DSA 2000ex |
| | July 23, 2008 |
Page _ 1 _ of _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1 _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.