EIGEN DSA 2000

K063846 · Eigen · LLZ · Jan 26, 2007 · Radiology

Device Facts

Record IDK063846
Device NameEIGEN DSA 2000
ApplicantEigen
Product CodeLLZ · Radiology
Decision DateJan 26, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.2050
Device ClassClass 2

Indications for Use

The Eigen DSA 2000 product is used in vascular imaging applications. During X-ray exposures, the DSA 2000 is used to acquire video images from the video display chain provided by the X-ray manufacturer's system. The images are stored in the DSA 2000 solid state memory, and written to the hard disk medium. Images are processed in real-time to provide increased image usability. The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, and pixel shifting. The Eigen DSA 2000 device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists and cardiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions.

Device Story

Eigen DSA 2000 is a real-time video acquisition device for existing X-ray systems; captures video signals from X-ray display chain; stores images in solid-state memory and hard disk. Device performs real-time image processing including digital subtraction, window/level adjustments, noise reduction, landscaping, and pixel shifting. Operated by radiologists and cardiologists in clinical labs; provides enhanced images for diagnostic review. Output transmitted to DICOM workstations or PACS. Benefits include improved image usability and streamlined capture, playback, and archiving workflows.

Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through module and system-level testing against established protocols.

Technological Characteristics

Platform: HP/Intel hardware; OS: Microsoft Windows XP. Connectivity: DICOM 3.0XA standard for lossless images. Features: Real-time video acquisition, digital subtraction, window/level adjustment, noise reduction, landscaping, pixel shifting. Data storage: Solid-state memory and hard disk.

Indications for Use

Indicated for vascular imaging applications in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists and cardiologists.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ eigen D63846 ## Special 510(k) Summary | Manufacturer: | Eigen LLC<br>13366 Grass Valley Avenue<br>Grass Valley, CA 95945<br>USA<br>Phone (530) 274-1240<br>(888) 924-2020<br>FAX (530) 274-3656 | |---------------|-----------------------------------------------------------------------------------------------------------------------------------------| |---------------|-----------------------------------------------------------------------------------------------------------------------------------------| JAN 26 2007 - Contact Person: Ms. Natalie J. Kennel Consultant NJK & Associates, Inc. 13721 Via Tres Vista San Diego, CA 92129 Phone: (858) 705-0350 Fax: (858) 764-9739 email: Natalie.Kennel@sbcglobal.net Date Prepared: December 21, 2006 ## DEVICE INFORMATION Trade/Proprietary Name: Eigen Digital Subtraction Angiography (DSA) 2000, also referred to as the Eigen DSA 2000 | Common/Classification Name: | Image Processing System<br>21 CFR 892.2050<br>Class II<br>Device Product Code: LLZ | |-----------------------------|------------------------------------------------------------------------------------| |-----------------------------|------------------------------------------------------------------------------------| #### Predicate Device: The Eigen DSA 2000 is substantially equivalent to another Eigen Imaging Device: the Digital 8. The DSA 2000 is a modification of Eigen's Digital 8, which was cleared under 510(k) K901956. The modifications made to the DSA 2000 do not alter the intended use or the fundamental scientific technology of the device. Both the Eigen DSA 2000 and its predicate, the Digital 8, chhance images by performing the following functions: Sales Comment Eigen DSA 2000 - Special 510(k) December 21, 2006 Company Confidential Page 19 of 76 {2}------------------------------------------------ - Noise Reduction - I Window/Level Feature - Edge Enhancement 트 - 트 Image Subtraction - I Still Image Display - 트 X-ray Lab Command Interface - 미 X-ray Lab Video Interface - Data Archiving 제 - 트 Data and Space Management - I Study Information (e.g., patient, site, and date) - 트 Configuration Menu - Still Image Acquisition and Display - 로 Sequence Image Acquisition and Display, and - Automated Contrast l The DSA 2000 device also offers the following additional features: Landscaping, Pixel Shifting, and Modality Work-List Support which are not in the predicate device. The DSA 2000 device does not offer the following functions which were in the original device: Magnification, Panning, Mix, Quantification and test patterns because it has been determined that there is not a customer demand for these features. #### Product Description: The Eigen Digital Subtraction Angiography® (DSA) is a real-time video acquisition device that can be added to an existing standard line-rate X-ray system and provides creation of photos and real-time digital subtraction taken from a mask image. The DSA acquires and transmits data to a DICOM workstation or PAC system. The data will then be available for display. The DSA output conforms to the DICOM 3.0XA Standard for lossless images. The Eigen DSA is assembled on a Hewlett Packard (HP)/Intel platform and uses the Microsoft Windows XP® operating system. #### Intended Use: The Eigen DSA 2000 product is used in vascular imaging applications. During X-ray exposures, the DSA 2000 is used to acquire video images from the video display chain provided by the X-ray manufacturer's system. The images are stored in the DSA 2000 solid state memory, and written to the hard disk medium. Images are processed in real-time to provide increased image usability. The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, and pixel Shifting. The Eigen DSA 2000 device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists and cardiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions. Eigen DSA 2000 Special 510(k) December 21, 2006 Company Confidential Page 20 of 76 {3}------------------------------------------------ ## Test Discussion Testing was performed at the module and system level according to written test protocols established before the testing was conducted. Test results were reviewed by designated technical professionals before release of the software. ## Conclusion: The test results support the conclusion that the DSA 2000 is substantially equivalent to its predicate device, D8. Actual device performance as tested internally conforms to the system requirements. Eigen DSA 2000 Special 510(k) December 21, 2006 Company Confidential Page 21 of 76 {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol with three curved lines, resembling a person with outstretched arms, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and is oriented to follow the curve of the circle. Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 Eiden, LLC % Ms. Natalie J. Kennel Principal NJK & Associates, Inc. 13721 Via Tres Vista SAN DIEGO CA 92129 JAN 2 6 2007 Re: K063846 Trade/Device Name: Eigen Digital Subtraction Angiography (DSA) 2000 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: II Product Code: LLZ and IZI Dated: December 21, 2006 Received: December 27, 2006 Dear Ms. Kennel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the radications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, or and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act, Tro general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulterviton. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/3/Picture/10 description: The image shows a circular logo with the text "1906-2006" at the top and "FDA Centennial" at the bottom. The letters "FDA" are prominently displayed in the center of the logo. There are three stars below the word "Centennial". The logo appears to be a commemorative emblem for the centennial anniversary of the Food and Drug Administration. Protecting and Promoting Public Health {5}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincerely vours. Nancy C. Hogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use Statement 14 01 3846 510(k) Number: Device Name: Eigen Digital Subtraction Angiography (DSA) 2000 #### Indications For Use: The Eigen DSA 2000 product is used in vascular imaging applications. During X-ray exposures, the DSA 2000 is used to acquire video images from the video display chain provided by the X-ray manufacturer's system. The images are stored in the DSA 2000 solid state memory, and written to the hard disk medium. Images are processed in real-time to provide increased image usability. The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, and pixel shifting. The Eigen DSA 2000 device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists and cardiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over the Counter Use (per 21 CFR 801.109) David h. Ingram (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...