The OrthoPediatrics PediPlate system is used for adult and pediatric patients as indicated for pelvic, small and long bone fractures. Indications for use include fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerous, ulna, middle hand and middle foot bones, treatment of the calcaneal; hip arthodesis, and provisional hole fixation; as well as for redirecting the angle of growth of long bones. This is useful for gradually correcting angular deformities in growing children. Specific pediatric conditions/diseases for which the devices will be indicated include: - Valgus, varus, or flexion, extension deformities of the knee (femur and/or tibia) . - Valgus, varus, or plantar flexion deformities of the ankle . - Valgus or varus deformities of the elbow (humerus) . - Radial or ulnar deviation, flexion or extension deformities of the wrist (radius) .
Device Story
System consists of contoured, low-profile metallic plates (O, I, and H-plates) and screws (cortical and cannulated) for pediatric and adult bone fixation. Plates feature central alignment marks for radiographic placement relative to growth plates. Used by surgeons in clinical settings to provide immediate stability and temporary fixation during bone healing or to redirect long bone growth in children. Output is physical stabilization of bone segments; clinical decision-making relies on surgeon assessment of fracture or deformity. Benefits include gradual correction of angular deformities and stabilization of fractures.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Manufactured from 316L stainless steel meeting ASTM F138 or ASTM F139. System includes plates (O, I, H-profiles) and screws (cortical/cannulated). Mechanical fixation device; no software or electronic components.
Indications for Use
Indicated for adult and pediatric patients with pelvic, small, and long bone fractures (tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, middle hand/foot bones, calcaneal), hip arthrodesis, provisional hole fixation, and angular deformity correction in growing children (knee, ankle, elbow, wrist).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
## JUL 2 2 2008
### SUMMARY OF SAFETY AND EFFECTIVENESS
| NAME OF FIRM: | OrthoPediatrics, Corp.<br>210 N. Buffalo Street<br>Warsaw, Indiana 46580<br>Establishment Registration No.: 9102640 |
|-----------------|---------------------------------------------------------------------------------------------------------------------|
| 510(K) CONTACT: | Gary Barnett<br>VP-Regulatory, Engineering & Operations<br>Tel: (574) 268-6379<br>Fax: (574) 269-3692 |
| TRADE NAME: | OrthoPediatrics PediPlatesTM System |
| COMMON NAME: | Plate, Fixation, Bone |
| CLASSIFICATION: | Single/multiple component metallic bone fixation<br>appliances and accessories: Class II per 21 CFR<br>§888.3030 |
DEVICE PRODUCT CODE(S): HRS and HWC
#### SUBSTANTIALLY EQUIVALENT DEVICES:
- Guided Growth Plate (K031493), Marketed by Orthofix .
- TC-100 Plating and Screw System (K993106), Smith & Nephew .
- PediPlate TM Plating System (K073344), OrthoPediatrics .
- Distal Tibia T-Plates (K080522), Synthes .
#### DEVICE DESCRIPTION:
The two and four-hole plates feature a contoured mid-section and low profile for pediatric use. There is an alignment mark in the center of the O and I-Plates to aid accurate placement of the device relative to the growth plate that is localized radiographically. The plates are available in various sizes to accommodate variations in bone size and geometry. The plate is transfixed to bone using two to four screws, depending on which plate is selected.
- Materials: The devices are manufactured from 316L stainless steel, which meet . ASTM F138 or ASTM F139.
- . Function: The system functions to provide immediate stability and temporary fixation during the natural healing process.
The OrthoPediatrics PediPlate ™ System is intended for use in the treatment of pelvic, small and long bone fractures, as well as deformity corrections of pediatric patients' long
{1}------------------------------------------------
bones, which includes osteotomies and redirecting the angle of growth of children's long bones. The system includes cortical and cannulated screws, and specialty plates, referred to as O, I, and H -Plates.
#### INDICATIONS FOR USE:
The OrthoPediatrics PediPlate system is used for adult and pediatric patients as indicated for pelvic, small and long bone fractures. Indications for use include fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerous, ulna, middle hand and middle foot bones, treatment of the calcaneal; hip arthodesis, and provisional hole fixation; as well as for redirecting the angle of growth of long bones. This is useful for gradually correcting angular deformities in growing children.
Specific pediatric conditions/diseases for which the devices will be indicated include:
- Valgus, varus, or flexion, extension deformities of the knee (femur and/or tibia) .
- Valgus, varus, or plantar flexion deformities of the ankle .
- Valgus or varus deformities of the elbow (humerus) .
- Radial or ulnar deviation, flexion or extension deformities of the wrist (radius) .
#### BASIS FOR SUBSTANTIAL EQUIVALENCE:
OrthoPediatrics believes that this system is substantially equivalent to the legally marketed predicate devices based on similarities in design, materials, and indications.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is facing to the right, and its wings are composed of three thick lines. The text is in all capital letters and is evenly spaced around the circle.
# FEB 19 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OrthoPediatrics, Corporation % Mr. Gary Barnett Vice President, Regulatory, Engineering & Operations 210 N. Buffalo Street Warsaw, Indiana 46580
Re: K081407 Trade/Device Name: OrthoPediatrics PediPlates™ System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC, OBT Dated: May 14, 2008 Received: May 19, 2008
Dear Mr. Barnett:
This letter corrects our substantially equivalent letter of July 22, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Gary Barnett
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N Milken
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K081407
#### Device Name: OrthoPediatrics Plating System
The OrthoPediatrics PediPlate system is used for adult and pediatric patients as indicated for pelvic, small and long bone fractures. Indications for use include fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerous, ulna, middle hand and middle foot bones, treatment of the calcaneal; hip arthodesis, and provisional hole fixation; as well as for redirecting the angle of growth of long bones. This is useful for gradually correcting angular deformities in growing children.
Specific pediatric conditions/diseases for which the devices will be indicated include:
- . Valgus, varus, or flexion, extension deformities of the knee (femur and/or tibia)
- Valgus, varus, or plantar flexion deformities of the ankle .
- Valgus or varus deformities of the elbow (humerus) ●
- Radial or ulnar deviation, flexion or extension deformities of the wrist (radius) .
Prescription Use _____________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark N. Milken
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K081407
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.