K081082 · Medtronic, Inc. · FZP · May 7, 2008 · General, Plastic Surgery
Device Facts
Record ID
K081082
Device Name
U-CLIP UNI-FIRE, MODEL M65
Applicant
Medtronic, Inc.
Product Code
FZP · General, Plastic Surgery
Decision Date
May 7, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The U-CLIP™ Uni-Fire is intended for endoscopic and non-endoscopic general soft tissue and prosthetic material approximation/attachment and/or ligation and the creation of anastomosis in blood vessels, grafts and other tubular structures; including use in cardiovascular and coronary artery bypass grafting procedures.
Device Story
U-CLIP™ Uni-Fire is a self-closing implantable clip for tissue/prosthetic approximation and vascular anastomosis. Device consists of Nitinol clip constrained in open position within stainless steel hypotube. Surgeon places clip; releases from side slot of hypotube. Upon release, Nitinol clip returns to closed configuration, securing tissue layers. Eliminates knot tying; facilitates interrupted suture technique. Used in endoscopic and non-endoscopic procedures, including cardiovascular and coronary artery bypass grafting. Operated by surgeons in clinical settings.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed functional characteristics, including clip strength and deployment, are equivalent to the predicate device. All test data satisfied documented product and performance specifications.
Technological Characteristics
Implantable self-closing clip. Material: Implantable grade Nitinol. Delivery system: Stainless steel hypotube. Mechanism: Shape-memory alloy clip constrained in open position, self-closing upon release. Dimensions: Same wire diameter, length, and closed diameter as predicate. Non-powered, mechanical device.
Indications for Use
Indicated for general soft tissue and prosthetic material approximation, attachment, or ligation, and creation of anastomosis in blood vessels, grafts, and tubular structures, including cardiovascular and coronary artery bypass grafting. Contraindicated for tubal ligation.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
K081082 page 1/2
MAY - 7 2008
U-CLIP™ Uni-Fire, Model M65
## 510(k) SUMMARY of Safety and Effectiveness
## l. Applicant Information:
| Date Prepared: | April 14, 2008 |
|-----------------------------------|---------------------------------------------------------|
| Submitter: | Medtronic, Inc. |
| Address: | 710 Medtronic Parkway, NE<br>Minneapolis, MN 55432-5604 |
| Establishment<br>Registration No. | 2135394 |
| Contact Person: | David D. Cox, Ph.D.<br>Regulatory Consultant |
| Telephone Number: | (651) 247-7900 |
| Fax Number: | (763) 391-9259 |
## II. Device Information:
| Trade Name: | U-CLIP™ Uni-Fire |
|--------------|------------------|
| Common Name: | Implantable clip |
| Classification Name: | Clip, Implantable |
|----------------------|---------------------------|
| Classification: | Class II, 21 CFR 878.4300 |
| Product Code: | FZP |
- Medtronic U-CLIP™, Model NC65 Predicate Device: 510(k) No. K062057, Reg. No. 878.4300; Product Code: FZP
Device Intended Use: The U-CLIP™ device is intended for endoscopic and non-endoscopic general soft tissue and prosthetic material approximation/attachment and/or ligation and the creation of anastomosis in blood vessels, grafts and other tubular structures; including cardiovascular and coronary artery bypass grafting procedures; including use in cardiovascular and coronary artery bypass grafting procedures.
{1}------------------------------------------------
K081082
U-CLIP™ Uni-Fire, Model M65
Device Description: The U-CLIP™ is a self-closing clip for anastomosis and tissue and prosthetic material approximation or attachment applications. The Model M65 U-CLIP" Uni-Fire device consists of a self-closing Nitinol clip that is constrained in an open position in a stainless steel hypotube until released by the surgeon after placement. The clip is released from a slot in the side of the hypotube. After release, the arms of the clip attempt to regain their preferred, closed configuration, thereby holding the tissue layers together. This design allows precise placement of clips prior to closure, and facilitates an interrupted "suture" technique by eliminating knot tying. The device is manufactured with a standard implantable grade of Nitinol.
- The U-CLIP™ Uni-Fire is intended for endoscopic and non-Intended Use: endoscopic general soft tissue and prosthetic material approximation/attachment and/or ligation and the creation of anastomosis in blood vessels, grafts and other tubular structures; including use in cardiovascular and coronary artery bypass grafting procedures.
- Contraindications: Do not use for tubal ligation.
Comparison to
Predicate Device(s): The U-CLIP™ Uni-Fire device is substantially equivalent to the U-CLIP" Model NC65 device cleared in K062057 and other U-CLIPs cleared in K31623, K024366, K060400, K023125, K021407, K013664, K012317, K994160, and K971588 in terms of materials, use and application. The new clip is released from the side of the hypotube, just like the predicate device, but it has a flattened end to allow it to be mechanically held prior to release. The new clip is made of the same Nitinol alloy, with the same wire diameter, overall length and closed clip diameter as the predicate device.
- Test Data: Verification and validation testing confirms that functional characteristics are substantially equivalent to the predicate device cited. This included clip strength and clip deployment. All test data obtained satisfied the documented product and performance specifications.
- Based upon the technical information, intended use, in vitro, in Summary: vivo, and clinical performance information provided in previous pre-market notifications, the U-CLIP™ Uni-Fire, Model M65 described in this submission has been shown to be substantially equivalent to currently marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 7 2008
Medtronic, Inc. % Innovatrix, Inc. David D. Cox, Ph.D. 3051 S. Newcombe Way Lakewood, Colorado 80227
Re: K081082
Trade/Device Name: U-CLIP® Uni-Fire, Model M65 Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: II Product Code: FZP Dated: April 14, 2008 Received: April 16, 2008
Dear Dr. Cox:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - David D. Cox, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Milliman
- Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known):
KO81082
U-CLIP® Uni-Fire, Model M65 Device Name:
Indications for Use:
The U-CLIP™ Uni-Fire is intended for endoscopic and non-endoscopic general soft tissue and prosthetic material approximation/attachment and/or ligation and the creation of anastomosis in blood vessels, grafts and other tubular structures; including use in cardiovascular and coronary artery bypass grafting procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Phil R. Ogden for Mayor
Division of General, Restorative, and Neurological Devices
**510(k) Number** K081082
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.