The Duo-Max Blood pressure and Glucose Monitor, Model HMF-100 consists of a meter with wrist cuff and test strips. The system is intended for use in the quantitative measurement of glucose in whole blood taken from the fingertip. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes mellitus, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for use on neonates. Also the system measures systolic and diastolic blood pressure and pulse rate from adult's wrist in the home care environment. The device employs a wrist cuff and the oscillometric metric method of measurement.
Device Story
Duo-Max HMF-100 combines blood pressure (BP) monitor and blood glucose (BG) monitor in single wrist-mounted unit. BP component uses wrist cuff with oscillometric measurement method; BG component uses test strips for capillary whole blood analysis. User interface includes LCD displaying systolic/diastolic BP, pulse rate, date, and time. Device stores 180 BP and 270 BG readings. Used in home care or clinical settings by patients or healthcare professionals. Provides quantitative glucose and BP/pulse data to aid diabetes management and BP monitoring.
Clinical Evidence
Bench testing only. Device conforms to ANSI/AAMI SP-10 (BP), IEC 60601-1 (Safety), and IEC 60601-1-2 (EMC) requirements. No clinical trial data reported.
Technological Characteristics
Wrist-mounted unit with integrated cuff. Oscillometric BP measurement; test strip-based glucose measurement. Memory: 180 BP readings, 270 BG readings. Pulse rate range: 20-199 bpm. Battery-powered. Standards: ANSI/AAMI SP-10, IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for quantitative measurement of capillary whole blood glucose in adults for diabetes management; not for neonates, critically ill, dehydrated, or hyperosmolar patients; not for screening or diagnosis of diabetes. Also indicated for non-invasive measurement of systolic/diastolic blood pressure and pulse rate in adults in home care settings.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Blood Glucose and Blood Pressure Monitor System, Model BGP-100 Sein Electronics Co., Ltd (k052108)
Infopia GlucolabTM Blood Glucose Monitoring System (k051285)
Related Devices
K080853 — DUO-CARE BLOOD GLUCOSE AND BLOOD PRESSURE MONITOR SYSTEM · Genexel-Sein, Inc. · Feb 20, 2009
K092875 — ECHO TD-3227 DUAL BLOOD PRESSURE/BLOOD GLUCOSE METER · Taidoc Technology Corporation · Aug 13, 2010
K052108 — SEIN BLOOD GLUCOSE AND BLOOD PRESSURE MONITOR SYSTEM, MODEL BGP-100 · Sein Electronics Co., Ltd. · Dec 7, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
**EXHIBIT #1**
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: January 11, 2008
FEB 1 3 2009
#### 1. Submitter:
| Name: | HuBDIC Co., Ltd.<br>195-42, Anyang 7-dong, Manan-gu, Anyang-si,<br>Gyeonggi-do, Korea |
|----------|---------------------------------------------------------------------------------------|
| Contact: | In-Young An<br>Phone +82-31-441-8671<br>Fax +82-31-442-4994 |
#### 2. Device:
| Proprietary Name: | Duo-Max Blood Pressure and Glucose Monitor, Model<br>HMF-100 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Blood Pressure Monitor, Blood Glucose Monitor,<br>Blood glucose test strip and quality control material<br>System, test, blood glucose, over the counter |
| Classification Name: | Glucose oxidase, glucose<br>Single (specified) analyte controls |
| Classification: | 21 CFR 862.1345 |
| Product Code: | NBW, CGA, JJX, DXN |
#### 3. Predicate Device:
Blood Glucose and Blood Pressure Monitor System, Model BGP-100 Sein Electronics Co., Ltd K052108 Infopia Glucolab™ Blood Glucose Monitoring System K051285
#### 4. Description:
HuBDIC's Duo- Max Blood Pressure and Glucose Monitor System, Model HMF-100 combines the function of a blood pressure meter and a blood glucose monitoring system in one unit. Supplies with the meter are test strips, lancets, Lancing device, storage case, batteries and log book.
HMF-100 adopts the wrist type cuff for blood pressure meter part. The cuff and control unit are combined into a single wrist-mounted assembly. The user interface pane has power switch, timer switch, memory switch for blood pressure meter part and LCD for displaying the systolic blood pressure, diastolic blood pressure, pulse rate, date and time. This device has the memory function that permits memory and display of the 180 most recent measurement results.
HuBDIC Co., Ltd.
{1}------------------------------------------------
#### 5. Indications for use:
The Duo-Max Blood pressure and Glucose Monitor, Model HMF-100 consists of a meter with wrist cuff and test strips. The system is intended for use in the quantitative measurement of glucose in whole blood taken from the fingertip. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes mellitus, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for use on neonates. Also the system measures systolic and diastolic blood pressure and pulse rate from adult's wrist in the home care environment. The device employs a wrist cuff and the oscillometric metric method of measurement.
#### 6. Comparison of Technological Characteristics with Predicate:
The Duo-Max Blood Pressure and Glucose Monitor, Model HMF-100 has the same general design and performance characteristics as the predicate device from Sein. The main difference is the physical size, shape and weight as seen below:
| ltem | Duo-Max Blood Pressure and Glucose<br>Monitor, model HMF-100 | Sein, Blood Glucose and Blood Pressure<br>Monitor System, Model BGP-100 |
|-------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------------|
| Pulse Rate | 20 ~ 199 pulse/min | 40 ~ 199 pulse/min |
| Number of<br>Readings stored<br>in Memory | 180 times (Blood Pressure Part)<br>270 times (Blood Glucose Part) | 60 times (Blood Pressure Part)<br>150 times (Blood Glucose Part) |
| Dimension | 64(W) x 84(W) x 31(H)(mm) | 62(W) x 83(W) x 33(H)(mm) |
| Weight | 163g | 192g |
#### 7. Performance Data:
In terms of operating specification, Safety & EMC requirements, the device conforms to applicable standards including ANSI/AAMI SP-10, IEC 60601-1 and IEC 60601-1-2 requirements.
The Blood Pressure and Glucose Monitor, Model HMF-100 tested and met all relevant requirements of the aforementioned tests.
#### 8. Conclusion:
We have demonstrated that the Blood Pressure and Glucose Monitor, Model HMF-100 is as safe and effective as the predicate devices based on our performance testing results as well as with the risk analysis supplied with this submission.
This submitter concludes that the Blood Pressure and Glucose Monitor, Model HMF-100 is therefore substantially equivalent to the predicate devices.
HuBDIC Co., Ltd.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "Public" on the first line, "Health" on the second line, and "Service" on the third line. The text is black against a white background.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
HuBDIC Co. Ltd. c/o Maria Griffin, MDI Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021
FEB 1 3 2009
Re: k080636
Trade Name: Duo-Max Blood Pressure and Glucose Monitor, Model HMF-100 Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW, CGA, DXN, JJX Dated: January 30, 2009 Received: February 2, 2009
### Dear Ms. Griffin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to sect additional commis . Enisting major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CEA), Econo 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page - 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free nuïnber. (800) 638-2041 or (301) 443-6597 or at its Internet address - http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Corg C. He
Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{4}------------------------------------------------
## Indication for Use
:
上一篇:
### 510(k) Number (if known): k080636
Device Name: Blood Pressure and Glucose Monitor, Duo-Max™ (Model HMF-100)
Indication For Use:
2013
The Blood pressure and glucose monitor. Duo-Max™ consists of a meter with wrist cuff and glucose test strips. The glucose test system is intended for use in the quantitative measurement of glucose in capillary whole blood taken from the fingertip, dorsal hand, ventral palm, upper arm, forearm, calf and thigh. Glucose testing is done outside the body (In Vitro diabetes mellitus, or in clinical settings by healthcare professionals. Glucose testing system is not to be used for the diagnosis or screening of diabetes. It is not intended for use on neonates. Also the blood pressure test system measures systolic and diastolic blood pressure and pulse rate from adult's wrist in the home care environment. The device employs a wrist cuff and the oscillomentric metric method of measurement.
Prescription Use (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use_v (21 CFR Part 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carol C. Benson
Division Sign Off Office of In Vitro Diagnostic Device Evaluation and Safety .. .
510(k) K080636
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.