ECHO TD-3227 DUAL BLOOD PRESSURE/BLOOD GLUCOSE METER
Applicant
Taidoc Technology Corporation
Product Code
NBW · Clinical Chemistry
Decision Date
Aug 13, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Clever Chek TD-3217 Blood Glucose plus Blood Pressure Monitoring System is intended for use in the home or clinical setting for the quantitative measurement of glucose in fresh capillary whole blood from the finger, palm, forearm, upper arm, calf, and thigh. It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control programs. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The blood pressure monitoring function is intended for use by adults to measure systolic and diastolic blood pressure and pulse rate.
Device Story
Dual-function meter measuring blood glucose via electrochemical biosensor (glucose oxidase) and blood pressure/pulse via oscillometric wrist cuff. Input: fresh capillary whole blood (glucose) and wrist pressure oscillations (BP). Output: digital display of glucose concentration, systolic/diastolic BP, and pulse rate. Used by patients at home or healthcare professionals in clinics. Aids monitoring of diabetes control programs. Cuff circumference 5.31"-8.66".
Clinical Evidence
Bench testing only. Verification and validation activities performed per design control requirements (21 CFR 820.30) to confirm modifications met predetermined acceptance criteria. Risk analysis conducted using ALARP methodology.
Technological Characteristics
Electrochemical biosensor (glucose oxidase) for glucose; oscillometric technology for BP. Wrist-worn device with inflatable cuff (5.31"-8.66").
Indications for Use
Indicated for quantitative blood glucose measurement (finger/alternative sites) and non-invasive blood pressure/pulse rate monitoring in patients with diabetes mellitus. Intended for home and professional use. Contraindicated for neonates and for diagnosis/screening of diabetes or hypertension.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Clever Chek TD-3217 Blood Glucose plus Blood Pressure Monitoring System (k062800)
Related Devices
K080014 — CLEVER CHEK TD-3250C BLOOD GLUCOSE PLUS BLOOD PRESSURE MONITORING SYSTEM · Taidoc Technology Corporation · Jun 9, 2008
K042795 — CLEVER CHEK TD-3213 · Taidoc Technology Corporation · Mar 7, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
### . Section 10. 510(k) Summary
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. The Assigned 510(k) number is: K092875
1. Submitter's Identification:
TaiDoc Technology Corporation
3F, 5F, No.127, Wugong 2nd Rd., Wugu Township, Taipei County, 248, Taiwan Correspondence:
Debra Liang
Regulatory Affairs Specialist Tel: +886-2-6625-8188 #1198 Fax: +886-2-6625-0288 Email: debra.liang@taidoc.com.tw
Prepared Date: September 14, 2009 .
2. Device name:
Proprietary name: TD-3227 Dual Blood Pressure/Blood Glucose Meter. Regulatory information:
Regulation section: 21 CFR § 862.1345, Glucose Test System 21 CFR §870.1130, Noninvasive blood pressure measurement system Classification: Class II (Glucose Test System)
Class II (Blood Pressure Measurement System)
Product Code: NBW, System, Test, Blood Glucose, Over The Counter
CGA, Glucose Oxidase, Glucose
DXN, System, Measurement, Blood-Pressure, Non-Invasive Panel: 75, Clinical Chemistry - Glucose Test System
74. Cardiovascular - Blood Pressure Measurement System
Page 1 10-1 of 3
{1}------------------------------------------------
### 3. Intended Use:
ECHO TD-3227 Dual Blood Pressure/Blood Glucose Meter is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the following alternative sites: the palm, forearm, upper-arm, calf and thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. The alternative site testing in the systems can be used only during steady-state blood glucose conditions. The system is also intended to measure non-invasively the systolic and diastolic blood pressure and pulse rate. For ECHO TD-3227 Dual Blood Pressure/Blood Glucose Meter. the blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 5.31"- 8.66". The device is not to be used for the diagnosis or screening of diabetes, hypertension or for testing on neonates.
- 4. Device Description:
The kit of ECHO TD-3227 Dual Blood Pressure/Blood Glucose Meter consist of four main products: the meter features both the blood glucose and blood pressure measurement functions, test strips, control solutions (cleared under k012430), and the lancet device (cleared under k833344). These products have been designed and tested to work together as a system to produce accurate blood glucose test results.
5. Substantial Equivalence Information: Predicate device name:
Clever Chek TD-3217 Blood Glucose plus Blood Pressure Monitoring System. Predicate K number: K062800 Comparison with predicate:
The modified ECHO TD-3227 Dual Blood Pressure/Blood Glucose Meter has the following similarities to the predicate device:
□ same operating principle,
[ same fundamental scientific technology,
[ incorporate the same basic circuit design,
Page 2 10-2 of 3 Page 3 10-3 of 3
{2}------------------------------------------------
O incorporate the same materials.
□ same shelf life
O packaged using the same materials, and
O Manufactured by the same process.
The modifications encompass:
[] A modification in the software of the meter
□ Modification in the physical appearance
O Expansion of cuff size
[] Labeling change due to the software modification
6. Test Principle:
For blood glucose, the detection and measurement is by an electrochemical biosensor technology using glucose oxidase.
For blood pressure, the measurement is by using oscillometric, non-invasive blood pressure (systolic, diastolic and mean arterial pressure) measuring technology.
### 7. Performance Characteristics:
Software verification and validation testing confirmed that the performance, safety and effectiveness of ECHO TD-3227 Dual Blood Pressure/Blood Glucose Meter met acceptance criteria.
### 8. Conclusion:
Based on the information provided in this submission. ECHO TD-3227 Dual Blood Pressure/Blood Glucose Meter is substantially equivalent to the predicate Clever Clever Chek TD-3217 Blood Glucose plus Blood Pressure Monitoring Systems.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle image.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver-Spring - MD-20993-0002
TaiDoc Technology Coporation C/O Debra Liang 6F, No. 127, Wugong 2nd Rd. Wugu Township, Taipei, Taipei County, Taiwan 248
AUG 1 3 2010
Re: k092875
Trade/Device Name: Echo TD-3227 Dual Blood Pressure/Blood Glucose Meter Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: CGA, NBW, DXN Dated: July 23, 2010 Received: July 26, 2010
Dear Ms. Liang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
# Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
CFC
Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K092875
AUG 1 3 2010
## Attachment 6. Indications for Use
# Indications for Use
510(k) Number: K092875
Device Name: ECHO TD-3227 Dual Blood Pressure/Blood Glucose Meter
Indications for Use:
ECHO TD-3227 Dual Blood Pressure/Blood Glucose Meter is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the following alternative sites: the palm, forearm, upper-arm, calf and thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. The alternative site testing in the systems can be used only during steady-state blood glucose conditions.
The system is also intended to measure non-invasively the systolic and diastolic blood pressure and pulse rate. For ECHO TD-3227 Dual Blood Pressure/Blood Glucose Meter, the blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 5.31"- 8.66".
The device is not to be used for the diagnosis or screening of diabetes, hypertension or for testing on neonates.
| Prescription Use | |
|-----------------------------|--|
| (21 CFR Part 801 Subpart D) | |
Over the Counter Use And/Or X (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Signature
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)
Page 1 of 1
l of l
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.