DDS STABILIZATION SYSTEM

K080241 · Paradigm Spine, LLC · NQP · Nov 28, 2008 · Orthopedic

Device Facts

Record IDK080241
Device NameDDS STABILIZATION SYSTEM
ApplicantParadigm Spine, LLC
Product CodeNQP · Orthopedic
Decision DateNov 28, 2008
DecisionSESU
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DSS™ Stabilization System - Rigid is intended as a single-level system for noncervical pedicle fixation from the T4 to S1 vertebrae in skeletally mature patients to help provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion. The DSS™ Stabilization System - Rigid is intended to be used with autograft and/or allograft.

Device Story

DSS™ Stabilization System - Rigid is a spinal fixation device consisting of pedicle screws and rigid couplers acting as longitudinal spacers. Used in orthopedic surgery to provide immobilization and stabilization of spinal segments as an adjunct to fusion. Implanted by surgeons in clinical settings. Device provides mechanical support to the spine to facilitate bone fusion; benefits include stabilization of spinal segments in patients with degenerative or traumatic conditions. Safety and effectiveness for stabilization without fusion have not been established.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Fabricated from wrought Ti-6Al-4V (ISO 5832-3, ASTM F136). System comprises pedicle screws and rigid couplers. Mechanical fixation device; no energy source or software components.

Indications for Use

Indicated for skeletally mature patients requiring noncervical pedicle fixation (T4-S1) for degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, spinal curvatures, tumor, pseudarthrosis, or failed previous fusion. Used as an adjunct to fusion with autograft/allograft. Single-level system.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K0802411 NOV ≥ 8 2008 # 510(k) Summary | Contact: | Mr. Justin Eggleton<br>Musculoskeletal Clinical & Regulatory Advisers, LLC<br>1331 H Street NW, 12th Floor<br>Washington, DC 20005<br>202.552.5800 | |--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name: | DSS™ Stabilization System - Rigid Coupler | | Manufacturer: | Paradigm Spine, LLC<br>505 Park Ave. 14th Floor<br>New York, NY 10022<br>212.583.9700 | | Common Name: | Pedicle screw spinal system | | Classification: | 21 CFR §888.3070 | | Class: | III | | Product Code: | NKB, MNH, MNI | ### Indications For Use: The DSSTM Stabilization System - Rigid Coupler is intended for noncervical pedicle fixation from the T4 to S1 vertebrae in skeletally mature patients to help provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion. The DSS™ Stabilization System - Rigid Coupler is intended to be used with autografl and/or allograft. ### Device Description: The DSS™ Stabilization System is comprised of a variety of pedicle screws and Rigid Couplers that act as longitudinal spacers. This device is intended to be used with bone graft to provide immobilization and stabilization of a spinal segment as an adjunct to fusion. The DSS™ Stabilization System consists of: - DSS™ pedicle screw sets . - DSS™ Rigid Couplers . {2}------------------------------------------------ The DSS™ Stabilization System is fabricated from wrought Ti-6AI-4V (ISO 5832-3 and ASTM F136). #### Predicate Device(s): The DSS™ Stabilization System was shown to be substantially equivalent to previously cleared devices and has the same indications for usc, design, function, and materials used, including, but not limited to: - 1. DePuy MOSS Miami (K962628) - 2. Biomet Synergy VLS Open (K973836) - 3. Synthes Pangea (K052123) - 4. Zimmer Silhouctte (K993067) ### Performance Standards: Testing performed indicates the DSS™ Stabilization System is substantially cquivalent to predicatc devices. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The eagle is depicted in a simple, black-and-white design. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Paradigm Spine, LLC c/o Musculoskeletal Clinical Regulatory Advisors, LLC Mr. Justin Eggleton 1331 H Street Northwest, 12th Floor Washington, DC 20005 NOV 2 8 2008 Re: K080241 Trade/Device Name: DSS™ Stabilization System - Rigid Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: NOP, Dated: October 27, 2008 Received: October 29, 2008 Dear Mr. Eggleton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act. the following limitation must appear in the Warnings section of the device's labeling: The safety and effectiveness of this device for the indication of spinal stabilization without fusion have not been established, {4}------------------------------------------------ Page 2 - Mr. Eggleton Furthermore, the indication for use: The DSS™ Stabilization System – Rigid is intended as a single-level system for noncervical pedicle fixation from the T4 to S1 vertebrae in skeletally mature patients to help provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion. The DSS™ Stabilization System - Rigid is intended to be used with autograft and/or allograft. must be prominently displayed in all labeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print. Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to procced to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling. If your device is classified (see above) into either class II (Special Controls) or class III. (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mcan that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. {5}------------------------------------------------ Page 3 - Mr. Eggleton Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Suttle G. Machino m.D. Donna-Bea Tillman, Ph.D., M.P.A. JEOR SR. TRU MAN Director Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: DSS™ Stabilization System - Rigid The DSS™ Stabilization System - Rigid is intended as a single-level system for noncervical pedicle fixation from the T4 to S1 vertebrae in skeletally mature patients to help provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following indications: degenerative disc disease (defined as back opin of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal sterensis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); turnor; pseudarthrosis; and failed previous fusion. The DSS™ Stabilization System - Rigid is intended to be used with autograft and/or allograft. 1/08024 Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODE) Mark A. Milken (Division (Division of General, Restorative, and Neurological Devices 510(k) Number_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...