K073432 · Trauson (Jiangsu) Medical Instrument Co., Ltd. · HRS · Aug 13, 2008 · Orthopedic
Device Facts
Record ID
K073432
Device Name
TRAUSON BONE PLATE
Applicant
Trauson (Jiangsu) Medical Instrument Co., Ltd.
Product Code
HRS · Orthopedic
Decision Date
Aug 13, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Trauson Bone Plate is intended for the fixation of fractures of humerus.
Device Story
Trauson Bone Plate is a metallic bone fixation appliance used for stabilizing humerus fractures. The device consists of plates of varying lengths and hole counts, manufactured from 316L stainless steel. It is intended for use by surgeons in a clinical or operating room setting to provide mechanical fixation of bone segments. The device is provided non-sterile and requires sterilization prior to use. It functions as a traditional orthopedic implant to support bone healing through rigid fixation. No software or electronic components are involved.
Clinical Evidence
Bench testing only. Mechanical performance evaluated per ASTM F382 and ASTM F897 to determine mean bending proof load and run-out load.
Technological Characteristics
Material: 316L stainless steel (ASTM F138). Form factor: Metallic bone plates with varying lengths and hole counts. Non-sterile. Mechanical fixation principle. No energy source. No software.
Indications for Use
Indicated for the fixation of humerus fractures in patients requiring surgical stabilization.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
ARTHREX HUMERAL FRACTURE PLATES AND SCREWS (K041965)
Related Devices
K060041 — TRIMED BONE PLATES · TriMed, Inc. · Feb 7, 2006
K972219 — ORTHOPEDIC PLATE SYSTEM · Ferguson Medical · Nov 20, 1997
K063095 — HUMERUS LOCKING PLATING SYSTEM · Newclip Technics · Dec 22, 2006
K142836 — Zimmer Plates and Screws (ZPS), Forte Screws · Zimmer, Inc. · Nov 12, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
## pg | 2)
### Attachment II 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The Assigned 510(k) Number is: K073432
1. Applicant Device Information Trade/Proprietary Name: Trauson Bone Plate Common Name: Single/multiple component metallic bone fixation appliances and accessories Classification Name: Plate, fixation, bone Device Class: II Product Code: HRS Regulation Number: 888.3030 Review Panel: Orthopedic Intended Use: Trauson Bone Plate is intended for the fixation of fractures of humerus.
#### 2. Submitter Information
Establishment Registration Name: TRAUSON (JIANGSU) MEDICAL INSTRUMENT CO., LTD. 31 Houcun Road, Niutang Town Changzhou, Jiangsu CHINA 213163 Phone: +86-757-86280075 Fax: +86-757-86397179
#### Contact Person of the Submission:
Ms. Diana. Hong, Mr. Eric. Chen Shanghai Mid-link Consulting Co., Ltd. Suite 8D, No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, 200030, CHINA Phone: +86-21-64264467 x 152 Fax: +86-21-64264468 x 809 Email: Diana.hong@mid-link.net Please CC: Eric.chen@mid-link.net
{1}------------------------------------------------
#### 3. Predicate Device
K Number: K041965
Trade Name: ARTHREX HUMERAL FRACTURE PLATES AND SCREWS
### 4. Device Description
The applicant device of Trauson Bone Plate made of medic 11 grade 3 16L stainless steel that meet ASTM F138 is intended for the fixation of fractures of humerus. The plates vary essentially through different lengths, number of plate holes.
The applicant devices are not sterile. The materials are wildly used in the industry with well know biocompatibility. No new materials are used in the development of this implant. No surface modified or coated.
### 5. Test Data
Bench tests of the applicant device following ASTM F382 and ASTM F897 are conducted to determine the mean bending proof load and run-out load, please see the Appendix 2, ASTM Test Report
#### 6. Substantially Equivalence
The applicant device is Substantially Equivalent (SE) to the predicate device in terms of Effectiveness and Safety.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular pattern around the symbol. The caduceus is depicted in black, and the text is also in black. The overall design is simple and recognizable.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Trauson (Jiangsu) Medical Instrument Co., Ltd. % Shanghai Midlink Business Consulting Co., Ltd. Ms. Diana Hong Suite 8D, Zhongxin Zhongshan Mansion No. 19. Lane 999. Zhong Shan No.2 Road Shanghai, 200030, China
AUG 1 3 2008
Re: K073432
Trade/Device Name: Trauson Bone Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: July 29, 2008 Received: July 29, 2008
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Diana Hong
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-100.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M Mullerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Attachment III Indications for Use
510(k) Number:___K073432___ Device Name: __ Trauson Bone Plate
Indications for Use:
Trauson Bone Plate is intended for the fixation of fractures of humerus.
Prescription Use __ V_ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Saubare Buchnd former
1
Div vision of General, Restorative, and Neurological Devices
**510(k) Number** K073432
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.