POWER PORT-A-CATH, CATH II, AND P.A.S. PORT T2 POWER P.A.C. IMPLANTABLE VENOUS ACCESS SYSTEMS AND POWER GRIPPER PLUS

K072657 · Smiths Medical MD, Inc. · LJT · Dec 14, 2007 · General Hospital

Device Facts

Record IDK072657
Device NamePOWER PORT-A-CATH, CATH II, AND P.A.S. PORT T2 POWER P.A.C. IMPLANTABLE VENOUS ACCESS SYSTEMS AND POWER GRIPPER PLUS
ApplicantSmiths Medical MD, Inc.
Product CodeLJT · General Hospital
Decision DateDec 14, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5965
Device ClassClass 2
AttributesTherapeutic

Intended Use

Smiths Medical MD, Inc. POWER P.A.C. Implantable Venous Access Systems are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling. When used with a GRIPPER PLUS® POWER P.A.C. Safety Huber Needle or other power injectable huber needle, the Smiths Medical MD, Inc. POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec. The GRIPPER PLUS® POWER P.A.C. Safety Huber Needle is indicated for the administration into or withdrawal of fluids from implanted ports. It is designed to help protect against exposure to bloodborne pathogens caused by accidental needlestick injuries. When used with power injectable implantable venous access ports, the GRIPPER PLUS® POWER P.A.C. Safety Huber Needle is indicated for power injection of contrast media.

Device Story

System comprises implantable portal with self-sealing septa and single/dual lumen catheter; used for long-term venous access. Portal placed in chest or arm; accessed percutaneously via Huber needle. When paired with power-injectable needle, system supports high-flow contrast media injection (up to 5 ml/sec) for diagnostic imaging. Needle includes passive safety mechanism to reduce needlestick injury risk. Used by clinicians in clinical settings for parenteral medication delivery, nutritional solutions, and blood sampling. Benefits include reliable long-term vascular access and safe, high-pressure contrast delivery for imaging procedures.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing and design verification. Functional testing confirmed performance against established acceptance criteria for power injection and safety features.

Technological Characteristics

System consists of portal with self-sealing septa and catheter. Needle features passive safety mechanism for needlestick protection. Sterile, non-pyrogenic, single-patient use. Designed for power injection up to 5 ml/sec. No software or electronic components.

Indications for Use

Indicated for patients requiring repeated vascular access for injection/infusion therapy, venous blood sampling, or power injection of contrast media (max 5 ml/sec) when used with compatible power-injectable needles. Needle indicated for fluid administration/withdrawal from implanted ports and protection against accidental needlestick injuries.

Regulatory Classification

Identification

A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.

Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K072657 1 of 3 ## SMITHS MEDICAL MD, INC. 510(k) Premarket Notification ### 510(k) Summary #### I. Applicant (Sponsor) Name and Address Smiths Medical MD, Inc. 1265 Grey Fox Road St. Paul, MN 55112 Establishment Reg. No.: 2183502 #### II. Contact Name and Phone Brian L. Haugstad DEC 1 4 2007 Senior Regulatory Affairs Associate Company Phone: (651) 628-7513 Company Fax: (651) 628-7457 ### III. Device Trade/Proprietary Name #### Catheter/Port POWER P.A.C. Implantable Venous Access Systems ### Needle GRIPPER PLUS® POWER P.A.C. Safety Huber Needle # IV. Device Classification/Common Name/Panel ### Catheter/Port 21 CFR Reference: §880.5965 21 CFR Common Name: Implanted Infusion Port Classification: Class II Product Code: LJT Panel: General Hospital ### Needle 21 CFR Reference: \$880.5440 21 CFR Common Name: Huber Needle Intravascular Administration Set Classification: Class II Product Code: FPA Review Panel: General Hospital and Personal Use {1}------------------------------------------------ # SMITHS MEDICAL MD. INC. 510(k) Premarket Notification #### 510(k) Summary ### V. Identification of Predicate Device Smiths Medical MD, Inc. believes the POWER P.A.C. Implantable Venous Access Systems and GRIPPER PLUS® POWER P.A.C. Safety Huber Needles are substantially equivalent to the following devices and thus may market the devices under these equivalencies. #### Catheter/Port #### Device Name: - 1. Smiths Medical MD, Inc.: PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems and Gripper Plus® POWER P.A.C. Needle - 2. Smiths Medical MD, Inc.: P.A.S. Port® T2 Implantable Venous Access System - Smiths Medical MD, Inc.: PORT-A-CATH® II 7 French, Dual Lumen Low ProfileTM 3. Implantable Venous Access System - 4. Medcomp : Power Injectable Implantable Infusion Port #### Needle Device Name: - 1. Smiths Medical MD, Inc. PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems and Gripper Plus® POWER P.A.C. Needle ### VI. Device Description ------------------------------------------------------------------------------------------------------------------------------------------------------- # POWER P.A.C. Implantable Venous Access Systems The POWER P.A.C. implantable venous access systems are similar to the current PORT-A-CATH® and PORT-A-CATH® II implantable venous access systems and Power Injectable Implantable Infusion Port. They both are totally implantable venous access systems designed to permit repeated access to the venous system for the parenteral delivery of medications, fluids, nutritional solutions, and for the sampling of venous blood. However, when used with a power indicated huber needle, the POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media. The POWER P.A.C. implantable.vanous access systems can be placed in the chest or arm. The POWER P.A.C. implantable venous access systems are supplied sterile and non-pyrogenic. All POWER P.A.C. implantable venous access systems are designed and intended for single patient use only. A system consists of a portal with one or two self-sealing septa and a single or dual lumen catheter and is accessible by percutaneous puncture with a non-sopina neede. # GRIPPER PLUS® POWER P.A.C. Needle The GRIPPER PLUS® POWER P.A.C. Needle is similar to the current GRIPPER PLUS® needle, with the addition of an added indication. Both are supplied sterile and non-pyrogenic and are intended for the administration into or withdrawal of fluids from implanted ports. In addition, they both have a passive needle stick protection feature that is designed to help protect against exposure to bloodborne pathogens caused by accidental needlestick in its protection in help not protect against other routes of bloodborne pathogen transmission. The GRIPPER PLUS® POWER P.A.C. Needle is indicated for power injection of contrast media. {2}------------------------------------------------ 16072657 3 af 3 # SMITHS MEDICAL MD, INC. 510(k) Premarket Notification ### 510(k) Summary #### VII. Intended Use of the Device The intended use of each device has not changed from that of the predicate. #### POWER P.A.C. Implantable Venous Access Systems The POWER P.A.C. implantable venous access systems are implantable vascular access devices designed to provide long term repeated access to the vascular system. ## GRIPPER PLUS® POWER P.A.C. Needle The GRIPPER PLUS® POWER P.A.C. Safety Huber Needle is designed for the administration into or withdrawal of fluids from implanted ports. #### VIII. Indications for Use #### POWER P.A.C. Implantable Venous Access Systems Smiths Medical MD, Inc. POWER P.A.C. Implantable Venous Access Systems are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling. When used with a GRIPPER PLUS® POWER P.A.C. Safety Huber Needle or other power injectable huber needle, the Smiths Medical MD, Inc. POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media. For power injection of contrast media. the maximum recommended infusion rate is 5 ml/sec. ## GRIPPER PLUS® POWER P.A.C. Needle The GRIPPER PLUS® POWER P.A.C. Safety Huber Needle is indicated for the administration into or withdrawal of fluids from implanted ports. It is designed to help protect against exposure to bloodborne pathogens caused by accidental needlestick injuries. When used with power injectable implantable venous access ports, the GRIPPER PLUS® POWER P.A.C. Safety Huber Needle is indicated for power injection of contrast media. #### IX. Summary of Studies #### A. Functional Testing The POWER P.A.C. Implantable Venous Access Systems and GRIPPER PLUS® POWER P.A.C. Needle met all established acceptance criteria for performance testing and design verification testing. Packaging and Sterilization systems are unchanged from predicate devices. #### B. Clinical Studies Clinical studies for the POWER P.A.C. Implantable Venous Access Systems and GRIPPER PLUS® POWER P.A.C. Safety Huber Needle were deemed not necessary due to their similarity in materials, design and function to current Smiths Medical MD, INC. devices and other commercially available systems. #### C. Conclusions Drawn from the Studies Based upon the information provided; the POWER P.A.C. Implantable Venous Access Systems and GRIPPER PLUS® POWER P.A.C. Safety Huber Needle met all acceptance criteria for performance testing and design verification testing. Therefore, these products are considered acceptable for human use. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle. DEC 1. 4 2007 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Brian L. Haugstad Senior Regulatory Affairs Associate Smiths Medical MD, Incorporated 1265 Grey Fox Road St. Paul, Minnesota 55112 Re: K072657 Trade/Device Name: GRIPPER PLUS™ POWER P.A.C. Safety Huber Needle and POWER P.A.C. Implantable Venous Access Systems Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT, FPA Dated: September 18, 2007 Received: September 20, 2007 Dear Mr. Haugstad: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 -- Mr. Haugstad Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Century D. Nester her Christian, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # SMITHS MEDICAL MD, INC. 510(k) Premarket Notification # Indications for Use # 510(k) Number: (OOQLS7 # Device Name: Smiths Medical MD, Inc. POWER P.A.C. Implantable Venous Access Systems are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling. When used with a GRIPPER PLUS® POWER P.A.C. Safety Huber Needle or other power injectable huber needle, the Smiths Medical MD, Inc. POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec. Prescription Use _____________________________________________________________________________________________________________________________________________________________ OR Over-The Counter Use ___________Per 21 CFR 801.109) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Anten Omz (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 大472657 FORH, Office of Device Evaluation (ODE) {6}------------------------------------------------ # SMITHS MEDICAL MD, INC. 510(k) Premarket Notification # Indications for Use 2.67 1078657 510(k) Number: #### GRIPPER PLUSTM POWER P.A.C. Safety Huber Needle Device Name: The GRIPPER PLUS® POWER P.A.C. Safety Huber Needle is indicated for the administration into or withdrawal of fluids from implanted ports. It is designed to help protect against exposure to bloodborne pathogens caused by accidental needlestick injuries. When used with power injectable implantable venous access ports, the GRIPPER PLUS® POWER P.A.C. Safety Huber Needle is indicated for power injection of contrast media. Prescription Use OR Over-The Counter Use _ _ _ Per 21 CFR 801.109) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices ______________________________________________________________________________________________________________________________________________________________________________ 510(K) Numberoncurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...