CONFORMIS METAL BACKED TIBIAL COMPONENT

K072368 · Conformis, Inc. · HSX · Sep 20, 2007 · Orthopedic

Device Facts

Record IDK072368
Device NameCONFORMIS METAL BACKED TIBIAL COMPONENT
ApplicantConformis, Inc.
Product CodeHSX · Orthopedic
Decision DateSep 20, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3520
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ConforMIS metal backed tibial component is intended for use with the ConforMIS Unicondylar Knee Repair System and the ConforMIS BiCompartmental Knee Repair Systems, in patients with: - joint impairment due to osteoarthritis or traumatic arthritis of the knee . - previous tibial condyle or plateau fracture, creating loss of function . - valgus or varus deformity of the knee . The ConforMIS metal backed tibial component is intended only for use with bone cement

Device Story

Patient-matched tibial component for knee arthroplasty; used with ConforMIS Unicondylar or BiCompartmental Knee Repair Systems. Input: patient-specific anatomical data from MRI or CT scans. Transformation: device geometry designed to conform to patient anatomy based on imaging. Output: cobalt-chrome alloy (CoCrMo) tray with UHMWPE insert (6-14 mm thickness). Context: surgical implant; used by orthopedic surgeons in clinical settings. Benefit: provides alternative to all-polyethylene components with patient-specific fit.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via design verification and non-clinical performance testing.

Technological Characteristics

Materials: Cobalt-chrome alloy (CoCrMo) tray, UHMWPE insert (6-14 mm). Design: Patient-matched geometry based on MRI/CT imaging. Fixation: Cemented only. Form factor: Tibial component tray and insert.

Indications for Use

Indicated for patients with knee joint impairment due to osteoarthritis or traumatic arthritis, previous tibial condyle or plateau fracture with loss of function, or valgus/varus deformity. Intended for use with ConforMIS Unicondylar or BiCompartmental Knee Repair Systems and bone cement.

Regulatory Classification

Identification

A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K072368 # SECTION 2 SEP 2 0 2007 1/3 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of the 510(k) premarket notification for the ConforMIS, Inc. Unicondylar Knee Repair System is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. # Summary of Safety and Effectiveness | Submitted By: | ConforMIS, Inc.<br>2 Fourth Avenue<br>Burlington, MA 01803 | |-------------------------------|-------------------------------------------------------------------------------------| | Contact Person: | S. Michael Sharp, PhD<br>Sr. Vice President, Regulatory/Clinical & Quality | | Date: | August 22, 2007 | | Trade/Proprietary Name | Metal Backed Tibial Component ("mBT") | | Common Name | Metal-backed tibial component | | Reference/Classification Name | 21 CFR 888.3520 - Knee joint femorotibial<br>metal/polymer non-constrained cemented | prosthesis ## Predicate Devices | Technological<br>Characteristics | Design & Insert<br>Thickness | Indications for Use | |-----------------------------------------------------------------|--------------------------------------------------------------------|-----------------------------------------------------------------| | ConforMIS UniCondylar<br>Knee Repair System<br>(K043570) | DePuy Preservation<br>Unicondylar Knee<br>(K040268) | ConforMIS UniCondylar<br>Knee Repair System<br>(K043570) | | ConforMIS<br>BiCompartmental<br>Knee Repair System<br>(K053488) | ConforMIS Metal-backed<br>Tibial Component<br>("mBT")<br>(K063432) | ConforMIS<br>BiCompartmental<br>Knee Repair System<br>(K053488) | | | DePuy Preservation<br>Unicondylar Knee<br>(K040268) | DePuy Preservation<br>Unicondylar Knee<br>(K040268) | | | Encore Uni Knee<br>(K022437) | | | | Whiteside Biomechanics<br>Unicompartmental Knee<br>System (K01280) | | {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows a white square with a black border. Inside the square, the fraction "2/3" is written in black ink. The number 2 is above the fraction line, and the number 3 is below the fraction line. ## Intended Use: The ConforMIS metal backed tibial component is intended for use with the ConforMIS Unicondylar Knee Repair System and the ConforMIS BiCompartmental Knee Repair Systems, in patients with: - joint impairment due to osteoarthritis or traumatic arthritis of the knee . - previous tibial condyle or plateau fracture, creating loss of function . - valgus or varus deformity of the knee . The ConforMIS metal backed tibial component is intended only for use with bone cement ### Device Description The ConforMIS Metal Backed Tibial Component is to be used with either the ConforMIS Unicondylar Knee Repair System or the BiCompartmental Knee Repair ConforMIS System to provide the surgeon with an alternative tibial component in the event that prefers to use a metal backed he/she component rather than an all polyethylene component. It consists of a cobalt-chrome alloy (CoCrMo) tray and a polyethylene insert. The X/Y dimensions of the device (i.e. the twodimensional shape or "footprint"). are designed to conform to the patient's anatomy as closely as possible based on images (MRI or CT scan) of the patient's knee. It is available with UHMEPE inserts with minimal thickness from 6 to 14 mm. #### Comparison to Predicates The ConforMIS Metal Backed Tibial Component the tibial substantially equivalent to is components cleared for the ConforMIS Unicondylar and BiCompartmental Knee Repair Systems in is use of imaging data to design a patient-matched implant geometry, as well in terms of design and production process, as well as materials and indications. It is substantially equivalent to the cited predicate devices in terms of design, materials, mechanical safety All are intended for and intended use. cemented use only. Non-clinical Performance and Conclusions: Testing completed as part of the design verification procedure for the ConforMIS Metal 510(K) Submission: ConforMIS mBT Extended Thickness Range August 23, 2007: CONFIDENTIAL 8 ## Performance Data {2}------------------------------------------------ Backed Tibial Component found this device to be as safe and effective as the predicate devices. further confirming substantial equivalence. Clinical Performance: Clinical data and conclusions are not necessary to demonstrate substantial equivalence. ### SUMMARY Based on the similarities in design, materials, function, and intended use the ConforMIS - Metal Backed Tibial Component is substantially equivalent to the devices currently marketed under the Federal Food, Drug and Cosmetic Act. In addition, the ConforMIS Metal Backed Tibial Component raises no new safety or effectiveness issues. ## USE OF THE TERM "SUBSTANTIAL EQUIVALENCE" The term "Substantial Equivalence" is used in this submission within the confines of the statutory use in the FDA's evaluation of a Pre-Market Notification Submission. Any statement regarding Substantial Equivalence used in this submission relates only to whether the device that is the subject of this submission may be lawfully marketed in the United States without pre-market approval or reclassification, and should not be interpreted as an admission, or any kind or type of evidence, in any patent proceeding, including patent infringement litigation or proceeding before any Patent Office. The present submission and statements therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the product addressed in the submission, or its use, may be considered indistinct, from a patentability perspective, from any other device referred to in this sub mission. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird figure. The text is in all capital letters and is evenly spaced around the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ConforMIS, Inc. % S. Michael Sharp, Ph.D. Sr. VP, Regulatory and Clinical 2 Fourth Avenue Burlington, Massachusetts 01803 SEP 2 0 2007 K072368 Re: Trade/Device Name: Metal Backed Tibial Component Regulation Number: 21 CFR 888.3520 Regulation Name: Knee joint femorotibial metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: HSX Dated: August 22, 2007 Received: August 23, 2007 Dear Dr. Sharp: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - S. Michael Sharp, Ph.D. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Barbara Buehrig Mark N. Me Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K K072368 ConforMIS Metal Backed Tibial Component (mBT) Device Name: Indications For Use: The ConforMIS metal backed tibial component is intended for use with the ConforMIS Unicondylar Knee Repair System and the ConforMIS BiCompartmental Knee repair Systems, in patients with: - joint impairment due to osteoarthritis or traumatic arthritis of the knee . - previous tibial condyle or plateau fracture, creating loss of function . - . valgus or varus deformity of the knee The ConforMIS metal backed tibial component is intended only for use with bone cement. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) harbare friedm (Division Sign of General, Restorative, Division --------------------------------------------------------------------------------------------------------------------------------------------------------------------- **510(k) Number** K072368 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...