K072230 · C.R. Bard, Inc. · LJS · Oct 5, 2007 · General Hospital
Device Facts
Record ID
K072230
Device Name
POWERPICC SOLO
Applicant
C.R. Bard, Inc.
Product Code
LJS · General Hospital
Decision Date
Oct 5, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerPICC SOLO™ catheters are intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.
Device Story
PowerPICC SOLO™ is a clampless, proximally valved, radiopaque polyurethane PICC; available in 4Fr SL, 5Fr SL/DL, 6Fr DL/TL configurations. Device provides peripheral access to central venous system for therapy, power injection of contrast media, and CVP monitoring. Features reverse taper design and depth indicators for insertion reference. Purple colorant and blue luer hub distinguish device as power-injectable. Used in clinical settings by healthcare professionals. Output is venous access; enables infusion, sampling, and pressure monitoring. Benefits include reduced need for multiple sticks and compatibility with power injection protocols.
Clinical Evidence
Bench testing only. Testing conducted per ISO 10555-1, ISO 10555-3, ISO 594-2, and IEC 60601-2-34. Biocompatibility evaluated per ISO 10993. Performance testing confirmed radiopacity (ASTM F640-79), sterilization validation (AAMI/ANSI/ISO 11135), and suitability for invasive blood pressure monitoring (ANSI/AAMI BP22).
Technological Characteristics
Radiopaque polyurethane catheter; clampless, proximally valved design. Reverse taper shaft with depth markings. 4Fr to 6Fr sizes. Luer lock fittings per ISO 594-2. Sterilized via ethylene oxide (AAMI/ANSI/ISO 11135). Compatible with invasive blood pressure monitoring equipment (IEC 60601-2-34).
Indications for Use
Indicated for short or long-term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and central venous pressure (CVP) monitoring. Requires 4 French or larger catheter for blood sampling, infusion, or therapy. Max infusion rate 5ml/sec. CVP monitoring requires 20 gauge or larger lumen.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
5 Fr Single Lumen (SL) PowerPICC® catheter (K033389)
Vaxcel® PICC with PASV® Valve Technology (K021704)
Submission Summary (Full Text)
{0}------------------------------------------------
Page 20 of 112
K072230
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
## 510(k) Summary 21 CFR 807.92(a)
| OCT 5 | 2007<br>PowerPICC SOLOTM Catheter Family<br>Prepared August 9, 2007 | | | |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--|
| General<br>Provisions | Submitter of 510(k)<br>Premarket Notification: | Bard Access Systems, Inc. (BAS)<br>[Subsidiary of C.R. Bard, Inc.]<br>Salt Lake City, Utah 84116<br>Phone: (801) 595-0700, Ext. 7136<br>(801) 595-5425<br>Fax: | | |
| | Contact Person: | Lynn M. Kirchoff<br>Regulatory Affairs Specialist | | |
| | Device Trade Name:<br>Device Generic Name: | PowerPICC SOLO™ Catheter<br>Peripherally Inserted Central Catheter (PICC) | | |
| | Trade Name:<br>Common/Usual Name: Peripherally Inserted Central Catheter (PICC)<br>Classification Name:<br>CFR Reference:<br>Classification Panel:<br>Premarket Notification: See below | PowerPICC® Catheter<br>21 CFR §880.5970, Class II<br>General Hospital | 80 LJS- Long Term Intravascular Catheter | |
| | Predicate Device Name | 510(k) | Concurrence Date | |
| | 5 Fr Single Lumen (SL)<br>PowerPICC® catheter | K033389 | March 14, 2004 | |
| Predicate<br>Devices | 6 Fr Dual Lumen (DL)<br>PowerPICC® catheter | K050931 | June 15, 2005 | |
| | 5 Fr Dual Lumen (DL)<br>PowerPICC® catheter | K051672 | November 23, 2005 | |
| | 6 Fr Triple Lumen (TL)<br>PowerPICC® catheter | K053501 | January 13, 2006 | |
| | 4 Fr Single Lumen (SL)<br>PowerPICC® catheter | K070996 | May 8, 2007 | |
| | PowerPICC®, Poly Per-Q-Cath,<br>6 Fr TL Poly Per-Q-Cath,<br>PowerHohn and PowerLines | K051991 | October 20, 2005 | |
| | Trade Name:<br>Common/Usual Name: Peripherally Inserted Central Catheter (PICC)<br>Classification Name:<br>CFR Reference:<br>Classification Panel:<br>Premarket Notification: See below | PowerGroshong™ PICC Catheter<br>21 CFR §880.5970, Class II<br>General Hospital | 80 LJS- Long Term Intravascular Catheter | |
| | Predicate Device Name | 510(k) | Concurrence Date | |
| | 5 Fr SL PowerGroshong™<br>PICC Catheter | K063848 | April 4, 2007 | |
{1}------------------------------------------------
Page 21 of 112
| Predicate<br>Devices<br>Continued | Trade Name: Vaxcel® PICC with PASV® Valve Technology<br>Common/Usual Name: Peripherally Inserted Central Catheter (PICC)<br>Classification Name: 80 LJS-Long Term Intravascular Catheter<br>CFR Reference: 21 CFR §880.5970, Class II<br>Classification Panel: General Hospital<br>Premarket Notification: See below | K072230<br>Pg2 | | | | | | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|--|--|--|--|--|--|
| | Predicate Device Name 510(k) Concurrence Date Vaxcel® PICC with PASV®<br>Valve Technology K021704 June 6, 2002 | | | | | | | |
| Classification | 21 CFR 880.5970, Class II, 80LJS - Long Term Intravascular Catheter | | | | | | | |
| Performance<br>Standards | Performance standards have not been established by FDA under section 514<br>of the Federal Food, Drug and Cosmetic Act. | | | | | | | |
| Intended Use | The PowerPICC SOLO™ catheters are intended for short or long term<br>peripheral access to the central venous system for intravenous therapy and<br>blood sampling.<br>The intended use has not changed from the predicate PowerPICC® catheters. | | | | | | | |
| Indications for<br>Use | The PowerPICC SOLO™ catheter is indicated for short or long term peripheral<br>access to the central venous system for intravenous therapy, power injection of<br>contrast media and allows for central venous pressure monitoring. For blood<br>sampling, infusion or therapy, use a 4 French or larger catheter. The maximum<br>recommended infusion rate is 5ml/sec. For central venous pressure<br>monitoring, it is recommended that catheter lumen of 20 gauge or larger be<br>used.<br>The indications for use have not changed from the predicate PowerPICC®<br>catheters. | | | | | | | |
| Device<br>Description | The PowerPICC SOLO™ catheter is a clampless proximally valved catheter. The PowerPICC SOLO™ catheters are open-ended radiopaque polyurethane. The PowerPICC SOLO™ catheters are offered in 4 Fr Single Lumen (SL), 5 Fr Single Lumen (SL), 5 Fr Dual Lumen (DL), 6 Fr Dual Lumen (DL), and 6 Fr Triple Lumen (TL) configurations. Catheter usable length is 55 cm. The catheter has a reverse taper design Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point. Catheters are provided sterile in basic and full PICC configurations with legally marketed kit components. Purple colorants were added to the catheter materials to provide the catheter with an appearance that allows the end user to differentiate | | | | | | | |
{2}------------------------------------------------
Bard Access Systems
PowerPICC SOLO™ Catheter
Special 510(k) Premarket Notification
Section 5 – 510(k) Summary
Page 22 of 112
K072230
Pg3
| Device<br>Description<br>Continued | Bard's power injectable catheters from other manufacturers' catheters.<br>Lower portion of luer hub is blue to identify the catheter as PowerPICC<br>SOLO™ catheter.<br>The catheter extension leg, luer hub and junction were printed with<br>markings to identify the catheter as PowerPICC SOLO™ and to<br>include information to facilitate proper use of the device. |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | Technological similarities between the subject PowerPICC SOLO™ catheters<br>and the predicate devices remain identical. There are no new questions raised<br>regarding safety or efficacy of the PowerPICC SOLO™ catheters. |
| Safety and<br>Performance<br>Tests | Biocompatibility requirements of ISO 10993 Biological Evaluation of Medical<br>Devices Part-1: Evaluation and Testing and the FDA Modified ISO 10993 Test<br>Profile for externally communicating, blood contacting, long-term devices have<br>been met. |
| | Performance testing of the PowerPICC SOLO™ catheters were conducted in<br>accordance with the following FDA guidance documents and international<br>standards: |
| | Guidance on Premarket Notification [510(k)] Submission for Short-<br>Term and Long-Term Intravascular Catheters, March 16, 1995 BS/EN/ISO 10555-1: 1997, Sterile, single-use intravascular<br>catheters, Part 1. General requirements ISO 10555-1:2004, Sterile, single-use intravascular catheters, Part<br>1. General requirements, Amendment 2 ASTM F640-79 (reapproved 2000), Standard Test Methods for<br>Radiopacity of Plastics for Medical Use BS/EN/ISO 10555-3:1997, Sterile, single-use intravascular<br>catheters, Part 3. Central venous catheters ISO 594-2: 1998, Conical fittings with a 6% (Luer) taper for<br>syringes, needles and certain other medical equipment - Part 2:<br>Lock Fittings AAMI/ANSI/ISO 11135:1994, Medical Devices - Validation and<br>Routine Control of Ethylene Oxide Sterilization AAMI/ANSI/ISO 10993-1:2003, Biological Evaluation of Medical<br>Devices Part 1: Evaluation and Testing, and the FDA Modified ISO<br>10993 Test Profile IEC 60601-2-34: 2000-10, Medical electrical equipment -<br>Particular requirements for the safety, including essential<br>performance, of invasive blood pressure monitoring equipment AAMI TIR9: 1992, Evaluation of Clinical Systems for Invasive<br>Blood Pressure Monitoring ANSI/AAMI BP22: 1994, Blood Pressure Transducers |
| | Subject product testing has yielded acceptable safety & performance<br>outcomes. |
| | The results of these tests, in conjunction with the substantial equivalence<br>claims as outlined in the premarket notification, effectively demonstrate the<br>PowerPICC SOLO™ catheters' substantial equivalence to the cited predicate<br>devices. |
{3}------------------------------------------------
Bard Access Systems PowerPICC SOLO™ Catheter
Special 510(k) Premarket Notification Section 5 - 510(k) Summary
Page 23 of 112
K072230
Pg4
Summary of Substantial Equivalence
Based on the indications for use, technological characteristics, and safety and performance testing, the subject PowerPICC SOLO™ catheters met the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available catheters/cited predicates.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the seal for the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
## NOV 18 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lynn M. Kirchoff Regulatory Affairs Specialist C.R. Bard, Incorporated Bard Access Systems 605 North 5600 West Salt Lake City, Utah 84116
Re: K072230 Trade/Device Name: PowerPICC SOLO™ Catheter Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: September 6, 2007 Received: September 7, 2007
Dear Ms. Kirchoff:
This letter corrects our substantially equivalent letter of October 5, 2007. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the [kit/tray] have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Kirchoff
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains chloraprep applicator, alcohol wipes, and lidocaine 1% which are subject to regulation as drugs.
Our substantially equivalent determination does not apply to the drug component[s] of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:
Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (240) 276-0141. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (240) 276-3150, or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph. D. Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Bard Access Systems PowerPICC SOLO™ Catheter Special 510(k) Premarket Notification Section 4 - Indications for Use Statement
Page 18 of 112
510(k) Number (if known):
Device Name:
长072230 PowerPICC SOLO™ Catheter Familv
Indications for Use:
The PowerPICC SOLO™ catheter is indicated for short or long term peripheral access . to the central venous system for intravenous therapy, power injection of contrast media and allows for central venous pressure monitoring. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.
Prescription Use V (Part 21 CFR §801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR §801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
(Division Sign-Off) Division of Anesthesiology, General Hospital nfection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.