CONFIRMA BREAST MRI INTERVENTIONAL COMPONENTS

K071534 · Confirma, Inc. · MOS · Jul 9, 2007 · Radiology

Device Facts

Record IDK071534
Device NameCONFIRMA BREAST MRI INTERVENTIONAL COMPONENTS
ApplicantConfirma, Inc.
Product CodeMOS · Radiology
Decision DateJul 9, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2

Intended Use

The Confirma Breast MRI Interventional Components are intended to be used in conjunction with a magnetic resonance scanner and the Confirma Access Breast Coil designs to permit MRI-guided interventional procedures, such as breast biopsy and localization of lesions that can be performed and interpreted by a trained physician.

Device Story

Confirma Breast MRI Interventional Components consist of grid and compression plate system designed for use with MRI scanners and Confirma Access Breast Coils. System immobilizes breast tissue to facilitate MRI-guided interventional procedures, specifically breast biopsy and lesion localization. Clinician positions patient and secures components using integrated locking mechanisms; grid plate allows needle block insertion for medial or lateral approach. Components are non-sterile, non-disposable, and require sterilization before each procedure. Device provides stable platform for physician to perform interventions; aids in accurate targeting of lesions identified via MRI. Benefits include improved patient positioning, ergonomic comfort, and support for both unilateral and bilateral procedures.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Components manufactured from Ultem 1000 polyetherimide; autoclavable or sterilizable with formaldehyde. System includes compression and grid plates with locking mechanisms. Supports 40 cm field of view. Compatible with Confirma Access breast coil designs. Non-sterile, non-disposable. Mechanical design for patient immobilization and needle guidance.

Indications for Use

Indicated for patients requiring MRI-guided breast interventional procedures, including biopsy and lesion localization. For use by trained physicians in conjunction with magnetic resonance scanners and Confirma Access Breast Coil designs.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K071534 JUL - 9 2007 # 510(k) Summary of Safety and Effectiveness This 510(k) summary of safety and effectiveness information is submitted as part of the PreMarket Notification in accordance with the requirements of 21 CFR Part 807, Subpart E and Section 807.92. ## 1. Identification of Submitter: | Submitter: | Confirma, Inc. | |----------------|------------------------------------------------------| | Address: | 11040 Main Street, Suite 100 | | | Bellevue, WA 98004 | | Phone: | 425-691-1400 | | Fax: | 425-691-1599 | | Contact: | Patricia A. Milbank, JD | | Title: | Vice President, Regulatory Affairs/Quality Assurance | | Phone: | 425-894-9733 | | Fax: | 425-822-3648 | | Date Prepared: | May 29, 2007 | ## 2. Identification of Product: | Trade Name: | Confirma Breast MRI Interventional Components | |--------------------|----------------------------------------------------------------------| | Regulatory Number: | 21 CFR 892.1000 | | Regulation Name: | Magnetic Resonance Diagnostic Device | | Common Name: | System, Nuclear Magnetic Resonance Imaging | | Regulatory Class: | Class II | | Product Code: | 90 MOS | | Manufacturer: | Confirma, Inc.<br>11040 Main Street, Suite 100<br>Bellevue, WA 98004 | #### 3. Indications for Use The Confirma Breast MRI Interventional Components are intended to be used in conjunction with a magnetic resonance scanner and the Confirma Access Breast Coil designs to permit MRI-guided interventional procedures, such as breast biopsy and localization of lesions that can be performed and interpreted by a trained physician. {1}------------------------------------------------ ### 4. Device Description: The Confirma Breast MRI Interventional Components provide a grid system to immobilize the breast in an Access Breast Coil device. The grid system allows the physician to perform a biopsy through a needle block inserted in the grid plate. The system supports a standard medial-lateral approach using the Access Breast Coil. The design of the interventional components focuses on ease of patient positioning and set-up by the clinician, and patient ergonomics for enhanced comfort during interventional procedures. The design allows for unilateral or bilateral access, and offers maximum access to aid the physician in MRguided interventional procedures such as breast biopsy and localization of lesions, providing lateral and medial access to the breast. The components are provided non-sterile and are non-disposable. The components must be sterilized at the beginning of each procedure. The components are manufactured from Ultem 1000 polyetherimide which can be autoclaved or sterilized with formaldehyde. The compression plate and grid plate system incorporate locking mechanisms that secure the components once positioned. These locks provide stability during the interventional procedure. The compression and grid plates are bidirectional, allowing them to be placed in either a medial-facing or lateral-facing position. This allows the clinician to perform the interventional procedure from the medial or lateral side. For unilateral breast intervention, the system uses one grid plate and one compression plate. For bilateral breast intervention, the system uses 2 grid plates and 2 compression plates. The components are made of medical grade plastic materials, which are fire rated and have high impact and tensile strength. ## Features of the Confirma Breast MRI Interventional Components: - . Optimized for parallel imaging applications - . Lateral and medial interventional access - . Supports both unilateral and bilateral imaging exams - . Compatible with Confirma Access breast coil designs - . 40 cm field of view - . Extensive coverage of chest wall and axilla - Ergonomically designed for patient comfort to minimize motion . artifact {2}------------------------------------------------ ## 5. Comparison with Legally Marketed Devices The Interventional Components are substantially equivalent to the following legally marketed devices: | 510(k)<br>No. | Trade Name | Manufacturer | Product<br>Code | Regulation<br>Number | |---------------|---------------------------------------------------------------------------------------------|---------------------------------------------|-----------------|----------------------| | K001303 | Breast Immobilization<br>Device MR-BY 160 | MRI Devices<br>Corporation | 90 MOS | 892.1000 | | K020289 | MICS Intervention Aid | Machnet BV | 90 LNH | 892.1000 | | K052704 | Model BBD Breast<br>Immobilization and Biopsy<br>Device | Invivo Corporation | 90 MOS | 892.1000 | | K052987 | Breast Immobilization and<br>Biopsy Device, Models BI<br>160-0, BI 160-PA, and BI<br>160-CC | Noras Röntgen-und<br>Medizintechnik<br>GmbH | 90 MOS | 892.1000 | The predicate devices are designed for use with MR scanners and standard breast coils to perform interventional breast procedures. The Confirma Breast MRI Interventional Components and the predicate devices allow bilateral and unilateral imaging and access of the breast as well as medial and lateral access to the breast for biopsy and interventional procedures. ### 6. Conclusions The Confirma Breast MRI Interventional Components are substantially equivalent to the identified legally marketed devices. The potential hazards have been studied and controlled as part of the product development process. including risk analysis, test and design considerations, and planned verification and validation testing processes. The Confirma Breast MRI Interventional Components allow access and support for breast interventional procedures comparable to the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized caduceus-like symbol with three parallel lines curving into a serpentine shape. The text "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged around the emblem in a circular fashion. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 ## - 9 2007 Ms. Patricia A. Milbank VP RA/OA Confirma, Inc. 11040 Main Street, Suite 100 BELLEVUE WA 98004 Re: K071534 Trade/Device Name: Confirma Breast MRI Interventional Components Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: June 1, 2007 Received: June 5, 2007 #### Dear Ms. Milbank: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for. use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations offecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/3/Picture/10 description: The image shows a logo with the text "FDA Centennial 1906-2006" in a circular design. Below the logo, the text "Protecting and Promoting Public Health" is written in a cursive font. The logo appears to be a commemorative emblem for the FDA's centennial anniversary. The text emphasizes the FDA's role in safeguarding and improving public health. {4}------------------------------------------------ Enclosure Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincerely yours, Nancy C. Hogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ ## Indication(s) for Use Statement 510(k) Number: To be assigned by FDA Confirma Breast MRI Interventional Components Device Name: Indications for Use: The Confirma Breast MRI Interventional Components are intended to be used in conjunction with a magnetic resonance scanner and the Confirma Access Breast Coil designs to permit MRI-guided interventional procedures, such as breast biopsy and localization of lesions that can be performed and interpreted by a trained physician. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Huld Remun (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number K071534
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...