K052045 · Confirma, Inc. · MOS · Aug 24, 2005 · Radiology
Device Facts
Record ID
K052045
Device Name
ACCESS BREAST COIL
Applicant
Confirma, Inc.
Product Code
MOS · Radiology
Decision Date
Aug 24, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Access Breast Coil is a 7-channel, phased array, receive-only RF coil, used in conjunction with a magnetic resonance scanner to produce diagnostic and interventional images of the breast, chest wall, and axillary tissues that can be interpreted by a trained physician. The device is designed to aid the physician in MR-guided breast biopsy, localization of lesions, and interventional procedures, providing lateral, medial and cranial-caudal access to the breast.
Device Story
Access Breast Coil is a 7-channel, phased array, receive-only RF coil for use with Siemens 1.5T MRI scanners. Device captures MR signals from breast, chest wall, and axillary tissues; transforms signals into diagnostic and interventional images. Features open design with integrated lighting to facilitate lateral, medial, and cranial-caudal interventional access. Includes patient positioning accessories (headrest, armrest, torso pad) to enhance comfort and minimize motion artifacts. Operated by clinicians in a clinical setting; output viewed by physicians to guide biopsies and localize lesions. Benefits include improved imaging access and patient ergonomics during interventional MRI procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance validated through risk analysis, design verification, and validation testing to ensure image quality comparable to predicate devices.
Technological Characteristics
7-channel phased array, receive-only RF coil. Materials: rigid plastic housing (fire-rated, high impact/tensile strength). Form factor: bilateral open design with large apertures. Connectivity: compatible with Siemens 1.5T MRI systems. Integrated lighting for interventional procedures. Positioning accessories: headrest, armrest, torso pad.
Indications for Use
Indicated for patients requiring diagnostic or interventional MRI of the breast, chest wall, and axillary tissues. Used by trained physicians for MR-guided breast biopsy, lesion localization, and interventional procedures.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Ko52045
AUG 2 4 2005
#### 7
# 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is submitted as part of the PreMarket Notification in accordance with the requirements of 21 CFR Part 807, Subpart E and Section 807.92.
### 1. Identification of Submitter:
| Submitter: | Confirma, Inc. |
|----------------|-------------------------------------------|
| Address: | 821 Kirkland Avenue<br>Kirkland, WA 98033 |
| Phone: | 425-576-1226 |
| Fax: | 425-576-9295 |
| Contact: | Patricia A. Milbank |
| Title: | Regulatory Consultant |
| Phone: | 425-894-9733 |
| Fax: | 425-865-9023 |
| Date Prepared: | July 25, 2005 |
#### 2. Identification of Product:
| Trade Name: | Access Breast Coil |
|--------------------|--------------------------------------|
| Regulatory Number: | 21 CFR 892.1000 |
| Regulation Name: | Magnetic Resonance Diagnostic Device |
| Common Name: | Coil, Magnetic Resonance Specialty |
| Regulatory Class: | Class II |
| Product Code: | 90 MOS |
| Manufacturer: | Confirma, Inc. |
|---------------|---------------------|
| | 821 Kirkland Avenue |
| | Kirkland, WA 98033 |
#### 4. Indications for Use
The Access Breast Coil is a 7-channel, phased array, receive-only RF coil, used in conjunction with a magnetic resonance scanner to produce diagnostic and interventional images of the breast, chest wall, and axillary tissues that can be interpreted by a trained physician. The device is designed to aid the physician in MR-guided breast biopsy, localization of lesions, and interventional procedures, providing lateral, medial and cranial-caudal access to the breast.
{1}------------------------------------------------
#### 5. Device Description:
Access is a 7-channel, phased array breast coil optimized for parallel imaging of the chest wall, breast and axillary tissue. Access is a dual purpose coil, offering both diagnostic imaging and flexibility in interventional access. The coil's open design allows for flexibility in performing interventional procedures with lateral, medial and cranial-caudal access.
The design of the Access breast coil focuses on patient ergonomics, allowing for enhanced comfort during both diagnostic and interventional exams. Positioning accessories, including a headrest, armrest and torso pad, help to relieve pressure on the sternum and provide customized patient positioning during the acquisition of breast MRI studies and interventional procedures.
The coil consists of a supporting base and two insulating coil chambers, one for each breast. Each of the hollow coil chambers houses two coil elements that are insulated from the patient by a rigid plastic housing. The coil housing is made of plastic materials, which are fire rated and have high impact and tensile strength.
The bilateral, open breast coil design offers optimized imaging capabilities and maximum access to aid the physician in MR-guided breast biopsy, localization of lesions, and interventional procedures, providing lateral, medial and cranial-caudal access to the breast.
#### Features of the Access Breast Coil:
#### Imaging
- 7-channel phased array design uses multiple elements that surround . the breast tissue
- t Optimized for parallel imaging applications
- Supports both unilateral and bilateral imaging exams .
- Compatible with Siemens 1.5T MRI systems .
- 40 cm field of view .
- Extensive coverage of chest wall and axilla ●
- High SNR t
#### Ergonomics
- Large apertures to accommodate a major percentage of patient . population
- Ergonomically designed for patient comfort to minimize motion . artifact
{2}------------------------------------------------
- Positioning accessories include a flexible, adjustable headrest, arm . rest and torso pad
- Cushioned head rest with mirror to reduce patient claustrophobic . response
- Padded arm rests for increased patient comfort .
#### Interventional Access
- Coil design allows lateral, medial and cranial-caudal interventional . access
- Lighting is integrated into the coil for interventional procedures . that demand a well lit area
- Compatible with SureLoc software to support interventional MRI .
## 6. Comparison with Legally Marketed Devices
The Access Breast Coil is substantially equivalent to the legally marketed devices listed below:
| Model: | Liberty 9000 Breast Coil |
|-----------------|------------------------------------|
| Manufacturer: | USA Instruments, Inc. |
| 510 (k) Number: | K000993 |
| Model: | Machnet Bilateral Open Breast Coil |
| Manufacturer: | Machnet BV |
| 510 (k) Number: | K013985 |
| Model: | Biopsy Breast Coil BBC-127 |
| Manufacturer: | MRI Devices Corporation |
| 510 (k) Number: | K041481 |
All of these breast coils are designed as receive-only RF coils and intended for use in conjunction with a magnetic resonance scanner to produce diagnostic images of the breast, chest wall and axillary tissues that can be interpreted by a physician.
The predicate devices are designed for use with various MR scanners. The Access Breast Coil is designed for use with the Siemens 1.5T MR scanner, as is the predicate device marketed by MRI Devices.
The Access Breast Coil is a 7-channel, phased array coil that allows bilateral and unilateral imaging of the breast. All of these breast coils have an open coil design to optimize imaging of breasts of various sizes.
{3}------------------------------------------------
The predicate devices allow lateral and medial access to the breast for biopsy The producate devries and in rassisting the Access Breast Coil is designed to allow for lateral, medial and cranial-caudal access to the breast.
#### 7. Conclusions
The Access Breast Coil is substantially equivalent to the identified legally The Access Dreast Oon is eating have been studied and controlled and controlled as marketed devices. - 110 pornent process, including risk analysis, test and part of the product as relephanned verification and validation testing design considerations, and Coil provides images comparable to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the border. Inside the circle is a stylized image of three birds in flight, arranged in a diagonal line. The birds are simple, with curved lines representing their bodies and wings.
AUG 2 4 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Patricia A. Milbank Regulatory Consultant Confirma, Inc. 821 Kirkland Avenue, Suite100 KIRKLAND WA 98033
Re: K052045
Trade/Device Name: Access Breast Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device
Regulatory Class: II Product Code: MOS Dated: July 27, 2005 Received: July 28, 2005
Dear Ms. Milbank:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter with anow you'ls ought maining of substantial equivalence of your device to a legally prematice notification: "Thesults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you deathe specific access to one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | , | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, produs nother general information on your responsibilities under the Act from the 807.77). " Ou may obtain variational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K052045
# Indication(s) for Use Statement
510(k) Number:
6
To be assigned by FDA
Device Name: Access Breast Coil
Indications for Use:
The Access Breast Coil is a 7-channel, phased array, receive-only RF coil, used in conjunction with a magnetic resonance scanner to produce diagnostic and interventional images of the breast, chest wall, and axillary tissues that can be interpreted by a trained physician. The device is designed to aid the physician in MR-guided breast biopsy, localization of lesions, and interventional procedures, providing lateral, medial and cranial-caudal access to the breast.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy bardon
(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.